← Product Code [NJM](/productcode/NJM) · K183542

# Signature Orthodontic System (K183542)

_Signature Orthodontics · NJM · Jan 2, 2019 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K183542

## Device Facts

- **Applicant:** Signature Orthodontics
- **Product Code:** [NJM](/productcode/NJM.md)
- **Decision Date:** Jan 2, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.5470
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The Signature Orthodontic System is a treatment planning software and orthodontic appliance system used to correct malocclusions in orthodontic patients using patient-matched orthodontic appliances.

## Device Story

System comprises treatment planning software (TPS), patient-specific ceramic brackets, bracket placement jigs, and arch wire templates. Orthodontists use TPS to view, measure, and modify digital impressions (STL files) to generate prescriptions. Signature Orthodontics uses proprietary additive manufacturing to produce patient-specific brackets and jigs based on TPS designs. Orthodontists bond brackets to teeth using placement jigs and ligate wires to facilitate tooth movement. System does not include arch wires, ligatures, or adhesives. Used in clinical settings by orthodontists to plan and execute orthodontic treatment, potentially improving tooth alignment accuracy through patient-matched components.

## Clinical Evidence

No clinical performance testing was conducted. Substantial equivalence is supported by non-clinical software validation testing comparing the new TPS 3.1 against the predicate TPS (Meshmixer 3.4), demonstrating equivalent performance in diagnosis, viewing, and data handling workflows.

## Technological Characteristics

System includes ceramic brackets and additive-manufactured placement jigs. Software component (TPS 3.1) is a proprietary application for 3D visualization and treatment planning. Connectivity involves secure web-based data transfer of STL files. No specific material standards or energy sources are cited for the hardware components.

## Regulatory Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

## Predicate Devices

- Signature Orthodontics' Signature Orthodontic System ([K181271](/device/K181271.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Signature Orthodontics % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, Nw Buffalo, Minnesota 55313

Re: K183542

Trade/Device Name: Signature Orthodontic System Regulation Number: 21 CFR Orthodontic Plastic Bracket Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NJM Dated: December 17, 2018 Received: December 20, 2018

Dear Mark Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

January 2, 2019

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Marv S. Mary S. Runner -S3 Runner -S3 Date: 2019.01.02

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

Device Name Signature Orthodontic System

Indications for Use (Describe)

The Signature Orthodontic System is a treatment planning software and orthodonic appliance system used to correct malocclusions in orthodontic patients using patient-matched orthodontic appliances.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## TAB 6

# 510K SUMMARY

## 510(k) SUMMARY

## Signature Orthodontics' Signature Orthodontic System

### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Signature Orthodontics 1035 Cambridge Street Cambridge, MA 02141

Phone: 540-229-1236 Facsimile: N/A Email: alfred@soiboston.com

Contact Person: Alfred Charles Griffin III

Date Prepared: November 13, 2018

#### Name of Device and Name/Address of Sponsor

Signature Orthodontic System

Signature Orthodontics 1035 Cambridge Street Cambridge, MA 02141

Trade/Proprietary Name of Device: Signature Orthodontic System

Common or Usual Name: Orthodontic Ceramic Bracket and Accessory

Classification Name: Orthodontic Ceramic Bracket, 21CFR§872.5470

Regulatory Class: II

Product Code: NJM

Predicate Devices:

Primary Predicate: Signature Orthodontics' Signature Orthodontic System (K181271)

{4}------------------------------------------------

#### Device Description and Justification

The Signature Orthodontic System (SO System) is a treatment planning software (TPS) and orthodontic appliance system used to correct malocclusions in orthodontic patientmatched orthodontic appliances. The SO System consists of patient-specific ceramic brackets, patient-specific bracket placement jigs, arch wire templates, and a TPS for viewing, and modifying cases. Signature Orthodontics' (SO) operators and the orthodontists use the TPS to generate a prescription of their choosing. SO then manufactures the patient-specific brackets and placement jigs using proprietary additive manufacturing techniques. The orthodontist then bonds the brackets to the teeth using the optional placement jig and ligates wires to enable tooth movement. The SO System does not contain commercially-available or patient-specific shaped arch wires, ligatures, or adhesives that affixes the brackets to the teeth.

The change is to replace Meshmixer 3.4 as the software component of the SO System with Treatment Planning 3.1 The software component of the SO System was originally cleared in K181271. Meshmixer is Off-the-Shelf (OTS) Software, the new software has been developed by SO exclusively. The functional requirements of the software component are unchanged. A comparison of Treatment Planning 3.1 and Meshmixer 3.4 that both software provide the same features and functional workflows is evidence of substantial equivalence. The replacement software is identical in performance and function to the previously used software.

#### Intended Use / Indications for Use

The indications for use are the same between this submission and the primary predicate K181271.

The Signature Orthodontic System is a treatment planning software and orthodontic appliance system used to correct malocclusions in orthodontic patient-matched orthodontic appliances.

The Signature Orthodontic System (K181271) is a treatment planning software and orthodontic appliance system used to correct malocclusions in orthodontic patient-matched orthodontic appliances.

#### Non-clinical Performance Testing

Validation testing of the TPS was performed in accordance with SO's design control activities for software and to the software's Test Plan. Validation results, Table 6-1, show that the version of the TPS tested performed equivalent to the software component of the primary predicate (K181271).

| Primary PREDICATE K181271<br>Signature Orthodontic System;<br>TPS Off-the-shelf | Signature Orthodontic System;<br>TPS 3.1                          | Equivalence Result          |
|---------------------------------------------------------------------------------|-------------------------------------------------------------------|-----------------------------|
| 4.1 Diagnosis - viewing patient's<br>digital impression                         | 4.1 Diagnosis - viewing patient's<br>digital impression           | Equivalent (Test<br>Report) |
| 4.1 Diagnosis - successful diagnosis<br>of patient's malocclusion               | 4.1 Diagnosis - successful diagnosis<br>of patient's malocclusion | Equivalent (Test<br>Report) |

{5}------------------------------------------------

| 4.4 Data Handling - case data<br>delivered securely and un-corrupted  | 4.4 Data Handling - case data<br>delivered securely and un-corrupted  | Equivalent (Test<br>Report) |
|-----------------------------------------------------------------------|-----------------------------------------------------------------------|-----------------------------|
| 4.1 Diagnosis - viewing and<br>measuring patient's digital impression | 4.1 Diagnosis - viewing and<br>measuring patient's digital impression | Equivalent (Test<br>Report) |

#### Clinical Performance Testing

No clinical performance testing was conducted on SO System brackets.

#### Device Comparison Tables

The following tables, Tables 6-2 and 6-3, provide a comparison of the device features, functions, and performance characteristics between the Signature Orthodontic System and its predicate devices.

| Table 6-2 - SO System TPS Feature Comparison                                   |                                                                               |                                                                               |                                                                                                                                              |                                                                                                                                                             |
|--------------------------------------------------------------------------------|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Workflow<br>Function                                                           | SO System TPS -<br>Off-the-shelf                                              | SO System TPS -<br>TPS 3.1                                                    | Similarities                                                                                                                                 | Differences                                                                                                                                                 |
| 4.1 Diagnosis -<br>viewing patient's<br>digital<br>impression                  | Rendering of<br>impression using<br>triangle meshes<br>read from STL<br>files | Rendering of<br>impression using<br>triangle meshes<br>read from STL<br>files | Identical                                                                                                                                    |                                                                                                                                                             |
| 4.1 Diagnosis -<br>successful<br>diagnosis of<br>patient's<br>malocclusion     | Hide/show<br>individual arches                                                | Hide/show<br>individual arches                                                | Identical                                                                                                                                    | TPS 3.1 enables<br>hide/show arches<br>with a single<br>click. OTS<br>requires multiple<br>clicks.                                                          |
| 4.4 Data<br>Handling - case<br>data delivered<br>securely and un-<br>corrupted | Secure link used<br>to download STL<br>files, then File -><br>Open in device. | Open secure link<br>in web browser.                                           | Secure link<br>communicated<br>using the<br>orthodontist's<br>confirmed e-<br>mail addresses<br>Impression data<br>transmitted over<br>HTTPs | TPS 3.1 removes<br>the download<br>STL files step<br>and integrates<br>the transfer with<br>the loading of the<br>software by the<br>user's web<br>browser. |
| 4.1 Diagnosis -<br>viewing and<br>measuring<br>patient's digital<br>impression | Rotate impression<br>display                                                  | Rotate impression<br>display                                                  | Identical                                                                                                                                    |                                                                                                                                                             |
|                                                                                | Pan impression<br>display                                                     | Pan impression<br>display                                                     | Identical                                                                                                                                    |                                                                                                                                                             |
|                                                                                | Zoom impression<br>display                                                    | Zoom impression<br>display                                                    | Identical                                                                                                                                    |                                                                                                                                                             |

{6}------------------------------------------------

|  | Point-to-point<br>measurement | N/A | The requirements<br>specification for<br>TPS 3.1 does not<br>include point-to-<br>point<br>measurement. |
|--|-------------------------------|-----|---------------------------------------------------------------------------------------------------------|
|--|-------------------------------|-----|---------------------------------------------------------------------------------------------------------|

| Table 6-3 - Device Comparison                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |           |  |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|--|
| ITEM                                            | Primary PREDICATE K181271<br>Signature Orthodontic<br>System; TPS Off-the-shelf                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Signature Orthodontic<br>System; TPS 3.1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Comments  |  |
| Description                                     | The Signature Orthodontic<br>System (SO System) is a<br>treatment planning software<br>(TPS) and orthodontic<br>appliance system used to<br>correct malocclusions in<br>orthodontic patients using<br>patient-matched orthodontic<br>appliances. The SO System<br>consists of patient-specific<br>ceramic brackets, patient-<br>specific bracket placement jigs,<br>arch wire templates, and a TPS<br>for viewing, measuring, and<br>modifying cases. Signature<br>Orthodontics' (SO) operators<br>and the orthodontists use the<br>TPS to generate a prescription<br>of their choosing. SO then<br>manufactures the patient-<br>specific brackets and<br>placement jigs using proprietary<br>additive manufacturing<br>techniques. The orthodontist<br>then bonds the brackets to the<br>teeth using the optional<br>placement jig and ligates wires<br>to enable tooth movement. The<br>SO System does not contain<br>commercially-available or<br>patient-specific shaped arch<br>wires, ligatures, or adhesives<br>that affixes the brackets to the<br>teeth. | The Signature Orthodontic<br>System (SO System) is a<br>treatment planning software<br>(TPS) and orthodontic<br>appliance system used to<br>correct malocclusions in<br>orthodontic patients using<br>patient-matched orthodontic<br>appliances. The SO System<br>consists of patient-specific<br>ceramic brackets, patient-<br>specific bracket placement<br>jigs, arch wire templates, and<br>a TPS for viewing,<br>measuring, and modifying<br>cases. Signature<br>Orthodontics' (SO) operators<br>and the orthodontists use the<br>TPS to generate a<br>prescription of their choosing.<br>SO then manufactures the<br>patient-specific brackets and<br>placement jigs using<br>proprietary additive<br>manufacturing techniques.<br>The orthodontist then bonds<br>the brackets to the teeth<br>using the optional placement<br>jig and ligates wires to enable<br>tooth movement. The SO<br>System does not contain<br>commercially-available or<br>patient-specific shaped arch<br>wires, ligatures, or adhesives<br>that affixes the brackets to<br>the teeth. | Identical |  |
| Product Codes/<br>Regulations                   | NJM (Orthodontic Ceramic<br>Bracket, 21CFR§872.5470)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | NJM (Orthodontic Ceramic<br>Bracket, 21CFR§872.5470)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Identical |  |
| Indications for<br>Use                          | The Signature Orthodontic<br>System is a treatment planning<br>software and orthodontic<br>appliance system used to<br>correct malocclusions in<br>orthodontic patients using<br>patient-matched orthodontic<br>appliances.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | The Signature Orthodontic<br>System is a treatment<br>planning software and<br>orthodontic appliance system<br>used to correct malocclusions<br>in orthodontic patients using<br>patient-matched orthodontic<br>appliances.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Identical |  |
| Sequence of<br>Treatment Plan<br>or Mode of Use | Signature Orthodontics (SO)<br>receives the patient's digital<br>data via commercially available<br>communications tools from the<br>DDS/DMD. SO prepares a final<br>tooth setup and bracket<br>placement plan and returns the<br>digital data back to the<br>DDS/DMD. The orthodontist<br>can then view, measure and<br>modify the case using the TPS.<br>Once approved, brackets and<br>jigs are manufactured and<br>shipped to the DDS/DMD.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Signature Orthodontics (SO)<br>receives the patient's digital<br>data via commercially<br>available communications<br>tools from the DDS/DMD. SO<br>prepares a final tooth setup<br>and bracket placement plan<br>and returns the digital data<br>back to the DDS/DMD. The<br>orthodontist can then view,<br>measure and modify the case<br>using the TPS. Once<br>approved, brackets and jigs<br>are manufactured and<br>shipped to the DDS/DMD.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical |  |
| Manufacturing<br>Method                         | Final desired arrangement of<br>teeth, brackets, wires and jigs<br>are designed with the guidance<br>of SO internal computer<br>software using a 3-D model of<br>the patient. The Treatment<br>Planning Software allows the<br>clinician to review, measure,<br>and modify the case. Internal<br>SO software generates the 3D<br>image file that proprietary<br>additive manufacturing<br>equipment uses to create the<br>brackets and indirect bonding<br>(IDB) tray.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Final desired arrangement of<br>teeth, brackets, wires and jigs<br>are designed with the<br>guidance of SO internal<br>computer software using a 3-<br>D model of the patient. The<br>Treatment Planning Software<br>allows the clinician to review,<br>measure, and modify the<br>case. Internal SO software<br>generates the 3D image file<br>that proprietary additive<br>manufacturing equipment<br>uses to create the brackets<br>and indirect bonding (IDB)<br>tray.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical |  |

{7}------------------------------------------------

## Conclusion

The Signature Orthodontics' Signature Orthodontic System presented in this 510(k) submission and its predicate the Signature Orthodontic System (K181271) are substantially equivalent in materials, performance and safety characteristics.

---

**Source:** [https://fda.innolitics.com/device/K183542](https://fda.innolitics.com/device/K183542)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
