← Product Code MRW · K183589

# PMT Facet Screw (K183589)

_Providence Medical Technology, Inc. · MRW · Oct 11, 2019 · OR · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K183589

## Device Facts

- **Applicant:** Providence Medical Technology, Inc.
- **Product Code:** MRW
- **Decision Date:** Oct 11, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** OR
- **Attributes:** Therapeutic

## Intended Use

The PMT Facet Screw is intended to spine as an aid to fusion through bilateral immobilization of the facet joints. The screws are inserted posteriorly through the superior side of the facet joint, and into the pedicle. The PMT Facet Screw is indicated for bilateral facet fixation, with or without bone graft, at single or multiple levels, from C2 to S1. The PMT Facet Screw is indicated for treatment of any or all of the following: 1. Pseudoarthrosis and failed previous fusion 2. Spondylolisthesis 3. Spondylolysis 4. Degenerative disc disease (DDD) as defined by neck and/or back pain of discogenic origin as confirmed by radiographic studies 5. Degeneration of the facets with instability; and 6. Fracture The PMT Facet Screw is intended for conventional or tissue-sparing surgical placement.

## Device Story

PMT Facet Screw is a spinal fixation implant; used to stabilize spine via bilateral facet joint immobilization. Screws are inserted posteriorly through superior facet joint into pedicle. Device facilitates fusion; indicated for levels C2-S1. Used in surgical settings by surgeons for conventional or tissue-sparing procedures. Provides mechanical stability to treat degenerative conditions, fractures, and failed previous fusions. Benefits patient by restoring spinal stability and aiding fusion process.

## Clinical Evidence

No clinical testing was performed. Substantial equivalence is supported by bench testing, including static and dynamic cantilever bending (ASTM F2193-14), static torsion, static pull-out, and drive torque testing (ASTM F543-17), alongside pyrogenicity testing.

## Technological Characteristics

Constructed of Titanium-6 Aluminum-4 Vanadium ELI alloy (ASTM F136-13). Fully threaded configuration; diameters 3.5 mm, 3.75 mm, 4.0 mm; lengths 13 mm to 25 mm. Supplied sterile for single use. Mechanical fixation device.

## Predicate Devices

- NuVasive Triad Facet Screw System ([K020411](/device/K020411.md))
- Venus Facet Screw System ([K120340](/device/K120340.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 11, 2019

Providence Medical Technology, Inc. Janie Mandrusov, Ph.D. Vice President, Clinical and Regulatory Affairs 3875 Hopyard Road, Suite 300 Pleasanton, California 94588

Re: K183589

Trade/Device Name: PMT Facet Screw Regulatory Class: Unclassified Product Code: MRW Dated: September 20, 2019 Received: September 20, 2019

Dear Dr. Mandrusov:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)

{1}------------------------------------------------

regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ronald P. Jean, Ph.D. Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K183589

Device Name PMT Facet Screw

#### Indications for Use (Describe)

The PMT Facet Screw is intended to spine as an aid to fusion through bilateral immobilization of the facet joints. The screws are inserted posteriorly through the superior side of the facet joint, and into the pedicle. The PMT Facet Screw is indicated for bilateral facet fixation, with or without bone graft, at single or multiple levels, from C2 to S1. The PMT Facet Screw is indicated for treatment of any or all of the following:

1. Pseudoarthrosis and failed previous fusion

- 2. Spondylolisthesis
- 3. Spondylolysis

4. Degenerative disc disease (DDD) as defined by neck and/or back pain of discogenic origin as confirmed by radiographic studies

- 5. Degeneration of the facets with instability; and
6. Fracture

The PMT Facet Screw is intended for conventional or tissue-sparing surgical placement.

| Type of Use (Select one or both, as applicable)                                      |                                                                      |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over The Counter Use (21 CFR 801 Subpart C) |

|XJ Prescription Use (Part 21 CFR

Cunter Use (21 CFR 801 Subbart Ci

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY

#### Providence Medical Technology, Inc.'s PMT Facet Screw

| Date Prepared:          | September 20, 2019                                                                         |
|-------------------------|--------------------------------------------------------------------------------------------|
| Company:                | Providence Medical Technology, Inc.<br>3875 Hopyard Rd., Suite 300<br>Pleasanton, CA 94588 |
| Contact Person:         | Janie Mandrusov<br>Phone:<br>415.301.3128<br>Facsimile:<br>415.923.9377                    |
| Trade Name:             | PMT Facet Screw                                                                            |
| Common Name:            | Facet Screw                                                                                |
| Classification<br>Name: | System, Facet Screw Spinal Device                                                          |
| Product Code,<br>Class: | MRW, Unclassified                                                                          |
| Predicate Devices:      |                                                                                            |

Primary Predicate: NuVasive Triad Facet Screw System (K020411, cleared 3/12/2012)

Additional Predicate: Venus Facet Screw System (K120340, cleared 10/19/12)

#### Indications for Use:

The PMT Facet Screw is intended to stabilize the spine as an aid to fusion through bilateral immobilization of the facet joints. The screws are inserted posteriorly through the superior side of the facet, across the facet joint, and into the pedicle. The PMT Facet Screw is indicated for bilateral facet fixation, with or without bone graft, at single or multiple levels, from C2 to S1. The PMT Facet Screw is indicated for treatment of any or all of the following:

- 1. Pseudoarthrosis and failed previous fusion
- 2. Spondylolisthesis
- 3. Spondylolysis

{4}------------------------------------------------

- 4. Degenerative disc disease (DDD) as defined by neck and/or back pain of discogenic origin as confirmed by radiographic studies
- న్. Degeneration of the facets with instability; and
- 6. Fracture

The PMT Facet Screw is intended for conventional or tissue-sparing surgical placement.

## Technological Characteristics:

The PMT Facet Screw is available in fully threaded configurations with diameters of 3.5 mm, 3.75 mm and 4.0 mm and lengths of 13 mm to 25 mm (in 2 mm increments). The PMT Facet Screw is constructed of Titanium-6 Aluminum- 4 Vanadium ELI alloy (conforming to ASTM F136-13) and supplied sterile for single use only.

## Performance Testing

The following testing was performed using the subject device:

## Static and Dynamic Cantilever Bending

- ASTM F2193-14 Standard Specifications and Test Methods for Components . Used in the Surgical Fixation of the Spinal Skeletal System
#### Static Torsion Testing

- ASTM F543-17 Standard Specification and Test Methods for Metallic Medical . Bone Screws
## Static Pull-Out Testing

- · ASTM F543-17 Standard Specification and Test Methods for Metallic Medical Bone Screws
#### Drive Torque

- · ASTM F543-17 Standard Specification and Test Methods for Metallic Medical Bone Screws
In addition, the following testing was performed:

## Pyrogenicity Testing

- · Pyrogenicity Testing was conducted in compliance with the FDA Guidance Documents listed below:
	- "Submission and Review of Sterility Information in Premarket o Notification (510(k)) Submissions for Devices Labeled as Sterile: Guidance for Industry and Food and Drug Administration Staff"

{5}------------------------------------------------

## o "Pyrogen and Endotoxins Testing: Questions and Answers"

In all instances, the PMT Facet Screw functioned as intended and met all pre-determined acceptance criteria and do not raise new issues of safety or effectiveness; thus demonstrating that the PMT Screw is substantially equivalent to the legally marketed predicate device.

No clinical testing was deemed necessary to support substantial equivalence.

## Substantial Equivalence

The PMT Facet Screw has the same indication for use, technological characteristics, and principles of operation as the predicate device. The successful completion of the performance testing further supports the subject PMT Facet Screw's substantial equivalence to the predicate device. No issues of safety or effectiveness are raised by the PMT Facet Screw.

---

**Source:** [https://fda.innolitics.com/device/K183589](https://fda.innolitics.com/device/K183589)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
