EnSite Velocity Cardiac Mapping System v5.2, EnSite Precision Cardiac Mapping System v2.2

K183128 · Abbott · DQK · Dec 12, 2018 · Cardiovascular

Device Facts

Record IDK183128
Device NameEnSite Velocity Cardiac Mapping System v5.2, EnSite Precision Cardiac Mapping System v2.2
ApplicantAbbott
Product CodeDQK · Cardiovascular
Decision DateDec 12, 2018
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1425
Device ClassClass 2

Indications for Use

EnSite™ Velocity™ Cardiac Mapping System v5.2: The EnSite Velocity Cardiac Mapping System is a suggested Diagnostic tool in patients for whom electrophysiology studies have been indicated. When used with EnSite™ Array Catheter, the EnSite™ Velocity™ Cardiac Mapping System is intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping system alone. or When used with the EnSite™ Velocity™ Surface Electrode Kit, the EnSite™ Velocity™ Cardiac Mapping System is intended to display the position of conventional electrophysiology (EP) catheters in the heart. EnSite Precision™ Cardiac Mapping System v2.2: The EnSite Precision™ Cardiac Mapping System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated. The EnSite Precision™ System interfaces to either the MediGuide™ Technology System or the EnSite Precision™ Module to combine and display magnetic processed patient positioning and navigation mapping information. When used with the EnSite™ Array™ Catheter, the EnSite Precision™ Cardiac Mapping System is intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone. or When used with an EnSite Precision™ Surface Electrode Kit, the EnSite Precision™ Cardiac Mapping System is intended to display the position of conventional electrophysiology (EP) catheters in the heart. EnSite™ Verismo™ Segmentation Tool: The EnSite Verismo™ Segmentation Tool is indicated for use in generating 3D models from CT. MR or rotational angiography DICOM image data. Generated models are intended to be displayed on the EnSite Velocity System. EnSite™ Derexi™ Module: When used with EnSite Derexi ™ Module, the EnSite System interfaces to the EP-WorkMate™ System / WorkMate Claris™ System for synchronizing and display of patient information. EnSite™ Courier™ Module: When used with EnSite Courier Module allows the patient data to be archived to, and retrieved from, a DICOM conformant PACs server. EnSite™ Fusion™ Registration Module: EnSite Fusion is indicated for registering the EnSite Navigation system to anatomic models, derived from CT scans, of the four individual cardiac chambers. EnSite™ Contact Force Module: When used with the SJM Contact Force Unit, the EnSite™ Contact Force Module is intended to provide visualization of force information from compatible catheters. EnSite™ AutoMap Module: When used with the EnSite AutoMap Module, the EnSite System is intended to automatically collect mapping points based on criteria set by the user. AutoMark Module: When used with compatible hardware, the AutoMark Module is intended to automatically catalog and display various parameters associated with RF information on the 3D model in real-time.

Device Story

Catheter navigation/mapping system for EP labs; displays 3D position of EP catheters, cardiac electrical activity (waveforms), and 3D isopotential maps. Inputs: electrical signals from catheters via specialized input modules (CIMs); magnetic data (Precision version) from MediGuide/Precision Module. Processing: impedance-based navigation; magnetic field scaling (NavX SE); shift detection (EnGuide Stability Monitor); respiration gating. Outputs: 3D anatomical models, electrical maps, catheter positions, force visualization, and RF lesion marks. Used by physicians in EP labs to identify/treat arrhythmias. Benefits: improved visualization of complex arrhythmias and catheter placement. Optional modules: Verismo (3D segmentation from CT/MR/angiography), Derexi (EP-WorkMate synchronization), Courier (PACS archiving), Fusion (anatomic registration), Contact Force (catheter force visualization), AutoMap (automated point collection), AutoMark (RF lesion cataloging).

Clinical Evidence

No clinical data provided. Substantial equivalence supported by non-clinical risk management review and predicate comparison.

Technological Characteristics

System consists of display workstation (DWS), amplifier subsystem, and specialized catheter input modules. Operates via impedance-only or combined impedance/magnetic sensing. Connectivity: Ethernet for data transmission, DICOM for PACS. Software modules: Verismo (segmentation), Derexi (synchronization), Courier (archiving), Fusion (registration), Contact Force (visualization), AutoMap (automated collection), AutoMark (RF lesion marking).

Indications for Use

Indicated for patients undergoing electrophysiology (EP) studies. Used for right atrial mapping in complex arrhythmias (with Array Catheter) or displaying EP catheter positions (with Surface Electrode Kit).

Regulatory Classification

Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. December 12, 2018 Abbott Cody Johnson Regulatory Affairs Specialist II One St. Jude Medical Drive St. Paul, Minnesota 55117 Re: K183128 Trade/Device Name: EnSite Velocity Cardiac Mapping System v5.2 EnSite Precision Cardiac Mapping System v2.2 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: November 9, 2018 Received: November 13, 2018 Dear Cody Johnson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {2}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerelv Mude Jellman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K183128 ### Device Name EnSite™ VelocityTM Cardiac Mapping System v5.2 EnSite Precision™ Cardiac Mapping System v2.2 Indications for Use (Describe) EnSite™ Velocity™ Cardiac Mapping System v5.2: The EnSite Velocity Cardiac Mapping System is a suggested Diagnostic tool in patients for whom electrophysiology studies have been indicated. When used with EnSite™ Array Catheter, the EnSite™ Velocity™ Cardiac Mapping System is intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping system alone. Or When used with the EnSite™ Surface Electrode Kit, the EnSite™ Velocity™ Cardiac Mapping System is intended to display the position of conventional electrophysiology (EP) catheters in the heart. EnSite Precision™ Cardiac Mapping System v2.2: The EnSite Precision™ Cardiac Mapping System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated. The EnSite Precision™ System interfaces to either the MediGuide™ Technology System or the EnSite Precision™ Module to combine and display magnetic processed patient positioning and navigation mapping information. When used with the EnSite™ Array™ Catheter, the EnSite Precision™ Cardiac Mapping System is intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone. or When used with an EnSite Precision™ Surface Electrode Kit, the EnSite Precision™ Cardiac Mapping System is intended to display the position of conventional electrophysiology (EP) catheters in the heart. EnSite™ Verismo™ Segmentation Tool: The EnSite Verismo™ Segmentation Tool is indicated for use in generating 3D models from CT. MR or rotational angiography DICOM image data. Generated models are intended to be displayed on the EnSite Velocity System. EnSite™ Derexi™ Module: When used with EnSite Derexi ™ Module, the EnSite System interfaces to the EP-WorkMate™ System / WorkMate Claris™ System for synchronizing and display of patient information. ## EnSite™ Courier™ Module: When used with EnSite Courier Module allows the patient data to be archived to, and retrieved from, a DICOM conformant PACs server. {4}------------------------------------------------ EnSite™ Fusion™ Registration Module: EnSite Fusion is indicated for registering the EnSite Navigation system to anatomic models, derived from CT scans, of the four individual cardiac chambers. EnSite™ Contact Force Module: When used with the SJM Contact Force Unit, the EnSite™ Contact Force Module is intended to provide visualization of force information from compatible catheters. EnSite™ AutoMap Module: When used with the EnSite AutoMap Module, the EnSite System is intended to automatically collect mapping points based on criteria set by the user. AutoMark Module: When used with compatible hardware, the AutoMark Module is intended to automatically catalog and display various parameters associated with RF information on the 3D model in real-time. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the Abbott logo. The logo consists of a blue, stylized "a" on the left and the word "Abbott" in black, bold font on the right. The "a" is made up of three horizontal lines within a rounded square. | 510(k) Information | | |-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number | K183128 | | 510(k) Type | Special 510(k) | | Date Prepared | November 9, 2018 | | Submitter Information | | | Manufacturer<br>Name/Address | Abbott<br>One St. Jude Medical Drive<br>St. Paul, MN 55117 | | Contact Person | Cody Johnson<br>Regulatory Affairs Specialist II<br>cody.johnson@abbott.com | | Device Information | | | Trade Name | EnSite™ Velocity™ Cardiac Mapping System v5.2<br>EnSite Precision™ Cardiac Mapping System v2.2 | | Common Name | Programmable Diagnostic Computer | | Class | II | | Classification Name | 870.1425, computer, diagnostic, programmable | | Product Code | DQK | | Predicate Device | EnSite™ Velocity™ Cardiac Mapping System v5.2 (K182644) | | Secondary Predicates | EnSite™ Precision™ Cardiac Mapping System v2.2 (K182644) | | Reference Applications | N/A | | Device Description | The EnSite™ Velocity™ Cardiac Mapping System with software version 5.2 /<br>EnSite Precision™ Cardiac Mapping System with software version 2.2 is a<br>catheter navigation and mapping system capable of displaying the three-<br>dimensional (3D) position of conventional electrophysiology catheters, as well<br>as displaying cardiac electrical activity as waveform traces and as dynamic 3-D<br>isopotential maps of the cardiac chamber. The contoured surfaces of these three-<br>dimensional maps are based on the anatomy of the patient's own cardiac<br>chamber.<br><br>The EnSite™ Velocity™ Cardiac Mapping System is used as a diagnostic tool<br>in electrophysiology (EP) Studies. An EP study involves the introduction of one<br>or more electrode catheters into the heart to record its electrical activity. These<br>catheters connect to the EnSite™ Velocity™ Cardiac Mapping System through<br>specialized catheter input modules (CIMs). The EnSite™ Velocity™ Cardiac<br>Mapping System v5.2 is designed for use in the EP laboratory in conjunction<br>with other equipment. | | Device Description<br>(continued) | The EnSite Velocity™ Cardiac Mapping System consists of hardware and<br>software elements. The EnSite Velocity / EnSite Precision System consists of<br>software, a display workstation (DWS) subsystem (DWS, Monitors, DWS<br>Accessory Kit, and DWS Power Kit), and an amplifier subsystem (Amplifier<br>and Amplifier Accessory Kit). The DWS houses the system software and<br>connects all the components together. The amplifier contains electronic<br>circuitry and firmware responsible for collecting and transmitting the electrical<br>signal data of the patient to the DWS software. Its primary function is to collect<br>and transmit via Ethernet the electrical data detected from the patient. The<br>amplifier accepts signals from NavLink, ArrayLink, CathLink, ECG Cable,<br>RecordConnect, and GenConnect, converts these signals to a digital format, and<br>sends them to the workstation for processing. | | | The NavLink connects surface electrodes and the system reference surface<br>electrode to the Amplifier. The ArrayLink connects the EnSite Array Multi-<br>electrode Diagnostic Catheter to the Amplifier. It also has a connection for an<br>auxiliary unipolar reference electrode. The CathLink connects the diagnostic<br>catheters to the Amplifier. The GenConnect connects the ablation catheter and<br>dispersive surface electrodes to the Amplifier. The RecordConnect allows<br>simultaneous connection for catheters and surface ECG to a recording system<br>and to the Amplifier. The ECG cable connects standard ECG electrodes to the<br>Amplifier. | | | The system operates using impedance only or impedance plus magnetics based<br>upon its configuration. The EnSite™ Velocity™ Cardiac Mapping System<br>base software only collects impedance data. Adding EnSite Precision™<br>software to the base software allows the system to receive both magnetic data<br>from the MediGuide™ Technology System or the EnSite Precision™ Module<br>hardware and impedance data when using magnetic sensor enabled tools. The<br>EnSite Precision™ Module and EnSite Precision™ software (added to the base<br>software) together make up the EnSite Precision™ Cardiac Mapping System. | | | The EnSite Precision™ software interfaces to the MediGuide Technology<br>System or the EnSite Precision™ Module to collect magnetic position and<br>orientation information. The EnSite Precision™ software uses the magnetic data<br>for magnetic field scaling (NavX SE), shift detection (EnGuide Stability<br>Monitor), and respiration gating. NavX SE field scaling adjusts the dimensions<br>of the navigation field based on both the position and orientation of magnetic<br>sensors and the electrodes on Sensor Enabled™ (SE) tools, optimizing the<br>appearance of the model. The system uses EnGuide Stability Monitor to notify<br>the user of a potential shift based on a correlation of magnetic and impedance<br>locations when using any Sensor Enabled catheter. The system uses respiration<br>gating to compensate to the end-point of the respiration cycle using magnetic<br>data to determine respiration phase. | | | The EnSite Precision™ Module consists of hardware to support magnetic<br>navigation. The hardware components consist of the EnSite Precision™ Link,<br>EnSite Precision™ Field Frame, and EnSite Precision™ Patient Reference<br>Sensors. | | Expansion Module<br>Device Description | The EnSite™ Velocity™ Cardiac Mapping System v5.2 includes the following optional expansion software modules: | | 1. | EnSite™ Verismo™ Segmentation Tool - an optional expansion module used in generating 3D models from CT, MR or rotational angiography DICOM image data and displaying images on the EnSite™ Velocity™ Cardiac Mapping System. The EnSite™ Verismo™ Segmentation Tool accepts DICOM images from CT and MRI scanners and converts the images into a 3D model of cardiac structures. | | 2. | EnSite™ Derexi™ Module - an optional expansion module that that allows the EnSite Velocity System to interface with the WorkMate™ Recording System to support the exchange of mapping point data and patient setup information between the two systems. | | 3. | EnSite™ Courier™ Module - The EnSite™ Courier™ Module is an optional expansion module that allows the EnSite™ Velocity™ Cardiac Mapping System to communicate with the hospital PACS (Picture Archiving and Communication System) server for the purposes of storing and retrieving patient data in DICOM format. | | 4. | EnSite™ Fusion™ Registration Module - an optional expansion module that provides non-fluoroscopic navigation, mapping, and labeling on a Digital Image Fusion (DIF) model. The module is used with the EnSite™ NavX™ Navigation and Visualization Technology Surface Electrode Kit and CT or MR scans segmented into a compatible file format. 3D models created from digital images from CT and MRI data can be imported onto the EnSite™ Velocity™ System. | | 5. | EnSite™ Contact Force Module - an optional expansion module that provides the display of information from the TactiSys Quartz System. The EnSite Velocity System's EnSite Contact Force Module is intended to provide visualization of force information from compatible catheters. | | 6. | EnSite™ AutoMap Module - an optional module that automatically collects mapping points based on criteria set by the user | | 7. | AutoMark Module - module allows the user to set parameters and the software automatically displays the lesion marks on the EnSite Velocity model during RF ablation. The user set parameters is based on data from Ensite™ Contact Force Module, the Ampere Generator, and the WorkMate Claris™ System which is displayed on the AutoMark Module as lesion marks on the during RF ablation. The color, size, and ranges of the AutoMark are defined by the user. | | Indications for Use | EnSite™ Velocity™ Cardiac Mapping System v5.2 | | | The EnSite Velocity Cardiac Mapping System is a suggested Diagnostic tool in<br>patients for whom electrophysiology studies have been indicated. | | | When used with EnSite™ Array Catheter, the EnSite™ Velocity™ Cardiac<br>Mapping System is intended to be used in the right atrium of patients with<br>complex arrhythmias that may be difficult to identify using conventional<br>mapping system alone. | | | or | | | When used with the EnSite™ Velocity™ Surface Electrode Kit, the EnSite™<br>Velocity™ Cardiac Mapping System is intended to display the position of<br>conventional electrophysiology (EP) catheters in the heart. | | | EnSite Precision™ Cardiac Mapping System v2.2 | | | The EnSite Precision™ Cardiac Mapping System is a suggested diagnostic tool<br>in patients for whom electrophysiology studies have been indicated. | | | The EnSite Precision™ System interfaces to either the MediGuide™<br>Technology System or the EnSite Precision™ Module to combine and display<br>magnetic processed patient positioning and navigation mapping information. | | | When used with the EnSite™ Array™ Catheter, the EnSite Precision™ Cardiac<br>Mapping System is intended to be used in the right atrium of patients with<br>complex arrhythmias that may be difficult to identify using conventional<br>mapping systems alone. | | | or | | | When used with an EnSite Precision™ Surface Electrode Kit, the EnSite<br>Precision™ Cardiac Mapping System is intended to display the position of<br>conventional electrophysiology (EP) catheters in the heart. | | | EnSite™ Verismo™ Segmentation Tool | | | The EnSite Verismo™ Segmentation Tool is indicated for use in generating 3D<br>models from CT, MR or rotational angiography DICOM image data. Generated<br>models are intended to be displayed on the EnSite Velocity System. | | | EnSite™ Derexi™ Module | | | When used with EnSite Derexi ™ Module, the EnSite System interfaces to the<br>EP-WorkMate™ System / WorkMate Claris™ System for synchronizing and<br>display of patient information | | | EnSite™ Courier™ Module | | | When used with EnSite Courier Module allows the patient data to be archived<br>to, and retrieved from, a DICOM conformant PACs server. | | | EnSite™ Fusion™ Registration Module | | | EnSite Fusion is indicated for registering the EnSite NavX navigation system to<br>anatomic models, derived from CT scans, of the four individual cardiac<br>chambers | | <b>Indications for Use</b><br>(continued) | <b>EnSite™ Contact Force Module</b><br>When used with the SJM Contact Force Unit, the EnSite™ Contact Force<br>Module is intended to provide visualization of force information from<br>compatible catheters.<br><br><b>EnSite™ AutoMap Module</b><br>When used with the EnSite AutoMap Module, the EnSite System is intended to<br>automatically collect mapping points based on criteria set by the user<br><br><b>AutoMark Module</b><br>When used with compatible hardware, the AutoMark Module is intended to<br>automatically catalog and display various parameters associated with RF<br>information on the 3D model in real-time. | | <b>Submission History</b> | No prior submissions have been made to FDA for the device that is the subject<br>of this submission. | | <b>Predicate Comparison</b> | | | <b>Comparison</b> | Both the subject and predicate devices have the same intended use, indications<br>for use, and operate using the same fundamental scientific technology to<br>facilitate catheter position and orientation, as well as cardiac mapping and<br>model creation. The proposed labeling modifications to the subject device<br>pertain to warnings only. The labeling modifications include removal of the<br>black box warning from the cover of the Instructions for Use as well as the<br>addition of a warning in the warnings section of the Instructions for Use.<br>There are no new or increased risks that result from the proposed modifications<br>presented within the submission, and the changes do not raise any new<br>questions of safety and effectiveness in regards to the subject device. | | <b>Non-Clinical Testing<br/>Summary</b> | No new non-clinical testing was completed, nor relied upon, in support of this<br>Special 510(k) submission. | | | <b>Risk Management</b><br>The changes to the EnSite™ Velocity™ Cardiac Mapping System / EnSite<br>Precision™ Cardiac Mapping System were evaluated through review of risk<br>management to ensure no new hazards have been introduced by this change. | | <b>Statement of<br/>Equivalence</b> | The technological characteristics for the devices are the same as the predicate<br>devices. Based on this and the data provided in this pre-market notification, the<br>subject devices and predicate devices have been demonstrated to be<br>substantially equivalent. | {6}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the Abbott logo. The logo consists of a stylized blue letter 'a' on the left and the word 'Abbott' in black bold font on the right. The blue 'a' symbol is a modern, abstract design, and the word 'Abbott' is easily readable. {7}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the Abbott logo. The logo consists of a blue stylized letter "a" on the left and the word "Abbott" in black bold font on the right. The blue "a" is a rounded square shape with a horizontal line through the middle. {8}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for Abbott. The logo consists of a blue, stylized letter "a" on the left and the word "Abbott" in black, bold font on the right. The blue "a" is made up of three horizontal lines connected by a vertical line on the left side. {9}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the Abbott logo. The logo consists of a stylized letter "a" in blue, followed by the word "Abbott" in black. The logo is simple and modern, and it is easily recognizable.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...