ProRad 2FC and ProRad 3NC Stationary Radiographic Systems

K183541 · Prognosys Medical Systems Private Limited · KPR · Jan 24, 2019 · Radiology

Device Facts

Record IDK183541
Device NameProRad 2FC and ProRad 3NC Stationary Radiographic Systems
ApplicantPrognosys Medical Systems Private Limited
Product CodeKPR · Radiology
Decision DateJan 24, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1680
Device ClassClass 2
AttributesPediatric

Indications for Use

ProRad Series Stationary Radiographic System is intended for use by a qualified, trained doctor or technician on both adult and paediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

Device Story

ProRad 2FC (floor-mounted) and 3NC (ceiling-suspended) are stationary digital radiographic systems. Input: X-ray radiation attenuated by patient anatomy. Transformation: X-ray tube generates radiation; flat panel detector (FPD) captures signal; software processes raw data into diagnostic images. Output: Digital radiographic images. Used in clinical settings by physicians or technicians. Healthcare providers view images on console to support diagnostic decision-making. Benefits: Enables general-purpose diagnostic imaging of human body.

Clinical Evidence

No clinical study required. Bench testing only. Compliance with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-54, 60601-1-6, and ISO 14971. Clinical images acquired by the device were reviewed by a radiologist and found acceptable for diagnostic use.

Technological Characteristics

Stationary X-ray system; configurations: floor-mounted (2FC) or ceiling-suspended (3NC). Components: X-ray generator (CPI CMP 200DR), X-ray tubes (Toshiba/Varex), collimators (Ralco/Daesung), and various FPDs (Canon, Varex, DRTECH). Connectivity: DICOM 3.0, Ethernet or IEEE 802.11n wireless. Software: Canon CXDI NE, DROC, or Econsole1. Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-54, 60601-1-6, ISO 14971.

Indications for Use

Indicated for adult and pediatric patients for diagnostic radiographic exposures of the skull, spine, chest, abdomen, extremities, and other body parts. Contraindicated for mammography.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text. Prognosys Medical Systems Private Limited % Dr. Shruti Sancheti Manager - Quality and Regulatory Affairs IZiel Healthcare 14. Hadapsar Industrial Estate Hadapsar, Pune, 411013 INDIA January 24, 2019 Re: K183541 Trade/Device Name: ProRad 2FC and ProRad 3NC Digital Stationary Radiographic Systems Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR, MQB Dated: December 20, 2018 Received: December 21, 2018 Dear Dr.Sancheti: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Michael D. O'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K183541 Device Name ProRad 2FC and ProRad 3NC Digital Stationary Radiographic Systems #### Indications for Use (Describe) ProRad Series Stationary Radiographic System is intended for use by a qualified, trained doctor or technician on both adult and paediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------|---------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) SUMMARY 510(k) summary of Safety and Effectiveness for ProRad 2FC and ProRad 3NC Stationary Radiographic Systems is provided in accordance with 21 CFR 807.92. | Date: | October 5, 2018 | |-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter (Owner): | V. Krishna Prasad<br>President<br>Prognosys Medical Systems Pvt. Ltd.<br>No. 249, Ground Floor, Front Building, 4th Main<br>Road, Chamrajpet,<br>Bangalore - 560 018, Karnataka, India<br>Phone: +91-8026621544<br>E-mail Id: info@prognosysmedical.com | | 510(k) Contact Person: | Dr. Shruti Sancheti<br>Manager-Quality & Regulatory Affairs<br>IZiel Healthcare<br>14, Hadapsar Industrial Estate,<br>Hadapsar, Pune – 411013<br>India.<br>P: +91-72762 2555 M:+91-9518546814<br>Email: shruti.sancheti@izielhealthcare.com | | Device Trade Name: | ProRad 2FC and ProRad 3NC Stationary<br>Radiographic Systems | | Regulation Number: | 21 CFR 892.1680 | | Regulation Description: | Stationary X-Ray System | | Classification Panel: | Radiology | | Device Class: | Class II | | Product Code: | KPR, MQB | | Predicate Device: | 1. Jumong Series Stationary Radiographic System<br>(K150816)<br>2. Amrad Medical OTS Digital Radiography System,<br>Amrad Medical DFMTS Digital Radiography<br>System, Amrad Medical FRS Digital Radiography<br>System (K153119) | # Indications for Use: The ProRad Series Stationary Radiographic System is intended for use by a qualified, trained doctor or technician on both adult and paediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography. # Description of the Device: The ProRad series Stationary Radiographic System is a diagnostic x-ray system intended for general purpose radiographic imaging of the human body. There are two types of configurations (2FC and 3NC) for ProRad; the difference is in the {4}------------------------------------------------ mounting of the X-ray tube. For X-ray tube mounting the configuration is either the floor mounted (2FC) or ceiling suspension (3NC) assembly. The devices are a new combination of a previously cleared solid state digital x-ray acquisition panel and software with the diagnostic x-ray components (including Xray tube, high frequency X-ray generator, a tilting vertical bucky, X-ray table and collimator) required to make a complete system. The purchaser may select any of the digital panels and software based on the user's requirements. The other components are also available in different configurations to meet specific customer needs. The X-ray panel and imaging software have been previously cleared by the FDA, and most of the other components are used in previously cleared 510(k) devices (respective approval numbers are discussed in Section 11, Table 1). Details of the various configurations of all the components are given in Table 1 below: | Component<br>Category | Component Description and<br>Model | Manufacturing<br>Location | ProRad<br>2FC | ProRad<br>3NC | |------------------------------|-----------------------------------------------------------------------|---------------------------------|---------------|---------------| | Digital<br>Imaging<br>System | CANON Flat Panel Detector CXDI-<br>401C Compact | Canon Inc., Japan | √ | √ | | | CANON Flat Panel Detector CXDI-<br>410C Wireless | | √ | √ | | | CANON Portable Detector CXDI-<br>701C Wireless | | √ | √ | | | CANON Portable Detector CXDI-<br>710C Wireless | | √ | √ | | | VAREX PAXSCAN 4343R v3 | Varian Medical<br>Systems, Inc. | √ | √ | | | VAREX PAXSCAN 4336W v4<br>Wireless | | √ | √ | | | VAREX (PerkinElmer) Flat Panel<br>Detector XRpad 4343F MED | Perkin-Elmer, Inc | √ | √ | | | VAREX (PerkinElmer) Flat Panel<br>Detector XRpad 4336 MED<br>Wireless | | √ | √ | | | DRTECH Digital Flat Panel X-ray<br>Detector EVS 4343, EVS 4343G | DRTECH<br>Corporation | √ | √ | | | DRTECH Digital X-ray Detector<br>EVS 3643, EVS 3643G Wireless | | √ | √ | | Image<br>Acquisition | Canon CXDI NE | Canon Inc., Japan | √ | √ | | Software | DROC (Digital Radiography<br>Operator Console) | E-COM Technology<br>Ltd. | √ | √ | | | Econsole1 | DRTECH Corp. | √ | √ | Table 1: Available system configurations {5}------------------------------------------------ | Component<br>Category | Component Description and<br>Model | Manufacturing<br>Location | ProRad<br>2FC | ProRad<br>3NC | |------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|---------------|---------------| | X-ray<br>Generator | CPI CMP 200DR (32 kW, 40 kW, 50 kW, 65 kW, 80 kW) | Communications &<br>Power Industries,<br>Canada | √ | √ | | X-ray Tube | TOSHIBA (E7239FX, E7242FX, E7252X, E7254X, E7869X, E7884X, E7843X) | TOSHIBA<br>ELECTRON<br>TUBES &<br>DEVICES CO.,<br>LTD. | √ | √ | | | VAREX (RAD14, RAD60, RAD92) | Varian Medical<br>Systems, Inc. | √ | √ | | Collimator | Ralco (R221, R225ACS) | Ralco Italy | √ | √ | | | Daesung (M38) | Daesung<br>Corporation, Korea | √ | √ | | Mechanical<br>Tube Stand | Ceiling Mounted Tube Stand<br>- Full Auto- Single (SCS-FA)<br>- Auto Tracking- Single (SCS-AT)<br>- Manual- Single (SCS-MA)<br>- Full Auto- Dual (DCS-FA)<br>- Auto Tracking- Dual (DCS-AT)<br>- Manual- Dual (DCS-MA)<br><br>Floor Mounted Tube Stand | | -- | √ | | | - Full Auto- Single (SFT-FA)<br>- Auto Tracking- Single (SFT-AT)<br>- Manual- Single (SFT-MA)<br>- Auto Tracking (FTC-AT)<br>- Manual (FTC-MA)<br>- Full Auto- Dual (DFT-FA)<br>- Auto Tracking- Dual (DFT-AT)<br>- Manual- Dual (DFT-MA) | | √ | -- | | Mechanical<br>Bucky<br>Stand | Floor Mounted Vertical Bucky Stand<br>- VB-MO<br>- VB-MA<br>- VBNTL-MO<br>- VBNTL-MA | | √ | √ | | Patient<br>Table | Mobile Patient Table PRO-MT<br>- Mobile Patient Table with 4-way movement PRO-MT4<br>- Mobile Patient Table with 6-way movement PRO-MT6<br>- Fixed Patient Table PRO-FT<br>- Fixed Patient Table with 4-way movement PRO-FT4<br>- Fixed Patient Table with 6-way movement PRO-FT6 | | √ | √ | {6}------------------------------------------------ #### Safety and effectiveness, comparison to predicate device: This combination device has the same indications for use, is similar in technological characteristics to the predicate devices and employs already 510(k) cleared digital panels and software. In addition, most of the other components employed in the device have already been used in previously cleared 510(k) devices. The device has also been adequately tested to the relevant performance standards to prove that the integrated device is safe and effective for the intended use. | | | Predicate | Predicate | Comparison | |---------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------| | Comparable | Subject Device | Device 1 | Device 2 | Results | | Properties | | (K150816) | (K153119) | | | Intended Use | ProRad Series Stationary<br>Radiographic System is<br>intended for use by a<br>qualified, trained doctor or<br>technician on both adult and<br>paediatric subjects for taking<br>diagnostic radiographic<br>exposures of the skull, spinal<br>column, chest, abdomen,<br>extremities, and other body<br>parts. Applications can be<br>performed with the patient<br>sitting, standing, or lying in the<br>prone or supine position. Not<br>for mammography. | Jumong Series<br>is intended for<br>use by a<br>qualified,<br>trained doctor or<br>technician on<br>both adult and<br>pediatric<br>subjects for<br>taking<br>diagnostic<br>radiographic<br>exposures of<br>the skull, spinal<br>column, chest.<br>abdomen,<br>extremities, and<br>other body<br>parts.<br>Applications can<br>be performed<br>with the patient<br>sitting, standing,<br>or lying in the<br>prone or supine<br>position. Not for<br>mammography. | These<br>radiographic<br>systems are<br>intended for<br>use by a<br>qualified/train<br>ed physician<br>or technician<br>on both adult<br>and pediatric<br>subjects for<br>taking<br>diagnostic x-<br>rays. Not for<br>mammograp<br>hy,<br>angiography,<br>interventional<br>, or<br>fluoroscopy<br>use. | Equivalent | | Comparable<br>Properties | Subject Device | Predicate<br>Device 1<br>(K150816) | Predicate<br>Device 2<br>(K153119) | Comparison<br>Results | | Configuration<br>of Digital<br>Panels | Battery or AC operated<br>wireless IEEE 802.11n or<br>Wired Ethernet | Battery or AC<br>operated<br>wireless IEEE<br>802.11n or<br>Wired Ethernet | Battery or<br>AC operated<br>wireless<br>IEEE<br>802.11n or<br>Wired<br>Ethernet | Same | | Digital Panel<br>Models and<br>their<br>Clearance<br>Numbers | CANON Flat Panel<br>detector CXDI-401C<br>Compact (K103591)<br>CANON Flat Panel<br>detector CXDI-410C<br>Wireless (K171270)<br>CANON Portable detector<br>CXDI-701C Wireless<br>(K131106)<br>CANON Portable detector<br>CXDI-710C Wireless<br>(K170332)<br>VAREX PAXSCAN<br>4343R v3 (K172951)<br>VAREX PAXSCAN<br>4336W v4 (K161459)<br>VAREX (PerkinElmer)<br>XRpad 4343F MED Flat<br>Panel Detector (K142698)<br>VAREX (PerkinElmer)<br>XRpad 4336 MED Flat<br>Panel Detector (K140551)<br>DRTECH EVS 4343, EVS<br>4343G Digital Flat Panel<br>X-ray Detector (K162555)<br>DRTECH EVS 3643, EVS<br>3643G Digital X-ray<br>detector (K162552) | Vieworks<br>- Vivix-S<br>Wireless<br>(K122865)<br>- Vivix-S with<br>Vxvue<br>(K122866)<br>- Vivix-S<br>(K120020) | - VAREX<br>PAXSCAN<br>4343R<br>(K130318)<br> | Similar<br>functionality<br>(Note 1) | | Image<br>Acquisition<br>Panel<br>Specification<br>S | 3,320 x 3,408 Pixels 125<br>µm (CXDI 401C<br>Compact)<br>3,320 x 3,408 Pixels 125<br>µm (CXDI 410C<br>Compact)<br>2,800 x 3,408 Pixels 125<br>µm (CXDI 701C) | FXRD-<br>1717SA,<br>FXRD-<br>1717SB)<br>3,072 x<br>3,072,<br>140µm or<br>FXRD-<br>1417SA, | 3,072 x<br>3,072 Pixels<br>139 µm<br>(PAXSCAN<br>4343R)<br>2,560 x<br>3,072 Pixels<br>139 µm | Similar<br>functionality<br>(Note 1) | | Comparable<br>Properties | Subject Device | Predicate | Predicate | Comparison | | | | Device 1 | Device 2 | Results | | | | (K150816) | (K153119) | | | | 2,800 x 3,408 Pixels 125<br>µm (CXDI 710C)<br>3,072 x 3,072 Pixels 139<br>µm (PAXSCAN 4343R<br>v3)<br>3,072 x 2,560 Pixels 139<br>µm (PAXSCAN 4336W<br>v4)<br>4,318 x 4,320 Pixels 100<br>µm (XRpad 4343F)<br>3,556 × 4,320 Pixels 100<br>µm (XRpad 4336)<br>3,072 x 3,072 Pixels 140<br>µm (EVS 4343, EVS<br>4343G)<br>2,560 x 3,072 Pixels 140<br>µm (EVS 3643, EVS<br>3643G) | FXRD-<br>1417SB)<br>2,560 x<br>3,072,<br>140µm<br>Wireless:<br>FXRD-<br>1417WA,<br>FXRD1417<br>WB, 2,560 x<br>3,072,<br>140µm | (PAXSCAN<br>4336R) | | | DICOM | DICOM 3 | DICOM 3 | DICOM 3 | Same | | WiFi<br>Wireless<br>IEEE802.11n<br>(All<br>others are<br>Ethernet<br>Tethered.) | CANON CXDI 701C<br>Wireless<br>VAREX PAXSCAN<br>4336W v4<br>PerkinElmer XRpad 4336<br>MED Flat Panel Detector<br>DRTECH EVS 3643, EVS<br>3643G Digital X-ray<br>detector | K122865 Vivix-<br>S Wireless | Wireless<br>detector- Not<br>applicable | Similar<br>functionality | | Image<br>Acquisition<br>Software | CANON CXDI NE<br>(K153312, K171270)<br>DROC (Digital<br>Radiography Operator<br>Console) (K130883)<br>Econsole1 (K152172) | Vieworks<br>Vivix-S With<br>Vxvue<br>(K122866)<br>Vivix-S<br>(K120020) | ECOM<br>Software<br>(K130883) | Similar<br>functionality<br>(Note 2) | | Power<br>Source | AC Line, various voltages<br>available | AC Line,<br>various voltages<br>available | AC Line | Same | | X-ray<br>Generator | CPI CMP 200DR (32 kW, 40<br>kW, 50 kW, 65 kW, 80 kW)<br>(125 kV/150 kV for 32 kW, 40<br>kW)<br>(150kV for 50 kW, 65 kW, 80<br>kW) | CPI CMP200DR | CPI CMP<br>200DR HF,<br>150kV<br>(40kW,<br>50kW, 65<br>kW, 80 kW) | Same or<br>similar<br>functionality<br>(Note 3) | | Comparable<br>Properties | Subject Device | Predicate<br>Device 1<br>(K150816) | Predicate<br>Device 2<br>(K153119) | Comparison<br>Results | | X-ray Tubes | TOSHIBA<br>E7239FX: 125kV, LF2.0,<br>SF1.0, 140kHU<br>E7242FX:<br>125kV, LF1.5, SF0.6, 200kHU<br>E7252X: 150kV, LF1.2,<br>SF0.6, 300kHU<br>E7254X: 150kV, LF1.2,<br>SF0.6, 400kHU<br>E7869X: 150kV, LF1.2,<br>SF0.6, 600kHU<br>E7884X: 150kV, LF1.2,<br>SF0.6, 300kHU<br>E7843X: 150kV, LF1.2,<br>SF0.6,150kHU<br>VAREX<br>RAD14: 150kV, LF1.2, SF0.6,<br>300kHU<br>RAD60: 150kV, LF1.2, SF0.6,<br>400kHU<br>RAD92: 150kV, LF1.2, SF0.6,<br>600kHU | -- | CPI CMP<br>200DR HF,<br>125kV<br>(40kW)<br>150 kVp<br>0.6/1.2 mm<br>focal spots | Similar<br>functionality<br>(Note 4) | | Collimator | Ralco R221, R225ACS<br>Deasung M38 | Ralco R225 | Collimare | Similar<br>functionality<br>(Note 5) | | Performance<br>Standard | FDA 21 CFR 1020.30-31 | FDA 21 CFR<br>1020.30-31 | FDA 21 CFR<br>1020.30-31 | Same | | Electrical<br>Safety | Electrical Safety as per IEC<br>60601. | Electrical Safety<br>as per IEC<br>60601 | Electrical<br>Safety as per<br>IEC 60601 | Same | #### Substantial equivalence: {7}------------------------------------------------ {8}------------------------------------------------ {9}------------------------------------------------ | Note | Description | |--------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Note 1 | The subject device utilizes a different X-Ray flat panel detector; however,<br>the flat panel detectors used by the subject device are already previously<br>cleared by the FDA and the testing demonstrate that it does not raise the<br>level of safety concern and affect any effectiveness. The relevant 510(k)<br>approval numbers are K103591, K171270, K131106, K170332, K172951,<br>K161459, K142698, K140551, K162555 and K162552. | {10}------------------------------------------------ | Note 2 | Prognosys utilizes image processing software that has been previously<br>cleared by the FDA. The respective 510(k) approval numbers are K153312,<br>K171270, K130883 and K152172. Prognosys does not modify any<br>software utilized for image processing or display. | |--------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Note 3 | In addition to the various power ratings of x-ray generators used by<br>predicate device, the subject device also utilizes a similar X-ray generator<br>with different power ratings which does not raise any level of safety concern<br>and affect any effectiveness; the generator frequency does affect the X-ray<br>exposure parameters. In addition, the X-ray generators used by the subject<br>device have also been used in previously cleared 510(k) devices with<br>approval numbers K150816 and K153119. | | Note 4 | Both configurations (Toshiba and Varex) provide similar imaging resolution<br>on both focal spots. In addition, the X-ray tubes have been used in<br>previously cleared 510(k) devices with approval numbers K152767,<br>K173823, K103050 and K1881874. | | Note 5 | The inherent filtration for the collimator in the subject device is different from<br>the predicate device; this difference does not affect the safety and<br>effectiveness. | The ProRad series Stationary Radiographic System is a diagnostic x-ray system intended for general purpose radiographic imaging of the human body. This device utilizes previously cleared X-ray flat panel detectors and image processing software which are the main components of any X-Ray diagnostic system. The other components of the imaging system such as the X-Ray generator and X-Ray Tube are used in previously cleared 510(k) devices and are either identical or similar to the predicate devices. The remaining other components such as the Collimator, Mechanical Tube Stand, Mechanical Bucky Stand and Patient Table have no impact on the safety and effectiveness of the system. In addition, the subject device has been tested to the relevant performance standards as summarized below. # Summary of non-clinical testing: The ProRad 2FC and ProRad 3NC Digital Stationary Radiographic Systems comply with the following standards: - . IEC 60601-1: Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance - IEC 60601-1-2: Medical Electrical Equipment Part 1-2: General ● requirements for basic safety Collateral standard: Electromagnetic compatibility requirements and tests {11}------------------------------------------------ - IEC 60601-1-3: General requirements for basic safety and essential ● performance - Collateral Standard: Radiation protection in diagnostic Xray equipment - IEC 60601-2-54: Particular requirements for the basic safety and essential ● performance of X-ray equipment for radiography and radioscopy - IEC 60601-1-6: Medical electrical equipment Part 1-6: General ● requirements for basic safety and essential performance - Collateral standard: Usability - ISO 14971: Medical devices. Application of risk management to medical ● devices Furthermore, cybersecurity concerns were addressed based on the US FDA Guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, 2014" in the device labeling of these products. # Summary of clinical testing: Since the digital x-ray panels and software have previously received FDA clearance, a clinical study was not required as per the FDA guidance document. The following information was not necessary to demonstrate substantial equivalence but provides additional support that the device works as intended. Clinical images acquired by the device were reviewed by a radiologist. The images were found to be acceptable and allowed the radiologist to make an accurate diagnosis. # Conclusion: Technological differences from the predicate devices include different detectors, some with wireless functionality, different collimators and x-ray tubes. Although some of the components differ from the predicate devices, the detectors and other components have been cleared as part of other 510(k) submissions. Despite the few differences, after analysis of all test information, including the indications for use and test data, it can be concluded that the devices "ProRad 2FC and ProRad 3NC Stationary Radiographic Systems" are as safe and effective as the predicate devices, thus rendering them substantially equivalent to the predicate devices,
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...