K183254 · Siemens Medical Solutions USA, Inc. · LNH · Jan 18, 2019 · Radiology
Device Facts
Record ID
K183254
Device Name
Magnetom Vida
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LNH · Radiology
Decision Date
Jan 18, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
Your MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. Your MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Device Story
MAGNETOM Vida is an MRI system utilizing magnetic resonance to generate cross-sectional images and spectra of human anatomy. The system processes RF signals and magnetic field gradients to reconstruct images. New features include: nose marker-based inline motion correction for 3D head scans; GOLiver optimized pulse sequences for abdominal/liver imaging; TSE_MDME for multi-contrast acquisition; SEMAC for metal artifact reduction in orthopedic imaging; Compressed Sensing (CS) for accelerated TOF MRA and SPACE sequences; Simultaneous Multi-Slice (SMS) for RESOLVE and QDWI; and respiratory self-gating for abdominal/thoracic imaging. Operated by healthcare professionals in clinical settings. Output consists of diagnostic images/spectra interpreted by physicians to assist in clinical diagnosis and interventional guidance. Benefits include reduced scan times, improved image quality via motion/artifact correction, and consistent exam throughput.
Clinical Evidence
No clinical tests conducted. Evidence consists of bench testing, software verification/validation, and sample clinical images to demonstrate performance of new/modified features. Compliance with IEC 62304, IEC 60601-2-33, and NEMA standards confirmed.
Technological Characteristics
Magnetic resonance diagnostic device. Hardware includes nose marker for motion tracking. Software features: Compressed Sensing, Simultaneous Multi-Slice (SMS), and various pulse sequences (TSE_MDME, SEMAC, VIBE). Connectivity via DICOM. Conforms to IEC 60601-1, IEC 60601-2-33, ISO 14971, and IEC 62304 standards.
Indications for Use
Indicated for patients requiring diagnostic MR imaging of the head, body, or extremities to visualize internal structure and function. Used for diagnostic imaging and interventional procedures. No specific age or gender restrictions; contraindications include presence of non-MR compatible devices.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
syngo.via VB30A2 (based on syngo.via VB10A, K150843)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Siemens Medical Solutions, USA, Inc. % Cordell L. Fields, Esq. Regulatory Affairs Specialist 40 Liberty Boulevard, Mailcode 65-1A MALVERN PA 19355
January 18, 2019
#### Re: K183254
Trade/Device Name: MAGNETOM Vida Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH, LNI and MOS Dated: November 19, 2018 Received: November 21, 2018
Dear Mr. Fields:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a bright teal. The letters are evenly spaced and the word is horizontally oriented. The image appears to be a logo or branding element for the company Siemens.
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form |
|-----------------------------------------|------------|
| Food and Drug Administration | Expiration |
| Indications for Use | See P |
Approved: OMB No. 0910-0120 tion Date: 06/30/2020 RA Statement below.
510(k) Number (if known)
#### K183254
Device Name
MAGNETOM Vida
Indications for Use (Describe)
Your MAGNETOM system is indicated for use as a magnetic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectroscopic images and or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/ or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yicld information that may assist in diagnosis.
Your MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act 1990 and 21 CFR § 807.92.
## 1. General Information
| Establishment | Siemens Medical Solutions USA, Inc. |
|---------------|-------------------------------------|
| | 40 Liberty Boulevard |
| | Mail Code 65-1A |
| | Malvern, PA 19355, USA |
| | Registration Number: 2240869 |
- Date Prepared November 19, 2018
- Manufacturer Siemens Healthcare GmbH Henkestrasse 127 Erlangen, Bayern, Germany 91052 Registration Number: 3002808157
# 2. Contact Information
Cordell L. Fields, Esq. Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355, USA Phone: (610) 448-6469 Fax: (610) 640-4481 E-mail: cordell.fields@siemens-healthineers.com
# 3. Device Name and Classification
| Device Name: | MAGNETOM Vida |
|-----------------------|---------------------------------------------|
| Trade Name: | MAGNETOM Vida |
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR § 892.1000 |
| Classification: | Class II |
| Product Code: | Primary: LNH, Secondary: LNI, MOS |
{4}------------------------------------------------
### 4. Legally Marketed Predicate Device
| Trade Name: | MAGNETOM Vida |
|-----------------------|--------------------------------------------|
| 510(k) Number: | K181433, cleared October 19, 2018 |
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR § 892.1000 |
| Classification: | Class II |
| Product Code: | Primary: LNH, Secondary: LNI, MOS |
## 5. Intended Use
The indications for use for the subject device is the same as the predicate device:
Your MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Your MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
### 6. Device Description
MAGNETOM Vida with software syngo MR XA11B includes new and modified component, features and software compared to the predicate device, MAGNETOM Vida with syngo MR XA11A. A high level summary of the new and modified features is provided below:
### Hardware
New Hardware
- Nose Marker for Inline Motion Correction
### Software
New Features and Applications
- TFL with Inline Motion Correction: Tracking of motion of the head during 3D MPRAGE head scans with a nose marker and a camera system. The MR system uses the tracking information to compensate for the detected motion.
{5}------------------------------------------------
# SIEMENS
- i GOLiver: GOLiver is a set of optimized pulse sequence for fast and efficient imaging of the abdomen / liver. It is designed to provide consistent exam slots and to reduce the workload for the user in abdominal / liver MRI.
- TSE_MDME: A special variant of the TSE pulse sequence type which acquires several contrasts (with different TI and TE, i.e. Multi Delay Multi Echo) within a single sequence.
- SEMAC: SEMAC is a method for metal artifact correction in ortho imaging of patients i with whole joint replacement. Using Compressed Sensing the acquisition can be accelerated.
- Angio TOF with Compressed Sensing: The Compressed Sensing (CS) functionality is now available for TOF MRA within the BEAT pulse sequence type. Scan time can be reduced by an incoherent undersampling of k-space data. The usage of CS as well as the acceleration factor and further options can be freely selected by the user.
- SMS for RESOLVE and QDWI: Simultaneous excitation and acquisition of multiple i slices with the Simultaneous multi-slice (SMS) technique for readout-segmented echo planar imaging (RESOLVE) and quiet diffusion weighted imaging (QDWI).
- i SPACE with Compressed Sensing: The Compressed Sensing (CS) functionality is now available for the SPACE pulse sequence type. Scan time can be reduced by the incoherent under-sampling of the k-space data. The usage of CS as well as the acceleration factor and other options can be freely selected by the user.
- RT Respiratory self-gating for FL3D_VIBE: Non-contrast abdominal and thoracic i examination in free breathing with reduced blur induced by respiratory motion.
# Other Modifications and / or Minor Changes
- i Turbo Suite is a marketing bundle of components for accelerated MR imaging offered for the MAGNETOM Vida MR systems.
- i Noise masking: a mechanism to remove the noise floor in outer regions is now available for RESOLVE and QDWI.
# 7. Technological Characteristics
The subject device, MAGNETOM Vida with software syngo MR XA11B, is substantially equivalent to the predicate device with regard to the operational environment, programming language, operating system, and performance.
MAGNETOM Vida with software syngo MR XA11B conforms to the standard for medical device software (IEC 62304:2006) and other relevant IEC and NEMA standards.
While there are some differences in technological characteristics between the subject device and predicate device, including new and modified software applications and
{6}------------------------------------------------
hardware additions, these differences have been tested and the conclusions from the non-clinical data suggests that the features bear an equivalent safety and performance profile to that of the predicate device.
# 8. Nonclinical Tests
The following performance testing was conducted on the subject device:
- Sample clinical images were taken for the new and modified software features.
- Image quality assessments of all new/modified pulse sequence types and algorithms were completed. In some cases a comparison of the image quality was made between the new/modified features and the predicate device features.
- Software verification and validation testing was completed in accordance with the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", dated May 11, 2005.
- । Performance tests were completed in accordance with the FDA quidance document. "Submission of Premarket Notifications for Maqnetic Resonance Diagnostic Devices" dated November 18, 2016.
The results from each set of tests demonstrate that the device performs as intended and is thus substantially equivalent to the predicate device to which it has been compared.
# 9. Clinical Tests
No clinical tests were conducted to support substantial equivalence for the subject device; however, sample clinical images were provided to support the new/modified component and software features per the FDA quidance document "Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices", dated November 18, 2016. Clinical publications were referenced to provide information on the use of some features and functions.
# 10. Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings to ensure safe and effective use of the device.
Risk management is ensured via ISO 14971:2007 compliance to identify and provide mitigation of potential hazards in a risk analysis early in the design phase and continuously throughout the development of the product. These risks are controlled via measures realized in hardware and software development, testing, and product labeling. To minimize risks, Siemens adheres to recognized and established industry practices and standards, such as the IEC 60601-1 series, to minimize electrical and mechanical risks. Furthermore, the device is intended for healthcare professionals
{7}------------------------------------------------
# SIEMENS
familiar with and responsible for the acquisition and post processing of magnetic resonance images.
MAGNETOM Vida with software syngo MR XA11B conforms to the following FDA recognized and international IEC, ISO and NEMA standards:
| Recognition<br>Number | Product<br>Area | Title of Standard | Reference Number<br>and date | Standards<br>Development<br>Organization |
|-----------------------|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|------------------------------------------|
| 19-4 | General | Medical electrical equipment -<br>part 1: general requirements for<br>basic safety and essential<br>performance | ES60601-<br>1:2005/(R) 2012<br>and A1:2012 | AAMI / ANSI |
| 19-8 | General | Medical electrical equipment -<br>Part 1-2: General requirements<br>for basic safety and essential<br>performance - Collateral<br>Standard: Electromagnetic<br>disturbances - Requirements<br>and tests | 60601-1-2 Edition<br>4.0:2014-02 | IEC |
| 12-295 | Radiology | Medical electrical equipment -<br>Part 2-33: Particular<br>requirements for the basic safety<br>and essential performance of<br>magnetic resonance equipment<br>for medical diagnosis | 60601-2-33 Ed.<br>3.2:2015 | IEC |
| 5-40 | General | Medical devices - Application of<br>risk management to medical<br>devices | 14971 Second<br>edition 2007-10 | ISO |
| 5-96 | General | Medical devices - Application of<br>usability engineering to medical<br>devices | 62366 Edition 1.0<br>2015 | AAMI<br>ANSI<br>IEC |
| 13-32 | Software | Medical device software -<br>Software life cycle processes | 62304:2006 | AAMI<br>ANSI<br>IEC |
| 12-232 | Radiology | Acoustic Noise Measurement<br>Procedure for Diagnosing<br>Magnetic Resonance Imaging<br>Devices | MS 4-2010 | NEMA |
| 12-288 | Radiology | Characterization of Phased<br>Array Coils for Diagnostic<br>Magnetic Resonance Images<br>(MRI) | MS 9-2008 | NEMA |
| 12-300 | Radiology | Digital Imaging and<br>Communications in Medicine<br>(DICOM) Set 03/16/2012<br>Radiology | PS 3.1 - 3.20<br>(2016) | NEMA |
| Recognition<br>Number | Product<br>Area | Title of Standard | Reference Number<br>and date | Standards<br>Development<br>Organization |
| 2-156 | Biocompati<br>bility | biological evaluation of medical<br>devices - part 1: evaluation and<br>testing within a risk management<br>process. (Biocompatibility) | 10993-1:2009/(R)<br>2013 | AAMI<br>ANSI<br>ISO |
{8}------------------------------------------------
### 11. Substantial Equivalence
MAGNETOM Vida with software syngo MR XA11B is substantially equivalent to the following predicate device:
| Predicate Device | FDA Clearance<br>Number and Date | Product code | Manufacturer |
|--------------------------------------|-----------------------------------------|----------------|----------------------------|
| MAGNETOM Vida with syngo MR<br>XA11A | K181433,<br>cleared October<br>19, 2018 | LNH<br>LNI,MOS | Siemens Healthcare<br>GmbH |
MAGNETOM Vida with syngo MR XA11B includes features already cleared on the following reference devices:
| Reference Devices | FDA Clearance<br>Number and Date | Product<br>code | Manufacturer |
|----------------------------------------------|-----------------------------------------------------------|-------------------------|----------------------------|
| Biograph mMR with syngo MR<br>E11P | K163234,<br>cleared February<br>28, 2017 | OUO<br>LNH, LNI,<br>KPS | Siemens Healthcare<br>GmbH |
| syngo.via VB30A2<br>based on syngo.via VB10A | syngo.via VB10A:<br>K150843,<br>cleared April 24,<br>2015 | LLZ | Siemens Healthcare<br>GmbH |
# 12. Conclusion as to Substantial Equivalence
MAGNETOM Vida with software syngo MR XA11B has the same intended use and same basic technological characteristics than the predicate device system, MAGNETOM Vida with syngo MR XA11A, with respect to the magnetic resonance features and functionalities. While there are some differences in technical features
വ Change released by internal documentation based on syngo.via VB10 (K150843).
{9}------------------------------------------------
compared to the predicate device, the differences have been tested and the conclusions from all verification and validation data suggest that the features bear an equivalent safety and performance profile to that of the predicate device and reference devices.
Siemens believes that MAGNETOM Vida with software syngo MR XA11B is substantially equivalent to the currently marketed device MAGNETOM Vida with software syngo MR XA11A (K181433, cleared on October 19, 2018).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.