Combined Aerobika OPEP and VersaPAP device

K183108 · Trudell Medical International · BWF · Aug 16, 2019 · Anesthesiology

Device Facts

Record IDK183108
Device NameCombined Aerobika OPEP and VersaPAP device
ApplicantTrudell Medical International
Product CodeBWF · Anesthesiology
Decision DateAug 16, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5690
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Combined VersaPAP™ and Aerobika® OPEP device is intended for use as a Positive Airway Pressure (PAP) device and a Positive Expiratory Pressure (PEP) device. The combined device has the ability to provide supplemental oxygen when used with compressed oxygen. The combined device is for patients (ages 5 years and above) who are capable of following directions for Positive Airway Pressure Therapy and capable of generating exhalation flow of 10 lpm for 3 - 4 seconds. The combined device is a single patient, multiple use device intended to be used in a hospital environment under the supervision of a healthcare professional.

Device Story

Combined VersaPAP and Aerobika OPEP device functions as both PAP and OPEP therapy tool. Input: compressed gas source (air/oxygen) for PAP; patient exhaled breath for OPEP oscillations. Operation: PAP uses Venturi principle to create positive airway pressure; OPEP uses patient exhalation to generate oscillations for secretion clearance. Output: positive airway pressure, oscillations, and supplemental oxygen. Used in hospital settings under healthcare professional supervision. Provider monitors airway pressure via manometer and adjusts exhalation resistance. Benefits: aids secretion clearance, treats/prevents atelectasis, and provides supplemental oxygen.

Clinical Evidence

Bench testing only. Comparative performance testing conducted at adult and child breathing settings against predicates and reference device. Biocompatibility testing performed per ISO 18562 for new configurations.

Technological Characteristics

Materials are identical to predicates (K150173, K173918). Operates via Venturi principle (PAP) and patient-generated exhalation oscillations (OPEP). Includes manometer for pressure measurement. Single-patient, multi-use device. Hospital environment use. No software/electronics.

Indications for Use

Indicated for patients (ages 5+) capable of following PAP therapy directions and generating 10 lpm exhalation flow for 3-4 seconds. Used for secretion clearance and treatment/prevention of atelectasis. Provides supplemental oxygen when used with compressed oxygen.

Regulatory Classification

Identification

An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 16, 2019 Trudell Medical International c/o Paul Dryden Consultant 725 Third Street London, N5V 5G4 CA Re: K183108 Trade/Device Name: Combined VersaPAP device and Aerobika OPEP Regulation Number: 21 CFR 868.5690 Regulation Name: Incentive Spirometer Regulatory Class: Class II Product Code: BWF Dated: July 18, 2019 Received: July 19, 2019 ## Dear Paul Dryden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, James Lee, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K183108 Device Name ## Combined VersaPAP™ device and Aerobika® OPEP #### Indications for Use (Describe) The Combined VersaPAP™ and Aerobika® OPEP device is intended for use as a Positive Airway Pressure (PAP) device and a Positive Expiratory Pressure (PEP) device. The combined device has the ability to provide supplemental oxygen when used with compressed oxygen. The combined device is for patients (ages 5 years and above) who are capable of following directions for Positive Airway Pressure Therapy and capable of generating exhalation flow of 10 lpm for 3 - 4 seconds. The combined device is a single patient, multiple use device intended to be used in a hospital environment under the supervision of a healthcare professional. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------|---------------------------------------------| | <b>XX</b> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) SC Publishing Services (301) 443-6740 EF {3}------------------------------------------------ #### 510(k) Summary 16-Aug-19 Trudell Medical International 725 Third Street London, ON N5V 5G4 CANADA Tel – 519-455-7060 Official Contact: Marianne Tanton - Director, Quality and Regulatory Affairs | Proprietary or Trade Name: | Combined VersaPAPTM device and Aerobika® OPEP | | Proposed<br>Combined<br>Aerobika® OPEP and<br>VersaPAP™ device | Predicate<br>Aerobika® OPEP<br>K150173 | Predicate<br>VersaPAP™ device<br>K173918 | |----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|--------------------------------------------| | Common/Usual Name: | Incentive Spirometer | CFR<br>Classification | | 868.5690<br>BWF | | | Classification Code/Name: | BWF – Incentive Spirometer<br>21 CFR 868.5690<br>Class II | Classification name | | | | | Predicate Devices: | K150173 – Trudell – Aerobika® OPEP<br>K173918 – Trudell – VersaPAPTM device | Intended Use | Intended for use as a Positive<br>Expiratory Pressure (PEP) device<br>(exercise patient's lungs and to<br>improve secretion clearance) | Incentive Spirometer<br>Intended for use as a<br>Positive Expiratory<br>Pressure (PEP) device. | | | Reference Device: | K151689 – Hill-Rom – MetaNeb® | | Intended for use as a Positive<br>Airway Pressure (PAP) device<br>(Treatment and prevention of<br>atelectasis) | | Treatment and prevention of<br>atelectasis | | Indications for Use | The Combined VersaPAP™ and<br>Aerobika® OPEP device is<br>intended for use as a Positive<br>Airway Pressure (PAP) device and<br>a Positive Expiratory Pressure<br>(PEP) device. The combined<br>device has the ability to provide<br>supplemental oxygen when used<br>with compressed oxygen. The<br>combined device is for patients<br>(ages 5 years and above) who are<br>capable of following directions for<br>Positive Airway Pressure Therapy<br>and capable of generating<br>exhalation flow of 10 lpm for 3 -<br>4 seconds. The combined device<br>is a single patient, multiple use<br>device intended to be used in a<br>hospital environment under the<br>supervision of a healthcare<br>professional | The Aerobika® Oscillating<br>Positive Expiratory<br>Pressure device is intended<br>for use as a Positive<br>Expiratory Pressure (PEP)<br>device. The Aerobika®<br>Oscillating PEP device<br>may also be used<br>simultaneously with<br>nebulized aerosol drug<br>delivery. The device is<br>intended to be used by<br>patients capable of<br>generating exhalation flow<br>of 10 lpm for 3-4 seconds. | The VersaPAP™ device is<br>indicated for the treatment and<br>prevention of atelectasis. The<br>VersaPAP™ device also has<br>the ability to provide<br>supplemental oxygen when<br>used with compressed oxygen.<br>The VersaPAP™ device is for<br>patients (ages 5 years and<br>above) who are capable of<br>following directions for<br>Positive Airway Pressure<br>Therapy in a hospital<br>environment. The<br>VersaPAP™ device is a single<br>patient, multiple use device<br>intended to be used under the<br>supervision of a healthcare<br>professional. | | | | Patient Population | Patients capable of generating<br>exhalation flow of 10 lpm for 3-4<br>seconds.<br>Patients (ages 5 years and above)<br>who are capable of following | Patients capable of<br>generating exhalation flow<br>of 10 lpm for 3-4 seconds. | Patients (ages 5 years and<br>above) who are capable of | | | | | directions. | | following directions. | | | #### Device Description: Date Prepared The Aerobika® OPEP device and VersaPAP™ device can be used independently or combined. When combined for ease of use. The combined device creates oscillations during the exhalation phase (OPEP) to aid in secretion clearance and during the inhalation phase (PAP) helps to maintain positive airway pressure for the treatment and prevention of atelectasis. #### Indications for Use: The Combined VersaPAP™ and Aerobika® OPEP device is intended for use as a Positive Airway Pressure (PAP) device and a Positive Expiratory Pressure (PEP) device. The combined device has the ability to provide supplemental oxygen when used with compressed oxygen. The combined device is for patients (ages 5 years and above) who are capable of following directions for Positive Airway Pressure Therapy and capable of generating exhalation flow of 10 lpm for 3 - 4 seconds. The combined device is a single patient, multiple use device intended to be used in a hospital environment under the supervision of a healthcare professional. {4}------------------------------------------------ ## Table 1 – Comparison to Predicates {5}------------------------------------------------ | | Proposed<br>Combined<br>Aerobika® OPEP and<br>VersaPAP™ device | Predicate<br>Aerobika® OPEP<br>K150173 | Predicate<br>VersaPAP™ device<br>K173918 | |------------------------------|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------| | Environments | Hospital settings | Hospital and clinical<br>Home settings | Hospital settings | | Operating principles<br>OPEP | Oscillations generated by patient's exhaled breath.<br>Adjustable exhalation resistance<br>Visual feedback via manometer | Oscillations generated by patient's exhaled breath.<br>Adjustable exhalation resistance<br>Visual feedback only when used with the manometer | | | PAP | Venturi principle created by compressed gas source<br>Supplemental oxygen<br>Measures airway pressure with manometer | | Venturi principle created by compressed gas source<br>Supplemental oxygen<br>Measures airway pressure with manometer | | Single patient, multi-use | Components may be cleaned and reused by the same patient | | | {6}------------------------------------------------ | | Proposed<br>Combined<br>OPEP + PAP | Reference<br>Hill-Rom - MetaNeb®<br>K151689 | |---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Classification | BWF<br>868.5690<br>Incentive Spirometer | NHJ<br>868.5905<br>Device, positive pressure breathing, intermittent<br>(IPPB) | | Intended use | Intended for use as a Positive Expiratory<br>Pressure (PEP) device (exercise patient's<br>lungs and to improve secretion clearance)<br><br>Intended for use as a Positive Airway<br>Pressure (PAP) device (Treatment and<br>prevention of atelectasis) | Intended for use as a Positive Expiratory Pressure<br>(PEP) device.<br><br>Treatment and prevention of atelectasis | | Indications for Use | The Combined VersaPAPTM and<br>Aerobika® OPEP device is intended for<br>use as a Positive Airway Pressure (PAP)<br>device and a Positive Expiratory Pressure<br>(PEP) device. The combined device has<br>the ability to provide supplemental oxygen<br>when used with compressed oxygen. The<br>combined device is for patients (ages 5<br>years and above) who are capable of<br>following directions for Positive Airway<br>Pressure Therapy and capable of<br>generating exhalation flow of 10 lpm for 3<br>– 4 seconds. The combined device is a<br>single patient, multiple use device<br>intended to be used in a hospital<br>environment under the supervision of a<br>healthcare professional | The MetaNeb® System is indicated for<br>mobilization of secretions, lung expansion therapy,<br>the treatment and prevention of pulmonary<br>atelectasis, and also has the ability to provide<br>supplemental oxygen when used with compressed<br>oxygen. | | Patient Population | Patients capable of generating exhalation<br>flow of 10 lpm for 3-4 seconds.<br><br>Patients (ages 5 years and above) who are<br>capable of following directions. | Patient 5 years and older | | Environments of Use | Hospital | Hospital<br>sub-acute facilities<br>Nursing care<br>Homecare | ## Table 2 - Comparison of the Subject vs. the Reference Hill-Rom MetaNeb (K151689) {7}------------------------------------------------ | | | Reference | |----------------------|-------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------| | | Proposed<br>Combined | Hill-Rom - MetaNeb® | | | OPEP + PAP | K151689 | | Mode of<br>Operation | OPEP<br>Oscillations generated by patient's exhaled<br>breath.<br>Adjustable exhalation resistance<br>Visual feedback via manometer | Can deliver continuous expiratory pressure<br>(CPEP) combined with medicated aerosol.<br>Oscillations generated by device | | | PAP<br>Venturi vacuum created by compressed gas<br>source<br>Supplemental oxygen<br>Measures airway pressure with manometer | Venturi from compressed gas source continuous<br><br>Supplemental oxygen<br>Measures airway pressure with manometer | | Therapy | OPEP<br>PAP | CPEP<br>CHFO<br>Aerosol | | Patient Interface | Mouthpiece | Disposable circuit includes connection for in-line<br>nebulizer with mouthpiece | | Gas source | Compressed air or oxygen | Pneumatic and air or oxygen | | Control settings | OPEP resistance adjustable<br><br>PAP flow rate: 5-15 L/min. | Mode selection<br><br>Frequency selection for CHFO mode<br><br>Pressure adjustment for CPEP mode | ## Predicate, Reference and Proposed Device Comparison and Substantial Equivalence Discussion: The proposed and predicates are similar in the following categories: - Indications for Use ● - Patient Population ● - Technological Characteristics - Materials in patient contact - Shelf-life - . Cleaning methods The Combined Aerobika® OPEP and VersaPAP™ device is substantially equivalent to the predicate and reference devices because: #### Indications - The proposed indications for use for secretions clearance and the treatment and prevention of pulmonary atelectasis, and also has the ability to provide supplemental oxygen when used with compressed oxygen are identical to the predicate and reference devices. Discussion - There are no differences between the subject and predicate devices. #### Patient Population - The patient population is identical to the predicates. Discussion - There are no differences between the subject and predicate devices. {8}------------------------------------------------ ## Environment of Use - The environments of use are similar to the predicates. Noting that the OPEP also includes the home setting when used individually, but the home setting is not intended for the subject device. Discussion - There are no differences between the subject and predicate devices except the combined device will not be used in the home setting. ## Technology - The technology for generating OPEP and PAP is identical to the predicates. The technology of having a device which offers the combination of OPEP and PEP is similar to the reference device. Discussion - There are no differences between the subject and predicate devices. ## Performance - The pressure was compared to the reference device. While the pressure of the subject device is less than that of the reference device, it is higher than that of the predicates when they are tested as standalone devices. Discussion - The difference in performance is within the range of the separate predicate devices and the reference device and does not raise new safety or effectiveness concerns. The frequency of the combined device is higher than that of the reference MetaNeb device, however, the measured frequency is comparable to the Aerobika® OPEP predicate device and is within acceptable clinical frequency efficacy range. Discussion - The frequency of the subject device is comparable to the Aerobika® OPEP predicate device. As such, the frequency generated by the subject device does not add risk to the patient compared to other FDA cleared devices. #### Non-clinical Comparative Performance ## Biocompatibility - The materials that are patient contacting and gas pathway are same as K150173 - Trudell -Aerobika® OPEP and K173918 – Trudell – VersaPAP™ device Discussion - Additionally testing was performed. ISO 18562, which supported patient exposure in new configurations. ## Bench Testing - We performed comparative testing at adult and child breathing setting to the predicates and reference. The results demonstrate that the subject device is substantially equivalent to the predicate and reference devices. ## Discussion of Differences The only differences are: - . Environment is only hospital setting - Where the Aerobika® OPEP device when used individually can also be used in the о home setting - Certain performance categories differ when the devices are combined vs. when they are used ● separately. - These differences relate to pressure and frequency (oscillations) when they are o combined. When compared to the predicates and reference device, these differences in pressure and frequency are within the performance specifications of the cleared predicates and reference device. {9}------------------------------------------------ Any differences do not raise new or different concerns of safety and effectiveness. ## Substantial Equivalence Conclusion Based upon the comparative performance testing we have demonstrated that the proposed device compared to the predicate and reference devices can be found to substantially equivalent.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...