The Tractus CSC is intended to be used during interventional procedures in the peripheral vasculature to support a guidewire and facilitate access in discrete regions, allow for guidewire exchanges, and provide a conduit for delivering saline solutions and contrast media.
Device Story
Single-lumen, over-the-wire crossing support catheter; used in peripheral vasculature. Input: 0.035" guidewire. Operation: catheter tracked over guidewire to target lesion under fluoroscopic guidance; provides axial stability for guidewire crossing; facilitates guidewire exchanges; allows infusion of saline/contrast media via luer hub. Used in clinical interventional settings by physicians. Output: physical support for guidewire; conduit for fluids. Benefits: enables navigation of tortuous vasculature; assists in crossing discrete lesions for subsequent therapy (angioplasty/stenting).
Clinical Evidence
Bench testing only. Testing included biocompatibility (toxicity, sensitization, irritation, hemolysis, thrombogenicity, complement activation), mechanical performance (flexibility, trackability, torque, tensile/bond strength, coating integrity, flow rate, corrosion), and packaging/sterilization validation. All tests met pre-established acceptance criteria.
Indicated for use during interventional procedures in the peripheral vasculature to support a guidewire, facilitate access in discrete regions, allow for guidewire exchanges, and provide a conduit for saline or contrast media delivery.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 5, 2019
Tractus Vascular LLC Sam Gilbert Manager, Regulatory/Clinical/Quality 15 Christopher Way Eatontown, New Jersey 07724
Re: K183305
Trade/Device Name: Crossing Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: July 10, 2019 Received: July 11, 2019
Dear Sam Gilbert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Kenneth Cavanaugh Acting Director DHT2C: Division of Coronary and Peripheral Interventional Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K183305
Device Name
Tractus Crossing Support Catheter
Indications for Use (Describe)
The Tractus CSC is intended to be used during interventional procedures in the peripheral vasculature to support a guidewire and facilitate access in discrete regions, allow for guidewire exchanges, and provide a conduit for delivering saline solutions and contrast media.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY - K183305
## Tractus Vascular LLC's Crossing Support Catheter (CSC)
### Submitter's Name, Address, Telephone Number, Contact Person, and Date Prepared
Janet Burpee, CEO Tractus Vascular, LLC. 15 Christopher Way Eatontown. NJ 07724 Phone: 732-996-8513
Date Prepared: July 30, 2019
## Name of Device
Crossing Support Catheter (CSC)
#### Common or Usual Name
Percutaneous Crossing Catheter
## Classification Name:
21 CFR 870.1250, Class II, product code DQY
#### Predicate Devices
Quick Cross® Extreme Support Catheters (K082561) (Primary Predicate)
TOTAL across™ (K133539) (Reference Predicate)
Primi™ Support Catheter (K132701) (Reference Predicate)
#### Intended Use/Indications for Use
The Tractus CSC is intended to be used during interventional procedures in the peripheral vasculature to support a guidewire and facilitate access in discrete regions, allow for guidewire exchanges, and provide a conduit for delivering saline solutions and contrast media.
#### Device Description
The CSC consists of a family of single-lumen, over-the-wire catheters offered in a variety of sizes for compatibility with a range of guidewire and sheath sizes as well as effective lengths. The subject device is a 0.035" guidewire compatible catheter with a 4.5 Fr crossing profile and available effective lengths of 90, 135, 155, and 170 cm. The catheters are used to navigate tortuous peripheral vasculature while providing axial stability to enhance guidewire crossing of discrete lesions of the vasculature. The catheters are also used to allow for quidewire exchanges, and provide a conduit for delivering saline solutions and contrast media.
The CSC consists of an outer, inner, and center tubing configuration. The outer tubing of the CSC consists of Pebax with hydrophilic coating on the distal end of the catheter to reduce frictional forces and enhance tracking. The inner tubing consists of a PTFE liner and Pebax. The
{4}------------------------------------------------
center tubing is spiral laser cut stainless steel. This design provides catheter flexibility, kink resistance and torsional strength while also providing axial stability to support a guidewire and allow for quidewire placement across discrete lesions for further percutaneous intervention, such as angioplasty or stent placement. The distal tip of the catheter, which is stainless steel covered with Pebax and is continuous with the stainless steel center tubing of the catheter, has a smooth, rounded, and tapered profile that provides a seamless catheter-to-guidewire transition.
All effective lengths of the CSC include three radiopaque marker bands that are evenly spaced 10cm apart along the distal end of the catheter including one radiopaque marker band within 4mm of the distal end of the catheter for fluoroscopic visualization of the distal tip. In addition to providing hub-to-tip visualization of the catheter, the marker bands assist in catheter positioning and aide in estimating geometry within the vascular system for subsequent therapies, such as PTA and/or stenting. The proximal end of the catheter includes a luer hub to allow for quidewire passage, flushing saline solution or contrast media through the inner lumen of the catheter, and to facilitate guidewire exchanges.
To use the device, the inner lumen of the catheter is flushed with saline until the solution exits the distal end of the device prior to use. The appropriate size quidewire is inserted into the distal end of the catheter and the catheter is advanced over the guidewire and through a sheath using standard interventional techniques. Under fluoroscopic guidance, the catheter is tracked to the target site within the vasculature and utilized to support guidewire access across discrete areas of the vasculature using standard interventional techniques. For guidewire exchanges, standard interventional techniques are used. Saline or contrast media can be infused into the catheter during operation for flushing and fluoroscopic visualization. Once quidewire access across the target lesion is gained, the catheter is withdrawn using standard interventional techniques.
# Technological Characteristics
The CSC has similar technological characteristics as its predicate devices. Both the CSC and the primary predicate are single lumen, over-the-wire catheters that include a luer hub to allow for flushing of saline solutions and contrast media. Both the CSC and primary predicate facilitate guidewire exchanges and are compatible with 0.035" guidewires for use in the peripheral vasculature under fluoroscopy with 5F sheaths. Furthermore, both the CSC and the primary predicate allow for device visualization using three (3) visual bands and feature tapered distal tips.
The overall construction of both the CSC and the primary predicate are similar. Both catheters consist of an outer, inner, and center tubing configuration. The outer tubing of both the subject device and primary predicate consist of Pebax with hydrophilic coating on the distal end to reduce frictional forces and enhance tracking. The inner tubing of both the subject device and the primary predicate is a PTFE liner. Further, the material composition of the center tubing for both the subject and primary predicate is the same (i.e. stainless steel), with slightly different configurations (spiral laser cut versus braided). Despite this difference in configuration the principle of operation is the same. Furthermore, the spiral laser cut inner tubing design of the subject device has been cleared for the TOTAL across reference predicate (K133539) which was also cleared for substantially the same indications for use as the subject device, albeit marketed for a 0.014" quidewire compatible catheter only.
The overall dimensions of the subject device and the primary predicate are similar. The subject device has slightly longer working length range than the primary predicate (90 – 170 cm for the subject device versus 90 - 150 cm for the primary predicate). However, similar working lengths
{5}------------------------------------------------
have been cleared for the Primi Catheter reference predicate (90 cm to 170 cm) which was cleared for a nearly identical indication for use compared to the subject device.
Both the subject device and the primary predicate are provided sterilization and are non-pyrogenic. Further, both devices are packaged in Tyvek pouches.
Therefore, the subject CSC has very similar technological characteristics as the primary predicate, as well as the reference devices.
# Performance Data
The following nonclinical testing has been conducted to support the substantial equivalence of the CSC to its predicate devices. In all instances, the CSC functioned as intended.
- . Acute Systemic Toxicity
- Materials-Mediated Pyrogenicity
- Intracutaneous Irritation ●
- . Maximization Sensitization
- MEM Elution
- . Hemolysis - Extract Method
- Hemolysis - Direct Contact Method
- Nonanticoagulated Canine Thromobgenicity
- Anticoagulated Porcine Thrombogenicity ●
- Complement Activation (SC5b-9) ●
- Simulated Use Testing
- Flexibility ●
- Trackability ●
- Guidewire Compatibility
- Torque Tolerance
- . Tensile Strength
- . Bond Strength
- Coating Integrity (Blue Dye Test, Lubricity) ●
- Radiopacity
- Particulate Emission ●
- Flow Rate .
- Corrosion Resistance
- Package Integrity
- Shipping Simulation .
- Environmental Conditioning ●
- Bubble Leak ●
- . Seal Strength
# Substantial Equivalence
The CSC has the same intended use as the primary predicate device and substantially the same indications for use as the reference devices. The subject device also has substantially the same technological characteristics and principles of operation as the predicate devices. Any minor differences do not raise different questions of safety or effectiveness and performance testing
{6}------------------------------------------------
demonstrated that the subject device performs in a substantially equivalent manner. Therefore, the subject CSC is substantially equivalent to its predicate devices.
# Conclusions
Tractus Vascular LLC's CSC is a Percutaneous Catheter, Class II device that has been evaluated in nonclinical testing in accordance with FDA's recognized standards and pre-established acceptance criteria. Testing demonstrated that the device performs as intended. The CSC is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.