← Product Code [CBK](/productcode/CBK) · K183364

# Bellavista (K183364)

_Imtmedical AG · CBK · Sep 13, 2019 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K183364

## Device Facts

- **Applicant:** Imtmedical AG
- **Product Code:** [CBK](/productcode/CBK.md)
- **Decision Date:** Sep 13, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5895
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The bellavista 1000/1000e ventilator is intended to provide positive pressure ventilatory support to adult and pediatric patients and optionally infant and neonatal patients by qualified, trained personnel under the direction of a physician. Environment of use: hospitals, sub-acute care facilities and intra-hospital transfer When used on neonatal patients: The environment of use is the Neonatal Intensive Care Unit (NICU).

## Device Story

Electronically controlled pneumatic ventilator; provides positive pressure ventilatory support. Inputs: user-defined parameters via touchscreen; proximal flow sensor readings. Operation: turbine-based pneumatic system delivers controlled gas mixture to patient. Output: respiratory gas delivery; ventilation monitoring. Used in hospitals, sub-acute facilities, and intra-hospital transport by trained personnel. Healthcare providers use output to manage patient ventilation; supports respiratory function in adult, pediatric, infant, and neonatal populations.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing, comparison of ventilation modes, operating ranges, and waveforms against predicate and reference devices.

## Technological Characteristics

Turbine-based pneumatic ventilation system; AC/DC/internal battery power. Features proximal flow sensor. Added SpO2 and EtCO2 monitoring capabilities. Connectivity includes touchscreen interface. System is electronically controlled.

## Regulatory Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

## Predicate Devices

- bellavista 1000 ([K163127](/device/K163127.md))

## Reference Devices

- Hamilton C3 ([K161450](/device/K161450.md))
- Dräger V500 ([K093633](/device/K093633.md))
- CareFusion Avea ([K103211](/device/K103211.md))
- Pulmonetic Systems Palmtop Ventilator PTV-10 ([K070594](/device/K070594.md))
- Vyaire SiPAP ([K031745](/device/K031745.md))
- Maquet Servo-u ([K151814](/device/K151814.md))
- Puritan Bennett 840 ([K151252](/device/K151252.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 13, 2019

Imtmedical AG Colleen Watson Director, Regulatory Affairs 26125 N. Riverwoods Blvd. Mettawa, IL 60045 USA

Re: K183364

Trade/Device Name: bellavista 1000 and bellavista 1000e Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK Dated: August 13, 2019 Received: August 15, 2019

Dear Colleen Watson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

James Lee Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K183364

Device Name

bellavista 1000 and bellavista 1000e

#### Indications for Use (Describe)

The bellavista 1000/1000e ventilator is intended to provide positive pressure ventilatory support to adult and pediatric patients and optionally infant and neonatal patients by qualified, trained personnel under the direction of a physician.

Environment of use: hospitals, sub-acute care facilities and intra-hospital transfer

When used on neonatal patients: The environment of use is the Neonatal Intensive Care Unit (NICU).

| Type of Use (Select one or both, as applicable)                                                 |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label>             |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Vyaire Medical. The word "Vyaire" is in a dark blue color, with the "V" and "y" connected. The word "MEDICAL" is in a light blue color and is located below the word "Vyaire".

### 510(k) Summary

This summary of 510(k) information is being submitted in accordance with the requirements of 21 CFR 807.92

The assigned 510(k) number is:

| Submitted by:                         | imtmedical AG<br>Gewerbestrasse 8<br>9470 Buchs SG Switzerland                                                |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Establishment Registration<br>Number: | 3004553423                                                                                                    |
| Contact Person:                       | Colleen Watson<br>Director, Regulatory Affairs<br>Phone: (872) 757-0070<br>e-mail: colleen.okeeffe@vyaire.com |
| Date Summary Prepared:                | August 12, 2019                                                                                               |
| Reason for Premarket<br>Notification: | New Device                                                                                                    |
| Trade Name:                           | bellavista 1000 and bellavista 1000e                                                                          |
| Common/Usual Name:                    | Ventilator, continuous, facility use                                                                          |
| Product Code:                         | CBK                                                                                                           |
| Regulatory Class:                     | Class II                                                                                                      |
| Predicate Device:                     | bellavista 1000 (K163127)                                                                                     |

vyaire.com

Vyaire Medical, Inc.
26125 N. Riverwoods Blvd. Mettawa, IL 60045 USA

Vyaire Medical, Inc. 22745 Savi Ranch Pkwy Yorba Linda, CA 92887 USA

Vyaire Medical Oy Kuortaneenkatu 2 FI-00510 Helsinki Finland

Vyaire Medical GmbH Leibnizstrasse 7 97204 Hoechberg Germany

{4}------------------------------------------------

#### Reference devices:

Multiple reference devices have been used for the modifications to support substantial equivalence.

| Manufacturer / Tradename                        | 510(k) Number | Product Code |
|-------------------------------------------------|---------------|--------------|
| Hamilton C3                                     | K161450       | CBK, DQA     |
| Dräger V500                                     | K093633       | CBK          |
| CareFusion Avea                                 | K103211       | CBK          |
| Pulmonetic Systems Palmtop<br>Ventilator PTV-10 | K070594       | CBK          |
| Vyaire SiPAP                                    | K031745       | CBK          |
| Maquet Servo-u                                  | K151814       | CBK          |
| Puritan Bennett 840                             | K151252       | CBK          |

#### Device Description:

bellavista is an electronically controlled pneumatic ventilation system. It is powered by AC or DC and also provided with an internal battery. The bellavista pneumatic system supplies respiratory gas whilst the electrical systems control the pneumatics and provides the power supply.

The user can enter values or parameters in the bellavista via the touch screen. These inputs entail instructions for bellavista's pneumatic system to ventilate the patient with a precisely controlled gas mixture. bellavista gathers readings from the proximal flow sensor within the ventilator.

#### Indications for Use:

The bellavista 1000/1000e ventilator is intended to provide positive pressure ventilatory support to adult and pediatric patients and optionally infant and neonatal patients by qualified, trained personnel under the direction of a physician.

Environment of use: hospitals, sub-acute care facilities and intra-hospital transfer

When used on neonatal patients: The environment of use is the Neonatal Intensive Care Unit (NICU).

#### Comparison of bellavista 1000/1000e to the predicate:

We present in the following tables a comparison of the subject device compared to the predicate and reference devices.

{5}------------------------------------------------

|                    | Subject bellavista<br>1000 / bellavista<br>1000e                                                                                                                                                       | Predicate bellavista<br>1000 (K163127)                                                                                                                         | Remarks, Substantial equivalence                                                                                                                                                                                                                                                                       |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product code       | CBK                                                                                                                                                                                                    | CBK                                                                                                                                                            | The same classification                                                                                                                                                                                                                                                                                |
| Intended use       | The bellavista<br>1000/1000e ventilator<br>is intended to provide<br>positive pressure<br>ventilatory support to<br>adult and pediatric<br>patients and<br>optionally infant and<br>neonatal patients. | The bellavista 1000<br>ventilator is intended<br>to provide positive<br>pressure ventilator<br>support to adults and<br>pediatrics (IBW<br>greater than 6 kg). | Neonates are included in the<br>intended patient population of<br>the reference device Hamilton C3<br>(K161450)                                                                                                                                                                                        |
| Environment of use | Environment of use:<br>hospitals, sub-acute<br>care facilities and<br>intra-hospital transfer<br><br>When used on<br>neonatal patients: the<br>environment of use is<br>the NICU                       | Environment of use:<br>hospitals, sub-acute<br>care facilities and<br>intra-hospital transfer                                                                  | Both devices are<br>substantially equivalent for<br>environment of use for adult and<br>pediatric patient population.<br><br>The environment of use for neonatal<br>and infant patients is substantially<br>equivalent to the environment of use<br>of the reference device, Hamilton C3<br>(K161450). |
| Intended users     | It is intended for use<br>by qualified, trained<br>personnel under the<br>direction of a<br>physician.                                                                                                 | It is intended for use<br>by qualified, trained<br>personnel under the<br>direction of a<br>physician.                                                         | Both devices are<br>substantially equivalent                                                                                                                                                                                                                                                           |
| Patient population | Adults, pediatrics,<br>infants and neonates                                                                                                                                                            | Adults and pediatrics<br>(IBW greater than 6<br>kg)                                                                                                            | Neonates and infants are included in<br>the intended patient population of<br>the reference device Hamilton C3<br>(K161450)                                                                                                                                                                            |

## Table 1: Comparison of Indications for Use

{6}------------------------------------------------

# Comparison of Technological Characteristics

We have added SpO2 and EtCO2 monitoring to the device.

# Discussion of the Differences:

## Indications for Use

Indications for use are to provide positive pressure ventilator support to adults, pediatrics and neonates.

# Technology

The technology of a turbine based portable ventilator is identical to the predicate.

# Environment of Use

The environment of use – hospital and intra-hospital transport is equal to the predicate.

# Patient Population

- The patient population of adults and pediatrics is equal to the predicate. .
- . The patient population of neonates is similar to the reference device Hamilton C3 (K161450)

# Ventilation Modes

# Adult/Pediatric Ventilation Modes

Ventilation modes that are equivalent and not changed between subject and predicate device bellavista 1000 are:

The following ventilation modes have been cleared in the predicate device (K163127):

- СРАР ●
- PCV
- P-A/C
- PC-SIMV
- bsv
- ഗ
- S/T ●
- T
- beLevel
- APRV VCV
- V-A/C
- VC-SIMV

The following modes were not available in the predicate device bellavista 1000 (K163127) but are part of the subject device. We have utilized reference devices where applicable.

{7}------------------------------------------------

### Neonatal ventilation modes (modes in K163127 but not for neonates)

As neonatal ventilation had not been cleared in predicate device bellavista 1000 (K163127), all neonatal modes are compared to reference devices.

#### Summary of Ventilation Mode Comparison

Ventilation modes are either equivalent to the predicate – bellavista 1000 (K163127) or are similar to one of the reference devices as identified. We compared the ventilation modes which had similar features and their operating ranges as well as waveforms and they were found to be substantially equivalent.

### Substantial Equivalence Conclusion

We have compared the subject device bellavista 1000/1000e to the predicate device bellavista 1000 and where needed provided reference devices. The testing performed has demonstrated that the bellavista can be found to be substantially equivalent.

---

**Source:** [https://fda.innolitics.com/device/K183364](https://fda.innolitics.com/device/K183364)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
