← Product Code [OTD](/productcode/OTD) · K182957

# Heli-FX EndoAnchor System (K182957)

_Medtronic Vascular, Inc. · OTD · Nov 21, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K182957

## Device Facts

- **Applicant:** Medtronic Vascular, Inc.
- **Product Code:** [OTD](/productcode/OTD.md)
- **Decision Date:** Nov 21, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.3460
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Heli-FX EndoAnchor System is intended to provide fixation and sealing between endovascular aortic grafts and the native artery. The Heli-FX EndoAnchor system is indicated for use in patients whose endovascular grafts have exhibited migration or endoleak, or are at risk of such complications, in whom augmented radial fixation and/or sealing is required to regain or maintain adequate aneurysm exclusion. The EndoAnchor may be implanted at the initial endograft placement, or during a secondary (i.e., repair) procedure.

## Device Story

System provides mechanical fixation between endovascular aortic grafts and native artery; consists of helical nickel-cobalt wire EndoAnchors, catheter-based Applier, and deflectable Heli-FX Guide sheath. Used by physicians in clinical settings during initial endograft placement or secondary repair procedures. Applier delivers individual anchors through graft material into vessel wall; helical shape and crossbar prevent over-penetration. Guide sheath facilitates navigation through tortuous vasculature to target site. System enhances graft stability and sealing, reducing risk of migration or endoleak; benefits patient by maintaining aneurysm exclusion.

## Clinical Evidence

No clinical data. Substantial equivalence supported by existing performance testing and compatibility rationale.

## Technological Characteristics

EndoAnchor: medical-grade nickel-cobalt wire. Applier: 12Fr OD catheter with control handle. Guide: 12Fr ID sheath with control handle and obturator; 62cm (16Fr OD) or 90cm (18Fr OD) lengths. Compatible with 0.035" guidewire. Sterile, single-use, disposable.

## Regulatory Identification

An endovascular suturing system is a medical device intended to provide fixation and sealing between an endovascular graft and the native artery. The system is comprised of the implant device and an endovascular delivery device used to implant the endovascular suture.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The device should be demonstrated to be biocompatible;
(2) Sterility and shelf life testing should demonstrate the sterility of patient-contacting components and the shelf-life of these components;
(3) Non-clinical and clinical performance testing should demonstrate substantial equivalence in safety and effectiveness, including durability, compatibility, migration resistance, corrosion resistance, and delivery and deployment;
(4) Non-clinical testing should evaluate the compatibility of the device in an magnetic resonance (MR) environment;
(5) Appropriate analysis and non-clinical testing should validate electromagnetic compatibility (EMC) and electrical safety;
(6) The sale, distribution, and use of the device are restricted to prescription use in accordance with 21 CFR 801.109 of this chapter; and
(7) Labeling must bear all information required for the safe and effective use of the device as outlined in § 801.109(c) of this chapter, including a detailed summary of the non-clinical and clinical evaluations pertinent to use of the device.

In addition to the general controls of the Food, Drug & Cosmetic Act, the Endovascular Suturing System is subject to the following special controls:

(1) The device should be demonstrated to be biocompatible;

(2) Sterility and shelf life testing should demonstrate the sterility of patient-contacting components and the shelf-life of these components;

(3) Non-clinical and clinical performance testing should demonstrate substantial equivalence in safety and effectiveness, including durability, compatibility, migration resistance, corrosion resistance, and delivery and deployment;

(4) Non-clinical testing should evaluate the compatibility of the device in an MR environment;

(5) Appropriate analysis and non-clinical testing should validate electromagnetic compatibility (EMC) and electrical safety;

(6) The sale, distribution, and use of the device are restricted to prescription use in accordance with 21 CFR 801.109; and

(7) Labeling must bear all information required for the safe and effective use of the device as outlined in 801.109(c), including a detailed summary of the non-clinical and clinical evaluations pertinent to use of the device.

## Predicate Devices

- Heli-FX EndoAnchor System ([K171427](/device/K171427.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 21, 2018

Medtronic Vascular, Inc. Janelle Wong Keller Regulatory Affairs Specialist 3850 Brickway Boulevard Santa Rosa, California 95403

Re: K182957

Trade/Device Name: Heli-FX EndoAnchor System Regulation Number: 21 CFR 870.3460 Regulation Name: Endovascular Suturing System Regulatory Class: Class II Product Code: OTD Dated: October 22, 2018 Received: October 24, 2018

Dear Janelle Wong Keller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Brian D. Pullin -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K182957

Device Name Heli-FX EndoAnchor System

#### Indications for Use (Describe)

The Heli-FX EndoAnchor System is intended to provide fixation and sealing between endovascular aortic grafts and the native artery. The Heli-FX EndoAnchor system is indicated for use in patients whose endovascular grafts have exhibited migration or endoleak, or are at risk of such complications, in whom augmented radial fixation and/or sealing is required to regain or maintain adequate aneurysm exclusion.

The EndoAnchor may be implanted at the initial endograft placement, or during a secondary (i.e., repair) procedure.

| Type of Use (Select one or both, as applicable)                 |
|-----------------------------------------------------------------|
| <span>    Prescription Use (Part 21 CFR 801 Subpart D)  </span> |
| <span>    Over-The-Counter Use (21 CFR 801 Subpart C)  </span>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

510(k) Number: K182957

### General Information

#### Manufacturer

Medtronic 710 Medtronic Parkway Minneapolis, MN 55432 Tel: (763) 514-4000

#### Contact Person

Janelle Wong Keller Regulatory Affairs Specialist

#### Alternate Contact Person

Burt Goodson Senior Principal Regulatory Affairs Specialist

#### Date Prepared

15 November 2018

### Device Description

The Heli-FXTM EndoAnchor™ Systems are each comprised of the respective Heli-FX Applier with EndoAnchor Cassette, containing ten EndoAnchor implants; the respective Heli-FX Guides; and the Ancillary EndoAnchor Cassette, containing five EndoAnchor implants.

#### Generic/Common Name and Classification

Endovascular Suturing System (product code OTD) per 21 CFR 870.3460. Heli-FX™ EndoAnchor™ Systems (comprising the Heli-FX EndoAnchor system and Heli-FX thoracic EndoAnchor System) are Class II devices.

#### Trade Name

Heli-FX EndoAnchor System

- Heli-FX Applier with EndoAnchor Cassette containing ten (10) EndoAnchor implants ●
- Heli-FX Guide ●
- Ancillary EndoAnchor Cassette containing five (5) EndoAnchor implants ●

{4}------------------------------------------------

#### Predicate Devices

- Heli-FX EndoAnchor System per K171427 ●
#### Intended Use

The Heli-FX EndoAnchor System is intended to provide fixation and sealing between endovascular aortic grafts and the native artery. The Heli-FX EndoAnchor System is indicated for use in patients whose endovascular grafts have exhibited migration or endoleak, or are at risk of such complications, in whom augmented radial fixation and/or sealing is required to regain or maintain adequate aneurysm exclusion.

The EndoAnchor may be implanted at the time of the initial endograft placement, or during a secondary (i.e., repair) procedure.

# Product Description

The Heli-FX EndoAnchor System comprises the EndoAnchor implant, an intravascularly-applied suture, supplied with the Heli-FX Applier in a Cassette containing ten (10) EndoAnchor implants, or separately in an Ancillary Cassette containing five (5) EndoAnchor implants; the Heli-FX Applier, a catheter-based device for placement of the EndoAnchor; and the Heli-FX Guide, a deflectable sheath to position the Applier.

The EndoAnchor is an endovascularly-placed suture designed to attach aortic endografts to the native vessel wall. The EndoAnchor is manufactured from medical-grade nickel-cobalt wire and is wound in a helical shape. The leading end is sharpened to a conical point to act as an integral needle facilitating atraumatic deployment through the graft material and vessel wall. The proximal end of the EndoAnchor includes a diagonal crossbar, which functions as a suture anchor designed to prevent over penetration of the EndoAnchor. Ten (10) EndoAnchor implants are prepackaged into a cassette, which is supplied with the Heli-FX Applier. The cassette is designed to facilitate easy and accurate loading of the EndoAnchor into the Applier catheter. EndoAnchor implants are also supplied separately in an Ancillary Cassette containing five (5) additional EndoAnchor implants.

The Heli-FX Applier is designed to implant the EndoAnchor. The Applier implants one EndoAnchor at a time, and can be used to implant multiple EndoAnchor implants in a single patient. The Applier is designed for use with the Heli-FX Guide. The Applier is a 12Fr (OD) catheter with an integrated control handle. Two Applier lengths are available for anchoring in different regions of the aorta.

The Heli-FX Guide is a sterile, single use, disposable device designed to direct the Heli-FX Applier to the desired location for EndoAnchor implantation. The device is compatible with a 0.035" Guide wire. The Heli-FX Guide consists of a 12 Fr-compatible (inner diameter) Guide sheath with integrated control handle, and a matching 12 Fr OD obturator. Deflection of the distal tip of the catheter is accomplished by rotating the Control Knob located on the control handle. The Guide is available in both 62cm (16Fr OD) and 90cm (18Fr OD) working lengths. Multiple deflectable tip lengths are available to accommodate a range of aortic diameters. The Obturator is

{5}------------------------------------------------

used during vessel access and is designed to follow the Guide wire and provide access through tortuous vasculature.

# Special Controls

Special controls have been established for endovascular suturing systems per 21 CFR 870.3460(b). These special controls include specific requirements related to biocompatibility. sterility and shelf-life, performance testing, MR compatibility, electromagnetic compatibility and electrical safety, labeling, and the prescription-only status of the devices.

# Substantial Equivalence

The Heli-FX EndoAnchor System and its intended use are identical to that cleared under K171427. The only difference that constitutes this Special 510(k) is the revision of the Heli-FX EndoAnchor System Instructions for Use to state that the Heli-FX EndoAnchor System is compatible with the Valiant Navion Stent Graft System. Adequate rationale was provided to support the compatibility of these systems based on the previously provided testing with the endografts that are currently listed as compatible with the Heli-FX EndoAnchor System in its IFU. Thus, no new testing with the Valiant Navion Stent Graft System was needed.

### Summary

The data and information presented in this application, support a determination of substantial equivalence of the predicate device: Heli-FX EndoAnchor System.

---

**Source:** [https://fda.innolitics.com/device/K182957](https://fda.innolitics.com/device/K182957)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
