← Product Code [MUE](/productcode/MUE) · K182951

# Pristina Serena 3D (K182951)

_GE Healthcare · MUE · Jan 18, 2019 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K182951

## Device Facts

- **Applicant:** GE Healthcare
- **Product Code:** [MUE](/productcode/MUE.md)
- **Decision Date:** Jan 18, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1715
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

Pristina Serena 3D is an optional accessory of Senographe Pristina intended to provide accurate location of lesions in the breast in three dimensions.

## Device Story

Pristina Serena 3D is a software-based biopsy guidance option for the Senographe Pristina mammography system. It utilizes DBT images to determine 3D coordinates (X, Y, Z) of breast lesions. The system calculates the required position for a motorized biopsy device holder based on lesion coordinates and system geometry. The radiologist selects a vertical or horizontal needle approach; the needle is inserted until it reaches a mechanical stop, ensuring the tip aligns with the target. Used in clinical settings by radiologists/technicians for interventional procedures. The device reduces patient dose by using a single DBT acquisition for both positioning and targeting, replacing the multiple X-ray acquisitions required by traditional stereotactic procedures. Output is visual guidance for accurate needle placement, facilitating biopsy or localization.

## Clinical Evidence

Bench testing only. No clinical data. Evidence includes image quality and dose testing demonstrating performance equivalent to the Senographe Pristina 3D system, and design/usability validation confirming that the device functions according to specifications.

## Technological Characteristics

Software-based biopsy guidance option for Senographe Pristina. Uses DBT image chain and reconstruction. Motorized biopsy device holder. Supports vertical and horizontal needle approaches. Connectivity: Integrated into existing Senographe Pristina platform. No hardware modifications required. Complies with 21 CFR 820 and ISO 13485.

## Regulatory Identification

A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.

## Predicate Devices

- Pristina Serena ([K173576](/device/K173576.md))

## Reference Devices

- Senographe Pristina 3D ([P130020](/device/P130020.md)/S002 and S003)
- Hologic's Affirm Breast Biopsy Guidance System ([K122836](/device/K122836.md))
- Affirm Lateral Arm Upright Biopsy Accessory ([K161575](/device/K161575.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA) along with the Department of Health & Human Services. The FDA logo features a blue square with the letters 'FDA' in white, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue. To the left of the FDA logo is the Department of Health & Human Services seal, which includes an emblem with a caduceus-like symbol and the text 'DEPARTMENT OF HEALTH & HUMAN SERVICES-USA' arranged around it.

January 18, 2019

GE Healthcare Nicole Landreville Regulatory Affairs Manager 283 rue de la Miniere BUC, 78530 FRANCE

Re: K182951

Trade/Device Name: Pristina Serena 3D Regulation Number: 21 CFR 892.1715 Regulation Name: Full-Field Digital Mammography System Regulatory Class: Class II Product Code: MUE Dated: October 22, 2018 Received: October 23, 2018

Dear Nicole Landreville:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical 

{1}------------------------------------------------

device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and --------------------------------------------------------------------------------------------------------------------------------------------------------------------------CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara

For

Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K182951

Device Name Pristina Serena 3D

#### Indications for Use (Describe)

The Pristina Serena 3D option provides the three-dimensional location of target lesions, using information obtained from Digital Breast Tomosynthesis (DBT) images. This information provides guidance for a variety of minimally invasive or interventional procedures in the breast such as: vacuum assisted biopsy, core biopsy, pre-surgical localization (e.q. hookwire), and fine needle aspirations (FNA).

Type of Use (Select one or both, as applicable)

| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

| Date:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | January 18, 2018                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | GE Healthcare<br>GE Medical Systems SCS<br>283 RUE DE LA MINIERE<br>78530 BUC – FRANCE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Primary Contact Person:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Nicole Landreville, Eng, RAC, FRAPS<br>Regulatory Affairs Manager<br>380 Hampton Heath road<br>Burlington, ON, Canada<br>L7L 4R2<br>Phone : 289-208-2365<br>Email : nicole.landreville@ge.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Secondary Contact Person:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Gregory Pessato,<br>Regulatory Affairs Program Manager,<br>GE Medical Systems SCS<br>283 RUE DE LA MINIERE<br>78530 BUC – FRANCE<br>Phone : + 33 1 30 70 93 16<br>Email : gregorypessato@ge.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Device Trade Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Pristina Serena 3D                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Common/Usual Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Stereotaxy biopsy guidance application for a Digital Breast Tomosynthesis (DBT) Mammography System Biopsy System for Senographe Pristina                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Classification Names:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 21 CFR 892.1715, Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Product Code:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | MUE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Predicate Device:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Pristina Serena (K173576)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Reference devices:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Senographe Pristina 3D (P130020/S002 and S003) Hologic's Affirm Breast Biopsy Guidance System (K122836) and Affirm Lateral Arm Upright Biopsy Accessory (K161575)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device Description:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Pristina Serena 3D is DBT Biopsy System for Senographe Pristina. It is an additional option that builds upon the Pristina Serena device (GE Healthcare Stereotaxy biopsy option for Senographe Pristina platform). Pristina Serena was cleared on May 14, 2018 (K173576).<br><br>Pristina Serena 3D enables biopsy medical application to be done using Digital Breast Tomosynthesis (DBT) images.<br><br>The Pristina Serena 3D add-on includes the following items:  Software: A new software version for the Senographe Pristina platform which includes software to manage the Pristina Serena 3D option. Labeling for the Biopsy DBT (3D) Medical application.              Pristina Serena 3D option is compatible with previously installed Senographe Pristina systems. Pristina Serena 3D does not require any hardware modification on the Senographe Pristina platform. The hardware that was cleared on Pristina Serena 3D (K173576) was also not modified.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Intended Use:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Pristina Serena 3D is an optional accessory of Senographe Pristina intended to provide accurate location of lesions in the breast in three dimensions.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Indications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | The Pristina Serena 3D option provides the three-dimensional location of target lesions, using information obtained from Digital Breast Tomosynthesis (DBT) images. This information provides guidance for a variety of minimally invasive or interventional procedures in the breast such as: vacuum assisted biopsy, core biopsy, pre-surgical localization (e.g. hookwire), and fine needle aspirations (ENA)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Technology:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Pristina Serena 3D uses the DBT images to determine the three-dimensional (3D) location (X, Y and Z coordinates) of an object of interest in the breast (such as a suspicious lesion). The X and Y coordinates are determined directly from the displayed plane as it is parallel to the detector while the Z coordinate is determined with the height of this same plane in the whole reconstructed volume.<br>The target lesion 3D coordinates information, together with the complete geometry of the device, are used to compute the required position of a biopsy device holder that will allow intervention in the breast at the exact target position (biopsy of sample tissue or placement of a hook wire for guidance of surgical interventions). With Pristina Serena 3D, the positioning of the biopsy device holder is motorized and takes into account the geometry of the biopsy needles, so when the biopsy device holder is in place, the user introduces the needle in the breast until reaching the mechanical stop of the biopsy device holder. As the biopsy needle is fixed on and guided by the biopsy device holder (not handed), the needle tip (or notch) will then be at the target lesion 3D coordinates.<br>As it was for Pristina Serena option, the Pristina Serena 3D allows two different needle approaches for Biopsy: vertical and horizontal (left or right).  In vertical approach, the needle is introduced from the “top” of the compressed breast. In horizontal approach, the needle is introduced from the side of the compressed breast.              Depending on patient morphology and location of the lesion to be biopsied in the breast, a radiologist might choose an approach or the other. Usually an effort is made to use an approach that would limit the distance crossed by the needle in the breast to reach the target location.<br>Since Pristina Serena 3D option uses the Image chain of the Senographe Pristina platform, the image quality of images during the biopsy procedure is equivalent to that of standard screening/diagnostic images of Senographe Pristina 3D.<br>Regarding dose management, with Pristina Serena 3D, a single DBT acquisition allows for both confirming appropriate breast positioning and for lesion targeting (as opposed to a Stereotaxy procedure that requires three X-ray acquisitions: one scout image |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | plus a stereotaxic pair). This translates into dose reduction to<br>patient.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Comparison with<br>Reference Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | In regards with Image Quality / Image processing, the Proposed<br>device and the reference device Senographe Pristina 3D<br>(P130020/S002 and S003) both use the same DBT image chain<br>and DBT reconstruction.<br>Regarding dose management, the Proposed Pristina Serena 3D<br>uses a single AOP mode available when acquiring biopsy<br>images. This AOP mode provides the same parameters as the<br>Senographe Pristina 3D AOP mode called "STD+"<br>(P130020/S003 3D STD+ on Senographe Pristina 3D<br>submission).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | The Proposed Pristina Serena 3D and the reference device<br>Hologic's Affirm Breast Biopsy Guidance System (K122836)<br>and Affirm Lateral Arm Upright Biopsy Accessory (K161575),<br>both DBT features use tomosynthesis images to provide accurate<br>location of lesions in the breast in three dimensions.<br>Both 3D localization features calculate a three dimensional<br>location for percutaneous placement for biopsy, pre-surgical<br>localization or treatment devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | It is important to note also that the Senographe Pristina with<br>Pristina Serena and Pristina Serena 3D offer both the 2D<br>stereotactic or 3D tomosynthesis image acquisition.<br>Both DBT features use the same method of use, they both use a<br>Dedicated Breast Biopsy Positioner, same mechanism of action<br>and same stated accuracy. They have the same Stereotactic<br>Angle (±15°) but they differ slightly in DBT Tomographic angle<br>15° total (± 7.5 degrees) for Hologics reference device versus<br>25° total (± 12.5°) (which has no impact on the target coordinate<br>computation for biopsy applications).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Determination of<br>Substantial Equivalence:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | The device has successfully completed required design control<br>testing per GE Healthcare's quality management system. No<br>unexpected test results were obtained. The Pristina Serena 3D<br>option was designed and will be manufactured under the Quality<br>System Regulations of 21CFR 820 and ISO 13485. The<br>following quality assurance measures were applied to the<br>development of the system:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Risk Analysis<br>•<br>Design Reviews<br>•                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| •                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Software Development Lifecycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| •                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Testing on unit level (Module verification)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| •                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Integration testing (System verification)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| •                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Performance testing (Verification)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| •                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Safety testing (Verification)…

---

**Source:** [https://fda.innolitics.com/device/K182951](https://fda.innolitics.com/device/K182951)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
