K183180 · Aesculap, Inc. · GEI · Mar 12, 2019 · General, Plastic Surgery
Device Facts
Record ID
K183180
Device Name
Caiman 5 Maryland
Applicant
Aesculap, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Caiman Seal and Cut Technology consists of dedicated bipolar electrosurgical instruments intended for use in general surgery and gynecologic surgical procedures where ligation of vessels is desired. The instruments create a seal by the application of bipolar electrosurgical RF energy (coagulation) to vascular structure (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue. Instruments 12.5 cm, 17 cm, and 24 cm in length are indicated for open procedures and instruments 36 cm and 44 cm in length are indicated for laparoscopic procedures. The include general surgical procedures, (including urologic, vascular, thoracic, and thoracoscopic), and gynecological procedures where ligation of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, adhesiolysis, bowel resection, and oophorectomy etc., or any procedure where vessel ligation (seal and cut), tissue grasping, and dissection is performed. The devices up to and including 7mm and bundles as large as will fit in the jaws of the instrument. Caiman Seal and Cut Technology has not be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures.
Device Story
Caiman 5 Maryland is a sterile, single-use bipolar electrosurgical instrument for vessel sealing and tissue division. Used in open (12.5cm, 17cm, 24cm) and laparoscopic (36cm, 44cm) procedures. Device consists of a handle, stainless steel shaft, and curved jaws with bipolar electrodes. Operated by surgeons; connects to Lektrafuse RF generator. Surgeon places vessel/tissue between jaws; applies bipolar RF energy to seal; manually actuates cutting blade to divide tissue. Provides hemostasis via vessel ligation; facilitates dissection and grasping. Benefits include efficient vessel sealing and tissue division in various surgical specialties.
Clinical Evidence
Bench testing only. Performance validated through tissue-related performance tests for arteries, burst pressure testing for veins, jaw tissue stress testing, and usability validation. Electrical safety and EMC compliance verified per IEC 60601 series. Biocompatibility confirmed per ISO 10993-1.
Technological Characteristics
Bipolar electrosurgical instrument. Materials: molded thermoplastic handle, stainless steel shaft, polyester/ETFE insulation, PEEK dissection clip. Energy: Bipolar RF (via Lektrafuse generator). Dimensions: 5mm diameter, various lengths (12.5-44cm). Connectivity: Wired to generator. Sterilization: ISO 11137-2. Features: Curved jaw, non-articulating or articulating, single electrode pair.
Indications for Use
Indicated for general and gynecologic surgical procedures (including urologic, vascular, thoracic, and thoracoscopic) requiring vessel ligation, tissue grasping, and dissection. Applicable to vessels up to 7mm and tissue bundles. Not indicated for tubal sterilization or coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Aragon Surgical RF System - 5mm Laparoscopic Instrument (K110824)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
March 12, 2019
Aesculap, Inc. Ms. Jessica Stigliano Regulatory Affairs Associate 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K183180
Trade/Device Name: Caiman 5 Maryland Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: February 8, 2019 Received: February 9, 2019
Dear Ms. Stigliano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows a digital signature. The signature is by Long H. Chen -S. The date of the signature is March 12, 2019, and the time is 13:03:43 -04'00'.
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K183180
Device Name Caiman 5 Maryland
#### Indications for Use (Describe)
Caiman Seal and Cut Technology consists of dedicated bipolar electrosurgical instruments intended for use in general surgery and gynecologic surgical procedures where ligation of vessels is desired. The instruments create a seal by the application of bipolar electrosurgical RF energy (coagulation) to vascular structure (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue.
Instruments 12.5 cm, 17 cm, and 24 cm in length are indicated for open procedures and instruments 36 cm and 44 cm in length are indicated for laparoscopic procedures. The include general surgical procedures, (including urologic, vascular, thoracic, and thoracoscopic), and gynecological procedures where ligation of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, adhesiolysis, bowel resection, and oophorectomy etc., or any procedure where vessel ligation (seal and cut), tissue grasping, and dissection is performed. The devices up to and including 7mm and bundles as large as will fit in the jaws of the instrument.
Caiman Seal and Cut Technology has not be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures.
Type of Use (Select one or both, as applicable)
|×| Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
# K183180
## 510(k) SUMMARY (as required by 21 CFR 807.92)
# Caiman® 5 Maryland
March 11, 2019
| COMPANY: | Aesculap®®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 |
|--------------------|---------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Ms. Jessica Stigliano<br>610-984-9063 (phone)<br>610-791-6882 (fax)<br>jessica.stigliano@Aesculapimplants.com |
| TRADE NAME: | Caiman® 5 Maryland |
| COMMON NAME: | Electrosurgical, Cutting & Coagulation & Accessories |
| REGULATION NUMBER: | 21 CFR 878.4400 |
| PRODUCT CODE: | GEI |
| REVIEW PANEL: | General and Plastic Surgery |
#### PRIMARY PREDICATE
K151696 - Caiman Seal and Cut Technology
#### REFERENCE DEVICE
K130596 - Caiman Seal and Cut Technology K110824 - Aragon Surgical RF System - 5mm Laparoscopic Instrument
## SUBSTANTIAL EQUIVALENCE
Aesculap®, Inc. believes that the Caiman® 5 Maryland is substantially equivalent to the primary predicate, Caiman® Seal and Cut Technology (K151696) and reference devices, Caiman® Seal and Cut Technology (K130596) and Aragon Surgical RF System - 5mm Laparoscopic Instrument (K110824).
#### DEVICE DESCRIPTION
The Caiman® 5 Maryland instruments are seal and cut devices which are provided as sterile, single use devices. These devices are capable of vessel sealing, blunt dissection, grasping and dividing tissue enclosed within its jaws during open and laparoscopic procedures. The devices
{4}------------------------------------------------
## K 183180
are designed to be used with the dedicated Lektrafuse RF Generator and create vesselligation by the application of bipolar electrical RF energy and tissue division with a cutting blade.
## INDICATIONS FOR USE
Instruments 12.5 cm, 17 cm, and 24 cm in length are indicated for open procedures and instruments 36 cm and 44 cm in length are indicated for laparoscopic procedures. The indications for use include general surgical procedures, (including urologic, vascular, thoracic, and thoracoscopic), and gynecological procedures where ligation and division of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, bowel resection, and oophorectomy etc., or any procedure where vessel ligation (seal and cut), tissue grasping, and dissection is performed. The devices can be used on vessels up to and including 7mm and bundles as large as will fit in the jaws of the instrument.
Caiman Seal and Cut Technology has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures.
## TECHNOLOGICAL CHARACTERISTICS (compared to Predicate(s))
The modifications made to the Caiman instruments do not affect the fundamental scientific technology. The modifications made to these devices do not raise any new issues of safety and effectiveness, as confirmed by the testing and validation activities described in the submission.
{5}------------------------------------------------
| Attribute | Subject Device | Primary Predicate | Reference Device | Reference Device |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Caiman® 5 Maryland | Caiman Seal and Cut | Caiman Seal and Cut | Aragon Surgical RF System |
| | | Technology (K151696) | Technology (K130596) | Caiman 5 Laparoscopic |
| | | | | Instrument (K110824) |
| Indications for | Caiman Seal and Cut | Caiman Seal and Cut | Caiman Seal and Cut Technology | The Aragon Surgical Caiman 5 |
| Use | Technology consists of | Technology consists of | consists of dedicated bipolar | Laparoscopic Instrument is a |
| | dedicated bipolar | dedicated bipolar | electrosurgical instruments | dedicated bipolar electrosurgical |
| | electrosurgical instruments | electrosurgical instruments | intended for use in general surgery | instrument intended for use in |
| | intended for use in general | intended for use in general | and gynecologic surgical | general surgical and gynecologic |
| | surgery and gynecologic | surgery and gynecologic | procedures where ligation and | laparoscopic procedures where |
| | surgical procedures where | surgical procedures where | division of vessels is desired. The | ligation and division of vessels is |
| | ligation and division of vessels | ligation and division of vessels | instruments create a seal by the | desired. The instrument creates a |
| | is desired. The instruments | is desired. The instruments | application of bipolar | seal by the application of bipolar |
| | create a seal by the application | create a seal by the application | electrosurgical RF energy | electrosurgical RF energy to |
| | of bipolar electrosurgical RF | of bipolar electrosurgical RF | (coagulation) to vascular structure | vascular structures (vessels) |
| | energy (coagulation) to vascular | energy (coagulation) to vascular | (vessels) interposed between the | interposed between the jaws of |
| | structure (vessels) interposed | structure (vessels) interposed | jaws of the device. A cutting blade | the device. A cutting blade is |
| | between the jaws of the device. | between the jaws of the device. | is actuated for the division of | actuated for the division of tissue. |
| | A cutting blade is actuated for | A cutting blade is actuated for | tissue. | Indications for use include |
| | the division of tissue. | the division of tissue. | The Caiman 12 Plus (44cm) and | general laparoscopic procedures, |
| | Instruments 12.5cm, 17 cm, | Instruments 24 cm in length are | the Caiman 5 are indicated for | including urologic, vascular, |
| | and 24 cm in length are | indicated for open procedures | laparoscopic procedures and the | thoracic, and thoracoscopic, and |
| | indicated for open procedures | and instruments 36 cm and 44 | Caiman 12 Plus (24cm) is | gynecological laparoscopic |
| | and instruments 36 cm and 44 | cm in length are indicated for | indicated for open procedures. | procedures where ligation and |
| | cm in length are indicated for | laparoscopic procedures. The | The indications for use include | division of vessels is performed. |
| | laparoscopic procedures. The | indications for use include | general surgical procedures, | These procedures include: |
| | indications for use include | general surgical procedures, | (including urologic, vascular, | laparoscopically assisted vaginal |
| | general surgical procedures, | (including urologic, vascular, | thoracic, and thoracoscopic), and | hysterectomies, Nissen |
| | (including urologic, vascular, | thoracic, and thoracoscopic), | gynecological procedures where | fundoplication, colectomy, |
| | thoracic, and thoracoscopic), | and gynecological procedures | ligation and division of vessels is | adhesiolysis, oophorectomy etc. |
| | and gynecological procedures | where ligation and division of | performed. These procedures | The Aragon Surgical Caiman 5 |
| | where ligation and division of | vessels is performed. These | include: vaginal hysterectomies, | Laparoscopic Instrument can be |
| | vessels is performed. These | procedures include: vaginal | Nissen fundoplication, colectomy, | used on vessels up to and |
| | procedures include: vaginal | hysterectomies, Nissen | adhesiolysis, bowel resection, and | including 7mm, and tissue |
| | hysterectomies, Nissen | fundoplication, colectomy, | oophorectomy etc., or any | bundles as large as will fit in the |
| | fundoplication, colectomy, | adhesiolysis, bowel resection, | procedure where vessel ligation | jaws of the instrument. |
| | adhesiolysis, bowel resection,<br>and oophorectomy etc., or any<br>procedure where vessel ligation<br>(seal and cut), tissue grasping,<br>and dissection is performed.<br>The devices can be used on<br>vessels up to and including<br>7mm and bundles as large as<br>will fit in the jaws of the<br>instrument.<br>Caiman Seal and Cut<br>Technology has not been shown<br>to be effective for tubal<br>sterilization or tubal<br>coagulation for sterilization<br>procedures. Do not use the<br>system for these procedures. | and oophorectomy etc., or any<br>procedure where vessel ligation<br>(seal and cut), tissue grasping,<br>and dissection is performed.<br>The devices can be used on<br>vessels up to and including<br>7mm and bundles as large as<br>will fit in the jaws of the<br>instrument.<br>Caiman Seal and Cut<br>Technology has not been shown<br>to be effective for tubal<br>sterilization or tubal<br>coagulation for sterilization<br>procedures. Do not use the<br>system for these procedures. | (seal and cut), tissue grasping, and<br>dissection is performed. The<br>devices can be used on vessels up<br>to and including 7mm and bundles<br>as large as will fit in the jaws of<br>the instrument.<br>Caiman Seal and Cut Technology<br>has not been shown to be effective<br>for tubal sterilization or tubal<br>coagulation for sterilization<br>procedures. Do not use the system<br>for these procedures. | The Aragon Surgical Lektrafuse<br>RF System has not been shown to<br>be effective for tubal sterilization<br>or tubal coagulation for<br>sterilization procedures and<br>should not be used for these<br>procedures. |
| Material<br>Composition | Handle - molded thermoplastic<br>Shaft - stainless steel<br>Insulation - polyester tube<br>(Caiman 5)<br>ETFE (Caiman 5 Articulating)<br>Dissection clip - PEEK | Handle - molded thermoplastic<br>Shaft - stainless steel<br>Insulation - polyester tube<br>(Caiman 5)<br>EFTE (Caiman 5 Articulating)<br>Dissection clip - Valox HX 420<br>HP | Handle - molded thermoplastic<br>Shaft - stainless steel<br>Insulation - polyester tube<br>(Caiman 5)<br>Dissection clip – PEEK (Caiman<br>5) | Molded thermoplastic handle,<br>stainless steel shaft |
| Functional Use | Grasping Ligation and<br>Coagulation (unchanged) | Grasping Ligation and<br>Coagulation | Grasping Ligation and Coagulation | Grasping Ligation and<br>Coagulation |
| Surgical<br>Approach | 125mm and 170mm – Open<br>360mm and 440mm –<br>Laparoscopic (unchanged) | 240mm - Open<br>360 and 440mm - Laparoscopic | Caiman 12 Plus (44cm)-<br>Laparoscopic<br>Caiman 12 Plus (24cm)- Open<br>Caiman 5- Laparoscopic | Laparoscopic |
| Length | 125mm<br>170mm<br>360mm<br>440mm | 240mm<br>360mm<br>440mm | Caiman 12 Plus- 44cm<br>Caiman 12 Plus- 24cm<br>Caiman 5- 36cm | 36cm |
| Jaw Shape | Curved | Straight…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.