← Product Code [OUR](/productcode/OUR) · K182983

# iFuse Implant System® (K182983)

_SI-BONE, Inc. · OUR · Nov 27, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K182983

## Device Facts

- **Applicant:** SI-BONE, Inc.
- **Product Code:** [OUR](/productcode/OUR.md)
- **Decision Date:** Nov 27, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The iFuse Implant System is intended for sacroiliac fusion for conditions including sacroiliac joint dysfunction that is a direct result of sacroiliac joint disruption and degenerative sacroiliitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months.

## Device Story

iFuse Implant System consists of cannulated triangular titanium implants and manual surgical instruments; used for sacroiliac (SI) joint fusion. Principle of operation: triangular shape and porous surface minimize joint motion/micromotion; interference fit provides fixation and stabilization to facilitate fusion. Surgical placement performed by physicians using guide pins for accuracy. Implants provided sterile via gamma sterilization. System stabilizes SI joint to treat dysfunction and degenerative sacroiliitis, potentially reducing pain and improving patient mobility.

## Clinical Evidence

No clinical data or performance testing required for this submission as the device labeling modification did not alter technological characteristics or intended use.

## Technological Characteristics

Cannulated triangular implants; material: Ti 6Al 4V ELI (ASTM F136/F1580 for iFuse, ASTM F3001 for iFuse-3D); porous surface; manual instrument delivery system; gamma sterilized; no software or electronic components.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- iFuse Implant System® by SI-BONE, Inc. ([K162733](/device/K162733.md))

## Reference Devices

- VIPER® Systems (Pedicle Screw Spinal System) ([K121020](/device/K121020.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the bottom line.

November 27, 2018

SI-BONE, Inc. Roxanne J. Dubois Vice President, Regulatory Affairs 471 El Camino Real, Suite 101 Santa Clara, California 95050

Re: K182983

Trade/Device Name: iFuse Implant System® Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: OUR Dated: October 26, 2018 Received: October 29, 2018

Dear Ms. Dubois:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Ronald P. Jean -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K182983

Device Name iFuse Implant System®

Indications for Use (Describe)

The iFuse Implant System is intended for sacroiliac fusion for conditions including sacroiliac joint dysfunction that is a direct result of sacroiliac joint disruption and degenerative sacroilitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months.

| Type of Use (Select one or both, as applicable)                                    |                                                                                   |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY - iFuse Implant System®

#### I. SUBMITTER

SI-BONE, Inc. 471 El Camino Real, Suite 101, Santa Clara, CA 95050 Phone: 408-207-0700; Fax: 408-557-8312

Contact Person: Roxanne J. Dubois, VP, Regulatory Affairs, SI-BONE, Inc. Email: rdubois@si-bone.com Phone: 408-828-5019; Fax: 408-557-8312 Date Prepared: October 26, 2018

### II. DEVICE

| Trade Name of Device: | iFuse Implant System <sup>®</sup>                          |
|-----------------------|------------------------------------------------------------|
| Common or Usual Name: | Sacroiliac Joint Fixation                                  |
| Regulation Number:    | 21 CFR 888.3040, Smooth or threaded metallic bone fastener |
| Product Code:         | OUR                                                        |

### III. PREDICATE AND REFERENCE DEVICES

| Predicate:        | iFuse Implant System® by SI-BONE, Inc., K162733       |
|-------------------|-------------------------------------------------------|
| Reference Device: | VIPER® Systems (Pedicle Screw Spinal System), K121020 |

#### IV. DEVICE DESCRIPTION

The iFuse Implant System® consists of cannulated triangular, titanium (iFuse implants: Ti 6Al 4V ELI, ASTM F136/F1580 and iFuse-3D implants: Ti 6Al 4V ELI, ASTM F3001) implants with a porous surface and an instrument system. The principle of operation is that the triangular implant shape and porous surface are designed to prevent and minimize motion/micromotion of the sacroiliac (SI) joint, and thereby stabilize the joint. The mechanism of action is that the interference fit allows fixation, stabilization and fusion. The delivery system uses guide pins for accurate permanent surgical placement. The implants are available in varying lengths and diameters and are provided sterile (gamma sterilization).

#### V. INDICATIONS FOR USE

The iFuse Implant System is intended for sacroiliac fusion for conditions including sacroiliac joint dysfunction that is a direct result of sacroiliac joint disruption and degenerative sacroiliitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months.

{4}------------------------------------------------

### TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE VI.

The iFuse Implant System® consists of manual instruments and a series of triangular titanium implants intended for surgical implantation to provide immediate SI joint stabilization and allow long-term fusion. There are no changes to the technological characteristics of the device that is the subject of this 510(k).

# VII. PERFORMANCE DATA

No performance testing was required to support the modified labeling that is the subject of this 510(k).

## VIII. CONCLUSIONS

The intended use, indications for use and the technological characteristics are unchanged compared to the predicate device and support that the proposal contained within this 510(k) is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K182983](https://fda.innolitics.com/device/K182983)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
