OMNI-STAT Vascular (Rapid); CELOX Vascular Rapid
K190012 · Medtrade Product, Ltd. · QSY · Dec 18, 2019 · SU
Device Facts
| Record ID | K190012 |
| Device Name | OMNI-STAT Vascular (Rapid); CELOX Vascular Rapid |
| Applicant | Medtrade Product, Ltd. |
| Product Code | QSY · SU |
| Decision Date | Dec 18, 2019 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic |
Indications for Use
OMNI-STAT Vascular Rapid in intended to control surface bleeding from vascular access sites.
Device Story
OMNI-STAT Vascular Rapid is a chitosan-based hemostatic dressing for controlling surface bleeding at vascular access sites. The device consists of hemostatic granules and a bio-adhesive bonded to non-woven gauze, with a PU foam film backing. It is applied directly to the access site and point of compression while the introducer sheath is in place. Upon sheath removal, the dressing contacts blood, and with applied pressure, the bio-adhesive and granules form a gel-like plug that absorbs water from the blood to create a physical barrier. Used in clinical settings by healthcare providers, the device facilitates hemostasis, potentially reducing bleeding time and improving site management compared to traditional methods.
Clinical Evidence
Bench testing only. Performance evaluated via tack adhesion, absorbency, wet tensile strength, elongation, dimensions, and packaging integrity. Biocompatibility assessed per ISO 10993-1 via toxicology risk assessment using predicate data. Sterilization validated per ISO 11137-1. No new clinical or porcine studies were conducted; substantial equivalence is supported by bench data and historical data from predicate/reference devices.
Technological Characteristics
Chitosan-based hemostatic granules, bio-adhesive, non-woven gauze, PU foam film backing. Dimensions: 51mm x 51mm. Sterilization: Gamma irradiation (ISO 11137, SAL 10^-6). Single-use, prescription-only. Mechanical hemostasis via physical barrier/gel plug formation.
Indications for Use
Indicated for local management and control of surface bleeding from vascular access sites, percutaneous catheters or tubes utilizing introducer sheaths up to 16French.
Regulatory Classification
Identification
To temporarily control bleeding and cover external wounds.
Predicate Devices
Reference Devices
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
April 21, 2023
Medtrade Product Ltd Mina Patel Regulatory Affairs Manager Electra House, Electra way, Crewe Business Park Crewe. Cheshire CW1 6GL United Kingdom
Re: K190012 Trade/Device Name: OMNI-STAT Vascular (Rapid); CELOX Vascular Rapid Regulatory Class: Unclassified Product Code: QSY
Dear Mina Patel:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated December 18, 2019. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product code QSY.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Julie Morabito, OHT4: Office of Surgical and Infection Control Devices, 240-402-3839, Julie.Morabito@fda.hhs.gov.
Sincerely,
# Julie A. Morabito -S
Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 18, 2019
Medtrade Product Ltd Mina Patel Regulatory Affairs Manager Electra House, Electra way, Crewe Business Park Crewe, Cheshire CW1 6GL United Kingdom
Re: K190012
Trade/Device Name: OMNI-STAT Vascular (Rapid); CELOX Vascular Rapid Regulatory Class: Unclassified Product Code: FRO Dated: November 15, 2019 Received: November 18, 2019
Dear Mina Patel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Lixin Liu -S
Kimberly Ferlin, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K190012
Device Name:
OMNI-STAT Vascular Rapid
CELOX Vascular Rapid
Indications for Use (Describe)
OMNI-STAT Vascular Rapid is indicated for the local management and control of surface bleeding from vascular access sites, percutaneous catheters or tubes utilizing introducer sheaths up to 16French
CELOX Vascular Rapid is indicated for the local management and control of surface bleeding from vascular access sites, perculaneous catheters or tubes utilizing introducer sheaths up to 16French
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------------------|---------------------------------------------|
|-------------------------------------------------------------------------------------------------|---------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# SECTION 5
# 510(k) SUMMARY
This Traditional 510(k) Premarket Notification is to provide the basis for determining substantial equivalence of the Medtrade Products Ltd OMNI-STAT® Vascular Rapid to the predicate device presented below. The 510(k) # given is K190012.
| Submitted by: | Medtrade Product Ltd<br>Electra House, Crewe Business Park<br>Crewe, Cheshire<br>CW1 6GL<br>United Kingdom |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration No.: | 9614493 |
| Contact: | Ms Mina Patel<br>Regulatory Affairs Manager<br>Medtrade Product Ltd<br>Electra House, Crewe Business Park<br>Crewe, Cheshire<br>CW1 6GL<br>United Kingdom<br><br>Telephone: + 44 (0)1270 500019<br>Facsimile: + 44 (0)1270 500045<br>Email: mina.patel@medtrade.co.uk |
| | |
| Date prepared: | 18th Dec 2019 |
| Device Identification: | |
| Device Name: | OMNI-STAT® Vascular Rapid |
| Common Name: | Dressing - Vascular Closure |
| Trade/Proprietary Names: | OMNI-STAT® Vascular Rapid,<br>CELOX™ Vascular Rapid |
| Model Number: | FG08834311 Size: 2in x 2in / 51mm x 51mm |
| Product Code: | FRO |
| Classification Name: | Dressing, Wound, Drug |
| Classification: | Unclassified |
{5}------------------------------------------------
# Predicate Device
510(K) number K093519 – CELOX Vascular
# Reference Device
510(K) number K110386 – CELOX Rapid Z-Fold Gauze
# Device Description:
OMNI-STAT Vascular Rapid is a chitosan-based hemostatic dressing/ pad presented in a sealed foil pouch, consisting of hemostatic granules with a bioadhesive bonded to one side of non-woven gauze and with a PU foam film backing on the other side of the gauze. The dressing is designed to control surface bleeding from vascular access sites. The granules coated side is applied directly to the access site and pressure applied until hemostasis is achieved.
OMNI-STAT Vascular Rapid achieves its principle intended action (hemostasis) by creating a physical barrier or plug to stop the flow of blood. The granules coated sides is applied directly to the access site and the point of compression, while the sheath is still in place. When the sheath is removed, and the product comes into contact with blood, with pressure applied OMNI-STAT Vascular Rapid with the bio-adhesive quickly forms a gel like plug absorbing the water from the blood.
# Intended Use
OMNI-STAT Vascular Rapid in intended to control surface bleeding from vascular access sites.
# Indications for use:
OMNI-STAT Vascular Rapid / CELOX Vascular Rapid is indicated for the local management and control of surface bleeding from vascular access sites, percutaneous catheters or tubes utilizing introducer sheaths up to 16French.
# Comparison of technological characteristics with the predicate device:
The technological characteristics of both the OMNI-STAT Vascular Rapid and the predicate device, CELOX Vascular are described in the following table:
{6}------------------------------------------------
| Category | OMNI-STAT Vascular Rapid<br>Subject Device<br>#K190012 | CELOX Vascular<br>Predicate Device<br>#K093519 |
|-----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design | OMNI-STAT Vascular Rapid is a<br>chitosan-based hemostatic<br>dressing/ pad presented in a<br>sealed foil pouch, consisting of<br>hemostatic granules with a bio-<br>adhesive bonded to one side a<br>non-woven base gauze and with<br>a PU foam film backing the<br>other side of the gauze.<br>OMNI-STAT Vascular Rapid<br>achieves its principle intended<br>action (hemostasis) by creating<br>a physical barrier or plug to stop<br>the flow of blood. The granules<br>coated sides is applied directly<br>to the access site and the point<br>of compression, while the<br>sheath is still in place. When the<br>sheath is removed, and the<br>product comes into contact with<br>blood, with pressure applied<br>OMNI-STAT Vascular Rapid<br>with the bio-adhesive quickly<br>forms a gel like plug absorbing<br>the water from the blood. | CELOX Vascular is a chitosan-<br>based hemostatic dressing/<br>pad presented in a sealed foil<br>pouch, consisting of<br>hemostatic granules bonded to<br>one side a non-woven gauze<br>and with a PU foam film<br>backing the other side of the<br>gauze.<br>CELOX Vascular achieves its<br>principle intended action<br>(hemostasis) by creating a<br>physical barrier or plug to stop<br>the flow of blood. The granules<br>coated sides is applied directly<br>to the access site and the<br>point of compression, while the<br>sheath is still in place. When<br>the sheath is removed, and the<br>product comes into contact<br>with blood, with pressure<br>applied CELOX Vascular<br>quickly forms a gel like plug<br>absorbing the water from the<br>blood. |
| Material/<br>chemical<br>Composition | A chitosan-based hemostatic<br>granules coated dressing/pad,<br>with a bio-adhesive and PU<br>foam film backing | A chitosan-based hemostatic<br>granules coated dressing/pad<br>and PU foam film backing |
| Indications<br>for Use<br>(Prescription<br>/Rx only) | OMNI-STAT Vascular Rapid is<br>indicated for the local<br>management and control of<br>surface bleeding from vascular<br>access sites, percutaneous<br>catheters or tubes utilizing<br>introducer sheaths up to<br>16French | CELOX Vascular is indicated<br>for the local management and<br>control of surface bleeding<br>from vascular access sites,<br>percutaneous catheters or<br>tubes utilizing introducer<br>sheaths up to 16French |
| Category | OMNI-STAT Vascular Rapid<br>Subject Device<br>#K190012 | CELOX Vascular<br>Predicate Device<br>#K093519 |
| For single use | Yes | Yes |
| For prescription<br>use<br>(Rx only) | Yes | Yes |
| Method of<br>sterilisation | Gamma Irradiation in<br>accordance with ISO 11137 | Gamma Irradiation in<br>accordance with ISO 11137 |
| Sterility<br>assurance level | Achieve Sterility Assurance<br>10-6 | Achieve Sterility Assurance<br>10-6 |
| Shelf-life | 3 Years based on predicate<br>stability data | 5 Years |
| Biocompatibility<br>testing<br>completed in<br>accordance with<br>ISO 10993 | No testing was carried out on<br>subject device;<br>biocompatibility profile was<br>established based on<br>toxicology risk assessment<br>and data from CELOX<br>Vascular and CELOX Rapid<br>Gauze. | Yes |
{7}------------------------------------------------
Both the OMNI-STAT Vascular Rapid and the predicate device CELOX Vascular are composed of the same raw materials besides in addition of a bio-adhesive in OMNI-STAT Vascular Rapid. This leads to OMNI-STAT Vascular Rapid having a slightly better adhesion profile than CELOX Vascular.
# Legally marketed device(s) for substantial equivalence comparison:
The OMNI-STAT® Vascular Rapid is a modification to an existing device cleared in K093519 CELOX™ Vascular on 14th Jan 2010 with the addition of granules used in CELOX™ Rapid Gauze cleared in K110386 on 10th May 2011.
The reason for the modification and key benefit of OMNI-STAT Vascular Rapid to CELOX™ Vascular is the enhanced adhesion profile achieved with the addition of a bio-adhesive.
{8}------------------------------------------------
# Performance Data
#### Non-Clinical Performance Data:
The following performance data was carried out in line with Medtrade products Ltd Design Control process in support of the substantial equivalence for the modified device:
Bench testing of OMNI-STAT Vascular Rapid - the subject device and its packaging was conducted on all of the main performance characteristics. The performance of the OMNI-STAT Vascular Rapid was also analysed in bench testing against the predicate device, CELOX Vascular. The following comparison tests were performed:
- Tack Adhesion 1)
- 2) Absorbency
- Wet Tensile Strength and Elongation 3)
- Dimension 4)
- Packaging Integrity & Strength 5)
All testing completed on both the subject device and predicate devices meet the defined criteria.
No additional pre-clinical testing was deemed necessary to support substantial equivalence between the OMNI-STAT Vascular Rapid and the cleared predicate device. Testing in porcine bleeding models was performed for both the predicate devices CELOX Vascular and CELOX Rapid Gauze. The results demonstrate that they both support the claim with regards to the control of bleeding and hemostasis being achieved.
The biocompatibility of the OMNI-STAT Vascular Rapid has been assessed and a rationale compiled stating the device is deemed in compliance with the requirements of ISO 10993-1 (Biological Evaluation of Medical Devices) based on toxicology risk assessment and data from the predicate device CELOX Vascular and the reference device CELOX Rapid Gauze.
Sterilization validation has been assessed and evaluated in compliance with ISO 11137-1 - Sterilization of healthcare products - Radiation - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices.
Sterilisation dose audits are carried out using VDmax25 as defined in standard ISO 11137-1.
The biocompatibility and non-clinical performance evaluations carried out for OMNI-STAT Vascular Rapid has demonstrated that they were adequate to
{9}------------------------------------------------
support determination of substantial equivalent to the original unmodified CELOX Vascular predicate device for the intended indications of use.
# Summary Statement of Substantial Equivalence:
The OMNI-STAT Vascular Rapid device is similar in composition to predicate device CELOX Vascular cleared under K093519 on Jan 14, 2010 with the addition of CELOX Rapid Granules used in CELOX Rapid Gauze cleared under K110386 on May 10, 2011 both of which are manufactured by Medtrade Products Ltd. The intended use and indications of use of the two devices (subject and predicate) are identical for the subject device OMNI-STAT Vascular Rapid and the predicate device CELOX Vascular and the above data demonstrating that the subject device is substantially equivalent to the predicate device CELOX Vascular.