K183538 · Starmed Co., Ltd. · GEI · Jan 17, 2019 · General, Plastic Surgery
Device Facts
Record ID
K183538
Device Name
VIVA combo RF System
Applicant
Starmed Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jan 17, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VIVA combo RF System is intended for use in percutaneous and intraoperative coagulation and ablation of tissue.
Device Story
VIVA combo RF System is an electrosurgical unit for percutaneous and intraoperative tissue coagulation and ablation. System inputs include user-defined settings and real-time sensor data (impedance, current, power, temperature). The device transforms these inputs via an RF generator to deliver up to 200 W of RF power (480 kHz) to an active electrode. A peristaltic pump provides chilled water for electrode cooling via an enclosed system. The system monitors parameters and automatically adjusts RF output based on impedance and temperature thresholds. Output is displayed on a PC (MRFALogger) or tablet (VIVALogger) for clinician monitoring. The device is operated by physicians in clinical settings. The system benefits patients by enabling controlled tissue ablation and coagulation. Modifications include improved electrical isolation, updated pump components, and software logic changes to RF current cut-off parameters.
Clinical Evidence
No clinical data was performed. Substantial equivalence is supported by non-clinical bench testing, including electrical safety, electromagnetic compatibility, and software verification/validation.
Technological Characteristics
Radiofrequency electrosurgical generator (480 kHz); max power 200 W @ 50 ohms. Includes peristaltic pump for electrode cooling. Connectivity via USB/data communication. Electrical safety per AAMI ANSI ES60601-1 and IEC 60601-2-2. EMC per IEC 60601-1-2. Software-controlled power delivery. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for percutaneous and intraoperative coagulation and ablation of tissue in patients requiring electrosurgical intervention.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
January 17, 2019
STARmed Co., Ltd. Mr. Jun-Young Jung OMR/ Deputy General Manager B-dong, 4F&12F, 158, Haneulmaeul-ro, IlsanDong-gu Goyang-si. Gyeonggi-do.10355. Republic of Korea
Re: K183538
Trade/Device Name: VIVA combo RF System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: December 10, 2018 Received: December 19, 2018
Dear Mr. Jung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
for
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/7 description: The image shows a digital signature. The name "Long H. Chen -S" is written on the left side of the image. On the right side, it says "Digitally signed by Long H. Chen -S" and the date and time of the signature: "Date: 2019.01.17 14:18:53 -05'00'".
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K183538
Device Name VIVA combo RF System
Indications for Use (Describe)
The VIVA combo RF System is intended for use in percutaneous and intraoperative coagulation and ablation of tissue.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
[As required by 21 CFR 807.92]
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92
## 1. Date Prepared [21 CFR 807.92(a) (1)]
12/10/2018
## 2. Submitter's Information [21 CFR 807.92(a) (1)]
| Name of Sponsor: | STARmed Co.,Ltd. |
|-----------------------|-------------------------------------------------------------------------------------------------|
| Address: | B-dong, 4F&12F, 158, Haneulmaeul-ro, IlsanDong-gu, Goyang-si,<br>Gyeonggi-do, Republic of Korea |
| Contact Name: | Mr. Jun-Young Jung |
| | Telephone #: +82-70-4673-8657 |
| | Fax #: +82-31-816-4546 |
| | Email: jjy3412@starmed4u.com |
| | Job Title : QMR/Deputy General Manager |
| | Degree : Bachelor's degree in Business(BBA) |
| Registration Number: | 3013557681 |
| Name of Manufacturer: | Same as Sponsor |
## 3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]
| Model Name: | VIVA combo RF System |
|----------------------|----------------------------------------------------------------|
| Common Name: | Electrosurgical Cutting and Coagulation Device and Accessories |
| Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Regulation Number: | 21 CFR 878.4400 |
| Product Code: | GEI |
| Device Class: | II |
## 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]
| 510(k) Number: | K163450 |
|----------------------|----------------------------------------------------------------|
| Applicant: | STARmed Co., Ltd. |
| Model Name: | VIVA combo RF System |
| Common Name: | Electrosurgical Cutting and Coagulation Device and Accessories |
| Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Regulation Number: | 21 CFR 878.4400 |
| Product Code: | GEI |
| Device Class: | II |
{4}------------------------------------------------
#### 5. Description of the Modified Device [21 CFR 807.92(a) (4)]
#### <Modification>
- . VIVA Combo RF System
- Apply an isolation of USB PCB and Foot Switch PCB using DC-DC Converter and -Digital Isolators
- Change a foot switch connector type and a data communication connector type -: The connector type has been changed from male type to female type.
- Addition of accessories (hospital-grade power cable) -
- Change a fuse from T 5AL 250 V to F 5AH 250 V -
- Software
- S VCS F Software for VIVA Combo RF System -
- : Version Change (1.10 to 1.11F) In the temperature mode, RF current is cut-off at 15 ohms (previous version: cut off at 25 ohms)
- -MRFALogger for PC
- : Version Change (1.10 to 1.41) Interface Change (No functional changes)
- VIVALogger for Tablet PC -
- : Version Change (2.20 to 3.30) Support different screen size (No functional changes)
- VIVA Pump ●
- Change a SMPS board from ECS25-60 to EPL225PS12 -
- Change a motor from OEM124-GJ600D to G550-S10 -
- Addition a pump (VP01-1); VP01-1 is different only a header, compared to changed pump -(VP01)
The VIVA combo RF System consists of RF generator, active electrode, grounding pad and peristaltic pump for electrode cooling. This device is designed to produce local tissue heating at the tip of the electrodes causing the coagulation of tissue. The VIVA combo RF System is capable of delivering up to 200 W of RF power and the available power is limited through software control. This system monitors the power, resistance, current and temperature.
The active electrodes are a sterile, single-use, hand-held electrosurgical instrument designed for use with VIVA combo RF System. Cooling of the electrode is provided by chilled water which is pumped through the inflow tubing, the electrode and out through the outflow tubing. This is an enclosed system within the electrode and the water is not to be in contact with the patient.
The VIVA combo RF system consists of S VCS F, MRFALogger and VIVALogger software.
The S_VCS_F software continuously monitors impedance, current, power and temperature. The unit automatically monitors rises in impedance and adjust RF output accordingly.
MRFALogger software can be stored and monitored on PC the RF output parameter (power, impedance, current, temperature and energy).
VIVALogger software can be stored and monitored on a tablet PC the RF output parameter (power, impedance, current, temperature and energy).
#### 6. Intended Use [21 CFR 807.92(a)(5)]
The intended use has not changed as a result of the modification and is as follows:
The VIVA combo RF System is intended for use in percutaneous and intraoperative coagulation and ablation of tissue.
#### 7. Technological Characteristics [21 CFR 807.92(a)(6)]
510(k) Summary [008- 2/6]
{5}------------------------------------------------
The VIVA combo RF System is substantially equivalent to legally marketed predicate device with respect to indications for use and technology characteristics. The table below presents side by side comparisons for each major component of each device:
| Parameter | Subject Device | Predicate Device | Remark | |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|-----------------------------------|-----------|
| 510(K) Number | Not Known | K163450 | - | |
| Manufacturer | STARmed Co., Ltd. | STARmed Co., Ltd. | - | |
| Model Name | VIVA combo RF System | VIVA combo RF System | - | |
| Classification Name | Electrosurgical, Cutting & Coagulation & Accessories | Electrosurgical, Cutting & Coagulation & Accessories | same | |
| Classification Panel | General and Plastic Surgery | General and Plastic Surgery | same | |
| Classification Regulation | 21 CFR 878.4400 | 21 CFR 878.4400 | same | |
| Product Code | GEI | GEI | same | |
| Device Class | Class II | Class II | same | |
| Intended Use | The VIVA combo RF System is intended for use in percutaneous and intraoperative coagulation and ablation of tissue. | The VIVA combo RF System is intended for use in percutaneous and intraoperative coagulation and ablation of tissue. | same | |
| Prescription or OTC | Prescription | Prescription | same | |
| Energy Used | Radiofrequency | Radiofrequency | same | |
| Electrosurgical Unit | | | | |
| Output Frequency | 480 kHz ± 10 % | 480 kHz ± 10 % | same | |
| Drive on Time | Up to 30 minutes | Up to 30 minutes | same | |
| Maximum Power Output | Up to 200 watts @ 50 ohms | Up to 200 watts @ 50 ohms | same | |
| Impedance Monitoring | Available | Available | same | |
| Temperature Monitoring | Available | Available | same | |
| Dimension<br>(W x L x H) | 260 x 348 x 115 mm | 260 x 348 x 115 mm | same | |
| Isolation of USB/Foot Switch PCB | Apply | Not apply | different | |
| Fuse | F 5AH 250 V | T 5AL 250 V | different | |
| Miscellaneous accessories | | | | |
| Peristaltic Pump | Model<br>SMPS<br>Motor | VP01 and VP01-1<br>EPL225PS12<br>G550-S10 | VP01<br>ECS25-60<br>OEM124-GJ600D | different |
| Hospital-grade power cable | | Included | Not included | different |
| Foot Switch Connect Type | | Female | Male | different |
| Software | | | | |
| | Version | 1.11F | 1.10 | different |
| S_VCS_B | Cut-Off in<br>temperature mode | at 15 ohms | at 25 ohms | different |
| MRFALog<br>ger | Version | 1.41 | 1.10 | different |
| VIVALog<br>ger | Version | 3.30 | 2.20 | different |
| Active Accessory | | | | |
| Monopolar/Bipolar | | Monopolar | Monopolar | Same |
| Diameter | | 17 Gauge | 17 Gauge | Same |
| Sterilization | | Ethylene Oxide | Ethylene Oxide | Same |
| Shelf Life | | 3 years | 3 years | Same |
| Biocompatibility | | ISO 10993-1 | ISO 10993-1 | Same |
{6}------------------------------------------------
| - Cytotoxicity<br>- Intracutaneous Irritation<br>- Sensitization<br>- Systemic Toxicity<br>- Pyrogen test | - Cytotoxicity<br>- Intracutaneous Irritation<br>- Sensitization<br>- Systemic Toxicity<br>- Pyrogen test | |
|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|--|
|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|--|
#### 8. Substantial Equivalence [21 CFR 807.92(b)]
When compared to the predicate device (K163450), the VIVA combo RF System presented in this submission has the same:
- Intended Use
- Technological characteristics
- · Operating principle
- · Energy Used
- · Design features
- · Physical specifications
- · Sterilization
- Shelf Life
- · Biocompatibility
A few differences are as follows
- Isolation of USB/Foot Switch PCB
- · Fuse
- Peristaltic Pump
- Foot switch Connect Type
- Software
There are no significant differences between the VIVA combo RF System and the predicate device that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended use.
#### Difference in the isolation of USB PCB and Foot Switch PCB
The changed subject device is applied an isolation of USB PCB and Foot Switch PCB using DC-DC Converter and Digital Isolators. The change in the isolation affects the electrical safety and electromagnetic compatibility. Therefore, the electromagnetic compatibility testing were conducted, the changed subject device was qualified with electrical safety according to IEC 60601-1, IEC 60601-2-2 and electromagnetic compatibility according to the IEC 60601-1-2.
#### Difference in the Fuse
The main fuse has been changed. The change in the fuse affects the electrical safety and electromagnetic compatibility. Therefore, the electrical safety testing was conducted, the changed subject device was qualified with electrical safety according to IEC 60601-1 and IEC 60601-2-2.
{7}------------------------------------------------
#### Difference in the peristaltic pump
The SMPS board and motor of the pump have been changed, and another pump (VP01-1) which is different only a pump header has been added. The change in the peristaltic pump affects the electrical safety and electromagnetic compatibility. Therefore, the electrical safety and electromagnetic compatibility testing were conducted, the changed subject device was qualified with electrical safety according to IEC 60601-1, IEC 60601-2-2 and electromagnetic compatibility according to the IEC 60601-1-2
#### Difference in the foot switch connect type
The connect type of foot switch has been changed from male to female type. The connect type does not affect the electrical safety and efficacy because the female connect type cannot be contacted with an operator; the female connect is not accessible part.
#### Difference in the software
S VCS F software has been changed that RF current is cut-off at 15 ohms in the temperature mode (previous version: cut off at 25 ohms). The change in the software affects the function of subject device. Therefore, the software verification and validation were conducted and performance test which verify the automatically cut-off at 15 ohms. The result of test provide that the change does not raise new safety or effective issues.
Also, the interface of MRFALogger software has been changed and VIVALogger software has been changed to support different screen size for tablet PC. The change in the software does not affect the function of subject device.
Due to these identical clauses and high similarities, the VIVA combo RF System is at least as safe as and as effective as the predicate device. The differences between the subject device and the predicate device do not raise new safety or effectiveness issues as explained above. Based on the electromagnetic compatibility and electrical safety testing, performance testing, and the comparison to predicate devices, the VIVA combo RF System is substantially equivalent to the previously cleared predicate devices.
#### 9. Summary of Non-Clinical Data [21 CFR 807.92(b)(1)]
- . Electrical Safety, Electromagnetic Compatibility and Performance: Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
- Testing to confirm compliance with the safety requirements of standard AAMI ANSI ES60601-1
- Testing to confirm compliance with particular requirements of standard IEC 60601-2-2
- Testing to confirm compliance with EMC requirements of standard IEC 60601-1-2
- Software Validation
The VIVA combo RF System contains MODERATE level of concern software. The software was designed and developed according to a software development process and was verified and validated. The software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005.
The VIVA combo RF System complies with the following international and FDA-recognized consensus standards:
| AAMI ANSI 60601-1: | Medical electrical equipment - Part 1: General requirements for basic safety<br>and essential performance (IEC 60601-1:2005, MOD, Edition 3.1) |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-2-2: | Medical electrical equipment - Part 2-2: Particular requirements for the basic<br>safety and essential performance of high frequency surgical equipment and<br>high frequency surgical accessories (Edition 6) |
{8}------------------------------------------------
- Medical Electrical Equipment Part 1-2: General Requirements For Basic IEC 60601-1-2: Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests (Edition 4)
#### 10. Summary of Clinical Data [21 CFR 807.92(b)(2)]
No clinical studies were considered necessary and performed.
#### 11. Conclusion [21 CFR 807.92(b)(3)]
The subject device is substantially equivalent to the currently marketed and predicate device in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.
Additionally, substantial equivalence was demonstrated through the non-clinical performance, which complied with the requirements specified in the international and FDA-recognized consensus standards, IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2 and bench testing, which complied with the requirements specified in the CDRH's Guidance for Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery.
The results of these tests demonstrate that VIVA combo RF System meets the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, electrical safety and electromagnetic compatibility testing demonstrates that the device is substantially equivalent to the previously cleared predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.