K183187 · Sfm Medical Devices GmbH · LHI · Mar 15, 2019 · General Hospital
Device Facts
Record ID
K183187
Device Name
nextaro Transfer System, nextaro va
Applicant
Sfm Medical Devices GmbH
Product Code
LHI · General Hospital
Decision Date
Mar 15, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The nextaro® Transfer System and nextaro® va is indicated for the transfer and mixing of drugs contained in vials.
Device Story
The nextaro® Transfer System and nextaro® va are sterile, single-use devices designed for the transfer and reconstitution of drugs in vials. The Transfer System consists of two components with plastic spikes to perforate vial seals, connected by a Luer adapter to create a closed fluid path. It utilizes vacuum pressure for solvent transfer between a solvent vial and a pharmaceutical vial. The nextaro® va is a standalone component used to connect a syringe to a single pharmaceutical vial for solvent transfer. Both devices feature an in-line filter to retain particles from vial seals or undissolved medication. After reconstitution via gentle swirling, the user aspirates the solution into a standard syringe. These devices are intended for use by healthcare professionals in clinical settings to facilitate safe drug preparation and reduce particulate contamination during the transfer process.
Clinical Evidence
Bench testing only. No clinical data provided. Performance was validated through testing against ISO 22413, ISO 594-2, ISO 8536-4, ISO 11607, and various ASTM standards for seal integrity, packaging, and transport. Biocompatibility was evaluated per ISO 10993-1, and sterility was validated per ISO 11135:2014.
Technological Characteristics
Materials: Polypropylene, TPE, and Polyethylene terephthalate (PET) filter. Sensing/Actuation: Manual operation using vacuum pressure or syringe-applied pressure. Dimensions: Designed for 20 mm vials. Connectivity: None. Sterilization: Ethylene oxide (EtO) to SAL 10^-6. Software: None.
Indications for Use
Indicated for the transfer and mixing of drugs contained in vials.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.
March 15, 2019
sfm medical devices GmbH % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313
Re: K183187
Trade/Device Name: nextaro® Transfer System, nextaro® va Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: LHI Dated: February 18, 2019 Received: February 21, 2019
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
# K183187 - Mark Job
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Sapana Patel -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183187
Device Name
The nextaro® Transfer System and nextaro® va
Indications for Use (Describe) The nextaro® Transfer System and nextaro® va is indicated for the transfer and mixing of drugs contained in vials.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510(k) 510k Summary
# 510K SUMMARY K183187
| Submitter's Name<br>and Address | sfm medical devices GmbH<br>Brueckenstrasse 5<br>Waechtersbach Hessen<br>GERMANY 63607 |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | sfm medical devices GmbH<br>Brueckenstrasse 5<br>Waechtersbach Hessen<br>GERMANY 63607 |
| Contact Person: | Phil Triolo<br>Phone: 801-699-9846<br>Facsimile: 801-328-2399 |
| Date Prepared: | March 14, 2019 |
| Device: | nextaro® Transfer System and nextaro® va |
| FDA Product Code: | LHI |
| Class: | II |
| Common or Usual Name: | I.V. Fluid Transfer Set |
| Regulation Description: | Intravascular Administration Set |
| Regulation Number: | 21 CFR 880.5440 |
| Legally Marketed Predicate: | Mix2Vial® Transfer Device (K031861)<br>West Pharmaceutical Services, Inc. |
| Reason for Submission: | New Device |
| Device Description: | The nextaro® Transfer System and nextaro® va is indicated for the transfer and<br>mixing of drugs contained in vials. The nextaro® Transfer System and nextaro® va is |
practically one system with two variants which are to be used independently.
{4}------------------------------------------------
| Technological<br>Characteristics: | The nextaro® Transfer System consists of two components already assembled<br>(screwed together) in the delivery condition. Each of the components has a plastic<br>spike which is used to perforate the seals of a solvent vial, or lyophilizated<br>pharmaceutical vial, respectively. The components have a male and female Luer<br>adapter to create an air-tight inner media-carrying system with open ends for fluid<br>transfer. Both components are equipped with a filter element to prevent particles<br>(fragments of vial seals, undissolved pharmaceutical) from being administered into<br>the patient's circulatory system. |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | To be used as intended, the spike of the blue ("upper") part of the nextaro®<br>Transfer System is used to perforate the seal of the solvent vial. The assembly is<br>flipped vertically and the spike of the white ("lower") part of the device is used to<br>perforate the seal of the pharmaceutical vial. The solvent transfer process starts<br>immediately due to the vacuum in the pharmaceutical vial. |
| | After solvent transfer and reconstitution of the pharmaceutical by gentle swirling,<br>the blue component of the device is removed from the white part by unscrewing, a<br>standard syringe with a male Luer is attached to the exposed female Luer of the<br>device to aspirate the ready-to-use solution. |
| | The nextaro® va is a standalone device. Basically, it is the lower part of the<br>nextaro® Transfer System, but without the thread which the latter has to connect<br>the upper part. After connecting the nextaro® va to a pharmaceutical vial by<br>pushing it downwards, a solvent can be transferred by a syringe via the female Luer.<br>After solvent transfer and reconstitution of the pharmaceutical by gentle swirling, a<br>standard syringe with a male Luer is attached to the female Luer of the device to<br>aspirate the ready-to-use solution. |
Indications for The nextaro® Transfer System and nextaro® va is indicated for the transfer and Use: mixing of drugs contained in vials.
| Areas for Comparison | Predicate Device<br>Mix2Vial<br>K031861 | Proposed Device<br>nextaro®<br>Transfer System | Proposed Device<br>nextaro® va | Comparison |
|-----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------|-------------------------------------------|
| Indications for Use | | | | |
| Intended use or<br>indications | Intended for<br>transferring and<br>mixing drugs<br>contained in two<br>vials | Indicated for the<br>transfer and mixing<br>of drugs contained<br>in vials | Indicated for the<br>transfer and mixing<br>of drugs contained<br>in vials | See Discussion of<br>Differences |
| Regulatory Information | | | | |
| Device Classification<br>Name | Set I.V. Fluid<br>Transfer | Set I.V. Fluid<br>Transfer | Set I.V. Fluid<br>Transfer | Equivalent |
| Regulation Number | 880.5440 | 880.5440 | 880.5440 | Equivalent |
| Regulation Description | Intravascular<br>Administration Set | Intravascular<br>Administration Set | Intravascular<br>Administration Set | Equivalent |
| Product Code | LHI | LHI | LHI | Equivalent |
| Areas for Comparison | Predicate Device<br>Mix2Vial<br>K031861 | Proposed Device<br>nextaro®<br>Transfer System | Proposed Device<br>nextaro® va | Comparison |
| Regulatory Medical<br>Specialty | General Hospital | General Hospital | General Hospital | Equivalent |
| Device Class | 2 | 2 | 2 | Equivalent |
| GMP Exempt | No | No | No | Equivalent |
| Design Features | | | | |
| Uses vacuum pressure | Yes | Yes | No | See Discussion of<br>Differences |
| Vial size the product is<br>to be used with | 20 mm /<br>14 mm | 20 mm | 20 mm | Equivalent<br>to 20 mm model |
| Vial adapter / vial access<br>component | Yes | Yes | Yes | Equivalent |
| In-line liquid filter | Yes | Yes | Yes | Equivalent |
| Mesh opening | 15 μm | 11 μm | 11 μm | See Discussion of<br>Differences |
| Open area | 9 % | 5 % | 5 % | See Discussion of<br>Differences |
| Female Luer adapter for<br>connection to syringes | Yes | Yes | Yes | Equivalent |
| Luer adapter and vial<br>adapter are integrated<br>in one component | Yes | Yes | Yes | Equivalent |
| Needleless access to vial | Yes | Yes | Yes | Equivalent |
| Sterile, biocompatible<br>fluid path | Yes | Yes | Yes | Equivalent |
| Manufactured by plastic<br>injection molding | Yes | Yes | Yes | Equivalent |
| Single use, sterile | Yes | Yes | Yes | Equivalent |
| Sterilization | Gamma / EtO | EtO | EtO | Equivalent to one of<br>the methods (EtO) |
| Sterility Assurance Level | SAL 10-6 | SAL 10-6 | SAL 10-6 | Equivalent |
| Biocompatibility | Pass | Pass | Pass | Equivalent |
| Shelf Life | > 3 years | 5 years | 5 years | Equivalent |
| Principles of Operation | | | | |
| Manually operated | Yes | Yes | Yes | Equivalent |
| Mechanically connected<br>to a drug vial | Yes | Yes | Yes | Equivalent |
| Transfers solvent<br>manually using a syringe<br>plunger rod | No | No | Yes | similar |
| Drug reconstitution<br>through manual<br>agitation of vial while<br>connected to device | Yes | Yes | Yes | Equivalent |
| Mixed drug is manually<br>aspirated into a syringe<br>barrel using the syringe<br>plunger rod | Yes | Yes | Yes | Equivalent |
| Areas for Comparison | Predicate Device<br>Mix2Vial<br>K031861 | Proposed Device<br>nextaro®<br>Transfer System | Proposed Device<br>nextaro® va | Comparison |
| Single Use | Yes | Yes | Yes | Equivalent |
| Materials | | | | |
| Transfer device | Polycarbonate | Polypropylene, TPE | Polypropylene | Similar |
| In-line Filter | Polyethylene<br>terephthalate | Polyethylene<br>terephthalate | Polyethylene<br>terephthalate | Equivalent |
| Packaging | Blister Pack: PETG<br>Lid: Tyvek | Blister Pack: PETG<br>Lid: Tyvek | Blister Pack: PETG<br>Lid: Tyvek | Equivalent |
# Substantial Equivalence Comparison:
{5}------------------------------------------------
{6}------------------------------------------------
## Discussion of Differences
- Difference in While the indications for use for the nextaro® Transfer System and nextaro® va are worded slightly different than that of the predicate device, both the subject devices Indications: and predicate device are used for the same purpose: The transfer of solvents into a drug vial / the aspiration of the ready-to-use solution of the dissolved pharmaceuticals which are contained in vials. The nextaro® Transfer System is intended to be used with two vials (solvent and pharmaceutical agent vial) just as is the predicate device, while the nextaro® va is intended to be used with one vial (pharmaceutical agent vial; the solvent transferred into pharmaceutical agent vial by a syringe with male Luer adapter filled with solvent). There is no critical difference in the intended therapeutic use of the subject and predicate devices.
- Uses Vacuum The nextaro® Transfer System, the nextaro® va and the Mix2Vial all are used to Pressure: transfer solvent into a vial which contains a freeze-dried pharmaceutical agent (lyophilizate). All of the devices use pressure to transfer the solvent: In the case of the nextaro® Transfer System, the devices use the negative pressure within the pharmaceutical agent vial to transfer the solvent from the solvent vial which was connected to the device beforehand, just as is the case with the predicate device. The nextaro® va also uses pressure to transfer the solvent, but due to the design of the nextaro® va, the solvent must be applied by means of a syringe. However, the direction of the solvent flow is the same for all predicate and subject devices.
There is no critical difference in the use of vacuum pressure or manual pressure (to the plunger of the syringe) to transfer the solvent. The solvent transfer methods of the subject and predicate devices can be regarded as equivalent.
Filter The nextaro® Transfer System, the nextaro® va and the Mix2Vial all are equipped with Elements: filter elements, to retain fragments which could be punched out during perforation of the vial seals from being transferred or aspirated into the ready-to-use solution. The same applies for undissolved pharmaceutical agent particles.
> The filter mesh used in the nextaro® and nextaro® va are capable to retain particles of ≥ 15 um size. A retention rate of > 95 % of particles sized ≥ 15 um has been confirmed by testing. The performance of the filters (retention) of the subject and predicate devices has been shown to be substantially equivalent.
{7}------------------------------------------------
- Shelf life: The predicate device has a shelf-life of > 3 years while the subject devices have a shelflife of 5 years. The packaging materials are equivalent. The subject devices are substantially equivalent with regard to their shelf-life. This does not raise different questions of safety or effectiveness
- Both the predicate and the nextaro® devices are made of biocompatible thermoplastic Materials: polymer materials. Although the basic materials for the manufacturing of the devices is not the same, the performance tests show that the subject devices are substantially equivalent regarding the performance of the devices. Both the subject device and predicate use the same filter material (PET). The biocompatibility tests show that the subject devices are biocompatible and can therefore be regarded as substantially equivalent in regard to biocompatibility.
## Testing
The following tests were conducted to demonstrate that the nextaro® Transfer System and nextaro® va perform as intended.
| Standard | Test Performed |
|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 22413 | • Fragmentation<br>• Penetration Force<br>• Piercing<br>• Verification of Design Specifications for Transfer Devices with Housing |
| ISO 594-2 | • Male Conical Fitting / Luer Connector |
| ISO 8536-4 | • Flow Rate of Infusion Fluid<br>• Fluid Filter (Retention Rate)<br>• Leakage (Aging/ Submersion Test)<br>• Metal Ions<br>• Particulate Contamination<br>• Penetration Force<br>• Reduction of Oxidizable Matter<br>• Residue on Evaporation<br>• Tensile Strength<br>• Titration Acidity or Alkalinity<br>• UV Absorption of Extract Solution |
{8}------------------------------------------------
| Standard | Test Performed |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 11607 | • Sterile Barrier and Packaging Systems |
| ASTM D3078 | • Bubble Emissions |
| ASTM F 1886 | • Integrity of Seals |
| ASTM F 1929-12 | • Seal Leakage |
| ASTM F 1980-07 | • Accelerated Aging |
| ASTM F 2054-07, EN 868-5 | • Seal Strength |
| ISTA 2A | • Transport Testing |
| USP <85> , USP <161> | • Bacterial Endotoxin |
| N/A | • Roll Inhibition (to design specification)<br>• Transfer performance for residual volume (to design specification)<br>• Opening torque (to design specification) |
# Biocompatibility
Tests were conducted to show that the nextaro® Transfer System and the nextaro® va were evaluated for biocompatibility in accordance with ISO 10993-1. Filter particulate testing was also conducted (see above table).
Material Mediated Pyrogenicity LAL Testing
## Sterility
The sterility of the nextaro® Transfer System and nextaro® va were validated in accordance with ISO 11135:2014 Sterilization of health care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices. The maximum levels of ethylene oxide (EtO) and ethylene chlorohydrin (ECH) residuals remaining on the products were within the guidelines for limited and tolerable contact limit devices according to ISO 10993-7:2008.
## Summary / Conclusion
The evaluation of the proposed nextaro® devices has demonstrated through performance testing to be substantially equivalent to the predicate device. Based on the device type, indications, intended use, technological characteristics, physical characteristics, the nextaro® Transfer Device and nextaro® va are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.