← Product Code [LZA](/productcode/LZA) · K183280

# PowderFree White Blue Sterilized Nitrile Copolymer Examination Gloves Tested for use with chemotherapy drugs (K183280)

_Smart Glove Corporation Sdn. Bhd. · LZA · Jul 26, 2019 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K183280

## Device Facts

- **Applicant:** Smart Glove Corporation Sdn. Bhd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Jul 26, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A Patient examination glove is a disposable device intended for medical purposes that worn on the examiner's hand or finger to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs as listed below.

## Device Story

Disposable nitrile copolymer examination gloves; worn by healthcare examiners to prevent cross-contamination between patient and examiner. Device provides a physical barrier; tested for permeation resistance against specific chemotherapy drugs. Used in clinical settings; operated by healthcare professionals. Output is the physical barrier protection provided during patient examination or chemotherapy drug handling. Benefits include reduced risk of contamination and exposure to hazardous chemotherapy agents.

## Clinical Evidence

Bench testing only. Permeation testing performed for 14 chemotherapy drugs to determine breakthrough detection times. Results indicate most drugs have breakthrough times >240 minutes, while Carmustine and Thio-Tepa showed lower breakthrough times, resulting in a contraindication for those specific agents.

## Technological Characteristics

Nitrile copolymer material; powder-free; sterilized; disposable; form-fitting glove design. No electronic components or software.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services-USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

July 26, 2019

Smart Glove Corporation Sdn. Bhd. Khon Foo Managing Director Lot 6487, Batu 5 3/4, Sementa, Jalan Kapar Klang, 42100 My

Re: K183280

Trade/Device Name: PowderFree White Blue Sterilized Nitrile Copolymer Examination Gloves Tested for use with chemotherapy drugs

Regulation Number: 21 CFR 880.6250 Regulation Name: Patient examination glove Regulatory Class: Class I Product Code: LZA Dated: June 18, 2019 Received: June 24, 2019

Dear Khon Foo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K183280

## Device Name

PowderFree White Blue Sterilized Nitrile Copolymer Examination Gloves Tested for Use with Chemotherapy Drugs.

Indications for Use (Describe)

A Patient examination glove is a disposable device intended for medical purposes that worn on the examiner's hand or finger to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs as listed below.

Chemotherapy Drug permeation and concentration ( Breakthrough Detection Time) in Minutes. The following chemicals have been tested with these gloves.

|                                                   | Breakthrough Detection Time in Minutes |            |
|---------------------------------------------------|----------------------------------------|------------|
| Chemotherapy Drugs:                               | White Glove                            | Blue Glove |
| 1. Carmustine ( BCNU)3.3 mg/ml(3,300 ppm)         | 7.1 min                                | 23.4 min   |
| 2. Cisplatin 1.0 mg/ml (1,000ppm)                 | >240 min                               | >240 min   |
| 3. Cyclophosphamide (Cytoxan) 20mg/ml(20,000ppm)  | >240 min                               | >240 min   |
| 4. Dacarbazine (DTIC)10.0 mg/ml(10,000ppm)        | >240 min                               | >240 min   |
| 5. Doxorubicin Hydrochloride 2.0 mg/ml (2,000ppm) | >240 min                               | >240 min   |
| 6. Etoposide (Toposar) 20.0 mg/ml (20,000ppm)     | >240 min                               | >240 min   |
| 7. Fluorouracil 50.0 mg/ml (50,000ppm)            | >240 min                               | >240 min   |
| 8. Ifosfamide 50.0 mg/ml (50,000ppm)              | >240 min                               | >240 min   |
| 9. Methotrexate 25 mg/ml (25,000ppm)              | >240 min                               | >240 min   |
| 10.Mitomycin C 0.5 mg/ml (500ppm)                 | >240 min                               | >240 min   |
| 11. Mitoxantrone 2.0 mg/ml (2,000ppm)             | >240 min                               | >240 min   |
| 12. Paclitaxel (Taxol) 6.0 mg/ml (6,000ppm)       | >240 min                               | >240 min   |
| 13. Thio-Tepa 10.0 mg/ml (10,000ppm)              | 78.6 min                               | 39.3 min   |
| 14.Vincristine Sulfate 1.0 mg/ml (1,000ppm)       | >240 min                               | >240 min   |

WARNING: Do not use with Carmustine or Thio-Tepa.

Please note that Carmustine and Thio-Tepa have low permeation time of less than 240 minutes.

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

---

**Source:** [https://fda.innolitics.com/device/K183280](https://fda.innolitics.com/device/K183280)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com