AXS Catalyst 7 Distal Access Catheter (068 x 115cm, 068 x 125cm, 068 x 132cm)
Applicant
Stryker
Product Code
NRY · Cardiovascular
Decision Date
Mar 13, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K183464 · Mar 13, 2019
AXS Catalyst 7 Distal Access Catheter (068 x 115cm, 068 x 125cm, 068 x 132cm)
Stryker
Published clinical study articles
A review of published clinical literature regarding predicate devices was used to support the safety, performance, and suitability of the subject device for the proposed indication.
Literature review; Clinical outcomes; Predicate device performance
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Review of published clinical study articles; Literature review
Patients treated with predicate devices for acute ischemic stroke
Not applicable for this study
Clinical outcomes
Indications for Use
The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment.
Device Story
AXS Catalyst 7 Distal Access Catheter is a sterile, single-lumen, variable-stiffness catheter for neurovascular thrombus removal via continuous aspiration. Input: vacuum pressure from external Medela Dominant Flex Pump (-68 kPa). Operation: physician advances catheter through vasculature under fluoroscopic guidance over a guidewire to the occlusion site; aspiration is applied via the catheter lumen to remove the clot. Output: thrombus removal. Used in clinical settings by interventionalists. Benefits: restores blood flow in acute ischemic stroke patients. Device includes RHV, Tuohy Borst valve, and peel-away introducer sheaths.
Clinical Evidence
No clinical study conducted. Evidence consists of bench testing (dimensional, tip configuration, surface integrity, buckling, lubricity, particulate, tensile, leak resistance, torsional bond, flexural fatigue, kink radius, aspiration flow rate) and animal testing (in-vivo safety/navigation/aspiration). Published clinical literature on predicate devices was reviewed to support the line extension.
Indicated for revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (internal carotid, middle cerebral M1/M2, basilar, vertebral arteries) within 8 hours of symptom onset. Candidates include patients ineligible for or failed IV t-PA.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Penumbra System ACE 68 Reperfusion Catheter (K161640)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 13, 2019
Stryker Neurovascular Germaine Fu. PhD Associate Project Manager, Regulatory Affairs 47900 Bayside Parkway Fremont, California 94538
Re: K183464
Trade/Device Name: AXS Universal Aspiration System (AXS Catalyst 7 Distal Access Catheter) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: December 13, 2018 Received: December 14, 2018
Dear Germaine Fu, PhD:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed
by Xiaolin Zheng
Date: 2019.03.13
18:49:44 -04'00'
Lin Zheng-s
Carlos L. Peña. PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183464
#### Device Name
AXS Universal Aspiration System (AXS Catalyst 7 Distal Access Catheter)
Indications for Use (Describe)
The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary K183464
#### Introduction:
According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for a determination of substantial equivalence.
#### Submitter Name, Address and Content:
| Submitter: | Stryker Neurovascular<br>47900 Bayside Parkway<br>Fremont, CA 94538-6515<br>(FDA Registration Number: 3008853977) |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Germaine Fu<br>Associate Project Manager, Regulatory Affairs<br>Phone: 510-413-2862<br>Fax: 510-413-2588<br>Email: germaine.fu@stryker.com |
| Date Prepared: | December 13, 2018 |
| Device Name and Classification: | |
| Trade/Proprietary Name: | AXS Universal Aspiration System (AXS Catalyst 7 Distal Access<br>Catheter) |
| Common Name: | Catheter, Thrombus Retriever |
| Classification Name: | Percutaneous Catheter, 21CFR 870.1250 – Class II |
| Product Code: | NRY |
{4}------------------------------------------------
| Legally Marketed<br>Predicate<br>Device(s): | Primary Predicate Device | Additional Predicate Device |
|---------------------------------------------|-----------------------------------------------------|-------------------------------------------------------------|
| | AXS Catalyst Distal Access<br>Catheter<br>(K173841) | Penumbra System ACE 68<br>Reperfusion Catheter<br>(K161640) |
#### Device Description
The AXS Catalyst 7 Distal Access Catheter is a sterile, single lumen, variable stiffness catheter designed for use in the removal of thrombus from the neuro vasculature using continuous aspiration. The catheter shaft has a hydrophilic coating to reduce friction during use. The catheter includes a radiopaque marker on the distal end for angiographic visualization and a luer hub on the proximal end allowing attachments for flushing and aspiration. It is packaged with a Rotating Hemostasis Valve (RHV), Tuohy Borst Valve with Sideport, and two peel-away introducer sheaths. The RHV and Tuohy Borst valve with sideport are used for flushing, insertion of catheters, and aspiration. The peel away introducer sheaths are designed to protect the distal tip of the catheter during insertion into the RHV or Tuohy Borst.
When used as part of the aspiration system, the AXS Catalyst Distal Access Catheter requires a minimum vacuum pressure of -68 kPa [-20.08 in Hg] from an external device (Medela Dominant Flex Pump). Bench and Animal testing utilizing the Medela Dominant Flex Pump and Suction Jar, AXS Universal Aspiration Tubing, and AXS Universal Liner Set were conducted to successfully demonstrate aspiration.
## Indications for Use
The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment.
## Technological Characteristics and Product Feature Comparison
Stryker Neurovascular has demonstrated the AXS Catalyst 7 Distal Access Catheter when used as part of the aspiration system is substantially equivalent to the Primary Predicate device (K173841) and Additional Predicate device (K161640) based on the same or similar materials, similar design, and the same fundamental operating principles. A comparison of the Subject device with the Predicate devices is summarized in Table 1 below.
{5}------------------------------------------------
| Table 1: Product Feature Comparison of Subject Device to Predicate Devices | | | |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Detail | Submission Subject Device | Primary Predicate Device | Additional Predicate Device |
| | AXS Catalyst 7 Distal Access Catheter | AXS Catalyst 5 and 6 Distal Access Catheter | Penumbra System ACE 68 Reperfusion Catheter |
| Manufacturer | Stryker Neurovascular | Stryker Neurovascular | Penumbra, Inc. |
| 510(k) Number | K183464 | K173841 | K161640 |
| Device Trade Name | AXS Catalyst® 7 Distal Access Catheter | AXS Catalyst® 5 and 6 Distal Access Catheter | Penumbra System ACE 68 Reperfusion Catheter |
| Regulation Number | 21 CFR 870.1250 | Same | Same |
| Regulation Name | Percutaneous Catheter | Same | Same |
| Classification | II | Same | Same |
| Product Code | NRY | Same | Same |
| Intended Use/ Indication for Use | The AXS Catalyst Distal Access Catheter as part of the AXS Universal Aspiration System is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 | Same | As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, |
| Table 1: Product Feature Comparison of Subject Device to Predicate Devices | | | |
| Detail | Submission Subject Device | Primary Predicate Device | Additional Predicate Device |
| | AXS Catalyst 7 Distal Access Catheter | AXS Catalyst 5 and 6 Distal Access Catheter | Penumbra System ACE 68 Reperfusion Catheter |
| | hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment. | | and vertebral arteries) within 8 hours of symptom onset. |
| Device Description | The AXS Universal Aspiration System is composed of the following components:<br>• AXS Catalyst Distal Access Catheter<br>• AXS Universal Aspiration Tubing<br>• AXS Universal Liner Set<br>• Medela Dominant Flex Pump<br><br>The AXS Universal Aspiration System is designed to remove thrombus from the neurovasculature using continuous aspiration. | Same | The Penumbra System ACE 68 Reperfusion Catheter is a component to the currently available Penumbra System. The ACE 68 Reperfusion Catheter is used with the Aspiration Pump to aspirate thrombus from an occluded vessel in the neurovasculature. ACE 68 is provided sterile, non-pyrogenic, and intended for single use only. |
| Detail | Submission Subject<br>Device | Primary Predicate<br>Device | Additional Predicate<br>Device |
| | AXS Catalyst 7 Distal<br>Access Catheter | AXS Catalyst 5 and 6<br>Distal Access Catheter | Penumbra System<br>ACE 68 Reperfusion<br>Catheter |
| Accessory<br>Devices<br>Provided (not<br>in direct<br>contact with<br>patient) | The AXS Catalyst<br>Distal Access Catheter<br>delivers aspiration<br>from the Medela<br>Dominant Flex Pump<br>directly to the site of<br>the occlusion to<br>remove the clot. The<br>AXS Catalyst Distal<br>Access Catheter is a<br>sterile, single lumen,<br>variable stiffness<br>catheter. | Same | Peelable sheath,<br>rotating hemostasis<br>valve, shaping<br>mandrel |
| Materials | | | |
| Outer Jacket | Pebax with Nylon,<br>Tecoflex | Same | Vestamid, Pebax,<br>Tecoflex, Pellethane |
| Reinforcement | Nitinol wire and<br>polymer fiber | Stainless Steel,<br>Nitinol wire, and<br>polymer fiber | Stainless steel and<br>Nitinol wire |
| Strain Relief | Thermoplastic rubber<br>(Polyolefin) | Thermoplastic rubber<br>(Santoprene) | Stainless steel, 304 |
| Inner Layer | PTFE | Same | Same |
| Catheter Hub | Nylon | Same | Proximal Hub<br>Grilamid<br>(TR55-LX) |
| Table 1: Product Feature Comparison of Subject Device to Predicate Devices | | | |
| Detail | Submission Subject<br>Device | Primary Predicate<br>Device | Additional Predicate<br>Device |
| | AXS Catalyst 7 Distal<br>Access Catheter | AXS Catalyst 5 and 6<br>Distal Access Catheter | Penumbra System<br>ACE 68 Reperfusion |
| | | | Catheter |
| Marker Band | Platinum/Iridium | Same | Same |
| Adhesive | Cyanoacrylate | Same | Not available |
| Outer Jacket<br>Coating | Hydrophilic Coating | Same | SRDX Harmony<br>(proprietary) coating |
| Features | | | |
| Labeled Shaft | Distal OD: | Distal OD: | Distal OD: |
| Outer<br>Diameter | 6.2F (0.082 in.) | 5.3F (0.0696 in.) | 0.084 in. |
| | Proximal OD: | 5.4F (0.0709 in.) | Proximal OD: |
| | 6.3F (0.0825 in.) | | 0.084 in. |
| | | Proximal OD: | |
| | | 5.6F (0.0735 in.) | |
| | | 6.0F (0.0787 in.) | |
| Effective | 115cm | CAT 5:<br>115cm, 132cm | 115cm |
| Lengths | 125cm<br>132cm | | 120cm |
| | | CAT 6: 132cm | 125cm |
| | | | 127cm |
| | | | 132cm |
| Distal ID | 0.068 in. | CAT 5: 0.058 in.<br>CAT 6: 0.060 in. | 0.068 in. |
| Proximal ID | 0.068 in. | CAT 5: 0.058 in.<br>CAT 6: 0.060 in. | 0.068 in. |
| Packaging<br>Materials | Polyethylene Tube<br>and HDPE<br>Packaging Card | Same | Polyester/<br>Polyethylene/Tyvek and…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.