Ammonia II

K183517 · Roche Diagnostics Operations (Rdo) · JIF · Feb 8, 2019 · Clinical Chemistry

Device Facts

Record IDK183517
Device NameAmmonia II
ApplicantRoche Diagnostics Operations (Rdo)
Product CodeJIF · Clinical Chemistry
Decision DateFeb 8, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1065
Device ClassClass 1

Intended Use

The Ammonia II assay is an enzymatic in vitro test for the quantitative determination of ammonia in human plasma on Roche/Hitachi cobas c systems. Ammonia measurements are used in the diagnosis and treatment of severe liver disorders, such as cirrhosis, hepatitis, and Reye's syndrome.

Device Story

Ammonia II is an enzymatic in vitro diagnostic assay for quantitative ammonia determination in human plasma. It utilizes glutamate dehydrogenase (GLDH) to catalyze the reductive amination of 2-oxoglutarate with NH4+ and NADPH, producing glutamate and NADP+. The reaction is monitored by measuring the decrease in absorbance at 340 nm, which is directly proportional to ammonia concentration. The device is intended for use on Roche/Hitachi cobas c systems in clinical laboratory settings by trained laboratory personnel. Results are provided to clinicians to aid in the diagnosis and management of severe liver conditions like cirrhosis, hepatitis, and Reye's syndrome. The assay requires K2- or K3-EDTA plasma samples.

Clinical Evidence

Bench testing only. Performance evaluated via precision (CLSI EP05-A3), sensitivity (LoB/LoD/LoQ per CLSI EP17-A2), linearity (CLSI EP06-A), and interference studies. Method comparison against the predicate device using 112 human plasma samples yielded a regression of y = 1.001x – 1.90 µmol/L with a correlation coefficient (r) of 1.000. Matrix comparison between K2-EDTA and K3-EDTA plasma showed high correlation (r=1.000).

Technological Characteristics

Enzymatic assay using glutamate dehydrogenase (microbial). Reagents include BICINE buffer, 2-oxoglutarate, and NADPH. Measures decrease in absorbance at 340 nm. Designed for use on Roche/Hitachi cobas c 501 analyzer. Measuring range: 10-1000 µmol/L (17-1703 µg/dL). Sample matrix: K2- or K3-EDTA plasma.

Indications for Use

Indicated for quantitative determination of ammonia in human plasma for patients requiring diagnosis and treatment of severe liver disorders, including cirrhosis, hepatitis, and Reye's syndrome.

Regulatory Classification

Identification

An ammonia test system is a device intended to measure ammonia levels in blood, serum, and plasma, Ammonia measurements are used in the diagnosis and treatment of severe liver disorders, such as cirrhosis, hepatitis, and Reye's syndrome.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION". February 8, 2019 Roche Diagnostics Operations (RDO) Noel Mencias Principal, Regulatory Affairs 9115 Hague Road Indianapolis, IN 46250 Re: K183517 Trade/Device Name: Ammonia II Regulation Number: 21 CFR 862.1065 Regulation Name: Ammonia test system Regulatory Class: Class I, reserved Product Code: JIF Dated: December 17, 2018 Received: December 18, 2018 Dear Noel Mencias: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Kellie B. Kelm -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) k183517 Device Name Ammonia II Indications for Use (Describe) The Ammonia II assay is an enzymatic in vitro test for the quantitative determination of ammonia in human plasma on Roche/Hitachi cobas c systems. Ammonia measurements are used in the diagnosis and treatment of severe liver disorders, such as cirrhosis, hepatitis, and Reye's syndrome. Type of Use (Select one or both, as applicable) | <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |-------------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## Ammonia II # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. In accordance with 21 CFR 807.87, Roche Diagnostics hereby submits official notification as required by Section 510(k) of the Federal Food, Drug and Cosmetics Act of our intention to market the device described in this Premarket Notification 510(k). The purpose of this Traditional 510(k) Premarket Notification is to obtain FDA review and clearance for the Ammonia II assay | Submitter Name | Roche Diagnostics | |--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Address | 9115 Hague Road<br>Indianapolis, IN 46250-0457 | | Contact | Noel B. Mencias<br>Phone: (317) 521-3172<br>FAX: (317) 521-2324<br>Email: noel.mencias@roche.com | | Date Prepared | December 17, 2018 | | Proprietary Name | Ammonia II | | Common Name | Ammonia | | Classification Name | Ammonia test system | | Product Codes,<br>Regulation Numbers | JIF, 21 CFR 862.1065 | | Predicate Devices | Beckman Coulter Ammonia | | Establishment<br>Registration | For the Ammonia II assay, the establishment registration number<br>for Roche Diagnostics GmbH in Mannheim, Germany is 9610126,<br>and for Penzberg, Germany, 9610529. The establishment<br>registration number for Roche Diagnostics in the United States is<br>1823260. | {4}------------------------------------------------ #### DEVICE DESCRIPTION 1. The Ammonia II (NH3L2) assay is an enzymatic in vitro test for the quantitative determination of ammonia in human plasma on Roche/Hitachi cobas c systems. The Ammonia II assay is an enzymatic method, with glutamate dehydrogenase. #### 2. INDICATIONS FOR USE The Ammonia II assay is an enzymatic in vitro test for the quantitative determination of ammonia in human plasma on Roche/Hitachi cobas c systems. Ammonia measurements are used in the diagnosis and treatment of severe liver disorders, such as cirrhosis, hepatitis, and Reye's syndrome. #### TECHNOLOGICAL CHARACTERISTICS 3. The Ammonia II assay is an enzymatic method, with glutamate dehydrogenase. Glutamate dehydrogenase (GLDH) catalyzes the reductive amination of 2-oxoglutarate with NH4+ and NADPH to form glutamate and NADP+. The concentration of the NADP+ formed is directly proportional to the ammonia concentration. It is determined by measuring the decrease in absorbance. The following tables compare the Ammonia II (NH3L2) assay with its predicate device, SYNCHRON Systems Ammonia Reagent (k003196). | Feature | SYNCHRON Systems<br>Ammonia Reagent (k003196) | Ammonia II | |--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | AMM reagent, when used in<br>conjunction with UniCel®<br>DxC 600/800 System(s) and<br>SYNCHRON® Systems<br>Ammonia<br>Calibrators, is intended for the<br>quantitative determination of<br>ammonia concentration in<br>human plasma. | The Ammonia II assay is an<br>enzymatic in vitro test for the<br>quantitative determination of<br>ammonia in human plasma on<br>Roche/Hitachi cobas c systems. | | Feature | SYNCHRON Systems<br>Ammonia Reagent (k003196) | Ammonia II | | Test Principle | The SYNCHRON® System(s)<br>automatically proportions the<br>appropriate sample and reagent<br>volumes into a cuvette. The<br>ratio used is one part sample to<br>6 parts reagents. The system<br>monitors the change in<br>absorbance at 340 nanometers.<br>This change in absorbance is<br>directly proportional to the<br>concentration of ammonia in<br>the sample and is used by the<br>SYNCHRON® System(s) to<br>calculate and express the<br>ammonia concentration. | Enzymatic method, with<br>glutamate dehydrogenase.<br>Glutamate dehydrogenase<br>(GLDH) catalyzes the reductive<br>amination of<br>2-oxoglutarate with NH4+ and<br>NADPH to form glutamate and<br>NADP+.<br>The concentration of the NADP+<br>formed is directly proportional to<br>the ammonia concentration. It is<br>determined by measuring the<br>decrease in absorbance. | | Instrument | UniCel DxC 600/800<br>System(s) and SYNCHRON<br>Systems | cobas c 501 | | Reagent Composition | REAGENT CONSTITUENTS<br>α-Ketoglutarate 3.23 mmol/L<br>ADP 1.9 mmol/L<br>NADPH 0.22 mmol/L<br>GLDH (Beef liver) >10 U/L | R1 BICINEa) buffer: 300<br>mmol/L, pH 8.3; GLDH<br>(microbial): ≥ 16.7 µkat/L;<br>detergents; preservative<br>R3 GLDH (microbial): ≥ 5.0<br>µkat/L; 2-oxoglutarate: 78<br>mmol/L;<br>NADPH: ≥ 1.3 mmol/L;<br>nonreactive buffer<br>a) BICINE = N,N-bis(2-<br>hydroxyethyl)-glycine | | Sample Type/Matrix | Sodium Heparin<br>EDTA | K2- and K3-EDTA plasma | | Calibrator | SYNCHRON Systems<br>Ammonia Calibrators | Ammonia/Ethanol/CO2<br>Calibrator | | Calibration Interval | Under typical operating<br>conditions the AMM reagent<br>cartridge must be calibrated<br>every 5 days and also with<br>certain parts replacement or<br>maintenance procedures, as<br>defined in the UniCel DxC<br>600/800 System Instructions<br>For Use (IFU) manual. | Calibration frequency 2-point<br>calibration<br>- after lot change<br>- automatically every 14 days<br>- as required following quality<br>control<br>procedures | | Feature | SYNCHRON Systems<br>Ammonia Reagent (k003196) | Ammonia II | | Controls | At least two levels of control<br>material | Ammonia/Ethanol/CO2 Control<br>Normal<br>Ammonia/Ethanol/CO2 Control<br>Abnormal | | Traceability/Standardization | Ammonia measurand (analyte)<br>in this calibrator is traceable to<br>the manufacturer's selected<br>measuring method. The<br>traceability process is based on<br>prEN ISO 17511. | This method has been<br>standardized against a primary<br>standard. | | Reagent Stability | AMM reagent, when stored<br>unopened at +2°C to +8°C,<br>will remain stable until the<br>expiration date printed on the<br>label. | Shelf life at 2-8 °C: See<br>expiration date on cobas c pack<br>label. | | Reagent On-Board Stability | Once opened, the reagent is<br>stable for 30 days at +2°C to<br>+8°C unless the expiration date<br>is exceeded. | On-board in use and refrigerated<br>on the analyzer: 16 weeks | | Measuring Range | 16 - 1700 µg/dL (9 – 1000<br>µmol/L) | 10-1000 µmol/L (17-1703<br>µg/dL) | | Lower Limits of<br>Measurement | lower limit of 9 µmol/L (16<br>µg/dL) | Limit of Blank = 10 µmol/L (17<br>µg/dL)<br>Limit of Detection = 10 µmol/L<br>(17 µg/dL)<br>Limit of Quantitation = 10<br>µmol/L (17 µg/dL) | | Sample Stability | Tubes should be filled<br>completely, mixed gently by<br>inversion, placed on ice,<br>centrifuged immediately for 10<br>minutes at an RCF of 1500G<br>and analyzed within 30<br>minutes. Samples should not be<br>frozen. The tubes should be<br>tightly stoppered at all times. | Stability in plasma:<br>30 min at 15-25 °C<br>2 hours at 2-8 °C<br>3 days at -20 ± 5 °C<br>4 weeks at (-60)-(-90) °C (at<br>least) | ## Table 1: Assay Comparison {5}------------------------------------------------ {6}------------------------------------------------ #### NON-CLINICAL PERFORMANCE EVALUATION 4. The following performance data were provided in support of the substantial equivalence determination: Precision according to CLSI EP05-A3 Detection Limit: LoB, LoD, LoQ according to CLSI EP17-A2 {7}------------------------------------------------ Linearity according to CLSI EP06-A Endogenous Interferences Exogenous Interferences – Drugs Method Comparison to Predicate Matrix Comparison - Anticoagulants. #### Precision 4.1. - Repeatability and Intermediate Precision 4.1.1. Precision experiments were performed in accordance with CLSI Guideline EP5-A3. Two aliquots per run, two runs per day for ≥ 21 days were performed on the same analyzer using 3 lots of reagent. Repeatability (within run precision) and intermediate precision (within lab precision) were calculated. The samples were randomized in each run separately. For each sample, the following were calculated: Mean, Repeatability and intermediate precision as CV and SD values, and the upper 95% confidence interval for SD and CV values. | Specimen | Mean (µmol/L) | SD (µmol/L) | CV (%) | |----------------|---------------|-------------|--------| | AMM-N | 66.6 | 1.40 | 2.1 | | AMM-P | 243 | 3.45 | 1.4 | | Human Plasma 1 | 26.0 | 1.26 | 4.8 | | Human Plasma 2 | 57.7 | 1.63 | 2.8 | | Human Plasma 3 | 110 | 1.62 | 1.5 | | Human Plasma 4 | 492 | 4.12 | 0.8 | | Human Plasma 5 | 863 | 9.54 | 1.1 | Table 3: Intermediate Precision Summary | Specimen | Mean (umol/L) | SD (umol/L) | CV (%) | |----------------|---------------|-------------|--------| | AMM-N | 67.9 | 1.61 | 2.4 | | AMM-P | 243 | 4.26 | 1.8 | | Human Plasma 1 | 26.0 | 1.29 | 4.9 | | Human Plasma 2 | 57.7 | 1.72 | 3.0 | {8}------------------------------------------------ | Specimen | Mean ( $\u03bcmol$ /L) | SD ( $\u03bcmol$ /L) | CV (%) | |----------------|------------------------|----------------------|--------| | Human Plasma 3 | 110 | 1.92 | 1.7 | | Human Plasma 4 | 480 | 6.30 | 1.3 | | Human Plasma 5 | 853 | 12.4 | 1.5 | All data passed the predetermined acceptance criteria. #### Analytical Sensitivity 4.2. LoB, LoD, and LoQ were determined according to CLSI EP17-A2. - Limit of Blank (LoB) 4.2.1. Limit of Blank determines the highest observed measurement values for samples free of analyte. For determination of LoB one analyte free sample was measured with three lots in 10-fold determination in 6 runs, distributed over 3 days, on one cobas c 501 analyzer. In total, 60 measurements were obtained per lot. Data analysis is based on determination of the 95th percentile of the 60 measured values #### Limit of Detection (LoD) 4.2.2. The LoD determines the lower limit for samples with analyte. The LoD was determined as the lowest amount of analyte in a sample that can be detected with a 95% probability. For determination of LoD five samples with low-analyte concentration (approximately up to 4 times the LoB) were measured with three lots in two-fold determination in 6 runs, distributed over 3 days, on one cobas c 501 analyzer. In total 60 measurements were obtained per lot. #### Limit of Quantitation (LoQ) 4.2.3. The limit of quantitation (LoQ), according to EP17-A2 is the lowest analyte concentration that can be quantitatively determined with a stated acceptable precision and trueness under stated experimental conditions. {9}------------------------------------------------ A low level sample Set was prepared by diluting 7 human plasma samples with water. The low level sample set was tested in 5 replicates per sample on 5 days, one run per day on one cobas c 501 analyzer. | | Result (µmol/L) | Claim (µmol/L) | |-----------------------------|-----------------|----------------| | Limit of Blank (LoB) | 1.80 | ≤ 10 µmol/L | | Limit of Detection (LoD) | 3.46 | ≤ 10 µmol/L | | Limit of Quantitation (LoQ) | 9.36 | ≤ 10 µmol/L | Table 4: LoB, LoD, and LoQ Experimental Determination All data passed the predetermined acceptance criteria. #### Linearity/Assay Reportable Range 4.3. #### Regression Analysis 4.3.1. The linearity study was conducted to demonstrate that measurements across the claimed measuring range for each parameter are linear. The study was performed according to CLSI guideline EP06-A. A linearity check was performed with first order (linear) regression and then with higher order models (quadratic and cubic). ## Table 5: Linearity Results | Reagent Lot | Linear Regression | |-------------|--------------------------------------------------------------| | 1 | $y = 1.003x - 2.19$<br>correlation coefficient (r²) = 0.9999 | | 2 | $y = 1.002x - 1.30$<br>correlation coefficient (r²) = 1 | | 3 | $y = 1.002x -1.56$<br>correlation coefficient (r²) = 0.9999 | All data passed the predetermined acceptance criteria. {10}------------------------------------------------ #### Endogenous Interferences 4.4. #### Hemolysis/Bilirubin/Lipemia/Albumin/IgG 4.4.1. The effects of interference by hemoglobin, lipemia (Intralipid), Albumin, Immunoglobulin (IgG) and Bilirubin on the NH3L2 test system were determined on the cobas c 501 analyzer using pooled human plasma samples spiked with varying levels of interferent. The resulting sample series (10 dilution steps per sample) were tested in triplicate and the median values used to calculate % recovery, by comparing the measured concentration to the expected concentration (which is the NH3L2 concentration when no interferent was added). | Substance tested | Tested Substance<br>Approximate Concentration | Ammonia concentrations in<br>umol/L | |------------------------|-----------------------------------------------|-------------------------------------| | Hemolysis | Level 1: 114 mg/dL<br>Level 2: 146 mg/dL | Level 1: 36.1<br>Level 2: 91.8 | | Unconjugated Bilirubin | Level 1: 69 mg/dL<br>Level 2: 68 mg/dL | Level 1: 46.4<br>Level 2: 113 | | Conjugated Bilirubin | Level 1: 64 mg/dL<br>Level 2: 64 mg/dL | Level 1: 40.2<br>Level 2: 89.1 | | Lipemia (Intralipid) | Level 1: 764 mg/dL<br>Level 2: 771 mg/dL | Level 1: 51.5<br>Level 2: 92.7 | | Albumin | Level 1: 77.5 g/L<br>Level 2: 77.2 g/L | Level 1: 47.2<br>Level 2: 108 | | Immunoglobulin (IgG) | Level 1: 71.7 g/L<br>Level 2: 71.4 g/L | Level 1: 41.5<br>Level 2: 83.9 | Table 6: Endogenous Interference Results Listed are the highest levels of interferent which passed specification at the analyte concentration levels. All data passed the predetermined acceptance criteria. #### Exogenous Interferences – Drugs 4.5. The purpose of this study was to evaluate drugs for potential interference with NH3L2 assay measured on the cobas c 501 analyzer. Two sample pools, containing a low and high concentration of NH3L2 were used. These sample pools were divided into an appropriate number of aliquots. One aliquot was not spiked with the drugs and it was used as the reference {11}------------------------------------------------ sample for NH3L2 concentration. The NH3L2 concentration in the sample was determined with n = 3 measurements on a cobas c 501 analyzer. The other sample aliquots, with either the high or low NH3L2 concentrations, are spiked with the respective amount of drug. The NH3L2 concentration of the spiked aliquots are determined in triplicate and the mean of the triplicate determinations is compared to the NH3L2 concentration determined for the reference aliquot (mean of n=3). No interference was found at therapeutic concentrations using common drug panels with the exceptions of Cefoxitin, Sulfasalazin, and Temozolomid which were found to interfere. #### 4.6. Sample Matrix Comparison The effect of the presence of anticoagulants on analyte recovery was determined by method comparison, obtained from samples drawn into different types of plasma collection tubes (K2 EDTA and K3 EDTA). For K2 EDTA and K3 EDTA 52 tubes for Lot 1 and 53 tubes for Lot 2 and 53 tubes for Lot 3 were collected and filled completely. Method comparison K2 EDTA versus K3 EDTA were calculated. Slope, Intercept and Correlation were calculated. | Reagent Lot | Regression | Correlation (Pearson(r)) | |-------------|---------------------|--------------------------| | 1 | $y = 1.002x -1.18$ | 1.000 | | 2 | $y = 0.987x - 0.77$ | 1.000 | | 3 | $y = 1.005x - 1.39$ | 1.000 | #### 4.7. Method Comparison to Predicate A total of 112 human plasma samples were tested in singlicate with the AMM test kit of Beckmann Coulter on Beckmann Synchron DxC 800 and the NH3L2 reagent on cobas c 501. The results were calculated using Passing/Bablok, and Linear regression. Regression analysis results: y = 1.001x – 1.90 µmol/L, r = 1.000 {12}------------------------------------------------ #### 5. CLINICAL PERFORMANCE EVALUATION Not applicable. #### 6. ADDITIONAL INFORMATION #### 6.1. Other Devices Marketed with This Assay The Ammonia II assay continues to use: Ammonia/Ethanol/CO2 Calibrator (k031880) Ammonia/Ethanol/CO2 Control Normal (k031880) Ammonia/Ethanol/CO2 Control Abnormal (k031880) #### 7. CONCLUSIONS The submitted information in this premarket notification supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...