Carestream DRX-1 System with DRX Plus 2530 Detectors
K183245 · Carestream Health, Inc. · MQB · Feb 8, 2019 · Radiology
Device Facts
Record ID
K183245
Device Name
Carestream DRX-1 System with DRX Plus 2530 Detectors
Applicant
Carestream Health, Inc.
Product Code
MQB · Radiology
Decision Date
Feb 8, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is intended to capture for display radiographic images of human anatomy including both pediatric and adult patients. The device is intended for use in general projection radiographic applications wherever conventional screen-film systems or CR systems may be used. Excluded from the indications for use are mammography, fluoroscopy, and angiography applications.
Device Story
Carestream DRX-1 System with DRX Plus 2530/2530C Detectors is a digital radiography (DR) diagnostic imaging system. It captures x-ray projections using flat panel digital detectors (25cm x 30cm) and transmits data via tethered or wireless connection to an operator console. The system utilizes Eclipse II image processing software to generate, display, and transmit digital images for diagnostic viewing. Used in clinical settings by healthcare professionals, the device replaces conventional screen-film or CR systems. The system features an All-in-One (AIO) electronics board and updated readout integrated circuit (ROIC) for improved noise and resolution performance. The output allows radiologists to perform diagnostic assessments of human anatomy. Benefits include high-resolution digital imaging, improved workflow, and diagnostic equivalence to predicate systems.
Clinical Evidence
Clinical study conducted at University of Rochester Medical Center using 2 adult cadavers and pediatric phantoms. 162 images acquired (facial bones, skull, C-Spine, extremities, chest) and evaluated by 3 board-certified radiologists using a 4-point RadLex scale. Results showed 98% (DRX Plus 2530) and 96% (DRX Plus 2530C) of images rated as Diagnostic (3) or Exemplary (4). Statistical equivalence tests confirmed mean ratings were equivalent to the predicate device.
Technological Characteristics
Flat panel digital imager; aluminum housing (picture frame design); 25cm x 30cm active area; 98-micron pixel pitch; All-in-One (AIO) electronics board; updated ROIC; tethered/wireless connectivity; uses Eclipse II software; stationary x-ray system class.
Indications for Use
Indicated for pediatric and adult patients requiring general projection radiographic imaging. Excludes mammography, fluoroscopy, and angiography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Carestream DRX-1 System (with DRX Plus 3543 Detectors) (K150766)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in blue, followed by the words "U.S. Food & Drug Administration" in blue.
Carestream Health, Inc. % Ms. Gina Maiolo Regulatory Affairs Manager 150 Verona Street ROCHESTER NY 14608
February 8, 2019
# Re: K183245
Trade/Device Name: Carestream DRX-1 System with DRX Plus 2530 Detectors Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: January 14, 2019 Received: January 17, 2019
Dear Ms. Maiolo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing
{1}------------------------------------------------
# safety reporting (21 CFR 4. Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183245
Device Name
Carestream DRX-1 System with DRX Plus 2530 Detectors
Indications for Use (Describe)
The device is intended to capture for display radiographic images of human anatomy including both pediatric and adult patients. The device is intended for use in general projections wherever conventional screen-film systems or CR systems may be used. Excluded from the indications for use are mammography, fluoroscopy and angiography applications.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Carestream
# "510(k) Summary"
| 510(k) Owner Name: | Carestream Health, Inc. |
|------------------------|-----------------------------------------------------------------------------------------------|
| 510(k) Owner Address: | 150 Verona Street<br>Rochester, New York 14608 |
| 510(k) Owner Phone: | 585 627-6543 |
| 510(k) Owner Fax: | 585 627-8802 |
| Contact Person & Info: | Gina Maiolo<br>Regulatory Affairs Manager, US&C<br>gina.maiolo@carestream.com<br>585-627-6543 |
| Date Summary Prepared: | February 6, 2019 |
| Device Trade Name: | Carestream DRX-1 System with DRX Plus 2530<br>Detectors |
| 510(k) No.: | K183245 |
| Device Common Name: | Flat Panel Digital Imager |
| Classification Name: | Stationary x-ray system |
| Device Class: | Class II |
| Device Code: | MQB |
| Regulation Number: | 21 CFR 892.1680 |
| Predicate Device: | Carestream DRX-1 System (with DRX Plus<br>3543 Detectors) |
| Manufactured by: | Carestream Health, Inc. |
| 510(k) No.: | K150766 (June 24, 2015) |
| Regulation Number: | 21 CFR 892.1680 |
| Classification Name: | Stationary x-ray system |
| Primary Product Code: | MQB |
#### Device Description:
The Carestream DRX-1 System is a diagnostic imaging system utilizing digital radiography (DR) technology that is used with diagnostic x-ray systems. The system consists of the Carestream DRX-1 System Console (operator console), flat panel digital imager (detector), and optional tether interface box. The system can be configured to register and use any of the two new DRX Plus 2530 and DRX Plus 2530C Detectors. Images captured with a flat panel digital detector can be communicated to the operator console via tethered or wireless connection.
{4}------------------------------------------------
### Indications for Use / Intended Use:
The Indications for Use for the device, as described in its labeling, are:
"The device is intended to capture for display radiographic images of human anatomy including both pediatric and adult patients. The device is intended for use in general projection radiographic applications wherever conventional screen-film systems or CR systems may be used. Excluded from the indications for use are mammography, fluoroscopy, and angiography applications."
The intended use for this device, as determined by descriptions and the proposed labeling contained in this submission, is similar to the Indications for Use statement provided above. The Carestream DRX-1 System with DRX Plus 2530 and DRX Plus 2530C Detectors is a diagnostic imaging system utilizing digital radiography (DR) technology that is used to capture x-rays for diagnostic procedures. We believe that the Carestream DRX-1 System with DRX Plus 2530 and DRX Plus 2530C Detectors and the predicate device have the same intended use.
The Indications for Use for the subject device is the same as for the predicate device and the intended use remains unchanged. Any variation in features or technical specifications have been identified and addressed through testing (described below) to support a substantial equivalence determination.
# Comparison of Technological Characteristics:
Based upon information provided within this submission, we believe that the Carestream DRX-1 System with DRX Plus 2530 and DRX Plus 2530C Detectors is substantially equivalent to the legally marketed Carestream DRX-1 System with the DRX Plus 3543 Detectors (predicate device) (K150766). Both the currently marketed DRX Plus 3543 Detectors and the new DRX Plus 2530 and DRX Plus 2530C Detectors are used in combination with the image processing software (Eclipse II; K180809) and user interface resident on the DRX-1 System Console component of the Carestream DRX-1 System. The system is used to directly capture conventional projected x-rays to generate digital images, regardless of which detector is being used. An image can be displayed on a preview monitor for viewing with any of the detectors. The system can transmit diagnostic images through a digital network for diagnostic viewing and printing regardless of which detector is used.
The predicate (DRX Plus 3543) detector shares similar design specifications except for the smaller physical size of the detector (25cm X 30cm) and the smaller pixel pitch of (98 X 98 microns) for the DRX Plus 2530 detectors versus (139 X 139 microns) for the DRX Plus 3543 detector.
The predicate device electronics contained separate circuit boards for the embedded controller, power, radio and image boards. The subject device electronics were redesigned to combine these into an All-in-One (AIO) board. In addition, the subject device has an updated readout integrated circuit (ROIC) on the chip on flex (COF). This is the next generation ROIC designed to produce lower noise and higher resolution images. Reuse of the DRX Plus electronics architecture allows for the reuse of the mature and field tested
{5}------------------------------------------------
embedded firmware from the predicate device. Reliability and performance improvements have resulted from this design change included in the subject device.
Both the predicate device and the subject device have housings constructed of aluminum. The predicate device contains a deep dish aluminum housing while the subject device contains an aluminum picture frame design. The change from the deep dish to the picture frame design was a necessary improvement to allow the service engineers full access to the back of the smaller sized detector. The design improvement has no impact to image quality or performance.
# Discussion of Testing:
The performance characteristics and operation / usability of the Carestream DRX-1 System with DRX Plus 2530 and DRX Plus 25303C Detectors were evaluated in clinical and nonclinical (bench) testing in accordance with FDA guidance document "Guidance for the Submission of 510(k)'s for Solid State Imaging Devices".
Non-clinical (bench) test results have demonstrated that the device conforms to its specifications. Acceptance criteria were determined based on desired performance with respect to image quality, intended use, workflow related performance, shipping performance, and general functionality and reliability, including both hardware and software requirements. Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device; therefore supporting a substantial equivalence determination. Acceptance criteria were identified for weight, pixel size, resolution, pixel pitch, total pixel area, usable pixel area, MTF (at various spatial resolutions), DQE (at various spatial resolutions), sensitivity, ghosting, boot-up time, operating temperature, exposure latitude, signal uniformity, and dark noise (ADC). Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device.
Additionally, a clinical study was performed to investigate the imaging performance of the Carestream DRX Plus 2530 Detector and the DRX Plus 2530C Detector (referred to as the "investigational devices within the clinical documentation") as compared to the currently marketed Carestream DRX Plus 3543 Detector (GOS) JK150766 (referred to as the "predicate device")]. The study was designed in accordance with the FDA Guidance titled "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices".
The predicate and subject devices were used for image capture at the University of Rochester Medical Center located in Rochester, NY. Adult cadavers (2) were used to perform triplicate acquisitions using the subject devices vs. the predicate device. Onehundred and sixty two (162) acquired images (cadaver and pediatric phantom) were included in this reader study. The cadaver images targeted exam types such as facial bones, nasal bones, skull, C-Spine, shoulder, clavicle, hip and extremity radiographs. Pediatric phantom images targeted exam types such as the chest and extremities. The images were acquired according to clinical protocol (AF3458), and were used to confirm diagnostic image quality using a Radlex rating scale. The images were evaluated by
{6}------------------------------------------------
three (3) board certified radiologists using a graduated 4 point scale based on diagnostic image quality.
The mean RadLex rating for the predicate and both subject devices are of Diagnostic (3) image quality and show very little variability between devices. 98% of the DRX Plus 2530 Detector RadLex rating responses are rated Diagnostic (3) or Exemplary (4). 96% of the DRX Plus 2530C Detector RadLex rating responses are rated Diagnostic (3) or Exemplary (4). In direct comparison to the predicate 71% of the DRX Plus 2530 Detector RadLex rating responses and 68% of the DRX Plus 2530C Detector RadLex rating responses are equivalent to the predicate or favor the subject device.
The two-sample equivalence tests yield results that confirm the equivalence between the mean ratings of the subject devices to the predicate as well as equivalence in either beam detect mode ("On" and "Off").
#### Summary:
The statistical test results and graphical summaries demonstrate the DRX Plus 2530 Detector and DRX Plus 2530C Detector deliver quality images that are equivalent in diagnostic quality compared with images obtained using a commercially available predicate device.
#### Conclusion:
In conclusion, the new detectors are equivalent to the DRX Plus 3543 Detectors in all applicable parameters recommended by the "Guidance for the Submission of 510(k)'s for Solid State Imaging Devices". Image quality parameters such as DQE, sensitivity, and MTF of the DRX Plus 2530 and DRX Plus 2530C Detectors demonstrate this. The tests used to determine these parameters were performed using industry standards.
The new detectors have been tested to conform to applicable existing specifications of the predicate DRX Plus 3543 Detectors. Ouality Assurance tests with traceable links to design inputs were performed to verify conformance to specifications. In addition, tests were included to confirm mitigation of risks identified in an extensive product hazard risk analysis based on results of a Failure Mode Effects Analysis.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.