Stryker Navigation System with SpineMap Go Software Application, Fluoroscopy Trackers and Fluoroscopy Adapters, SpineMask Tracker
K183196 · Stryker Corporation · OLO · Feb 14, 2019 · Neurology
Device Facts
| Record ID | K183196 |
| Device Name | Stryker Navigation System with SpineMap Go Software Application, Fluoroscopy Trackers and Fluoroscopy Adapters, SpineMask Tracker |
| Applicant | Stryker Corporation |
| Product Code | OLO · Neurology |
| Decision Date | Feb 14, 2019 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 882.4560 |
| Device Class | Class 2 |
Intended Use
The Stryker Navigation System, when used with the SpineMap Go software application, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery. The system is indicated for any medical condition in which computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a rigid anatomical structure such as the pelvis, skull, or spine can be identified. The system assists in the positioning of instruments for procedures on the spine and pelvis, including the following procedures: - Screw Placement in the spine and pelvis - Needle Placement in the spine and pelvis
Device Story
Stryker Navigation System with SpineMap Go software provides intraoperative guidance for spinal/pelvic surgery. System inputs include 2D fluoroscopic images (via C-arm/analog video/NaviLink2D) and spatial data from wireless optical trackers (SpineMask Tracker, Fluoroscopy Trackers). The system calculates spatial relationships between C-arm, patient, and surgical instruments; displays instrument position relative to anatomy on a monitor. Used in OR by surgeons. Benefits include precise instrument/implant positioning. SpineMask Tracker (flexible, adhesive) enables patient registration/tracking for minimally invasive lumbar/thoracic procedures. Fluoroscopy adapters/trackers correct image distortion and compute C-arm geometry. Software provides augmented navigation (virtual ruler, 3D models).
Clinical Evidence
No clinical testing was performed. Bench testing included design verification, validation in cadaver labs/simulated use, software V&V per IEC 62304, electrical safety (IEC 60601-1), and EMC (IEC 60601-1-2).
Technological Characteristics
Infrared optical active sensing; wireless tracking. Platform includes computer, camera, monitor, IO tablet. Connectivity: DICOM, Ethernet (NaviLink 2D), analog video. Software: SpineMap Go 1.0-9/009 on Windows 8.1. Complies with IEC 60601-1:2012 (3.1 ed) and IEC 60601-1-2:2014 (4th ed).
Indications for Use
Indicated for patients requiring open or percutaneous computer-assisted surgery on the spine or pelvis, including screw and needle placement, where rigid anatomical structures can be identified. Applicable to any medical condition where computer-assisted planning/surgery is appropriate.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
- Stryker Navigation System - Spine and Fluoroscopy System with Fluoroscopy 2.3 Software (K012380)
- SpineMask Tracker (K141941)
Reference Devices
- SpineMap 3D 3.1 software application (K172034)
Related Devices
- K172034 — Stryker Spine Navigation System with SpineMap 3D software application, OrthoLock, nGenius Spine Clamp, Navigated Drill Guide Set, Navigated Xia 3 Awl Tap, Navigated Xia 3 Serrato Tap · Stryker Corporation · Oct 18, 2017
- K230783 — Anatase Spine Surgery Navigation System · Remex Medical Corp. · Apr 21, 2023
- K233513 — Remex Spine Surgery Navigation System II · Remex Medical Corp. · May 1, 2024
- K012380 — MODIFICATION TO: STRYKER NAVIGATION SYSTEM, STRYKER FLUOROSCOPY SYSTEM,VIRTUAL FLUOROSCOPY · Stryker Leibinger · Dec 3, 2001
- K251315 — Anatase Spine Surgery Navigation System and Remex Spine Surgery Navigation System II · Remex Medical Corp. · May 29, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below it.
February 14, 2019
Stryker Corporation Andrea Wallen-Gerding Principal Regulatory Affairs Specialist Bötzinger Straße 41 Freiburg, D-79111 DE
Re: K183196
Trade/Device Name: Stryker Navigation System with SpineMap Go Software Application, Fluoroscopy Trackers and Fluoroscopy Adapters, SpineMask Tracker Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: November 16, 2018 Received: November 19, 2018
Dear Andrea Wallen-Gerding:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jesse Muir-S
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183196
## Device Name
Fluoroscopy Adapters and Fluoroscopy Trackers
## Indications for Use (Describe)
The Fluoroscopy Adapters and Fluoroscopy Trackers are intended to be used as accessories to the Stryker Navigation System. The Fluoroscopy Tracker is connected to the Fluoroscopy Adapter which is attached to the C-arm fluoroscopy system. The Fluoroscopy Tracker transmits signals that are used by the Stryker Navigation System to calculate the spatial relationship between the C-arm and the patient.
The Fluoroscopy Adapter and Fluoroscopy Tracker may be used as part of the Stryker Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183196
Device Name
Stryker Navigation System with SpineMap Go software application
Indications for Use (Describe)
The Stryker Navigation System, when used with the SpineMap Go software application, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery.
The system is indicated for any medical condition in which computer assisted planning and surgery may be appropriate. The system can be used for intraoperative quidance where a rigid anatomical structure such as the pelvis, skull, or spine can be identified.
The system assists in the positioning of instruments for procedures on the spine and pelvis, including the following procedures:
- Screw Placement in the spine and pelvis
- Needle Placement in the spine and pelvis
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183196
Device Name SpineMask Tracker
Indications for Use (Describe)
The Stryker SpineMask™ Tracker is intended to be used as an accessory to the Stryker Spine Navigation Systems. It is placed onto the patient's skin dorsal to the spine.
· In combination with intraoperative imaging devices, it enables patient registration for open or percutaneous computer assisted surgery.
· When used for patient tracking, the Stryker SpineMask™ Tracker supports minimally invasive procedures on the lumbar and thoracic spine.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------|--------------------------------------------------------------|
| <div> </div> | |
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
### Submitter Information 6.0
### This Premarket Notification is submitted by: 6.1
Stryker Leibinger GmbH & Co. KG
Bötzinger Straße 41
79111 Freiburg, Germany
### 6.2 Contact Information
Contact Name: Andrea N. Wallen-Gerding Office Telephone: (269) 389-2884 Fax: (269) 389-5412 Email: andrea.wallengerding@stryker.com Date Prepared: November 16, 2018
### Device Name 6.3
# Table 6-1: Device Name
| Subject (Modified) Device Information | |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Trade/ Proprietary Name | Stryker Navigation System with SpineMap Go software<br>application, SpineMask Tracker, Fluoroscopy Adapters,<br>Fluoroscopy Trackers |
| Common Name | Spine Navigation System, Fluoroscopy Adapters,<br>Fluoroscopy Trackers, SpineMask Tracker |
| Classification | Class II |
| Classification Product Code | OLO |
| Classification Name | Orthopedic Stereotaxic Instrument |
| Classification Regulation | 21 CFR 882.4560 |
| Review Panel | Orthopedic |
### Predicate Devices 6.4
The following are the legally marketed predicate devices for the subject devices included in this Traditional 510(k):
| Subject Device | Predicate Device Trade Name | 510(k) | Predicate Device<br>Product Code |
|---------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|---------|----------------------------------|
| Stryker Navigation System<br>with the SpineMap Go<br>Software Application | Stryker Navigation System - Spine<br>and Fluoroscopy System with<br>Fluoroscopy 2.3 Software | K012380 | HAW |
| SpineMask Tracker | SpineMask Tracker | K141941 | OLO |
| Fluoroscopy Adapter (Philips<br>BV300, 9") | Fluoroscopy Adapter (Philips<br>BV300, 9") | K012380 | HAW |
Table 6-2: Predicate Device List
{6}------------------------------------------------
| Subject Device | Predicate Device Trade Name | 510(k) | Predicate Device<br>Product Code |
|-----------------------------------------------|-----------------------------------------------|---------|----------------------------------|
| Fluoroscopy Adapter (OEC<br>9800, 9") | Fluoroscopy Adapter (OEC 9800,<br>9") | K012380 | HAW |
| Fluoroscopy Adapter (Ziehm<br>Exposcope, 9") | Fluoroscopy Adapter (Ziehm<br>Exposcope, 9") | K012380 | HAW |
| Fluoroscopy Adapter (Siemens<br>ISO-C, 9") | Fluoroscopy Adapter (Siemens<br>ISO-C, 9") | K012380 | HAW |
| Fluoroscopy Adapter (Philips<br>BV300, 12") | Fluoroscopy Adapter (Philips<br>BV300, 12") | K012380 | HAW |
| Fluoroscopy Adapter (OEC,<br>12") | Fluoroscopy Adapter (OEC, 12") | K012380 | HAW |
| Fluoroscopy 3D Kit (Siemens<br>ISO-C, 9") | Fluoroscopy 3D kit (Siemens ISO-<br>C, 9") | K012380 | HAW |
| Fluoroscopy 3D Adapter<br>(Siemens ISO-C, 9") | Fluoroscopy 3D Adapter (Siemens<br>ISO-C, 9") | K012380 | HAW |
| 3D Installation Set (Siemens<br>ISO-C) | 3D Installation Set (Siemens ISO-<br>C) | K012380 | HAW |
| Fluoroscopy Tracker Kit, 9" | Fluoroscopy Tracker Kit, 9" | K012380 | HAW |
| Fluoroscopy Tracker Kit, 12" | Fluoroscopy Tracker Kit, 12" | K012380 | HAW |
### 6.5 Device Descriptions
### 6.5.1 Stryker Navigation System with the SpineMap Go™ Software Application Overview
The Stryker Navigation System with the SpineMap Go software application is intended for use as an image guided surgery system to enable open or percutaneous computer assisted spinal surgery. It assists the surgeon in positioning of instrumentation during spinal surgeries. The system provides intraoperative guidance to the surgeon using augmented fluoroscopy. Wireless optical tracking technology is used to estimate the position of navigated surgical instruments and displaying their position on medical fluoroscopic images from mobile C-arm.
The Stryker Navigation System with SpineMap Go software is comprised of a platform, SpineMap Go software, navigated instruments (e.g. patient/instrument trackers, pointers), and accessories. The system uses wireless optical tracking technology to display the intraoperative location of navigated surgical instruments relative to mobile C-arm fluoroscopy images. The platform consists of a computer, camera, monitor and IO (input/output) Tablet. The SpineMap Go software is dedicated for spinal procedures as defined in the Indications for Use. Required navigated instruments include instruments such as a patient tracker, an instrument tracker, and pointers. An instrument battery is also required when a battery powered navigated instrument or calibration device is used.
#### 6.5.1.1 SpineMap GoTM Software Application
The SpineMap Go software application is a required part of the Stryker Navigation System. It is installed by a Stryker representative on the platform. The SpineMap Go software application is used on a platform and interfaces with Stryker navigated instruments and accessories. It is compatible with the
{7}------------------------------------------------
Nav3i Platform family, which includes the NAV3i. NAV3. and NavSuite3.
SpineMap Go is an interactive software application that provides the functions necessary to conduct the indicated spinal procedures. The software application implements methods for image adjustment, image registration, and instrument navigation.
The SpineMap Go Software Application provides new features including the integration of the SpineMask Tracker and the complete Stryker navigation enabled spine instrument portfolio, an updated Image Acquisition component for 2D C-arm support, support for existing C-arm Fluoroscopy Trackers, an updated user interface for improved usability, an updated software architecture, and improved cybersecurity aspects.
#### 6.5.2 SpineMask Tracker
The SpineMask Tracker is a sterile, single use device. It is intended to be used as an accessory to the Stryker Spine Navigation Systems which include the SpineMap 3D and SpineMap Go software applications. (The SpineMask Tracker is currently included as an accessory to the SpineMap 3D 3.1 software application which was cleared per 510(k) K172034.)
The SpineMask Tracker is a flexible, patient tracker that is attached to a patient's skin and provides non-invasive patient tracking of the thoracic and lumbar spine. It is applied to the skin, dorsal to the patient's spine, via an integrated adhesive tape.
The SpineMask Tracker includes infrared LED's that are positioned on the flexible portion of the device. The LED positions are measured by the navigation system after the SpineMask Tracker is applied to the patient's back. Upon activation of the SpineMask Tracker, the LED positions are measured (i.e., captured) by the navigation system. While being tracked, the LED positions are measured and compared to their initial positions as determined during the capturing step. The software is able to detect whether the flexible SpineMask Tracker is deformed when the position of the corresponding LED pairs differ significantly. If this difference is beyond a specific threshold, the LEDs are not used for patient tracking.
#### 6.5.3 Fluoroscopy Trackers and Fluoroscopy Adapters
The Fluoroscopy Trackers and the Fluoroscopy Adapters serve to connect a C-arm with the Stryker navigation system. The Fluoroscopy Adapter is mounted onto the C-arm image intensifier to interface with the Fluoroscopy Tracker. The Fluoroscopy Tracker transmits signals that are received by the navigation camera. The navigation system uses the signals to calculate the spatial relationship between the C-arm and the patient. The position of the C-arm, relative to the patient's anatomy, is displayed on the navigation system monitor.
The radiopaque fluoroscopy markers within the Fluoroscopy Phantom on the tracker are used to correct the image distortion caused by the image intensifier sensor technology and to compute the geometry of the C-arm. The computed geometry of the C-arm is defined by the position of the focal
{8}------------------------------------------------
point relative to the image sensor. The position of the focal point varies based on the C-arm position and gravity deforming the C-arm. The focal point is used later to compute the correct projection of the 3D tool onto the 2D fluoroscopy image.
#### 6.6 Indications for Use
#### 6.6.1 Stryker Navigation System with the SpineMap Go Software Application
The Stryker Navigation System, when used with the SpineMap Go software application, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery.
The system is indicated for any medical condition in which computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure such as the pelvis, skull, or spine can be identified.
The system assists in the positioning of instruments for procedures on the spine and pelvis, including the following procedures:
- Screw Placement in the spine and pelvis
- Needle Placement in the spine and pelvis
#### 6.6.2 SpineMask Tracker
The Stryker SpineMask™ Tracker is intended to be used as an accessory to the Stryker Spine Navigation Systems. It is placed onto the patient's skin dorsal to the spine. In combination with intraoperative imaging devices, it enables automatic patient registration for open or percutaneous computer assisted surgery. When used for patient tracking, the Stryker SpineMaskTM Tracker supports minimally invasive procedures on the lumbar and thoracic spine.
#### 6.6.3 Fluoroscopy Adapters and Fluoroscopy Trackers
The Fluoroscopy Adapters and Fluoroscopy Trackers are intended to be used as accessories to the Stryker Navigation System. The Fluoroscopy Tracker is connected to the Fluoroscopy Adapter which is attached to the Carm fluoroscopy system. The Fluoroscopy Tracker transmits signals that are used by the Stryker Navigation System to calculate the spatial relationship between the C-arm and the patient.
The Fluoroscopy Adapter and Fluoroscopy Tracker may be used as part of the Stryker Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified.
#### 6.7 Comparison of Technological Characteristics
A comparison of the technological characteristics of the subject devices included in the scope of this Traditional 510(k) is included in the tables below.
{9}------------------------------------------------
- 6.7.1 Technological Comparison between the Stryker Navigation System with SpineMap Go Software and the Stryker Navigation System - Spine and Fluoroscopy System with Fluoroscopy 2.3 Software
The technological comparison between the subject device, Stryker Navigation System with SpineMap Go software application and the predicate device (Stryker Navigation System - Spine and Fluoroscopy System with Fluoroscopy 2.3 Software) is included in Table 6-3 below. The system accuracy for both the predicate and subject devices are based on non-clinical test data.
Table 6-3: Technological Comparison between Stryker Navigation System with SpineMap Go software application (Subject Device) and the Predicate Device
| Item | Subject Device:<br>Stryker Navigation System with SpineMap Go<br>Software Application | Predicate Device:<br>Stryker Navigation System – Spine and<br>Fluoroscopy System with Fluoroscopy 2.3<br>Software (K012380) |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Stryker Navigation System, when used with<br>the SpineMap Go software application, is intended<br>as a planning and intraoperative guidance system<br>to enable open or percutaneous computer assisted<br>surgery.<br>The system is indicated for any medical condition<br>in which computer assisted planning and surgery<br>may be appropriate. The system can be used for<br>intraoperative guidance where a reference to a<br>rigid anatomical structure such as the pelvis, skull,<br>or spine can be identified.<br>The system assists in the positioning of<br>instruments for procedures on the spine and pelvis,<br>including the following procedures:<br>• Screw Placement in the spine and pelvis<br>• Needle Placement in the spine and pelvis | The Stryker Navigation System is intended<br>as a planning and intraoperative guidance<br>system to enable open or percutaneous image<br>guided surgery. The system is indicated for<br>any medical condition in which the use of<br>image guided surgery may be appropriate,<br>and where a reference to a rigid anatomical<br>structure such as the skull, or vertebra can be<br>identified relative to medical images.<br>The Stryker Navigation System – Spine &<br>Fluoroscopy Module supports, but is not<br>limited to the following surgical procedures:<br>• Pedicle Screw Placement<br>• Navigation<br>• Precisely positioning instruments or<br>implants during orthopedic surgery,<br>such as operations performed with<br>spinal structure, hip and bones in the<br>upper and lower extremities and long<br>bones. |
| Main System<br>Components | • Platform<br>• SpineMap Go Software Application<br>• Navigated Instruments<br>• Accessories | • Platform<br>• Fluoroscopy 2.3 Software Application<br>• Navigated Instruments<br>• Accessories |
| Modes of<br>Operation | • Patient Preparation<br>• System Set-up<br>• Image Registration<br>• Navigation | • Patient Preparation<br>• System Set-up<br>• Image Registration<br>• Navigation |
| Localizing<br>and Tracking<br>Technology | Infrared optical active sensing technology:<br>Infrared light emitted by diodes placed in specific<br>locations on navigated surgical instruments is<br>sensed by a camera array (navigation camera) on<br>the platform, which allows for computation of the<br>spatial information | Infrared optical active sensing technology:<br>Infrared light emitted by diodes placed in<br>specific locations on navigated surgical<br>instruments is sensed by a camera array<br>(navigation camera) on the platform, which<br>allows for computation of the spatial<br>information |
| Operating<br>Principle | • The software is installed on the computer that is<br>part of the platform | • The software is installed on the computer<br>that is part of the platform |
{10}------------------------------------------------
| Item | Subject Device:<br>Stryker Navigation System with SpineMap Go<br>Software Application | Predicate Device:<br>Stryker Navigation System – Spine and<br>Fluoroscopy System with Fluoroscopy 2.3<br>Software (K012380) | | | | | | | | | | | | | | | | | | | | | |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| | • Images are imported in DICOM (Siemens C-arm with proprietary NaviLink 2D protocol) or analog video connection<br>• The software displays the images with navigational information on a monitor | • Images are imported in DICOM (Siemens C-arm with proprietary NaviLink 2D protocol) or analog video connection<br>• The software displays the images and planned items with navigational information on a monitor | | | | | | | | | | | | | | | | | | | | | |
| System<br>Accuracy | The system is designed to work in the working space with a mean accuracy of 2 mm point and 2º angular axis displacement within the registration zone<br>The 95th percentile of the point displacement is ≤3 mm and ≤3° for angular axis displacement. | Mean navigation accuracy of ± 2mm point (tip) displacement and ± 2° angular axis displacement. | | | | | | | | | | | | | | | | | | | | | |
| | Positional<br>displacement<br>(mm) Trajectory<br>Angle<br>Displacement<br>(Degrees) Median 0.54 0.51 Interquartile<br>Range 0.85 0.50 99th<br>Percentile 2.17 2.09 Mean 0.68 0.56 Standard<br>Deviation 0.59 0.37 99th<br>Confidence<br>Interval<br>Upper<br>Bound 2.14 1.52 | | | | | | | | | | | | | | | | | | | | | | |
| Supported<br>Imaging<br>Modalities | • C-arm with image intensifier | • C-arm with image intensifier | | | | | | | | | | | | | | | | | | | | | |
| Registration<br>Features | • Image Registration | • Image Registration | | | | | | | | | | | | | | | | | | | | | |
| Energy<br>Source | • Main Power: Alternating Current (AC) power supply, 100/240 V and 50/60 Hz<br>• Off-the-Shelf uninterruptable power supply for power interruptions ≤ 6 minutes | • Main Power: Alternating Current (AC) power supply, 100/240 V and 50/60 Hz<br>• Off-the-Shelf uninterruptable power supply for power interruptions ≤ 6 minutes | | | | | | | | | | | | | | | | | | | | | |
| Intended Use<br>Environment | Operating Room (OR) | Operating Room (OR) | | | | | | | | | | | | | | | | | | | | | |
| Input | Analog Video Input (BNC)<br>NaviLink2D (Ethernet) | Analog Video Input (BNC)<br>NaviLink2D (Ethernet) | | | | | | | | | | | | | | | | | | | | | |
| Output | • 2D fluoroscopy image,<br>• Augmented navigation information<br>- Instrument<br>- Screw (3D model)<br>- Virtual Ruler | • 2D fluoroscopy image,<br>• Augmented navigation information<br>- Instrument<br>- Screw (Cylinder)<br>• C-arm position | | | | | | | | | | | | | | | | | | | | | |
{11}------------------------------------------------
| Item | Subject Device:<br>Stryker Navigation System with SpineMap Go<br>Software Application | Predicate Device:<br>Stryker Navigation System – Spine and<br>Fluoroscopy System with Fluoroscopy 2.3<br>Software (K012380) |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| User Interface | • Black-style graphical user interface<br>• 16:9 screen ratio<br>• One tab per task concept from left to right on<br>top of screen<br>• Image toolbox for navigated tool visualization<br>• Independent image adjustment options for each<br>view<br>• Virtual keyboard<br>• Mouse<br>• IO Tablet with touch screen, USB ports, and<br>CD/DVD drive<br>• Buttons on Navigated Instruments | • Gray-style graphical user interface<br>• 4:3 screen ratio<br>• Movement between task concepts<br>controlled by forward and back arrows.<br>• Current task panel on the right<br>• Image toolbar with image tools<br>• Virtual keyboard<br>• Mouse<br>• IO Tablet with touch screen, USB ports,<br>and CD/DVD drive<br>• Buttons on Navigated Instruments |
| Camera<br>Working<br>Space | 1.25 m (meters) | 1.25 m (meters) |
## 6.7.2 Technological Comparison between the SpineMap Go Software and the Fluoroscopy 2.3 Software
The technological comparison between the subject device software (SpineMap Go software) and the predicate device is included in Table 6-4 below. The Fluoroscopy 2.3 software received 510(k) clearance per 510(k) number K012380.
## Table 6-4: Technological Comparison between SpineMap Go software (Subject Device) and the Predicate Device Software
| Item | Subject Device:<br>SpineMap Go Software | Predicate Device:<br>Fluoroscopy Software 2.3 (K012380) |
|------------------------------|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Computer Operating<br>System | • Windows 8.1 (SPC 3.1) | • Windows XP Embedded (SPC 3.0 - Stryker<br>Personal Computer)<br>• Windows 8.1 (SPC 3.1)<br>• Off-the-Shelf Service Pack 3 |
| Software Version | SpineMap Go 1.0-9/009 | Virtual Fluoroscopy 2.3-7/009 |
### 6.7.3 Technological Comparison of the SpineMask Tracker, the Fluoroscopy Trackers, and Fluoroscopy Adapter with their Predicate Devices
The design of the SpineMask Tracker and the Fluoroscopy Trackers were updated to meet IEC 60601-1:2012 edition 3.1 and IEC 60601-1-2:2014 4th edition requirements. No changes were made to the design of the Fluoroscopy Adapters. In addition, the Indications for Use statements are being updated for the SpineMask Tracker and the Fluoroscopy Trackers and Fluoroscopy Adapters. The SpineMask Tracker's indications for use statement is being updated to allow it to be used with 2D spine navigation. The indications for use for the Fluoroscopy Trackers and the Fluoroscopy
{12}------------------------------------------------
Adapters is being updated for clarification purposes only. No changes are being made to the performance or risk profile of these devices.
### Summary of Non-Clinical Testing 6.8
- The function and performance of the subject devices (Stryker Navigation 6.8.1 System with SpineMap Go software application, SpineMask Tracker, and Fluoroscopy Trackers and Fluoroscopy Adapters) have been evaluated through non-clinical design verification and validation testing. The results of the evaluation tests demonstrate that the subject devices successfully meet the requirements of their intended use.
- 6.8.2 Additional testing was performed on the subject devices to ensure the subject devices met their design requirements. A summary of the testing and the results are included in Table 6-5 below.
| Item | Summary of Testing | | | | | | | | | | | | | | | | | | | | | |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Intended Use/ User<br>Needs | The subject devices were validated with intended users in cadaver<br>labs or simulated use tests to ensure the user needs and intended use<br>requirements were met. All requirements were met and no new issues<br>of safety or effectiveness were raised. | | | | | | | | | | | | | | | | | | | | | |
| Accuracy | The system is designed to work in the working space with a mean<br>accuracy of 2 mm point and 2º angular axis displacement within the<br>registration zone<br>The 95th percentile of the point displacement is ≤3 mm and ≤3° for<br>angular axis displacement. | | | | | | | | | | | | | | | | | | | | | |
| | Positional<br>displacement<br>(mm) Trajectory Angle<br>Displacement<br>(Degrees) Median 0.54 0.51 Interquartile Range 0.85 0.50 99th Percentile 2.17 2.09 Mean 0.68 0.56 Standard Deviation 0.59 0.37 99th Confidence<br>Interval Upper<br>Bound 2.14 1.52 | | | | | | | | | | | | | | | | | | | | | |
| Safety | Verified the effectiveness of all risk controls determined in the device<br>risk analysis. No new issues of safety or effectiveness were raised. | | | | | | | | | | | | | | | | | | | | | |
| General Requirements<br>and Performance | Verified all components against their design specifications. All<br>requirements were met and no new issues of safety or effectiveness<br>were raised. | | | | | | | | | | | | | | | | | | | | | |
| Software | Software verification and validation testing was conducted as required<br>by IEC 62304 and FDA guidance on general principles of software<br>validation, January 11, 2002. All requirements were met and no new<br>issues of safety or effectiveness were raised. | | | | | | | | | | | | | | | | | | | | | |
| Electrical Safety | Verified conformance to IEC 60601-1:2012 Edition 3.1 | | | | | | | | | | | | | | | | | | | | | |
| Electromagnetic<br>Compatibility | Verified conformance to IEC 60601-1-2:2014 Edition 4 | | | | | | | | | | | | | | | | | | | | | |
## Table 6-5: Summary of Non-clinical Testing
{13}------------------------------------------------
### Summary of Clinical Testing 6.9
No clinical testing was performed.
# 6.10 Conclusion
The subject devices, Stryker Navigation System with SpineMap Go software application, SpineMask Tracker, Fluoroscopy Trackers and Fluoroscopy Adapters perform as intended and are substantially equivalent to their respective predicate device with respect to intended use, design, principles of operation, technology, materials, and performance. No new issues of safety or effectiveness have been raised.