K183543 · Kci USA, Inc. · OMP · Mar 20, 2019 · General, Plastic Surgery
Device Facts
Record ID
K183543
Device Name
ACTIV.A.C. Therapy Unit
Applicant
Kci USA, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
Mar 20, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ACTV.A.C.™ Neative Pressure Wound Therapy System is an integrated wound management system for use in acute, extended and home care settings. When used on open wounds, it is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts. When used on closed surgical incisions, it is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closed environment and removing exudates via the application of negative pressure wound therapy.
Device Story
Portable, battery-powered, reusable NPWT unit; provides continuous or intermittent negative pressure (-25 to -200 mmHg) to wound beds. System includes disposable canister, wound interface dressing, semi-occlusive drape, sensing pad, and lumen. Used in home, acute, or extended care settings by clinicians or patients. Software monitors target pressure; triggers alarms for blockages, full/missing canisters, inactive therapy, low battery, or leaks. Features include Seal Check for leak identification, Therapy Settings Guide, screen guard, USB therapy history export, and log tool. Device promotes healing by reducing edema, promoting granulation/perfusion, and removing exudate/infectious material. Clinicians use output to manage wound environment and track therapy adherence.
Clinical Evidence
Bench testing only. Testing included conformance to IEC 60601-1 standards, software assessment per FDA guidance, negative pressure maintenance system testing, and extreme environmental storage condition testing. No clinical data presented.
Technological Characteristics
Portable, battery-powered, software-controlled negative pressure pump. Uses brushless DC double diaphragm pump. Pressure range: -25 to -200 mmHg in 25 mmHg increments. Connectivity: USB data port for history export. Housing includes gaskets for fluid ingress protection. Software-controlled monitoring and alarm system.
Indications for Use
Indicated for patients with open wounds (chronic, acute, traumatic, subacute, dehisced, partial-thickness burns, ulcers, flaps, grafts) or closed surgical incisions that continue to drain following sutured or stapled closure.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The Department of Health and Human Services logo is a stylized representation of a human figure, while the FDA acronym and full name are written in blue.
September 12, 2019
KCI USA, Inc. Teri Feeley Sr. Regulatory Specialist 6203 Farinon Dr. San Antonio, Texas 78249
Re: K183543
Trade/Device Name: ACTIV.A.C. Therapy Unit Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: December 19, 2018 Received: December 20, 2018
Dear Teri Feeley:
This letter corrects our substantially equivalent letter of March 20, 2019.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cynthia Chang -S
Cynthia J. Chang, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
510(k) Number (if known) K183543
Device Name
#### ACTIV.A.C. ™ Therapy Unit
Indications for Use (Describe)
The ACTV.A.C.™ Neative Pressure Wound Therapy System is an integrated wound management system for use in acute, extended and home care settings.
When used on open wounds, it is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts.
When used on closed surgical incisions, it is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closed environment and removing exudates via the application of negative pressure wound therapy.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number
{3}------------------------------------------------
# 510(k) SUMMARY
# ACTIV.A.C.™ Therapy Unit
KCI USA, Inc. 6203 Farinon San Antonio, TX 78249
Phone: 210-515-4396 Facsimile: 210-255-6727 Contact Person: Teri Feeley Date Prepared: 20 Mar 2019
Name of Device: ACTIV.A.C.™ Therapy Unit
Common or Usual Name: Neqative Pressure Wound Therapy Unit
Classification Name: Negative Pressure Wound Therapy Powered Suction Pump (and components)
510(k) Number: K183543
Regulatory Number: 21 CFR 878.4780
Regulatory Class: Il
Product Code: OMP
Predicate Device
KCI USA, Inc., ACTIV.A.C. ™ Therapy Unit (K063692, K120033)
#### Device Description
The ACTIV.A.C.™ Therapy Unit is a component of the ACTIV.A.C.™ Negative Pressure Wound Therapy (NPWT) System. The ACTIV.A.C.™ Negative Pressure Wound Therapy (NPWT) system consists of:
- ACTIV.A.C. TM Therapy Unit (the subject of this submission) ●
- disposable canister which collects wound exudate
- a wound interface dressing
- . semi-occlusive wound drape
- sensing pad and lumen
The ACTIV.A.C. ™ Therapy Unit is a portable, battery-powered, reusable, software-controlled therapy unit that can provide continuous or intermittent applications of negative pressure to the wound bed in the selectable range of -25mmHg to -200mmHg. The ACTV.A.C.™ Therapy Unit is designed for the application of Negative Pressure Wound Therapy in the home, acute or extended care setting. The open cells of the dressing, to which the therapy unit is connected via pad and lumen, enables distribution of the negative pressure across the surface of the wound bed, while the
{4}------------------------------------------------
tubing transfers accumulated fluids to the canister. The software monitors and maintains target pressure and alarms as needed to help assure target pressure is maintained and constant therapy is delivered. The safety features of the system include additional alarms, such as those that signal for tubing blockages, a full or missing collection canister, inactive therapy, low battery, and leaks in the seal of the dressing.
Optional ancillary features include: Seal Check™ for identifying dressing leaks, a Therapy Settings Guide that contains preset therapy settings based on wound type, a screen guard feature that prevents unintentional screen changes, an exportable Therapy History Report via USB data port, and a Log Tool for recording canister changes, dressing changes and dressing pieces used.
#### Intended Use / Indications for Use
The ACTIV.A.C. ™ Negative Pressure Wound Therapy System is an integrated wound management system for use in acute, extended and home care settings.
When used on open wounds, it is intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts.
When used on closed surgical incisions, it is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy.
{5}------------------------------------------------
### Summary of Technological Characteristics
Controlled delivery of negative pressure to the wound site is the technological principle for both the subject and predicate devices. The software-controlled therapy unit applies negative pressure to the wound bed. The open cells of the dressing, to which the therapy unit is connected via pad and lumen, enables distribution of the negative pressure across the surface of the wound bed, while the tubing transfers accumulated fluids to the canister. At a high level, the subject and predicate devices are based on the following same technological elements:
- . Intended use
- . Indicated wound types
- ACTIV.A.C. Therapy System dressing components and canister .
- . Software controlled negative pressure pump and negative pressure specifications
- . Use environment is acute, extended and home care settings
The following minor design differences exist between the subject and the predicate device as cleared under K063692:
Changes associated with this submission:
- Additional and modified gaskets on the ACTIV.A.C.™ Therapy Unit housing for . improved fluid ingress protection for added safety.
- . Expanded storage temperature specifications to demonstrate the device functions as intended if exposed to more rigorous storage environments.
In addition, the submission summarized minor device modifications implemented since the time of predicate device clearance that could not significantly affect device safety or effectiveness and thereby did not require 510(k) submissions. These include routine component updates to address component obsolescence and device manufacturability, minor software updates, upgraded power supply and power connector, and canister modifications.
A table comparing the key features of the subject and predicate devices is provided below.
Summary of the technological characteristics of the device compared to the predicate device [21 CFR 807.92(a)(6)]
{6}------------------------------------------------
| Characteristic | Subject Device:<br>ACTIV.A.C. ™ Therapy<br>Unit | Predicate Device:<br>ACTIV.A.C. ™ Therapy Unit<br>(K063692) | Predicate Device:<br>ACTIV.A.C. ™ Therapy Unit<br>(K120033) |
|------------------------------------------------------|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Identical to K120033 | To deliver and maintain<br>negative pressure wound<br>therapy to the wound site. | To deliver and maintain<br>negative pressure wound<br>therapy to the open wound<br>site and closed surgical<br>incision. |
| Indicated Wound Types | Identical to K120033 | Chronic Acute Traumatic Subacute Dehisced wounds Partial-thickness burns Ulcers (such as diabetic, pressure or venous<br>insufficiency) Flaps and grafts | Chronic Acute Traumatic Subacute Dehisced wounds Partial-thickness burns Ulcers (such as diabetic, pressure or venous<br>insufficiency) Flaps and grafts Surgically closed incisions |
| Negative Pressure Wound<br>Therapy System Components | Identical | ACTIV.A.C. ™ Therapy Unit Disposable canister which collects wound exudate A wound interface dressing Semi-occlusive wound drape Sensing pad and lumen | ACTIV.A.C. ™ Therapy Unit Disposable canister which collects wound exudate A wound interface dressing Semi-occlusive wound drape Sensing pad and lumen |
| Use environment/Care<br>Setting | Identical | Acute, extended and home<br>care settings | Acute, extended and home<br>care settings |
| Negative Pressure Unit Pump | Identical | Brushless, DC powered,<br>double diaphragm pump | Brushless, DC powered,<br>double diaphragm pump |
| Negative pressure options | Identical | 25-200 mmHg, 25mmHg<br>increments | 25-200 mmHg, 25mmHg<br>increments |
# Performance Data
Summary of non-clinical tests conducted for determination of substantial equivalence:
- . Conformance to the most current General Requirements for Basic Safety and Essential Performance 60601-1 standards.
- Software has been assessed in accordance with FDA Guidance for the Content of . Premarket Submission for Software Contained in Medical Devices (May 11, 2005).
- ACTIV.A.C. ™ Negative Pressure Maintenance System Test demonstrates the ● ACTIV.A.C.™ Therapy Unit maintains negative pressure within specifications and manages fluid exudate without unexpected alarms.
{7}------------------------------------------------
- . ACTIV.A.C. ™ Therapy Unit Extreme Environmental Storage Conditions test demonstrates the therapy unit functions as intended when exposed to wider temperature specifications.
In all instances, the ACTIV.A.C.™ Therapy Unit functioned as intended and all test results observed were as expected.
Human factors engineering testing was not required since the subject device has the same user interface and use environment as the predicate.
#### Conclusions
The ACTIV.A.C.™ Therapy Unit is as safe and effective as the predicate ACTIV.A.C.™ Therapy Unit (K063692). The subject device's fundamental technology and principles of operation for the ACTIV.A.C.™ Therapy Unit are unchanged compared to the predicate device as cleared under K063692. The subject device's Intended Use remains unchanged from the predicate device as cleared under K063692. The Indications for Use for the ACTIV.A.C. was recently modified to include closed incisions under K120033, which was cleared in 2012. Therefore, the Intended Use/Indications for Use for the subject device remain unchanged
The minor technological differences between the ACTIV.A.C.™ Therapy Unit and its predicate device could not significantly affect the safety or effectiveness of the device, nor did they represent a major change in intended use safety. The performance data demonstrates that the ACTIV.A.C.™ Therapy Unit is as safe and effective as the predicate. Thus, the ACTIV.A.C.™ Therapy Unit is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.