K183385 · Angiodynamics · OAB · Jun 18, 2019 · General, Plastic Surgery
Device Facts
Record ID
K183385
Device Name
NanoKnife System
Applicant
Angiodynamics
Product Code
OAB · General, Plastic Surgery
Decision Date
Jun 18, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NanoKnife System with six outputs is indicated for the surgical ablation of soft tissue.
Device Story
NanoKnife System is a software-controlled low-energy direct-current (LEDC) generator for surgical soft tissue ablation. System uses up to six electrode probes placed in an array within tissue. Clinician sets parameters including volts/cm, pulse length, pulse count, and timing mode (90PPM or ECG synchronization). System delivers scheduled voltage pulses between matched probe pairs via foot-pedal activation. RFID technology in probe connectors ensures use of compatible, non-expired probes. Device used in clinical settings by physicians. Output allows for controlled tissue ablation. Modifications from predicate include updated electronic components, firmware, and user interface, plus improved EMC performance per IEC 60601-1-2 4th edition.
Clinical Evidence
Bench testing performed per consensus standards (IEC 60601-1, IEC 60601-1-2, ISO 10993, ISO 11135). Animal study conducted to characterize performance and thermal effects on tissue. Comparison of mean ablation volume between subject (NK 3.0) and predicate (NK 2.2) yielded p-value of 0.670, supporting substantial equivalence.
Technological Characteristics
LEDC generator; software-controlled. Six-output architecture. RFID-enabled probe connectivity. Materials biocompatible per ISO 10993-1:2009/(R)2013. Sterilization via validated cycles per ISO 11135. Compliance with IEC 60601-1-2 4th edition for EMC. Firmware/UI updated for component obsolescence.
Indications for Use
Indicated for surgical ablation of soft tissue in patients requiring such procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
June 18, 2019
Angiodynamics Troy Roberts Director, Regulatory Affairs 26 Forest Street Marlborough, Massachusetts 01752
Re: K183385
Trade/Device Name: NanoKnife System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: OAB Dated: May 22, 2019 Received: May 23, 2019
Dear Troy Roberts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K183385
Device Name NanoKnife System
Indications for Use (Describe)
The NanoKnife System with six outputs is indicated for the surgical ablation of soft tissue.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary – K183385
The following information is provided in accordance with 21 CFR 807.92 for the Premarket 510(k) Summary:
## Submitter Information
| Company: | Angiodynamics<br>26 Forest Street<br>Marlborough, MA 01752 USA<br>Telephone: 508-930-0266<br>Fax: 508-658-7934<br>troberts@angiodynamics.com |
|----------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Troy Roberts<br>Director Regulatory Affairs<br>Angiodynamics<br>26 Forest Street<br>Marlborough, MA 01752 USA<br>Telephone: 508-930-0266<br>Fax: 508-658-7934<br>troberts@angiodynamics.com |
Date Summary Prepared:
June 18, 2019
Name of the Device
| Trade Name: | NanoKnife System |
|----------------------|---------------------------------------------------|
| Common Name: | Low energy direct current ablation device |
| Classification Name: | Low Energy Direct Current Thermal Ablation System |
| Review Panel: | General & Plastic Surgery (SU) |
| Regulation: | 878.4400 |
| Class: | Class II |
| Product Code: | OAB |
#### Equivalence Claimed to Predicate Device
The NanoKnife System is equivalent to the NanoKnife System Version 2.2.1 (K150089), manufactured by Angiodynamics, Inc.
{5}------------------------------------------------
## Device Description
The subject device is substantially equivalent to the predicate device. The NanoKnife System is a software-controlled low-energy direct-current (LEDC) generator which surgically ablates soft tissue. Included for use with the system are the NanoKnife Single Electrode Probes and optional Probe Spacer. With the NanoKnife System, a voltage applied between pairs of probes in a series of pulses. The waveform of the voltage is adjustable as determined by clinician-chosen parameters. These parameters include volts/cm, pulse length, number of pulses to be delivered between electrode pairs, the distance between probes, and the timing mode (90PPM or ECG synchronization). Up to six probes may be placed in an array within the tissue. The probes of the array are matched as pairs by the system. When probes are activated via a foot-pedal, the scheduled voltage is delivered to tissue between subsequent pairs of probes. Soft tissue between the probes is ablated.
#### Intended Use
The intended use of the predicate device and the subject device are identical. The intended use is for the surgical ablation of soft tissue.
#### Indications for Use
The NanoKnife System with six outputs is indicated for the surgical ablation of soft tissue.
#### Summary of Similarities and Differences in Technological Characteristics and Performance between the Predicate and Subject Device
The principle of operation of the subject device is identical to the predicate device. For the subject device, the available ablation parameters (pulse amplitude, pulse length, maximum energy per pulse, etc.) are identical to the Predicate device. Two probe configurations options that were available in the predicate device NanoKnife 2.2.1 were deleted in the subject device: bipolar probe; star shaped six probe array.
The physical design of the device has been modified, primarily to replace electronic components which have become obsolete, and to ensure conformance with the 4th edition of IEC 60601-1-2. In addition, device firmware was updated to support the new components, and the User Interface software was updated. Except for improving the EMC performance to comply with the new revision of IEC 60601-1-2, these modifications did not alter the performance of the device.
The use of RFID to read tags within the probe connectors was introduced. The RFID function assures that only compatible NanoKnife Probes which have not expired are used with the NanoKnife 3.0 system.
#### Biocompatibility
The patient contact materials are identical to the predicate device, however additional biocompatibility testing in accordance with ISO 10993-1:2009/(R) 2013 was performed because
{6}------------------------------------------------
of changes that have occurred in the consensus standard since the date that the original testing included in the Predicate Device submission (K150089) was performed.
## Sterilization
The design of the sterilized disposable probes and probe spacer, the sterile packaging and the sterilization cycle are identical between the subject device and the predicate device, however additional testing was performed because of changes that have occurred in the relevant consensus standards (ISO 11135 2nd Edition, ISO 10993-7 2nd Edition, and ASTM F1980-07) since the date that the original testing included in the Predicate Device submission (K150089) was performed.
#### Safety Data
Device safety was evaluated against the following published consensus standards and FDA guidance documents.
- ES60601-1:2005/(R)2012 And A1:2012 ●
- IEC 60601-1-2 Edition 4.0 2014-02 .
- IEC 60601-1-6 Edition 3.1 2013-10 .
- ISO 10993-1:2009/(R)2013 .
- ISO 10993-7:2008-10-15 Second edition .
#### Performance Data
Device performance was evaluated against the following published consensus standards and FDA guidance documents.
- ANSI C63.27:2017 .
- ASTM D4169 .
- ASTM F1886-98 ●
- ASTM F1980-07 ●
- EN 301 489-1 ●
- EN 301 489-3 ●
- IEC 62304 Edition 1.1 2015-06 ●
- FDA guidance document: Premarket Notification (510(k)) Submissions for . Electrosurgical Devices for General Surgery, dated August 15, 2016
To address the recommendations contained in the FDA Guidance document "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", specifically section E (System Testing), subsection 1 (Thermal Effects on Tissue), an animal study was conducted. The purpose of the testing is to characterize how the device performs over the expected range of operating conditions. The overall performance of the subject device was similar to the performance of the predicate device. There was significant variability observed in ablation volume measurements, however this variability can be attributed to specific experimental factors that are not related to the version of the
{7}------------------------------------------------
NanonKnife system under test. The P-value (0.670) for the comparison of mean ablation volume between the Predicate device NK 2.2 and the Subject device NK 3.0, supports the substantial equivalence of the two systems.
#### Conclusion
The subject NanoKnife System is substantially equivalent to the predicate NanoKnife System (K150089). The device modifications included in this 510(k) Premarket Notification do not affect the cleared intended use or the fundamental technological characteristics of the predicate. The subject device is as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.