K183400 · Elucent Medical, Inc. · NEU · Mar 22, 2019 · General, Plastic Surgery
Device Facts
Record ID
K183400
Device Name
EnVisio Navigation Sytem
Applicant
Elucent Medical, Inc.
Product Code
NEU · General, Plastic Surgery
Decision Date
Mar 22, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Indications for Use
The SmartClip Soft Tissue Marker is intended to be placed percutaneously in soft tissue to permanently mark a biopsy site or a soft tissue site intended for surgical removal. Using imaging guidance (such as ultrasound, MRI, or radiography) or aided by non-imaging guidance (EnVisio Navigation System); the SmartClip is located and surgically removed with the target tissue. The EnVisio Navigation System is intended only for the non-imaging detection and localization (by navigation) of the SmartClip that has been implanted in a soft tissue biopsy site or a soft tissue site intended for surgical removal.
Device Story
EnVisio Navigation System detects and provides real-time 3D navigation to implanted SmartClip soft tissue markers during surgery. System components include console, heads-up display, patient pad, and foot pedal; sterile single-use navigator and calibration disk. Device acquires and analyzes electromagnetic signatures emitted by up to three unique SmartClips. Used in operating rooms by surgeons to locate and excise marked soft tissue. Provides real-time distance, depth, and direction data from electrosurgical tool tip to implanted markers. Visual and audible feedback assists surgeon in precise localization, facilitating accurate tissue removal and potentially improving surgical outcomes.
Clinical Evidence
Bench testing only. Performance testing included system design verification, navigation performance verification and validation, EMC/safety testing, packaging/shelf life, and sterilization validation. No clinical data presented.
Technological Characteristics
Electromagnetic wave technology for detection and navigation. System components: console, heads-up display, patient pad, foot pedal. Sterile single-use components: navigator, calibration disk. Sterilization: Ethylene oxide. Connectivity: Not specified. Software: Embedded navigation algorithm.
Indications for Use
Indicated for adult patients requiring percutaneous placement of a soft tissue marker at a biopsy or surgical site for subsequent non-imaging detection and surgical localization.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
Cianna Medical SAVI Scout Reflector and SAVI Scout System (K181007)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 22, 2019
Elucent Medical, Inc Ms. Debra Kridner Vice President of Regulatory and Quality 7480 Flying Cloud Drive, Suite 110 Eden Prairie, Minnesota 55344
Re: K183400
Trade/Device Name: EnVisio Navigation System Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: February 20, 2019 Received: February 22, 2019
Dear Ms. Kridner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by David Krause David Krause -S Date: 2019.03.22 14:41:33 -04'00' for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183400
Device Name EnVisio™ Navigation System
#### Indications for Use (Describe)
The SmartClip Soft Tissue Marker is intended to be placed percutaneously in soft tissue to permanently mark a biopsy site or a soft tissue site intended for surgical removal. Using imaging guidance (such as ultrasound, MRI, or radiography) or aided by non-imaging guidance (EnVisio Navigation System); the SmartClip is located and surgically removed with the target tissue. The EnVisio Navigation System is intended only for the non-imaging detection and localization (by navigation) of the SmartClip that has been implanted in a soft tissue biopsy site or a soft tissue site intended for surgical removal.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <svg height="15" width="15"> <rect height="15" style="fill:lightgrey;stroke-width:1;stroke:rgb(0,0,0)" width="15"></rect> <path d="M2,2 L13,13 M13,2 L2,13" stroke="black" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;"> <svg height="15" width="15"> <rect height="15" style="fill:white;stroke-width:1;stroke:rgb(0,0,0)" width="15"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
| Date Prepared: | March 20, 2019 |
|-------------------------|---------------------------------------------------------------------------------------------|
| 1. APPLICANT | Elucent Medical, Inc.<br>7480 Flying Cloud Drive, Suite 110<br>Eden Prairie, MN 55344 USA |
| Official Correspondent: | Debra Kridner<br>Tel: (844) 417 1700 / Fax: (952) 314-7105<br>debra.kridner@elucent.com |
| 2. DEVICE | |
| Trade Name: | EnVisio™ Navigation System |
| 510(k) | K183400 |
| Common Name: | Implanted, Radiographic, Tissue Marker |
| Classification Number: | 21 CFR 878.4300 |
| Classification Name: | Implantable Clip |
| Device Class: | Class II |
| Product Code: | NEU |
| Review Panel: | General and Plastic Surgery |
## 3. PREDICATE DEVICE (PRIMARY)
| Trade Name: | Cianna Medical SAVI Scout Reflector and SAVI Scout System |
|------------------------|-----------------------------------------------------------|
| 510(k) | K181007 |
| Common Name: | Implanted, Radiographic, Tissue Marker |
| Classification Number: | 21 CFR 878.4300 |
| Classification Name: | Implantable Clip |
| Device Class: | Class II |
| Product Code: | NEU |
| Review Panel: | General and Plastic Surgery |
### 4. DEVICE DESCRIPTION (for the device subject to this 510(k) notification)
The EnVisio™ Navigation System is a medical device that detects the presence of the SmartClip™ Soft Tissue Marker(s) and provides real-time three-dimensional navigation during surgery. The EnVisio™ Navigation System acquires and analyzes electromagnetic signatures that display the relative distance, depth, and direction from the tip of the electrosurgical tool to the implanted SmartClips™ within the surgeon's field of view
{5}------------------------------------------------
during the procedure. The EnVisio™ Navigation System will be utilized in an operating room where surgical excision of soft tissue is performed. The EnVisio™ system can detect and provide real-time three-dimensional navigation for up to three separate SmartClips™ each programmed to emit a unique electromagnetic signature.
The EnVisio™ Navigation System equipment components are the Console, Heads Up Display, Patient Pad and Foot Pedal. The EnVisio™ Navigation System Navigator and Calibration Disk are sterile, non-patient contacting single use devices which are available separately.
### 5. INDICATIONS FOR USE (for the Device Subject to this 510(k) Notification)
The SmartClip™ Soft Tissue Marker is intended to be placed percutaneously in soft tissue to permanently mark a biopsy site or a soft tissue site intended for surgical removal. Using imaging guidance (such as ultrasound, MRI, or radiography) or aided by non-imaging guidance (EnVisio™ Navigation System) the SmartClip™ is located and surgically removed with the target tissue. The EnVisio™ Navigation System is intended only for the non-imaging detection and localization (by navigation) of the SmartClip™ that has been implanted in a soft tissue biopsy site or a soft tissue site intended for surgical removal.
| Comparison of<br>Subject Device<br>to Predicate<br>Device | Subject Device<br>EnVisio™ Navigation System<br>(K183400) | Predicate Device:<br>Cianna Medical SAVI Scout Reflector<br>and SAVI Scout System (K181007) |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The SmartClip™ Soft Tissue Marker is<br>intended to be placed percutaneously in<br>soft tissue to permanently mark a biopsy<br>site or a soft tissue site intended for<br>surgical removal. Using imaging guidance<br>(such as ultrasound, MRI, or radiography)<br>or aided by non-imaging guidance<br>(EnVisio™ Navigation System) the<br>SmartClip™ is located and surgically<br>removed with the target tissue. The<br>EnVisio™ Navigation System is intended<br>only for the non-imaging detection and<br>localization (by navigation) of the<br>SmartClip™ that has been implanted in a<br>soft tissue biopsy site or a soft tissue site<br>intended for surgical removal. | The SAVI Scout Reflector is intended to be<br>placed percutaneously in soft tissue to mark<br>(>30 days) a biopsy site or a soft tissue site<br>intended for surgical removal. Using<br>imaging guidance (such as ultrasound, MRI,<br>or radiography) or aided by non-imaging<br>guidance (SAVI Scout System) the SAVI<br>Scout Reflector is located and surgically<br>removed with the target tissue. The SAVI<br>Scout System is intended only for the non-<br>imaging detection and localization of the<br>SAVI Scout Reflector that has been<br>implanted in a soft tissue biopsy site or a soft<br>tissue site intended for surgical removal. |
| Intended Use | Detect and Navigate to the implanted<br>SmartClip - Soft Tissue Marker | Detect and Navigate to the implanted<br>Reflector - Soft Tissue Marker |
| Use<br>Environment | Intra-operative, guided surgical procedure | Intra-operative, guided surgical procedure |
| Anatomical Site | Soft Tissue | Soft Tissue |
| Patient<br>Population | Adult | Adult |
| Comparison of Subject<br>Device to Predicate<br>Device | Subject Device<br>EnVisio™ Navigation System<br>(K183400) | Predicate Device:<br>Cianna Medical SAVI Scout<br>Reflector and SAVI Scout System<br>(K181007) |
| Manufacturer | Elucent Medical, Inc. | Cianna Medical, Inc. |
| Device Name | EnVisio™ Navigation System | SAVI Scout Reflector and Scout System |
| 510(k) Number | K183400 | K181007 |
| Principles of Operation | Electromagnetic wave technology to<br>detect and navigate to the soft tissue<br>marker | Electromagnetic wave technology to<br>detect and navigate to the soft tissue<br>marker |
| Intraoperative<br>Localization Energy | Electromagnetic, passive marker | Electromagnetic, passive marker |
| Audible Location<br>Indication | Yes | Yes |
| Visual Location<br>Indication | Yes | Yes |
| Visualization<br>Compatibility | Using imaging guidance (such as<br>ultrasound, MRI, or radiography) or<br>aided by non-imaging guidance<br>(EnVisio Navigation System) | Using imaging guidance (such as<br>ultrasound, MRI, or radiography) or<br>aided by non-imaging guidance (SAVI<br>Scout System) |
| System Components<br>(Equipment) | Console<br>• Heads-Up Display<br>• Patient Pad<br>• Foot Pedal (optional)<br>• Cart | • Console |
| System Components<br>(Sterile) | Navigator<br>• Calibration Disk | • Hand piece<br>• Sterile Sheath applied to<br>nonsterile Handpiece |
| Sterilization Method | Ethylene oxide | Ethylene oxide |
| Implanted Soft Tissue<br>Marker | Yes - SmartClip | Yes - Reflector |
| Biocompatibility | Not applicable for Navigator and<br>Calibration Disk - no direct or<br>indirect patient contact | Yes -Handpiece has direct patient<br>contact |
### Indications for Use Comparison:
{6}------------------------------------------------
## 6. TECHNOLOGICAL AND PERFORMANCE CHARACTERISTICS COMPARISON
### Performance Testing:
Performance testing was conducted to evaluate and characterize the performance of the EnVisio Navigation System. The following performance testing was conducted on the EnVisio Navigation System to support a determination of substantial equivalence to the predicate device.
- System Design Verification
- Navigation Performance Verification ●
- Navigation Performance Validation ●
{7}------------------------------------------------
- EMC/Safety Testing ●
- Packaging/Shelf Life ●
- . Sterilization Validation
### Conclusion:
The EnVisio Navigation System met all specified criteria and did not raise new safety or performance questions.
### Basis for Determination of Substantial Equivalence:
The Indications for Use, Intended Use, Fundamental Scientific Technology and Principles of Operation for the EnVisio Navigation System are the same as those described for the predicate device.
In summary, the EnVisio Navigation System has the following similarities to the predicate device which has previously received 510(k) clearance:
- Has substantially equivalent indications for use ●
- Has substantially equivalent intended use ●
- Used in substantially equivalent anatomical sites ●
- Uses substantially equivalent technological characteristics ●
- Uses substantially equivalent principles of operation
- Uses substantially equivalent sterilization methodology ●
Therefore, the conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device predicate. The EnVisio Navigation System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.