K183545 · The Anspach Effort, Inc. · HBE · Jul 19, 2019 · Neurology
Device Facts
Record ID
K183545
Device Name
Anspach Helix Dissection Tools
Applicant
The Anspach Effort, Inc.
Product Code
HBE · Neurology
Decision Date
Jul 19, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Anspach Helix™ Dissection Tools are intended for cutting and shaping bone including the spine and cranium.
Device Story
Anspach Helix™ Dissection Tools are one-piece M2 tool steel bone-cutting burs; designed for attachment to pneumatic or electric surgical drill systems via a patented Diamond Flat locking mechanism. Used by surgeons in clinical settings (e.g., neurosurgery, orthopedics) for cutting and shaping bone during cranial or spinal procedures. The device functions as a mechanical cutting accessory; the 2-fluted ball design reduces chatter and vibration compared to predicate multi-flute burs. Output is the physical removal or shaping of bone tissue. Benefits include improved surgical precision and reduced vibration during bone dissection.
Clinical Evidence
Bench testing only. Verification activities included tolerance analysis for mating clearance/anti-rotation, shaft compatibility, and performance testing using Bi-resin simulated bone. Results demonstrated the 2-flute design met acceptance criteria for cutting/shaping and exhibited statistically lower chatter/vibration compared to predicate devices. Biocompatibility evaluation performed per ISO 10993-1. No clinical or animal data provided.
Indicated for cutting and shaping bone, including spine and cranium, in patients requiring surgical bone removal or modification.
Regulatory Classification
Identification
Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
The Anspach Effort, Inc. Tamara West Regulatory Affairs Manager 4500 Riverside Drive Palm Beach Gardens, Florida 33410
#### Re: K183545
Trade/Device Name: Anspach Helix Dissection Tools Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, And Their Accessories Regulatory Class: Class II Product Code: HBE Dated: June 14, 2019 Received: June 17, 2019
#### Dear Tamara West:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Matthew Krueger Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
Device Name
Anspach Helix™ Dissection Tools
Indications for Use (Describe) Anspach Helix™ Dissection Tools are intended for cutting and shaping bone including the spine and cranium.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "ANSPACH" in large, bold, blue letters. To the left of the word is a blue square with a white bird inside. To the right of the word is a small, superscripted "®" symbol, indicating that the word is a registered trademark. The word is the main focus of the image and is easily readable.
## 510(k) SUMMARY
### I. SUBMITTER
The Anspach Effort, Inc. 4500 Riverside Drive Palm Beach Gardens, Florida 33410 USA
Registration Number: 1045834
Contact: Tamara West Phone: (561) 494-3712 Fax: (561) 624-8383 Email: twest1@its.jnj.com
Alternate Contact: Jeannette G. Dailey Phone: (561) 494-3710 Fax: (561) 624-8383 Email: jdailey5@its.jnj.com
Date Prepared: June 14, 2019
### II. DEVICE
Trade Name: Anspach Helix™ Dissection Tools
Common Name: Dissection Tools
Classification Information: 21 CFR 882.4310 Drills, burs, trephines, and accessories (simple, powered)
Regulatory Class: II
Product Code: HBE
Classification panel: Neurology
### III. PREDICATE DEVICE
K113476 Dissection Tools Owner: The Anspach Effort, Inc.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "ANSPACH" in large, bold, blue letters. To the left of the word is a blue square containing a white seagull. A small "R" in a circle is located to the upper right of the word.
## IV. DEVICE DESCRIPTION
The Anspach Helix™ Dissection Tools (otherwise known as burs) are part of a drill system that consists of a control console, a handpiece (motor), a control switch and various attachments and accessories. The Anspach Helix™ Dissection Tools are onepiece metal bone cutting and shaping devices that attach to the handpiece/attachment of a pneumatic or electric powered drilling system.
The Anspach Helix™ Dissection Tools are manufactured from M2 tool steel and feature a 2-fluted ball sharp cutting edge at the distal end. The proximal end, which secures the dissection tool to the handpiece of the power system, incorporates a patented Diamond Flat configuration locking mechanism.
The Anspach Helix™ Dissection Tools are provided sterile.
## V. INDICATIONS FOR USE
Anspach Helix™ Dissection Tools are intended for cutting and shaping bone including the spine and cranium.
| | Anspach Helix™ Dissection Tools | Dissection Tools |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| 510(k) Number | K183545 | K113476 |
| Indications for Use | Anspach Helix™ Dissection Tools are used for cutting and shaping bone including the spine and cranium. | Dissection tools are intended for cutting shaping bone including the spine and cranium. |
| Product Code | HBE | HBE |
| Class | 2 | 2 |
| Sterilization | Gamma irradiated | Gamma irradiated |
| Sterilization<br>Assurance Level<br>(ASL) | 10-6 | 10-6 |
| TECHNOLOGICAL CHARACTERISTICS | | |
| Head Styles | Fluted Ball | Fluted Ball, Fluted Matchstick,<br>Diamond Ball |
| Flutes | 2 Flutes | 2 Flutes (Matchstick)<br>4 – 12 Flutes (Ball) |
| Head Diameter | 2mm – 7mm | Fluted Ball:<br>0.5mm – 9mm |
| Material | M2 Tool Steel | M2 Tool Steel, Stainless Steel,<br>Carbide, Diamond coated |
| Locking mechanism | Diamond Flat Configuration | Diamond Flat Configuration |
| Labeling | Sterile, single-use | Sterile, single-use |
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Anspach Companies 4500 Riverside Drive, Palm Beach Gardens, Florida 33
{5}------------------------------------------------
| Packaging<br>Configuration | PETG tray in a sealed Tyvek pouch<br>placed into a carton | PETG tray in a sealed Tyvek pouch |
|----------------------------|-----------------------------------------------------------|-----------------------------------|
| Shelf life | 5 years | 5 years |
The predicate dissection tools (K113476) are part of an electric/pneumatic drill system and are available in many different styles and dimensions, including a fluted ball. A minor design modification was made to the head of the existing fluted ball style by offering a 2fluted style as compared to 4 or more flutes with the predicate device. The 2-fluted style was developed to provide a dissection tool that produces less chatter and vibration while cutting than the predicate dissection tools (K113476).
The proposed Anspach Helix™ Dissection Tools remain identical to the predicate dissection tools (K113476) in terms of operating principles, where it attaches to the handpiece/attachment of either an electric or pneumatic drill utilizing the diamond flat locking mechanism. The indications for use, material and manufacturing remain identical to the predicate dissection tools (K113476).
The technological characteristics of the proposed Anspach Helix™ dissection tools are similar to those of the predicate (K113476) device in the following ways:
- Both dissection tools use a patented locking mechanism on the distal end to secure to the handpiece
- · Both dissection tools are compatible with the same electric and pneumatic drill systems, including attachments
- Both dissection tools have a straight shaft and fluted ball style head, where the proposed dissection tools have 2 flutes opposed to 4 or more flutes of the predicate dissection tools
- · Both dissection tools are provided sterile
As with the predicate dissection tools (K113476), the Anspach Helix™ Dissection Tools have no impact on the safety and effectiveness of the device or the ability to perform as intended.
# VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility Testing:
The Anspach Helix™ dissection tools are classified as external communicating devices: tissue/bone/dentin with limited patient contact (< 24 hours).
A biocompatibility evaluation was conducted for the Anspach Helix™ Dissection Tools per the FDA guidance document Use of International Standard ISO 10993-1. "Biological
> Anspach Companies 4500 Riverside Drive, Palm Beach Gardens, Florida 33410
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "ANSPACH" in large, bold, blue letters. To the left of the word is a blue square containing a white bird in flight. A small circled "R" symbol is located in the upper right corner of the word.
evaluation of medical devices – Part 1: Evaluation and testing within a risk management process".
The Anspach Helix™ Dissection Tools are manufactured and packaged in the same manner with identical material as the predicate device (K113476). No new or additional processing steps were introduced; therefore, biocompatibility requirements have been met.
#### Design Control:
The following verification activities were performed to demonstrate that the device meets the performance requirements under its indications for use conditions for determination of substantial equivalence.
| Test | Test Method Summary | Results |
|----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Verification Analysis - Mating<br>Clearance & Anti-Rotation<br>Interference | Conduct a tolerance analysis to<br>determine that there is mating<br>clearance and anti-rotation<br>interference when engaged with<br>drive spindle of EG1, XMAX, &<br>EMAX2PLUS handpieces | The analysis shows that the<br>design output meets the design<br>input; the proposed Helix<br>dissection tools drive feature<br>meets the acceptance criteria of<br>mating clearance and anti-<br>rotation interference when<br>engaged with drive spindles of<br>EG1, XMAX, & EMAX2PLUS<br>handpieces. |
| Helix Dissection Tool and<br>Attachment Compatibility - Shaft<br>Diameter/Shaft Length | Conduct a tolerance analysis to<br>determine the minimum and<br>maximum clearance between the<br>shaft diameter of the dissection<br>tool and inner race bearing<br>diameter of the attachment. | Analysis was performed by<br>evaluating all Helix dissection<br>tools and attachment (predicate)<br>combinations to confirm<br>compatibility.<br>Tolerance analysis reflects 0.0001<br>to 0.0007 inches of clearance<br>between dissection tools and the<br>attachment which is in the<br>required range of scope with a<br>3/32-inch shaft diameter and the<br>inner race bearing diameter of<br>compatible attachments as listed<br>in the compatibility table.<br>Tolerance analysis reflects that<br>the minimum length in the lock<br>position exists, so the head will<br>not interfere with the attachment.<br>Tolerance analysis reflects that<br>the tapered section of the<br>dissection tool will not interfere<br>with the distal bearing inner<br>diameter of the attachment. |
Anspach Companies 4500 Riverside Drive, Palm Beach Gardens, Florida 33410
{7}------------------------------------------------
| Helix Ball Bur Chatter/Cut and<br>Shape Verification Report | | |
|-------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The % chatter for each group of<br>same-size Helix ball burs must<br>be less than the chatter of the<br>corresponding group of same-<br>size predicate G1<br>burs, as determined by a<br>hypothesis test with a<br>significance level (alpha) = 0.10. | All samples met the acceptance<br>criteria and exhibited statistically<br>lower% chatter than the<br>corresponding same size<br>predicate devices with the<br>exception of the S-25SB-HX-G1-1.<br>For this size, while lower, the<br>results were not statistically<br>significant as compared to the<br>corresponding same size<br>predicate. |
| | Each Helix ball bur used to cut<br>into the Bi-resin (simulated<br>bone) samples and then shall be<br>evaluated to confirm that they<br>cut and shape bone the same as<br>the predicate. | All samples met the acceptance<br>criteria to cut and shape bone,<br>the same as the predicate. |
| | Each Helix ball bur used to cut<br>into the Bi-resin (simulated<br>bone) samples and then shall be<br>visually inspected for cutter head<br>fracture with the acceptance<br>criteria of the bur head intact on<br>the cutter and not broken, the<br>same as the predicate. | All samples met the acceptance<br>criteria for the head of the cutter<br>to remain intact, the same as<br>the predicate. |
The above verification activities demonstrate the Anspach Helix™ Dissection Tools perform as intended in cutting and shaping bone in the same manner as the predicate (K113476) and are compatible with existing predicate devices (K113746). Verification activities also demonstrate that select Helix designs exhibited statistically significantly lower chatter and vibration while cutting than the corresponding same size burs of the predicate devices (K113476). Based on the results of the verification activities, the proposed Anspach Helix™ Dissection Tools are determined to be substantially equivalent to the predicate devices (K113476).
No animal or clinical testing was necessary for a determination of substantial equivalence.
# VII. CONCLUSIONS
The Anspach Helix™ Dissection Tools described in this submission have the same intended use, general design, material, technological characteristics, and principles of operation as the predicate device (K113476) and do not raise any new issues of safety or efficacy. In addition, the select 2-flute designs that are part of the marketing claims within this submission also do not raise any new issues of safety or efficacy and were found to be substantially equivalent to the predicate devices (K113476). Therefore, the proposed Anspach Helix™ Dissection Tools are found to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.