VITEK 2 AST-Gram Negative Meropenem/Vaborbactam is designed for antimicrobial susceptibility testing of Gram Negative bacilli and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The VITEK 2 Gram-negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed.
Device Story
VITEK 2 AST-Gram Negative Meropenem/Vaborbactam is an automated, miniaturized, growth-based antimicrobial susceptibility test (AST) system. It uses a 64-well card containing premeasured antimicrobial concentrations (0.5/8–64/8 µg/mL) and nutrient media. Input: standardized saline suspension of clinical bacterial isolates. Operation: VITEK 2 system automatically dilutes, fills, seals, and incubates cards; VITEK 2 Compact uses manual filling/sealing. Principle: optical scanner measures light attenuation (transmittance) every 15 minutes to monitor bacterial growth. Output: MIC values and interpretive category results (Susceptible/Intermediate/Resistant). Used in clinical microbiology labs to guide antibiotic therapy by determining susceptibility of aerobic Gram-negative bacilli.
Clinical Evidence
External evaluation conducted using fresh and stock clinical isolates plus challenge strains. Performance compared against CLSI broth microdilution reference method (16-20 hour incubation). Results: 98.2% overall Essential Agreement and 98.7% overall Category Agreement. Reproducibility and Quality Control testing met acceptance criteria. Data representative of both VITEK 2 and VITEK 2 Compact platforms.
Technological Characteristics
Miniaturized broth microdilution card; contains antimicrobial agents in concentrations equivalent to standard methods. Automated optical growth monitoring system. Connectivity: VITEK 2 and VITEK 2 Compact instrument platforms. Software-controlled incubation and analysis. No specific materials or sterilization standards provided.
Indications for Use
Indicated for antimicrobial susceptibility testing of clinically significant aerobic Gram-negative bacilli, including Enterobacter cloacae complex, Escherichia coli, Klebsiella pneumoniae, Citrobacter freundii, Citrobacter koseri, Klebsiella aerogenes, Klebsiella oxytoca, Morganella morganii, Proteus mirabilis, Providencia spp., and Serratia marcescens. For use in clinical laboratories by trained personnel.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
K211136 — VITEK 2 AST-Gram Negative Imipenem/Relebactam (<0.25/4 - >16/4 ug/ML) · bioMerieux, Inc. · Jun 14, 2021
K183415 — VITEK 2 AST-Gram Negative Imipenem (</=0.25->= 16 ug/mL) · bioMerieux, Inc. · Feb 20, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 26, 2019
bioMerieux, Inc Esther Hernandez Regulatory Affairs Specialist 595 Anglum Road Hazelwood, Missouri 63042
## Re: K183360
Trade/Device Name: VITEK 2 AST-Gram Negative Meropenem/Vaborbactam (≤0.5/8 ->64/8 µg/mL) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully automated short-term incubation cycle antimicrobial susceptibility system Regulatory Class: Class II Product Code: LON, LTW, LTT Dated: November 29, 2018 Received: December 4, 2018
Dear Esther Hernandez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle with a blue top half and a gradient bottom half that transitions from yellow to green. The word "BIOMÉRIEUX" is written in white, sans-serif font in the center of the blue portion of the circle.
VITEK® 2 AST-GN Meropenem/Vaborbactam Traditional 510(k) Submission
## 510(k) SUMMARY
# VITEK® 2 AST-GN Meropenem/Vaborbactam
### 510(k) Submission Information:
| Submitter's Name: | bioMérieux, Inc. |
|----------------------|---------------------------------------------------------------------------------------------------------------------------|
| Address: | 595 Anglum Road<br>Hazelwood, MO 63042 |
| Contact Person: | Esther Hernandez<br>Regulatory Affairs Specialist |
| Phone Number: | 314-731-8841 |
| Fax Number: | 314-731-8689 |
| Date of Preparation: | November 29, 2018 |
| B. Device Name: | |
| Formal/Trade Name: | VITEK® 2 AST- Gram Negative Meropenem/<br>Vaborbactam (≤0.5/8 -> 64/8 µg/mL) |
| Classification Name: | 21 CFR 866.1645<br>Fully Automated Short-Term Incubation Cycle<br>Antimicrobial Susceptibility System<br>Product Code LON |
| Common Name: | VITEK® 2 AST-GN Meropenem/Vaborbactam |
| C. Predicate Device: | VITEK® 2 AST-GN Amikacin (K172731) |
### D. 510(k) Summary:
VITEK® 2 AST-Gram Negative Meropenem/Vaborbactam is designed for antimicrobial susceptibility testing of Gram Negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 AST-Gram Negative Meropenem/Vaborbactam is a quantitative test. Meropenem/Vaborbactam has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle that is split into two halves. The top half of the circle is blue and contains the word "BIOMÉRIEUX" in white letters. The bottom half of the circle is yellow.
VITEK® 2 AST-GN Meropenem/Vaborbactam Traditional 510(k) Submission
Active in vitro and in clinical infections: Enterobacter cloacae complex Escherichia coli Klebsiella pneumoniae
In vitro data available but clinical significance unknown:
Citrobacter freundii Citrobacter koseri Klebsiella (Enterobacter) aerogenes Klebsiella oxytoca Morganella morganii Proteus mirabilis Providencia spp. Serratia marcescens
The VITEK® 2 Gram-negative Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed.
The antimicrobial presented in VITEK® 2 AST-GN Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/ml. The VITEK® 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.
The isolate to be tested is diluted to a standardized concentration in 0.45 - 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling and sealing operation. The VITEK® 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.
VITEK®
N AST-GN Meropenem/Vaborbactam substantially equivalent performance when compared with the CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle with a blue top half and a green bottom half. The word "BIOMÉRIEUX" is written in white letters in the center of the blue half of the circle.
The Premarket Notification (510[k]) presents data in support of VITEK® 2 AST-GN Meropenem/Vaborbactam. An external evaluation was conducted with fresh and stock clinical isolates, as well as a set of challenge strains. The external evaluations were designed to confirm the acceptability of VITEK® 2 AST-GN Meropenem/Vaborbactam by comparing its performance with the CLSI broth microdilution reference method incubated at 16-20 hours. The data is representative of performance on both the VITEK® 2 and VITEK® 2 Compact instrument platforms. VITEK® 2 AST-GN Meropenem/Vaborbactam demonstrated acceptable performance of 98.2% overall Essential Agreement and 98.7% overall Category Agreement with the reference method. Reproducibility and Quality Control demonstrated acceptable results.
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
Indications for Use (Describe)
Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.