K183012 · Elucid Bioimaging, Inc. · LLZ · Dec 21, 2018 · Radiology
Device Facts
Record ID
K183012
Device Name
vascuCAP
Applicant
Elucid Bioimaging, Inc.
Product Code
LLZ · Radiology
Decision Date
Dec 21, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K183012 · Dec 21, 2018
vascuCAP
Elucid Bioimaging, Inc.
Ex vivo histopathologic tissue specimens; Clinical CT images
The device performance was validated by comparing software-calculated tissue characteristics against ground truth established by expert pathologist annotations of ex vivo histopathologic specimens paired with clinical CTA images.
Performance validation using clinical images and histopathologic specimens; Retrospective validation study comparing software output to expert histopathologic annotation
Carotid artery tissue specimens
Not applicable for this study
Accuracy of tissue characterization (Calcified Area, LRNC Area, Matrix Area) compared to pathologist annotation
Indications for Use
vascuCAP is a medical image analysis system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired with contrast from CT imaging devices. vascuCAP is intended to assist trained physicians in the stratification of patients identified to have atherosclerosis. The software post processes images obtained using a multidetector CT. The package provides tools for the measurement and visualization (color coded maps) of arterial vessels. Clinicians can select any artery to view the following anatomical references: the highlighted vessel in 3D, two rotatable curved MPR vessel views displayed at angles orthogonal to each other, and cross sections of the vessel. Cross-sectional measurements can be obtained using standard vascuCAP software measuring tools. Clinicians can semi-automatically determine contrasted lumen boundaries, stenosis measurements, and maximum lumen diameters. In addition, clinicians can edit lumen boundaries and examine Hounsfield unit or signal intensity statistics. Clinicians can also manually measure vessel length along the centerline in standard curved MPR views. The measurements provided by vascuCAP are not intended to provide a diagnosis or clinical recommendations. vascuCAP is intended as a tool to complement standard of care.
Device Story
vascuCAP is an image analysis software package for evaluating contrast-enhanced multidetector CT images of arterial vessels. It processes, reviews, and analyzes multidimensional digital images; provides 3D visualization and color-coded maps of arterial vessels. Clinicians use the software to select arteries, view 3D highlights, rotatable curved multi-planar reconstruction (MPR) views, and cross-sections. The system enables semi-automatic determination of contrasted lumen boundaries, stenosis measurements, and maximum/minimum lumen diameters; allows manual editing of lumen boundaries and examination of Hounsfield unit or signal intensity statistics; supports manual vessel length measurement. Output is used by physicians to complement standard of care in atherosclerosis patient stratification; measurements are not for diagnosis or clinical recommendations.
Clinical Evidence
Bench and clinical validation testing performed. Anatomic structure calculations validated against anthropomorphic phantoms; tissue characteristics validated against histopathologic specimens annotated by board-certified pathologists. Metrics include bias, intercept, slope, and R² for lumen area, wall area, stenosis, wall thickness, plaque burden, calcified area, LRNC area, and matrix area. R² values ranged from 0.5222 to 0.9987. Testing accounted for pathologist discordance and radiologist user variability. Results indicate the device is as safe and effective as the predicate.
Technological Characteristics
Software-based image analysis system for CT data. Features multi-dimensional visualization, semi-automated lumen boundary detection, and quantitative measurement tools (stenosis, diameter, length, Hounsfield unit statistics). Operates on standard clinical workstations. Refined processing algorithms compared to predicate. No specific hardware materials or energy sources; software-only device.
Indications for Use
Indicated for trained physicians to assist in the stratification of patients identified to have atherosclerosis using contrast-enhanced multidetector CT images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
December 21, 2018
Elucid Bioimaging, Inc. Andrew Buckler President 225 Main Street, Suite 15 WENHAM, MA 01984
Re: K183012
Trade/Device Name: vascuCAP Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: December 17, 2018 Received: December 17, 2018
Dear Andrew Buckler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hse 2. Nils
for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183012
Device Name vascuCAP
#### Indications for Use (Describe)
vascuCAP is a medical image analysis system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired with contrast from CT imaging devices.
vascuCAP is intended to assist trained physicians in the stratification of patients identified to have atherosclerosis. The software post processes images obtained using a multidetector CT. The package provides tools for the measurement and visualization (color coded maps) of arterial vessels.
Clinicians can select any artery to view the following anatomical references: the highlighted vessel in 3D, two rotatable curved MPR vessel views displayed at angles orthogonal to each other, and cross sections of the vessel. Cross-sectional measurements can be obtained using standard vascuCAP software measuring tools. Clinicians can semi-automatically determine contrasted lumen boundaries, stenosis measurements, and maximum lumen diameters. In addition, clinicians can edit lumen boundaries and examine Hounsfield unit or signal intensity statistics. Clinicians can also manually measure vessel length along the centerline in standard curved MPR views.
The measurements provided by vascuCAP are not intended to provide a diagnosis or clinical recommendations. vascuCAP is intended as a tool to complement standard of care.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### K183012
### 510(K) SUMMARY
| 510(k) SUMMARY | |
|-----------------------|--------------------------------------------------------------------------|
| 510(k) submitter: | Elucid Bioimaging, Inc.<br>225 Main Street, Suite 15<br>Wenham, MA 01984 |
| | Ph. 978-468-0508<br>Fax: 978-468-0527 |
| Contact person: | Andrew J. Buckler, President and CTO, Elucid Bioimaging Inc. |
| Date prepared: | |
| Device: | |
| Name of device: | vascuCAP™ |
| Common or usual name: | Image processing system |
| Classification name: | Picture archiving and communications system |
| Regulatory class: | II |
| Product code: | LLZ |
### Predicate device:
Elucid Bioimaging Inc. vascuCAP A.1.1 (K163071)
#### Device Description:
vascuCAP is an image analysis software package for evaluating CT images of arterial vessels. It allows the processing, review, analysis, communication, and media interchange of multidimensional digital images acquired from CT scanners. vascuCAP provides multi-dimensional visualization of digital images to aid clinicians in their analysis of anatomy and tissue characteristics. The vascuCAP software application user interface follows typical clinical workflow patterns to process, review, and analyze digital images.
#### Intended Use:
vascuCAP is a medical image analysis system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired with contrast from CT imaging devices.
vascuCAP is intended to assist trained physicians in the stratification of patients identified to have atherosclerosis. The software post processes images obtained using a multidetector CT. The package provides tools for the measurement and visualization (color coded maps) of arterial vessels.
Clinicians can select any artery to view the following anatomical references: the highlighted vessel in 3D. two rotatable curved MPR vessel views displayed at angles orthogonal to each other, and cross sections of the vessel. Cross-sectional measurements can be obtained using standard vascuCAP software measuring tools. Clinicians can semi-automatically determine contrasted lumen boundaries, stenosis measurements, and maximum and minimum lumen
{5}------------------------------------------------
diameters. In addition, clinicians can edit lumen boundaries and examine Hounsfield unit or signal intensity statistics. Clinicians can also manually measure vessel length along the centerline in standard curved MPR views.
The measurements provided by vascuCAP are not intended to provide a diagnosis or clinical recommendations. vascuCAP is intended as a tool to complement standard of care.
# Technological Characteristics Comparing to the Predicate:
vascuCAP A.1.2 has all the same technological characteristics and features as vascuCAP A.1, but refines processing algorithms to improve measurement performance.
# Performance Data:
Software verification and validation: Software verification consistent with FDA quidance on "General Principles of Software Validation" was conducted, comprising quality planning, requirements analysis, design reviews, software construction, and testing. Verification testing addressed installation and operation qualification, demonstrating that the product meets defined system requirements and features.
Performance testing: Validation testing using phantom and clinical images was conducted to address performance qualification of the subject device under typical operating conditions. Clinical images were evaluated using vascuCAP. Objectives evaluated included calculations of anatomic structure (compared to anthropomorphic phantoms) and calculations of tissue characteristics (compared to expert annotation by board certified pathologists of histopathologic specimens). As a result of this testing, the following analytic performance metrics have been established*:
| Structure | Lumen Area,<br>tested range 0.3 - 290.1mm² | Bias: 0.81mm² [0.3, 1.9], Intercept: 0.65mm² [-0.6, 0.9],<br>Slope: 1.01 [0.9, 1.0], Quadratic term: 0.0 [0.0, 0.0], R²: 0.9987 |
|-------------|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Wall Area,<br>tested range 9.4 - 448.6mm² | Bias: 0.50mm² [-1.08, 1.29], Intercept: -0.59mm² [-4.1, 2. 8.0],<br>Slope: 1.0 [0.99, 1.04], Quadratic term: 0.0 [0.0, 0.0], R²: 0.9974 |
| | Stenosis**,<br>tested range 33-69% | Vessels ≥5.9mm: Bias: 3.7% [1.29, 4.47], Intercept: 5.99% [-0.81, 9.93],<br>Slope: 0.96 [0.84, 1.1], Quadratic term: -0.01 [-0.02, 0.01], R²: 0.8034<br><br>Vessels <5.9mm: Bias: 9.3% [2.14, 12.72], Intercept: 34.0% [-2.3, 38.9],<br>Slope: 0.55 [0.42, 1.21], Quadratic term: 0.001 [-0.02, 0.06], R²: 0.9549 |
| | Wall Thickness,<br>tested range 1.0 - 9.0mm | Bias: 0.5mm [0.3, 0.6], Intercept: 0.27mm [-0.1, 0.5],<br>Slope: 1.05 [1.01, 1.1], Quadratic term: -0.008 [-0.02, 0.01], R²: 0.9855 |
| | Plaque Burden,<br>tested range 0.4 -1.0 (ratio) | Bias: -0.01 [-0.01, .004], Intercept: 0.01 [-0.1, 0.04],<br>Slope: 0.99 [0.9, 1.1], Quadratic term: 0.03 [-0.1, 0.3], R²: 0.9794 |
| Composition | Calcified Area,<br>tested range 0.0 - 51.2mm² | Difference: 0.15mm² [-0.5, 0.97], Intercept: 0.4mm² [-0.02, 1.6],<br>Slope: 0.9 [0.6, 1.1], Quadratic term: -0.01 [-0.1, 0.04], R²: 0.875 |
| | LRNC Area,<br>tested range 0.0 - 26.8mm² | Difference: 0.8mm² [-0.7, 2.6], Intercept: 1.44mm² [0.2, 3.4],<br>Slope: 0.8 [0.2, 1.1], Quadratic term: 0.004 [-0.1, 0.3], R²: 0.5222 |
| | Matrix Area,<br>tested range 2.6 - 57.1mm² | Difference: -1.6mm² [-3.6, 0.32], Intercept: 2mm² [-3, 5],<br>Slope: 0.83 [0.7, 1.0], Quadratic term: -0.01 [-0.04, 0.01], R²: 0.7469 |
*brief explanatory notes to help interpret the table:
- Range indicates the smallest and largest true value for the measurand tested. .
{6}------------------------------------------------
- . Each metric is presented as a point estimate followed by a 95% confidence interval (CI). The CI is computed from the statistics of the observed data. It is acknowledged that wide confidence intervals make the established metric quite uncertain, and in general stem from the number of tested data points and metric specific factors.
- Bias for structural measurands and plaque burden are derived from phantom experiments such that ground truth is assessed using micrometer measurements on anthropomorphic objects. Width of confidence intervals follow from the relative difficulty of each phantom geometry and typical variation experienced across clinically-accepted scanning protocols. The mean tested phantom vessel size is 8.7mm [3.9mm]. For stenosis, the mean tested vessel size of the vessels ≥5.9mm bin was 5.2mm [3.9mm], and for the vessels <5.9mm bin was 11.9mm [7.9mm, 23.9mm].
- . Systematic difference from histopathology for tissue types is estimated relative to pathologist annotation of ex vivo tissue specimens with paired CTA such that ground truth is assessed based on expert interpretation that the relevant scientific and clinical community relies upon for diagnosis or other specific categorization of the studied tissue. The mean tested specimen vessel size is 7.9mm [3.6mm, 12.9mm]. The tissue specimens are from the carotid artery, and that as a result, may not account for errors due to motion that may be present in imaging of small vessels depending on the use of ECG gating. Width of confidence interval follows from:
- agreement of pathologists (three independent annotations were used for these results to O account for acknowledged discordance in histopathology interpretation),
- certainty of positioning of annotated sections into 3D radiology volume (four combinations o resulting from two unique positioners crossed with two independent radiologist users were used for these results to account for differences in judgment on where the annotated section data applies within the in vivo volume, blinded to vascuCAP results),
- relative difficulty of physiologic presentation, and O
- o typical variation experienced across clinically-accepted scanning protocols.
**important note regarding stenosis by diameter: given the reliance of stenosis by diameter as being computed from lumen diameters, and the relative difficulty of accurately estimating lumen diameter as the lumens become appreciably smaller than the finite voxel size, the stenosis may be overestimated. This issue is not unique to vascuCAP but rather a known issue for any interpretation of CTA as lumen size decreases. It is important to follow current clinical training to disregard quantitative calculations of stenosis by diameter from CTA when the lumen is not readily visualized and instead for it to be judged qualitatively. Use of such calculations as %stenosis by area, also available from vascuCAP, mitigates but does not completely avoid this issue.
See User Guide for tables of scanner makes, models, and settings used in the testing as well as patient characteristics of the tested population.
# Conclusions:
Based on software verification and validation comprising bench and clinical testing under typical operating conditions, Elucid Bioimaging concludes that vascuCAP A.1.2 is as safe and effective as the predicate device for the intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.