← Product Code [GCJ](/productcode/GCJ) · K183264

# Flex-THOR scope (K183264)

_KARL STORZ Endoscopy-America, Inc. · GCJ · Jan 18, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K183264

## Device Facts

- **Applicant:** KARL STORZ Endoscopy-America, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jan 18, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Flex-THOR System is intended for visualization purposes during abdominal and thoracic diagnostic and therapeutic procedures.

## Device Story

Flex-THOR System consists of a flexible endoscope and a Camera Control Unit (CCU). The scope features a 2.9 mm outer diameter insertion shaft, 675 mm working length, and a 1.2 mm working channel for instrument access. It utilizes an internal LED light source and a CMOS imager to capture visual data from within thoracic or abdominal cavities. The device is used by clinicians in surgical settings to perform diagnostic and therapeutic procedures. The clinician views the output on a display connected to the CCU, facilitating real-time visualization of internal anatomy to guide surgical interventions. The system is reusable and requires reprocessing between uses.

## Clinical Evidence

No clinical data. Substantial equivalence was established through non-clinical bench testing, including electrical safety, electromagnetic compatibility, and reprocessing validation.

## Technological Characteristics

Flexible endoscope with CMOS imager and internal LED light source. Distal tip: 3.2 mm x 2.4 mm; shaft length: 675 mm; working channel: 1.2 mm. Reprocessing standards: AAMI TIR 12:2010, ISO 15883-5:2005, AAMI TIR 30:2011, AAMI/ANSI/ISO 11737-1:2006/(R)2011, ASTM E1837-96:2014. Electrical safety/EMC standards: ANSI/AAMI ES:60601-1:2005, IEC 60601-1-2:2007.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Vibrynt Inc's VISTA Endoscope ([K100533](/device/K100533.md))

## Reference Devices

- XLTF-VAW Laparo-Thoraco Videoscope ([K053382](/device/K053382.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a partial view of the words "& DRUG" in a bold, blue font. The text is cropped, suggesting it is part of a larger title or heading. The background is plain white, which makes the blue text stand out.

January 18, 2019

KARL STORZ Endoscopy-America, Inc. Nozomi Yagi Sr. Regulatory Affairs Specialist 2151 E. Grand Avenue EI Segundo, California 90245

Re: K183264

Trade/Device Name: Flex-THOR scope Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: November 21, 2018 Received: November 23, 2018

Dear Nozomi Yagi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809: medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long H. Chen Digitally signed by Long H.
Chen-s Date: 2019.01.18 17:48:16 -5 -05'00' for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) k183264

Device Name Flex-Thor System

Indications for Use (Describe)

The Flex-THOR System is indicated for use in providing access to, and visualization of, the thoracic and abdominal cavities, to allow for the performance of various diagnostic and therapeutic surgical procedures.

Type of Use (Select one or both, as applicable)

| <div> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)    </div> |
|----------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)    </div>  |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K183264

## 510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.

| Submitter:                               | KARL STORZ Endoscopy-America, Inc.<br>2151 E. Grand Avenue<br>EI Segundo, CA 90245                                                                                                                                                                                                                                                      |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                                 | Nozomi Yagi<br>Sr. Regulatory Affairs Specialist<br>Phone: (424) 218-8351<br>Fax: (424) 218-8519                                                                                                                                                                                                                                        |
| Date of<br>Preparation:                  | November 21, 2018                                                                                                                                                                                                                                                                                                                       |
| Type of 510(k)<br>Submission:            | Traditional                                                                                                                                                                                                                                                                                                                             |
| Device<br>Identification:                | Trade Name: Flex-THOR System<br>Classification Name: Endoscope and accessories (21 CFR Part<br>876.1500)                                                                                                                                                                                                                                |
| Regulatory<br>Class:                     | II                                                                                                                                                                                                                                                                                                                                      |
| Product Code<br>and Common<br>Name:      | GCJ (Laparoscope, General & Plastic Surgery)                                                                                                                                                                                                                                                                                            |
| Guidance<br>Document:                    | Not Applicable for GCJ product code                                                                                                                                                                                                                                                                                                     |
| Recognized<br>Consensus<br>Standards:    | Not Applicable for GCJ product code                                                                                                                                                                                                                                                                                                     |
| Predicate and<br>Reference<br>Device(s): | Primary Predicate Device: Vibrynt Inc's VISTA Endoscope<br>(K100533)<br>Reference Device: XLTF-VAW Laparo-Thoraco<br>Videoscope (K053382)<br>**The above predicate and reference device have not been subjected to any<br>recall**                                                                                                      |
| Device<br>Description:                   | The Flex-THOR System includes two main components: (1) the Flex-<br>THOR Scope (Part Number: 11292VS(U)A-THOR), and (2) the<br>Camera Control Unit (CCU). The insertion shaft of the Flex-THOR<br>Scope (Part Number: 11292VS(U)A-THOR) has an outer diameter of<br>2.9 mm and a working length of 675 mm with 8.5 Fr elliptical shaped |

{4}------------------------------------------------

| distal tip (major diameter of 3.2 mm and minor diameter of 2.4 mm).<br>Users can access the 1.2 mm working channel through the Luer ports. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                      |                            |
|--------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|----------------------------|
| Intended Use:                                                                                                                              | The Flex-THOR System is intended for visualization purposes during<br>abdominal and thoracic diagnostic and therapeutic procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                      |                            |
| Indications For<br>Use:                                                                                                                    | The Flex-THOR System is indicated for use in providing access to,<br>and visualization of, the thoracic and abdominal cavities, to allow for<br>the performance of various diagnostic and therapeutic surgical<br>procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                      |                            |
| Technological<br>Characteristics:                                                                                                          | Comparison Table: Subject vs. Primary Predicate Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                      |                            |
|                                                                                                                                            | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Primary Predicate Device,<br>K100533 |                            |
|                                                                                                                                            | Type of Scope                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Flexible                             | Same as the subject device |
|                                                                                                                                            | Distal End<br>Diameter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 3.2 mm x 2.4 mm                      | 5.0 mm                     |
|                                                                                                                                            | Insertion Shaft<br>Length                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 675 mm                               | 530 and 890 mm             |
|                                                                                                                                            | Type of Imager                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | CMOS                                 | Fiber                      |
|                                                                                                                                            | Direction of View                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 0°                                   | 0°                         |
|                                                                                                                                            | Light Source                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Internal LED                         | External                   |
|                                                                                                                                            | Cleaning                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                  | Yes                        |
|                                                                                                                                            | Sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                  | Yes                        |
| Non-Clinical<br>Performance<br>Data:                                                                                                       | Electrical Safety and Electromagnetic Compatibility Summary<br>The electrical safety and EMC data included in the submission is in compliance<br>with the following FDA recognized standards:<br>• ANSI/AAMI ES:60601-1:2005<br>• IEC 60601-1-2:2007<br>Reprocessing Validation Summary<br>The Flex-THOR Scope (Part Number: 11292VS(U)A-THOR) is provided non-<br>sterile and is reusable. The users are required to reprocess it for initial and after<br>each use. The subject device contacts intact mucosal membranes so it is a semi-<br>critical device per Spaulding Classification. We performed validation activities for<br>cleaning and sterilization according to the FDA Guidance. The reprocessing data<br>submitted is in compliance with the following standards:<br>• AAMI TIR 12:2010<br>• ISO 15883-5:2005<br>• AAMI TIR 30:2011<br>• AAMI/ANSI/ISO 11737-1:2006/ (R)2011<br>• ASTM E1837-96:2014 |                                      |                            |
| Clinical<br>Performance<br>Data:                                                                                                           | Clinical testing was not required to demonstrate the substantial<br>equivalence to the predicate devices. Non-clinical bench testing was<br>sufficient to establish the substantial equivalence of the<br>modifications.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                      |                            |
| Conclusion:                                                                                                                                | The conclusions drawn from the non-clinical performance data and<br>comparison of device characteristics demonstrated that the subject<br>device is as safe as and as effective as the predicate device. As such,<br>we concluded that the substantial equivalence of the subject and the                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                      |                            |

{5}------------------------------------------------

| predicate device has been met, and the differences between the        |
|-----------------------------------------------------------------------|
| subject and the predicate device do not raise new questions of safety |
| and effectiveness.                                                    |

---

**Source:** [https://fda.innolitics.com/device/K183264](https://fda.innolitics.com/device/K183264)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).