K190005 · Sterilucent, Inc. · MLR · Sep 13, 2019 · General Hospital
Device Facts
Record ID
K190005
Device Name
Sterilucent HC 80TT Hydrogen Peroxide Sterilizer
Applicant
Sterilucent, Inc.
Product Code
MLR · General Hospital
Decision Date
Sep 13, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6860
Device Class
Class 2
Indications for Use
The Sterilucent HC 80TT Hydrogen Peroxide Sterilizer is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in healthcare facilities. The two pre-programmed sterilization cycles, the Lumen Cycle and the Flexible Cycle, operate at low pressure and low temperature and are thus suitable for processing medical devices that are sensitive to heat and moisture.
Device Story
The Sterilucent HC 80TT is a standalone, low-temperature, low-pressure vaporized hydrogen peroxide sterilizer for reusable medical devices. It uses a 59% aqueous hydrogen peroxide solution, which is heated and injected into an 80L aluminum chamber. The device features two pre-programmed cycles: Lumen and Flexible. It is operated by healthcare staff in clinical settings via a 7-inch capacitive touch TFT LCD GUI. The system monitors chamber temperature, pressure, and sterilant concentration using thermistors, transducers, and a real-time vapor monitor. Output includes cycle status and process verification via biological/chemical indicators and an integrated printer. The device ensures sterility of heat/moisture-sensitive instruments, including complex endoscopes, by maintaining a Sterility Assurance Level (SAL) of 10^-6. It benefits patients by providing effective terminal sterilization for delicate surgical instruments.
Clinical Evidence
Bench testing only. Validation included overkill method testing using Geobacillus stearothermophilus spores on various materials, mated surfaces, and lumen configurations. Results demonstrated an SAL of at least 10^-6. Simulated use testing with organic/inorganic challenge soils and in-use testing with clinically soiled instruments confirmed sterilization efficacy. No clinical trials were performed.
Technological Characteristics
Self-contained, standalone, deep vacuum sterilizer. Chamber: 80L aluminum. Sterilant: 59% aqueous hydrogen peroxide. Heating: electric silicone-rubber resistive heaters. Monitoring: chamber/vaporizer thermistors, pressure transducers, real-time H2O2 vapor monitor. Interface: 7-inch capacitive touch TFT LCD. Connectivity: USB port for external communication. Firmware-controlled, non-programmable.
Indications for Use
Indicated for terminal sterilization of cleaned, rinsed, and dried reusable metal and non-metal medical devices in healthcare facilities. Lumen Cycle: lumen/non-lumen devices with diffusion-restricted spaces; rigid/semi-rigid stainless steel endoscopes (≥0.77mm ID/≤410mm L, ≥1.33mm ID/≤430mm L, or ≥1.00mm ID/≤310mm L). Flexible Cycle: rigid/semi-rigid non-lumen devices; flexible endoscopes (single channel ≥1.00mm ID/≤1280mm L; dual channel ≥0.80mm ID/≤1000mm L).
Regulatory Classification
Identification
An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
September 13, 2019
Sterilucent, Inc. Peter Kalkbrenner Director of Engineering 1400 Marshall St. NE Minneapolis, Minnesota 55413
Re: K190005
Trade/Device Name: Sterilucent HC 80TT Hydrogen Peroxide Sterilizer Regulation Number: 21 CFR 880.6860 Regulation Name: Ethylene Oxide Gas Sterilizer Regulatory Class: Class II Product Code: MLR Dated: August 8, 2019 Received: August 13, 2019
Dear Peter Kalkbrenner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K190005
Device Name
Sterilucent HC 80TT Hydrogen Peroxide Sterilizer
Indications for Use (Describe)
Indications for Use:
The Sterilucent HC 80TT Hydrogen Peroxide Sterilizer is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterlization of cleaned, rinsed and dried reusable metal medical devices used in healthcare facilities. The two pre-programmed sterilization cycles, the Lumen Cycle and the Flexible Cycle, operate at low temperature and are thus suitable for processing medical devices that are sensitive to heat and moisture.
The HC 80TT Lumen Cycle can sterilize:
o Lumen and non-lumen medical devices with diffusion-restricted spaces or mated surfaces such as the hinged portion of forceps and scissors;
O Single or dual channeled rigid and semi-rigid endoscopes, with stainless steel lumens that are:
- · ≥ 0.77 mm internal diameter (ID) and ≤ 410 mm long, or
- · ≥ 1.33 mm ID and ≤ 430 mm long; and,
o Triple channeled rigid and semi-rigid endoscopes, with stainless steel lumens that are:
- ≥ 1.00 mm ID and ≤ 310 mm long.
The validation testing for all lumen sizes was conducted using a maximum of fifteen (15) Jumens per load. Hospital loads should not exceed the maximum number offumens validated by this testing. The validation studies were performed using two(2) separate validation loads as described in the following table.
| HC-80TT-Lumen-Cycle-Validation-Load-Descriptions | | |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Load# | Description | Load-<br>Weight |
| 1 | Two-wrapped-trays-with-silicone-mats, one-(1)-triple-channel-<br>hysteroscope, three-(3)-dual-channel-rigid-ureteroscopes, -six-(6)-suction-<br>tubes-(15-total-lumens), -and-general-medical-instruments-such-as-<br>clamps, forceps, shears, mallets, scissors-and-retractors. | 20.2-lb. |
| 2 | Wrapped-heavy-drill-set-with-ten-(10)-lumens-and-pouched-batteries. | 28.1-lb. |
Excluding the weight of the wrap.
- 1 Excluding the weight of the wrap.
The HC 80TT Flexible Cycle can sterilize:
- · Rigid or semi-rigid non-lumen medical devices including non-lumen devices with metallic diffusion-restricted spaces such or mated surfaces such as the hinged portion of forceps or scissors;
- · Single channel flexible endoscopes with flexible Iumens that are:
- · ≥ 1.00 mm ID and ≤ 1280 mm long; and,
- · Dual channel flexible endoscopes with flexible Iumens that are:
- · 2: 0.80 mm ID and :S 1000 mm long.
The validation studies were performed using four (4) separate validation loads as described in the following table.
FORM FDA 3881 (7/17) Page 1 of 2 psc Publishing Services (30)) 433-6740 EF
{3}------------------------------------------------
-
| HC-80TT-Flexible-Cycle-Validation-Load-Descriptionsα | | |
|------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Load#α | Descriptionα | Load-Weight¹α |
| 1α | Two-wrapped-trays-with-silicone mats, one-(1)-single-channel-flexible-<br>angioscope (1.0mm-ID-x-1280mm-L), -one-(1)-single-channel-flexible-<br>ureteroscope-(1.2mm-ID-x-800mm-L), -and-general-medical-instruments-<br>such-as-clamps, forceps, shears, rongeurs, mallets, scissors, retractors-<br>and-spreaders.α | 31.9-lb.α |
| 2α | Two-wrapped-trays with silicone-mats-and-general-medical-instruments-<br>such-as-clamps, forceps, shears, rongeurs, mallets, scissors, retractors-<br>and-spreaders.α | 30.6-lb.α |
| 3α | Two-wrapped-trays with silicone mats, one-with-one-(1)-dual-channel-<br>flexible-ureteroscope-(1.2/0.8-mm-ID-x-850/1000-mm-L)-with-light-cord-<br>and-accessories, the other with one-(1)-single-channel-flexible-<br>angioscope (1.0mm-ID-x-1280mm-L)-with-light-cord.α | 14.4-lb.α |
| • 4α | Two-wrapped-trays, one-with-a-silicone-mat-and-general-medical-<br>instruments-such as forceps, shears, mallets, and retractors, the other-<br>with eight-(8)-batteries.α | 17.6-lb.α |
· 1 Excluding the weight of the wrap. T
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 2 of 2
{4}------------------------------------------------
# 510(K) SUMMARY
| Submitted by: | Sterilucent, Inc.<br>1400 Marshall Street NE<br>Minneapolis, MN 55413<br>Phone: 612-767-3260<br>Fax: 612-767-3261 |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Peter R. Kalkbrenner<br>Director of Engineering<br>peter.kalkbrenner@sterilucent.com<br>Phone: 612-767-3253<br>Fax: 612-767-3261 |
| Date of Summary: | September 11, 2019 |
| Device Trade Name: | Sterilucent HC 80TT Hydrogen Peroxide Sterilizer |
| Common or Usual Name: | Vapor Phase Hydrogen Peroxide Sterilizer |
| Classification: | 21 CFR 880.6860 |
| Class: | II |
| Product Code: | MLR |
| Predicate Device(s): | Sterilucent PSD-85 Hydrogen Peroxide Sterilizer (K140464) |
| Device Description: | The Sterilucent HC 80TT Hydrogen Peroxide Sterilizer (HC 80TT)<br>is a self-contained standalone device, using vaporized hydrogen<br>peroxide as the sterilant. The HC 80TT is intended for use in the<br>terminal sterilization of cleaned, rinsed, and dried, reusable metal and<br>non-metal medical device used in healthcare facilities. The<br>sterilization cycle operates at low pressure and temperatures and is<br>therefore suitable for processing medical devices that are sensitive to<br>heat and moisture. The hydrogen peroxide vapor is generated by<br>heating aqueous hydrogen peroxide and injecting into a chamber. |
| Indications for Use: | The Sterilucent HC 80TT Hydrogen Peroxide Sterilizer is a<br>vaporized hydrogen peroxide sterilizer intended for use in the<br>terminal sterilization of cleaned, rinsed and dried reusable metal and<br>nonmetal medical devices used in healthcare facilities. The two pre-<br>programmed sterilization cycles, the Lumen Cycle and the Flexible<br>Cycle, operate at low pressure and low temperature and are thus<br>suitable for processing medical devices that are sensitive to heat and<br>moisture.<br>The HC 80TT Lumen Cycle can sterilize:<br>• Lumen and non-lumen medical devices with diffusion-<br>restricted spaces or mated surfaces such as the hinged<br>portion of forceps and scissors; |
{5}------------------------------------------------
| ●<br>● | Single or dual channeled rigid and semi-rigid endoscopes,<br>with stainless steel lumens that are:<br>≥ 0.77 mm internal diameter (ID) and ≤ 410 mm<br>O<br>long, or<br>≥ 1.33 mm ID and ≤ 430 mm long; and,<br>O<br>Triple channeled rigid and semi-rigid endoscopes, with<br>stainless steel lumens that are:<br>≥ 1.00 mm ID and ≤ 310 mm long.<br>O | | |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|--|
| table. | The validation testing for all lumen sizes was conducted using<br>a maximum of fifteen (15) lumens per load. Hospital loads<br>should not exceed the maximum number of lumens validated<br>by this testing. The validation studies were performed using<br>two (2) separate validation loads as described in the following | | |
| | HC 80TT Lumen Cycle Validation Load Descriptions | | |
| Load<br># | Description | Load<br>Weight1 | |
| 1 | Two wrapped trays with silicone mats, one<br>(1) triple channel hysteroscope, three (3) dual<br>channel rigid ureteroscopes, six (6) suction<br>tubes (15 total lumens), and general medical<br>instruments such as clamps, forceps, shears,<br>mallets, scissors and retractors. | 20.2 lb. | |
| 2 | Wrapped heavy drill set with ten (10) lumens<br>and pouched batteries. | 28.1 lb. | |
| ●<br>● | 1 Excluding the weight of the wrap.<br>The HC 80TT Flexible Cycle can sterilize:<br>Rigid or semi-rigid non-lumen medical devices including non-lumen<br>devices with metallic diffusion-restricted spaces such or mated<br>surfaces such as the hinged portion of forceps or scissors;<br>Single channel flexible endoscopes with flexible lumens that<br>are:<br>≥ 1.00 mm ID and ≤ 1280 mm long; and,<br>O<br>Dual channel flexible endoscopes with flexible lumens that<br>are:<br>≥ 0.80 mm ID and ≤ 1000 mm long.<br>O<br>The validation studies were performed using four (4) separate<br>validation loads as described in the following table. | | |
| HC 80TT Flexible Cycle Validation Load Descriptions | | | |
| Load<br># | Description | Load<br>Weight1 | |
| 1 | Two wrapped trays with silicone mats, one (1) single channel flexible angioscope (1.0mm ID x 1280mm L), one (1) single channel flexible ureteroscope (1.2mm ID x 800mm L), and general medical instruments such as clamps, forceps, shears, rongeurs, mallets, scissors, retractors and spreaders. | 31.9 lb. | |
| 2 | Two wrapped trays with silicone mats and general medical instruments such as clamps, forceps, shears, rongeurs, mallets, scissors, retractors and spreaders. | 30.6 lb. | |
| 3 | Two wrapped trays with silicone mats, one with one (1) dual channel flexible ureteroscope (1.2/0.8 mm ID x 850/1000 mm L) with light cord and accessories, the other with one (1) single channel flexible angioscope (1.0mm ID x 1280mm L) with light cord. | 14.4 lb. | |
| 4 | Two wrapped trays, one with a silicone mat and general medical instruments such as forceps, shears, mallets, and retractors, the other with eight (8) batteries. | 27.7 lb. | |
| 1 Excluding the weight of the wrap. | | | |
| Technological<br>Characteristics | The Sterilucent HC 80TT Hydrogen Peroxide Sterilizer (HC 80TT) is compared to the predicate device Sterilucent PSD-85 Hydrogen Peroxide Sterilizer (K140464) using the information supplied below. | | |
{6}------------------------------------------------
{7}------------------------------------------------
| | Subject Device:<br>Sterilucent HC 80TT Hydrogen Peroxide Sterilizer (HC<br>80TT) | Predicate Device (K140464):<br>Sterilucent PSD-85 Hydrogen Peroxide Sterilizer<br>(PSD-85) |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | 21 CFR 880.6860 | 21 CFR 880.6860 |
| Product Code | MLR | MLR |
| Indications for Use | The Sterilucent HC 80TT Hydrogen Peroxide Sterilizer is a<br>vaporized hydrogen peroxide sterilizer intended for use in the<br>terminal sterilization of cleaned, rinsed and dried reusable<br>metal and nonmetal medical devices used in healthcare<br>facilities. The two pre-programmed sterilization cycles, the<br>Lumen Cycle and the Flexible Cycle, operate at low pressure<br>and low temperature and are thus suitable for processing<br>medical devices that are sensitive to heat and moisture.<br>The HC 80TT Lumen Cycle can sterilize:<br>Lumen and non-lumen medical devices with diffusion-<br>restricted spaces or mated surfaces such as the hinged<br>portion of forceps and scissors; Single or dual channeled rigid and semi-rigid<br>endoscopes, with stainless steel lumens that are:<br>$\geq$ 0.77 mm internal diameter (ID) and $\leq$ 410 mm long, or $\geq$ 1.33 mm ID and $\leq$ 430 mm long; and, Triple channeled rigid and semi-rigid endoscopes, with<br>stainless steel lumens that are:<br>$\geq$ 1.00 mm ID and $\leq$ 310 mm long. The validation testing for all lumen sizes was conducted using<br>a maximum of fifteen (15) lumens per load. Hospital loads | The Sterilucent PSD-85 Hydrogen Peroxide Sterilizer<br>(PSD-85) is a vaporized hydrogen peroxide sterilizer<br>intended for use in the terminal sterilization of<br>cleaned, rinsed and dried reusable metal and nonmetal<br>medical devices used in healthcare facilities. The two<br>pre-programmed sterilization cycles, the Lumen and<br>the Non-Lumen Cycles, operate at low<br>pressure and low temperature and are thus suitable for<br>processing medical devices sensitive to heat and<br>moisture.<br>The PSD-85 Lumen Cycle can sterilize*:<br>Instruments with diffusion-restricted spaces<br>such as the hinged portion of forceps and<br>scissors Medical devices with a single stainless steel<br>lumen with: An inside diameter of 1 mm or larger<br>and a length of 60 mm or shorter An inside diameter of 2 mm or larger<br>and a length of 250 mm or shorter An inside diameter of 3 mm or larger<br>and a length of 350 mm or shorter |
{8}------------------------------------------------
| Load # | Description | Load Weight1 |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| 1 | Two wrapped trays with silicone mats, one (1) triple channel hysteroscope, three (3) dual channel rigid ureteroscopes, six (6) suction tubes (15 total lumens), and general medical instruments such as clamps, forceps, shears, mallets, scissors and retractors. | 20.2 lb. |
| 2 | Wrapped heavy drill set with ten (10) lumens and pouched batteries. | 28.1 lb. |
| 1Excluding the weight of the wrap.<br><br>The HC 80TT Flexible Cycle can sterilize:<br>Rigid or semi-rigid non-lumen medical devices including non-lumen devices with metallic diffusion- restricted spaces such or mated surfaces such as the hinged portion of forceps or scissors; Single channel flexible endoscopes with flexible lumens that are: $\geq$ 1.00 mm ID and $\leq$ 1280 mm long; and, Dual channel flexible endoscopes with flexible lumens that are: $\geq$ 0.80 mm ID and $\leq$ 1000 mm long. | | |
{9}------------------------------------------------
| The validation studies were performed using four (4) separate<br>validation loads as described in the following table. | | |
|------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| <b>HC 80TT Flexible Cycle Validation Load Descriptions</b> | | |
| Load # | Description | Load Weight1 |
| 1 | Two wrapped trays with silicone mats, one (1) single channel flexible angioscope (1.0mm ID x 1280mm L), one (1) single channel flexible ureteroscope (1.2mm ID x 800mm L), and general medical instruments such as clamps, forceps, shears, rongeurs, mallets, scissors, retractors and spreaders. | 31.9 lb. |
| 2 | Two wrapped trays with silicone mats and general medical instruments such as clamps, forceps, shears, rongeurs, mallets, scissors, retractors and spreaders. | 30.6 lb. |
| 3 | Two wrapped trays with silicone mats, one with one (1) dual channel flexible ureteroscope (1.2/0.8 mm ID x 850/1000 mm L) with light cord and accessories, the other with one (1) single channel flexible angioscope (1.0mm ID x 1280mm L) with light cord. | 14.4 lb. |
| 4 | Two wrapped trays, one with a silicone mat and general medical | 27.7 lb. |
{10}------------------------------------------------
| | instruments such as forceps, shears,<br>mallets, and retractors, the other with<br>eight (8) batteries.<br>1 Excluding the weight of the wrap. | |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Single Use/Reusable | Reusable | Reusable |
| Type of Process | Terminal Sterilization | Terminal Sterilization |
| Sterility Assurance Level | $10^{-6}$ | $10^{-6}$ |
| Sterilant | 59% aqueous solution of hydrogen peroxide (H2O2) | 59% aqueous solution of hydrogen peroxide (H2O2) |
| Sterilization Cycles | Two pre-programmed cycles:<br>Lumen (approximately 65 min) and<br>Flexible (approximately 35 min) | Two pre-programmed cycles:<br>Lumen (approximately 72 min) and<br>Non-Lumen (approximately 40 min) |
| Physical Characteristics | Self-contained, stand-alone device | Self-contained, stand-alone device |
| Design and Construction | Welded frame onto which is mounted the sterilization chamber<br>along with a variety of instruments and components, controls,<br>piping, and vacuum pump, all of which is housed in an<br>enclosed frame | Welded frame onto which is mounted the sterilization<br>chamber along with a variety of instruments and<br>components, controls, piping, and vacuum pump, all<br>of which is housed in an enclosed frame |
| • Chamber Volume | 80 L | 85 L |
| • Weight | 306 lbs (139 kg) | 440 lbs (200 kg) |
| • Maximum Power | 2400 Watts | 1650 Watts |
| Overall design | Deep vacuum sterilizer<br>Firmware controlled, non-programmable<br>Single door<br>Two shelves | Deep vacuum sterilizer<br>Firmware controlled, non-programmable<br>Single door<br>One shelf |
| Chamber Materials | Aluminum | Stainless steel |
| Chamber Heating system | Electric silicone-rubber resistive heaters | Electric silicone-rubber resistive heaters |
| Internal Process Monitors | | |
| | | |
| • Temperature | Chamber and vaporizer thermistors | Chamber and vaporizer thermistors |
| • Pressure | Chamber pressure transducers | Chamber pressure transducers |
| • Sterilant Concentration | Real-time hydrogen peroxide vapor monitor | Real-time hydrogen peroxide vapor monitor |
| User Interface | 7 inch capacitive touch thin-film-transistor (TFT) liquid-crystal<br>display (LCD) graphical user interface (GUI) | 40 character x 4 line dot matrix liquid crystal display<br>(LCD) |
| External Process Monitors | • Electronic controls<br>• USB port for communicating with separate computer<br>• Printer | • Electronic controls<br>• USB port for communicating with separate<br>computer |
| Operational Principle | Sterilization of the intended devices by exposure under<br>controlled conditions of pressure, temperature, and time | Sterilization of the intended devices by exposure<br>under controlled conditions of pressure, temperature,<br>and time |
| Operational Parameters | Low pressure and temperature | Low pressure and temperature |
| Pre-processing requirements | Cleaned, rinsed and dried medical devices | Cleaned, rinsed and dried medical devices |
| Amount of sterilant per injection | Variable, based on actual measured vapor concentration | Variable, based on actual measured vapor<br>concentration |
| Monitoring – biological and chemical indicators | | |
| • Biological Indicator | Self-contained biological indicator, Geobacillus<br>stearothermophilus | Self-contained biological indicator, Geobacillus<br>stearothermophilus |
| • Process Challenge<br>Device/Routine Test Pack | Self-contained biological indicator, Geobacillus<br>stearothermophilus | Self-contained biological indicator, Geobacillus<br>stearothermophilus |
| • Chemical Indicator | Sterilucent CI Strips, Labels | Sterilucent CI Strips, Labels and Tape |
| Miscellaneous | | |
| Materials Sterilized | Reusable metal and non-metal medical devices that are used in<br>healthcare facilities, including those that are sensitive to heat<br>and moisture | Reusable metal and non-metal medical devices that<br>are used in healthcare facilities, including those that<br>are sensitive to heat and moisture |
Sterilucent, Inc.
{11}------------------------------------------------
{12}------------------------------------------------
Comparison to the Both the HC 80TT and the predicate device have the same: Predicate device: . Intended use Processes validated to a Sterility assurance level (10-6) ● . Sterilant The HC 80TT design differs slightly from the predicate; the HC 80TT is slightly smaller and lighter that the PSD-85. Using the Lumen Cycle, the HC 80TT can sterilize single-, dual-, or triplechanneled rigid and semi-rigid endoscopes with stainless steel lumens as specified in the Indications for Use, while the PSD-85 is indicated to sterilize medical devices with only a single stainless-steel lumen of larger diameter and shorter length. The HC 80TT's "Flexible Cycle" and the PSD-85's "Non-Lumen Cycle'' both sterilize non-lumen instruments with diffusion-restricted spaces, but the HC 80TT is also indicated to sterilize single- and dual-channel flexible endoscopes. Performance Testing The HC 80TT sterilization cycle parameters have been shown to provide a (Bench): level of safety and efficacy at least equivalent to that of the predicate device. Testing was performed using the "overkill" method. Pre-Validation Testing Test Organism: Geobacillus stearothermophilus Process Variables and Parameters Testing was conducted to characterize the effect of process parameters on lethality. The four critical process parameters are chamber wall temperature, vaporizer temperature, injection pressure and vaporized hydrogen peroxide concentration. The level tested for each parameter was selected to provide a worst-case situation for the test series and to be outside the abort levels or settings for the sterilizer. The study showed that process lethality was unaffected over the range of tested process parameters. HC 80TT Sterilization Process Validation Demonstration of Dose-Response Relationship to Increasing Hydrogen Peroxide Concentration Dose response testing was performed using various materials (representative of materials used in medical devices) as spore carriers. Geobacillus stearothermophilus death kinetics data obtained for each material tested demonstrate a positive "dose response" to increasing concentration of hydrogen peroxide injected under half cycle conditions in the HC 80TT. There were no spore survivors on any material at concentrations lower than the standard exposure concentration for the Lumen and Flexible Cycles. The result
{13}------------------------------------------------
demonstrates that the dose response observed is not limited to a single substrate and in each case >6 Spore Log Reduction (SLR) was observed for a half cycle exposure.
## Surface Sterilization
The purpose of the study was to demonstrate sterilization of medical device surfaces. Geobacillus stearothermophilus spores were inoculated on a wide variety of material coupons that were representative of materials used in reusable medical devices. The coupons were inoculated with at least 106 Geobacillus stearothermophilus spores and then exposed to ½ cycle sterilization parameters.
Results from this testing demonstrate a Sterility Assurance Level (SAL) of at least 10-6 for medical device surface sterilization in the HC 80TT for all materials listed as recommended for use in the HC 80TT.
# Mated Surface Sterilization
The purpose of the study was to demonstrate sterilization of mated medical device surfaces using the HC 80TT. Geobacillus stearothermophilus spores were inoculated on a variety of medical device materials. The materials were than mated to the same material and exposed to ½ cycle sterilization parameters. An SAL of least 10° was demonstrated for mated material sterilization.
## Lumen Sterilization
The purpose of this validation test was to demonstrate that the HC 80TT could effectively sterilize specific dimensions of rigid stainless steel and flexible lumened medical devices. Testing was completed by placing at least 10° Geobacillus stearothermophilus spores in the center of the lumens and exposing them to ½ cycle sterilization parameters. There were no spore survivors after multiple ½ cycle exposure tests.
Results from this testing demonstrate an SAL of at least 10-6 for lumen sterilization in the HC 80TT for the lumen dimensions listed as indicated for use in the HC 80TT.
# Simulated Use Testing
Simulated use testing was performed to confirm sterilization of medical device surfaces and lumens after processing in the HC 80TT. Representative lumen and non-lumen devices were inoculated with at least 10-6 Geobacillus stearothermophilus spores, suspended in an inorganic and organic challenge soil and exposed to a full sterilization cycle. The results showed sterility of all tested devices.
# In-Use Testing
{14}------------------------------------------------
Devices representative of surface features and lumen claims for the HC 80TT were selected for sterility testing. Devices including stainless steel devices with open surfaces, mated or hinged surfaces, representative stainless steel lumen devices, and representative flexible scopes were soiled with clinical soils. The soiled devices were manually cleaned and/or washed per the Manufacturer's Instructions for Use, dried, packaged and processed in the applicable HC 80TT cycle (Lumen or Flexible). The results of the in-use testing demonstrated that the HC 80TT successfully sterilizes surgical instruments used in clinical settings.
# Supporting Microbiological Testing
Sterilizing agent efficacy (Sporicidal Activity of a Disinfectant) testing was performed in the HC 80TT in accordance with the guidelines provided in AOAC Official Method 966.04, Sporicidal Activity of Disinfectants. None of the carriers demonstrated growth.
## Software Validation
Software validation was performed according to FDA's moderate level of concern recommendations provided in the document "Guidance for the Content for Premarket Submissions for Software Contained in Medical Devices".
# Conclusion:
The nonclinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.