Luna 3D GEN2 Interbody Fusion System

K183560 · Benvenue Medical, Inc. · MAX · May 23, 2019 · Orthopedic

Device Facts

Record IDK183560
Device NameLuna 3D GEN2 Interbody Fusion System
ApplicantBenvenue Medical, Inc.
Product CodeMAX · Orthopedic
Decision DateMay 23, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Luna 3D GEN2 Interbody Fusion System consists of the Luna 3D Implant and associated accessories. This system is indicated for spinal fusion procedures in skeletally mature patients with symptomatic degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to grade I spondylolisthesis or retrolisthesis at the involved level(s). The Luna 3D GEN2 Interbody Fusion System is to be used with autogenous bone graft and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft to facilitate fusion. Patients receiving the device should have had at least six months of nonoperative treatment proor to receiving the Luna 3D GEN2 Implant. The Luna 3D GEN2 Interbody Fusion System is to be used with supplemental fixation.

Device Story

The Luna 3D GEN2 Interbody Fusion System is an intervertebral body fusion device used in lumbar spinal fusion procedures. The implant is inserted via a straight cannula and, upon placement into the disc space, transitions into a closed, fixed, circular shape. The device is then expanded to its final height to create a bone graft pocket, with surface teeth engaging the vertebral endplates for stability. It is intended for use by surgeons in a clinical setting to treat degenerative disc disease. The device facilitates spinal fusion by providing structural support and a space for bone graft, potentially reducing discogenic back pain and stabilizing the spinal segment.

Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F2077-14 and F2267-04 (static/dynamic compression, torsion, shear, expulsion, subsidence). Corrosion testing per ASTM F2129 and nickel leaching analysis via ICP/MS were conducted. Surgical technique validation was performed using cadaver specimens.

Technological Characteristics

Materials: PEEK (ASTM F2026), nitinol wire (ASTM F2063), and tantalum rod (ASTM F560). The device features a flexible, expandable geometry that transitions from a straight insertion profile to a circular, height-expanded shape. It is provided sterile in a single-use insertion tool.

Indications for Use

Indicated for spinal fusion in skeletally mature patients with symptomatic degenerative disc disease (DDD) at one or two contiguous levels from L2-S1, including patients with up to grade I spondylolisthesis or retrolisthesis. Requires at least six months of prior nonoperative treatment. Must be used with bone graft and supplemental fixation.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 23, 2019 Benvenue Medical, Inc. % Justin Eggleton Senior Director, Spine Regulatory Affairs Musculoskeletal Clinical Regulatory Advisers LLC 1050 K Street NW, Suite 1000 Washington, District of Columbia 20001 Re: K183560 Trade/Device Name: Luna 3D GEN2 Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: April 25, 2019 Received: April 26, 2019 Dear Justin Eggleton: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for CAPT Raquel Peat, PhD, MPH, USPHS Director OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K183560 #### Device Name Luna 3D GEN2 Interbody Fusion System ### Indications for Use (Describe) The Luna 3D GEN2 Interbody Fusion System consists of the Luna 3D Implant and associated accessories. This system is indicated for spinal fusion procedures in skeletally mature patients with symptomatic degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to grade I spondylolisthesis or retrolisthesis at the involved level(s). The Luna 3D GEN2 Interbody Fusion System is to be used with autogenous bone graft and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft to facilitate fusion. Patients receiving the device should have had at least six months of nonoperative treatment proor to receiving the Luna 3D GEN2 Implant. The Luna 3D GEN2 Interbody Fusion System is to be used with supplemental fixation. Type of Use (*Select one or both, as applicable*) | <div> <span style="font-size:14px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size:14px">☐</span> Over-The-Counter Use </div> | |----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------| |----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------| Jse (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 4. 510(k) Summary | Device Trade Name: | Luna 3D GEN2 Interbody Fusion System | |--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer: | Benvenue Medical, Inc.<br>4590 Patrick Henry Dr.,<br>Santa Clara, California 95054 USA | | Contact: | Laurent Schaller<br>CTO and Founder<br>Benvenue Medical, Inc.<br>4590 Patrick Henry Dr.,<br>Santa Clara, California 95054 USA | | Prepared by: | Mr. Justin Eggleton<br>Senior Director, Spine Regulatory Affairs<br>Musculoskeletal Clinical Regulatory Advisers, LLC<br>1050 K Street NW, Suite 1000<br>Washington, DC 20001<br>jeggleton@mcra.com | | Date Prepared: | December 20, 2018 | | Classifications: | 21 CFR §888.3080, Intervertebral body fusion device | | Class: | II | | Product Codes: | MAX | # Primary Predicate Device: Benvenue Medical Luna 3D Interbody Fusion System (K162431) # Indications for Use: The Luna 3D GEN2 Interbody Fusion System consists of the Luna 3D Implant and associated accessories. This system is indicated for spinal fusion procedures in skeletally mature patients with symptomatic degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to grade I spondylolisthesis or retrolisthesis at the involved level(s). The Luna 3D GEN2 Interbody Fusion System is to be used with autogenous bone graft and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft to facilitate fusion. Patients receiving the device should have had at least six months of nonoperative treatment prior to receiving the Luna 3D GEN2 Implant. The Luna 3D GEN2 Interbody Fusion System is to be used with supplemental fixation. {5}------------------------------------------------ # Device Description: The Benvenue Luna 3D GEN2 Interbody Fusion System consists of the Luna 3D GEN2 Implant and associated accessories set of disposable and re-usable accessories for use in lumbar fusion procedures to treat degenerative disc disease. The proposed indications for use for the Luna 3D GEN2 Interbody Fusion System are identical to the primary predicate device. The Luna 3D GEN2 Implant is provided pre-loaded and sterile within a single-use Insertion Tool. The Luna 3D GEN2 Implant is available in heights ranging from 10mm to 14mm with 2mm increments and a 6° lordotic angle and from 12mm with 2mm increments and a 12° lordotic angle. A series of vertically oriented slots allows the device to flex and enables it to be inserted from a straight cannula and then attain a closed, fixed, and circular shape upon being placed into the disc space. Once the implant is in the desired position, the device is expanded into its ultimate height and forming a bone graft pocket. Teeth on the outer surfaces of the top and bottom components engage the implant into the adjacent endplates. The Luna 3D GEN2Implant that is the subject of this 510(k) is manufactured from polyetheretherketone (VESTAKEEP i4R, conforming to ASTM F2026), nitinol wire (superelastic alloy SE508, conforming to ASTM F2063) and tantalum rod conforming to ASTM F560. This 510(k) was submitted in support of the Generation 2 device. # Performance Testing: The worst-case devices were subjected to mechanical testing. Testing included static compression, static compression-shear, static torsion, static modified compression-shear, dynamic compression, dynamic compression-shear, dynamic torsion, expulsion, and subsidence per ASTM F2077-14 and F2267-04. In addition, corrosion testing was completed according to ASTM F2129 and nickel leaching quantitative ICP/MS analysis was performed. A surgical technique validation in cadaver specimens was also performed. The results demonstrate that the Luna 3D GEN2 Interbody Fusion System device is substantially equivalent to the predicate devices. # Substantial Equivalence Summary: Comparative information presented in the 510(k) supports the substantial equivalence of the Luna 3D GEN2 Interbody Fusion System to the primary predicate device. Comparisons were designed to show the indications, intended use, design, and performance are equivalent between the Benvenue Luna 3D GEN2 Interbody Fusion System and primary predicate device and additional predicate devices. # Conclusion: The information and performance data demonstrate that the device is as safe, as effective, and performs as well as or better than the primary predicate device. This 510(k) was submitted on behalf of the Luna 3D GEN2 Interbody Fusion System in support of the next generation device design. Substantial equivalence was determined in response to sufficient comparisons to the primary predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...