K183526 · Kls Martin GmbH + Co. KG · GEI · Apr 11, 2019 · General, Plastic Surgery
Device Facts
Record ID
K183526
Device Name
Codman Electrosurgical Generator, Foot Pedal
Applicant
Kls Martin GmbH + Co. KG
Product Code
GEI · General, Plastic Surgery
Decision Date
Apr 11, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Generator is indicated for use for cutting and coagulation of soft tissue in neurosurgical procedures.
Device Story
Codman Electrosurgical Generator is a compact RF energy source for cutting and coagulating soft tissue during neurosurgical procedures. Device operates at 4.0 MHz frequency; provides monopolar and bipolar output waveforms. User selects operating modes (CUT/COAG) and power levels via front-panel pushbuttons and knobs; digital displays indicate power levels and system status. Activation occurs via instrument finger switch or foot pedal. Used in clinical settings by healthcare professionals. Output affects tissue via thermal effects; enables precise hemostasis and tissue resection. Benefits include controlled surgical cutting and coagulation.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety, electromagnetic compatibility, risk management (ISO 14971), software performance (IEC 62304), system performance, thermal effects on tissue, usability, and transport simulation.
Technological Characteristics
Compact RF generator; 4.0 MHz operating frequency; 80W max output (CUT/COAG). Materials: metal and plastics. Connectivity: monopolar/bipolar instrument ports, foot pedal. Standards: IEC 60601-1 (electrical safety), IEC 60601-2-2 (electrosurgical safety), IEC 60601-1-2 (EMC). Software: compliant with IEC 62304.
Indications for Use
Indicated for cutting and coagulation of soft tissue in neurosurgical procedures. Prescription use only.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 11, 2019
KLS Martin GmbH + Co. KG Hanna Huttmeyer Regulatory Affairs Manager Am Flughafen 18 Freiburg im Breisgau, 79108 DE
Re: K183526
Trade/Device Name: Codman Electrosurgical Generator, Foot Pedal Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: March 8, 2019 Received: March 12, 2019
Dear Hanna Huttmeyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Integra LifeSciences Corporation Traditional 510(k) Premarket Submission Codman® Electrosurgical Generator
Image /page/2/Picture/2 description: The image shows the Integra logo. The word "INTEGRA" is written in gray, with the tagline "LIMIT UNCERTAINTY" written in a smaller font size below it. To the right of the text is a graphic of several green squares of varying sizes.
# Section 6 - 510(k) Summary
# 510(k) Summary Codman Electrosurgical Generator K183526
## 1. Submission Sponsor (Owner)
Integra LifeSciences Corporation 11 Cabot Boulevard Mansfield, MA 02048
Phone: +1 781-971-5692 Contact: Jennifer Siu Title: Sr. Regulatory Affairs Specialist Email: Jennifer.siu@integralife.com
#### 2. Submission Correspondent
KLS Martin GmbH + Co. KG Am Flughafen 18 79108 Freiburg, Germany
Phone: +49 761 557 997-0 Fax: +49 761 557 997-137
Contact: Hanna Hüttmeyer Title: Regulatory Affairs Email: hanna.huettmeyer@klsmartin.com
## 3. Date Prepared
December 17th, 2018
## 4. Device Identification
| Trade/Proprietary Name: | Codman Electrosurgical Generator |
|-------------------------|------------------------------------------------------|
| Common/Usual Name: | Electrosurgical Generator |
| Classification Name: | Electrosurgical, cutting & coagulation & accessories |
| Regulation Number: | 21 CFR 878.4400 |
| Product Code: | GEI |
| Device Class: | Class II |
| Classification Panel: | General and Plastic Surgery |
## 5. Legally Marketed Predicate Device
Sutter CURIS® RF Generator by Sutter Medizintechnik GmbH; K171869
{3}------------------------------------------------
Integra LifeSciences Corporation Traditional 510(k) Premarket Submission Codman® Electrosurgical Generator
Image /page/3/Picture/2 description: The image shows the word "INTEGRA" in gray font. Below the word "INTEGRA" is the phrase "LIMIT UNCERTAINTY" in a smaller, green font. To the right of the word "INTEGRA" are four green squares of varying sizes. The squares are arranged in a diagonal line, with the largest square at the top and the smallest square at the bottom.
#### 6. Indication for Use Statement
The Generator is indicated for use for cutting and coagulation of soft tissue in neurosurgical procedures.
#### 7. Device Description
The Codman Electrosurgical Generator is a compact source of high radiofrequency RF energy to be used with monopolar and bipolar instruments to cut and coagulate tissue during surgical procedures. This is achieved by the selection of electrosurgical accessories (e. g. monopolar or bipolar instruments), different operating modes and power levels on the front panel of the device. All selections are effected through pushbuttons and knobs. Indicator lights give the operator feedback on the status.
The power level for each mode is indicated by front-panel digital displays. During self-test or for error messages these digital displays also serve as indicators. Activation of output power is performed through an instrument's finger switch and/or a foot pedal.
The maximum power output of the device in CUT mode is 80 watts (at 600 Ohms load), and 80 watts in COAG mode (at 400 Ohms load). Operating frequency of the device is maintained at 4.0 MHz.
The device offers both monopolar and bipolar output wave forms. Both monopolar mode and bipolar mode offers one CUT and one COAG wave form, translating into a total of 4 operating modes for the user.
The Codman Electrosurgical Generator is packaged together with power cord and instruction manual. Typical accessories are a foot pedal with cable (part of this 510(k) submission) and appropriate monopolar and bipolar cables, electrodes and instruments, which are cleared and available separately.
The Codman Electrosurgical Generator is a prescription device.
## 8. Substantial Equivalence Discussion
The following table compares the Codman Electrosurgical Generator to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
| Manufacturer | Integra LifeSciences Corporation | Sutter Medizintechnik GmbH |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Codman® Electrosurgical<br>Generator<br>(Subject Device) | CURIS® RF Generator<br>(Predicate Device) |
| Manufacturer | Integra LifeSciences Corporation | Sutter Medizintechnik GmbH |
| Trade Name | Codman® Electrosurgical<br>Generator<br>(Subject Device) | CURIS® RF Generator<br>(Predicate Device) |
| 510(k) Number | TBD | K171869 |
| Product Code | GEI | GEI |
| Regulation Number | 21 CFR 878.4400 | 21 CFR 878.4400 |
| Regulation Name | Electrosurgical, cutting & coagulation<br>& accessories | Electrosurgical, cutting & coagulation &<br>accessories |
| Indications for Use | The Generator is indicated for use for<br>cutting and coagulation of soft tissue<br>in neurosurgical procedures. | Orthopedic, arthroscopic, spinal and<br>neurosurgical:<br>For resection and coagulation of soft<br>tissues and hemostasis of blood vessels in<br>orthopedic, arthroscopic, spinal and<br>neurosurgical procedures<br>For soft tissue resection and coagulation<br>during arthroscopic surgical procedures of<br>knee, shoulder, ankle, elbow, hip and wrist.<br><br>Cutting:<br>Snoring, submucosal palatal shrinkage,<br>traditional uvulopalatoplasty (RAUP),<br>myringotomy with effective hemorrhage<br>control, epistaxis treatment and turbinate<br>shrinkage, skin incisions, biopsy, cysts,<br>abscesses, tumors, cosmetic repairs,<br>development of skin flaps, skin tags,<br>blepharoplasty<br><br>Blended Cutting and Coagulation:<br>Snoring, submucosal palatal shrinkage,<br>traditional uvulopalatoplasty (RAUP),<br>myringotomy with effective hemorrhage<br>control, epistaxis treatment and turbinate<br>shrinkage, skin tags, papilloma keloids,<br>keratosis, verrucae, basal cell carcinoma,<br>nevi, fistulas, epithelioma, cosmetic repairs,<br>cysts, abscesses, development of skin<br>flaps<br><br>Hemostasis and nonablative Coagulation:<br>Control of bleeding, epilation,<br>teleangiectasia<br><br>Bipolar:<br>Pinpoint, precise coagulation, pinpoint<br>hemostasis in any field (wet or dry),<br>snoring, submucosal palatal shrinkage,<br>traditional uvulopalatoplasty (RAUP),<br>myringotomy with effective hemorrhage<br>control, epistaxis treatment and turbinate<br>shrinkage |
| Premarket Submission<br>surgical Generator | | |
| | Integra LifeSciences Corporation | Sutter Medizintechnik GmbH |
| | Codman® Electrosurgical<br>Generator<br>(Subject Device) | CURIS® RF Generator<br>(Predicate Device) |
| | Metal & plastics | Metal & plastics |
| | No | no |
| | Yes | yes |
| | IEC 60601-1 all applicable<br>requirements met | IEC 60601-1 all applicable requirements<br>met |
#### Table 6A - Comparison of Characteristics
{4}------------------------------------------------
# Integra LifeSciences Corporation
Traditional 510(k) Premarket Submission Codman® Electrosurgical Generator
Image /page/4/Picture/3 description: The image shows the Integra logo. The word "INTEGRA" is in large, gray, sans-serif font. Below it, in smaller green font, is the phrase "LIMIT UNCERTAINTY". To the right of the text are three green squares of varying sizes, arranged in a column.
{5}------------------------------------------------
Manufacturer Trade Name
Battery Operated AC Powered Electrical Safety
Electromagnetic
Compatibility
HF Electrical Safety
Material
met
met
IEC 60601-1-2 all applicable requirements
IEC 60601-2-2 all applicable requirements
## 9. Non-Clinical Performance Data
IEC 60601-1-2 all applicable
IEC 60601-2-2 all applicable
requirements met
requirements met
As part of demonstrating safety and effectiveness of Codman Electrosurgical Generator and in showing substantial equivalence to the predicate device that is subject to this 510(k) submission, Integra LifeSciences Corporation completed a number of non-clinical performance tests against applicable standards.
| Test | Pass / fail criteria | Results | |
|------|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|--------|
| 1 | Electrical safety | Compliance to IEC 60601-1:2005 + A1:2012;<br>ANSI AAMI ES60601-1:2005/(R)2012 and<br>A1:2012, C1:2009/(R)2012 and<br>A2:2010/(R)2012 | Passed |
| 2 | Electrical safety | Compliance to IEC 60601-2-2:2017 (Generator)<br>Compliance to IEC 60601-2-2:2009 (Foot Pedal) | Passed |
| 3 | Electromagnetic compatibility | Compliance to IEC 60601-1-2:2014 | Passed |
| 4 | Risk Management | Compliance to ISO 14971:2007 | Passed |
| 5 | Software Performance | Compliance to IEC 62304:2006 | Passed |
Table 6B - Performance Standards Testing Summary
To demonstrate that the Codman Electrosurgical Generator meets all design specifications and performance requirements, nonclinical bench testing has been performed in accordance with the internal R&D process in compliance with the proposals and recommendations of FDA guidance: "Premarket Notification [510(k)] Submissions for Electrosurgical Devices for General Surgery" - Guidance for Industry and Food and Drug Administration Staff, August 15, 2016.
In particular, tests have been carried out with respect to the following subject areas:
- Electromagnetic Compatibility and Electrical Safety
- System Performance Testing ●
- Thermal effects on tissue ●
- Usability
- Transport simulation ●
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the text "K183526 - Page 5 of 5" in blue font. Below the text is the word "INTEGRA" in gray font, with the words "LIMIT UNCERTAINTY" in smaller font below it. To the right of the word "INTEGRA" are four green squares of varying sizes.
Internal verification and validation testing confirms that product specifications are met which are equivalent to the predicate device in technological characteristics. The testing results support that the requirements for performance and electrical safety testing were met for the acceptance of the device. The Codman Electrosurgical Generator passed all testing and supports the claims of substantial equivalence to the predicate device.
## 10. Statement of Substantial Equivalence
Based upon the intended use, design, operating principles, patient/user interface, comparison to the predicate device, and testing conducted, it is concluded that the subject device, Codman Electrosurgical Generator, is substantially equivalent to the predicate device, Sutter CURIS® RF Generator (K171869), and therefore does not raise different questions of safety and effectiveness.
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183526
Device Name Codman Electrosurgical Generator
Indications for Use (Describe)
The Generator is indicated for use for cutting and coagulation of soft tissue in neurosurgical procedures.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.