K183203 · Aesculap, Inc. · HBE · Aug 9, 2019 · Neurology
Device Facts
Record ID
K183203
Device Name
ELAN 4 Tools
Applicant
Aesculap, Inc.
Product Code
HBE · Neurology
Decision Date
Aug 9, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Aesculap ELAN 4 Tools are intended for high speed cutting, sawing, and drilling of bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery
Device Story
ELAN 4 Tools are surgical cutting accessories designed for use with the ELAN 4 Electro Motor System. Tools include fluted, ball, conical, oval, barrel, acorn, pin, and twist drills, cutters, and discs. Manufactured from stainless steel, some variants feature a diamond chip coating. Operated by surgeons in clinical settings (Spine, ENT, Neuro, Maxillofacial), these tools attach to the motor system to perform mechanical bone resection. The device facilitates bone cutting, sawing, and drilling, aiding in surgical access or bone shaping. Benefits include high-speed performance and compatibility with existing motor systems.
Clinical Evidence
Bench testing only. Testing included rotating performance tests (up to 80,000 rpm) and bone temperature increase evaluation during milling/drilling/cutting under constant pressure and irrigation. Biocompatibility testing included cytotoxicity, sensitization, intracutaneous irritation, acute systemic toxicity, hemolysis, and pyrogen/endotoxin testing per ISO 10993-1. All tests met acceptance criteria.
Technological Characteristics
Materials: Stainless steel and diamond coating. Energy: Powered by ELAN 4 Electro Motor System. Dimensions: 0.5 mm to 8 mm. Sterilization: Gamma irradiation per ISO 11137-1/2. Packaging validated per ISO 11607; 5-year shelf life validated per ASTM F1980.
Indications for Use
Indicated for high speed cutting, sawing, and drilling of bone in Spine, ENT, Neuro, and Maxillofacial surgical procedures.
Regulatory Classification
Identification
Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 9, 2019
Aesculap, Inc. Kathy A. Racosky Sr. Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, PA 18034
Re: K183203
Trade/Device Name: Aesculap ELAN 4 Tools Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, And Their Accessories Regulatory Class: Class II Product Code: HBE Dated: July 11, 2019 Received: November 19, 2018
Dear Ms. Racosky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Matthew Krueger, M.S.E. Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K183203 Device Name Aesculap ELAN 4 Tools
Indications for Use (Describe)
Aesculap ELAN 4 Tools are intended for high speed cutting, sawing, and drilling of bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery
Type of Use (Select one or both, as applicable):
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Page 1 of 4
#### B. 510(k) SUMMARY (as required by 21 CFR 807.92)
| | Aesculap ELAN 4 Tools<br>July 17, 2019 |
|----------------------|--------------------------------------------------------------------------------------------------------------------|
| COMPANY: | Aesculap®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 |
| CONTACT: | Kathy A. Racosky<br>610-984-9291 (phone)<br>610-791-6882 (fax)<br>kathy.racosky@aesculapimplants.com |
| TRADE NAME: | ELAN 4 Tools |
| COMMON NAME: | Drills, Burrs, Trephines & Accessories (Simple, Powered) |
| CLASSIFICATION: | Class II |
| CLASSIFICATION NAME: | Powered simple cranial drills, burrs, trephines and their<br>accessories |
| REGULATION NUMBER: | 882.4310 |
| PRODUCT CODE: | HRE |
#### SUBSTANTIAL EQUIVALENCE
The ELAN 4 Tools are substantially equivalent to the primary predicate, ELAN 4 Tools (K172709) and predicate Anspach Dissection Tools (K113476).
### DEVICE DESCRIPTION
The ELAN 4 Tools are cutting devices designed for use with the ELAN 4 Electro Motor System. The ELAN 4 Tools are designed for cutting, sawing and drilling of bone. The ELAN 4 Tools have an attachment mechanism designed specifically for the type of motors and attachments with which they will be used. The ELAN 4 Tools are manufactured from stainless steel with some tools containing a coated layer of diamond chips. The ELAN 4 Tools are components of the ELAN 4 Electro Motor System (K152960).
#### INDICATIONS FOR USE
The ELAN 4 Tools are intended for high speed cutting, sawing, and drilling of bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.
{5}------------------------------------------------
Page 2 of 4
#### TECHNOLOGICAL CHARACTERISTICS (compared to predicate)
The ELAN 4 Tools are substantially equivalent to the predicate, ELAN 4 Tools (K172709) and the reference predicate Anspach Dissection Tools (K113476). The subject device is shown to be substantially equivalent and has the same performance characteristics to its primary predicate device through comparison in design, principles of operation and indications for use. The subject device has the same sizes and manufacturing materials as the reference predicate. A comparison table summarizing these characteristics can be found below.
The ELAN 4 Tools offer similar components when compared to the primary predicate, ELAN 4 Tools (K172709). The proposed device has some differences from the primary predicate device. The subject device ranges in sizes from 0.5 mm to 8 mm versus 0.6 mm to 9 mm for the predicate device. The subject device is manufactured from stainless steel with some tools containing a layer of diamond chips whereas the predicate is made from stainless steel. Similar to the devices that are subject to this submission, the predicate. Anspach Dissection Tool (K113476) are manufactured from the same materials; stainless steel with some tools containing a layer of diamond chips and range in the same sizes as the subject device offers.
| K# | Aesculap Inc.<br>ELAN 4 Tools | Aesculap Inc.<br>ELAN 4 Tools | Anspach Dissection Tool |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| | Proposed device | K172709 - Primary | K113476 |
| Indications | The ELAN Tools are<br>intended for high speed<br>cutting, sawing, and<br>drilling of bone in the<br>fields of Spine, ENT,<br>Neuro, and Maxillofacial<br>Surgery. | The ELAN Tools are<br>intended for high speed<br>cutting, sawing, and<br>drilling of bone in the<br>fields of Spine, ENT,<br>Neuro, and Maxillofacial<br>Surgery. | Dissection tools are<br>intended for cutting and<br>shaping bone including<br>spine and cranium. |
| Regulation number | 882.4310 | 882.4310 | 882.4310 |
| Product Code | HBE | HBE | HBE |
| Regulatory Panel | Neurology | Neurology | Neurology |
| Tools | | | |
| Burrs & drills | | | |
| --Size | 0.5 mm to 8 mm | 0.6 mm to 9 mm | 0.5 mm to 24.5 mm |
| --Type | fluted, ball, conical, oval,<br>barrel, acorn, pin, twist,<br>drills, cutters, and discs | fluted, ball, conical, oval,<br>barrel, acorn, pin, twist,<br>drills, cutters, and discs | fluted, ball, conical, oval,<br>barrel, acorn, pin, twist,<br>drills, cutters, and discs |
| Materials | Stainless steel & diamond | Stainless steel | Stainless steel, diamond,<br>carbide |
| Attachment mechanism | Yes | Yes | Yes |
| Sterile Single Use | Yes | Yes | Yes |
| Sterilization | GAMMA irradiation | GAMMA irradiation | unknown |
{6}------------------------------------------------
# PERFORMANCE DATA
# Biocompatibility Testing
Biocompatibility testing in accordance to ISO 10991-1 and Use of International Standard ISO 10993-1. "Biological evaluation of medical devices –1 Part 1: Evaluation and testing within a risk management process", Guidance for Industry and Food and Drug Administration Staff has been performed to demonstrate that the ELAN 4 Tools are substantially equivalent to other predicate devices.
Biocompatibility testing within this submission includes the following: Cytotoxicity, Sensitization, Intracutaneous Irritation, Acute Systemic Toxicity, Hemolysis, and Pyrogen testing. In addition, endotoxin testing was conducted with all testing meeting the acceptance criterion.
## Sterilization Validation
Product sterilization is validated in accordance with the ANSI/AAMI/ISO 11137-1:2006 -Sterilization of health care products -- Radiation -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices and ISO 11137-2:2006 - Sterilization of health care products -- Radiation -- Part 2: Establishing the sterilization dose.
The ELAN 4 Tools are packaged and sterilized by gamma irradiation.
Packaging is validated per ISO 11607. Testing per ASTM F1980 validated the 5 year shelf life. Accelerated aging data (5 years) for ELAN 4 Tools has been generated to support this submission.
Bacterial Endotoxin testing per Endotoxin ISO 10993-5 met the acceptance criterion.
# Bench Testing
The following bench testing was performed to demonstrate that the ELAN 4 Tools perform as intended and are substantially equivalent to the predicate devices.
| Test | Test Summary | Results |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|---------|
| Rotating performance test for<br>rotating tools | Design suitable for speeds up<br>to 80,000 rpm for 2 minutes | Pass |
| Measurement of temperature | With constant pressure and<br>regular irrigation, evaluated<br>bone temperature increase<br>after milling, drilling or<br>cutting | Pass |
{7}------------------------------------------------
Page 4 of 4
# CONCLUSION:
The biocompatibility and bench testing results along with a comparison between the technology, materials and intended use for the ELAN 4 Tools and the predicate devices demonstrate that the ELAN 4 Tools are as safe, as effective, and perform as well as the predicate devices. The biocompatibility and bench testing results and technological comparison demonstrates that the ELAN 4 Tools are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.