The TRAUS Dental Handpieces, TRAUS CRB46LN and TRAUS CRB46NN are intended for a wide range of dental procedures including: A. Implant placement , including 1. Preparation of the osteotomy site 2. Bone contouring , osteoplasty B. Periodontal surgeries 1. Bone contouring & alveoplasty around living teeth 2. Removal of exostosis C. Bone grafting 1. Preparation of the donor site (for e.g. symphysis and ascending rames etc.) 2. Harvesting autogen living bone 3. Sinus elevation & grafting of alveolar sockets D. Removal and sectioning of teeth and teeth bone for e.g. impacted third molars and complicated extractions
Device Story
TRAUS Dental Handpieces (models CRB46LN, CRB46NN) are gear-driven, hand-held dental instruments with 20:1 gear ratio. Devices attach to torque-adjustable electrical motors via ISO 3964 coupling. Head clamp accepts instruments per ISO 1797-1. Used by dental professionals in clinical settings for surgical procedures including implant placement, bone grafting, periodontal surgery, and tooth extraction. Handpieces facilitate access to difficult-to-reach areas for cavity preparation and cleaning. Output is mechanical rotation of dental burs/instruments; healthcare providers use visual and tactile feedback during operation to perform surgical tasks. Benefits include precise bone and tooth manipulation during complex dental surgeries.
Clinical Evidence
Bench testing only. Compliance with ISO 14457:2017 (Dentistry - Handpieces and Motors) verified. Biocompatibility testing performed per ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization/irritation). Sterilization validated per ISO 17665-1/2 and FDA reprocessing guidance.
Technological Characteristics
Gear-driven handpiece; 20:1 gear ratio; 40,000 rpm max speed (2,000 rpm reduced). Materials: SUS303F (head/nozzle, hard chrome coated), SUS420F (chuck/gear, vacuum heat treated), SUS304 (pipe, hard chrome coated). ISO 3964 coupling; ISO 1797-1 chuck. Push-button locking. Steam sterilization (132°C/4 min).
Indications for Use
Indicated for dental procedures including implant placement (osteotomy preparation, bone contouring/osteoplasty), periodontal surgeries (bone contouring, alveoplasty, exostosis removal), bone grafting (donor site preparation, autogenous bone harvesting, sinus elevation, alveolar socket grafting), and tooth/bone removal/sectioning (impacted third molars, complicated extractions).
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 26, 2019
Saeshin Precision Co., Ltd. Choi Jong Woo Quality Manager 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun Daegu, 42921 Republic of Korea
Re: K182892
Trade/Device Name: TRAUS Dental Handpieces Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: Class I Product Code: EGS Dated: August 21, 2019 Received: September 3, 2019
Dear Choi Jong Woo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182892
Device Name TRAUS Dental Handpieces
Indications for Use (Describe)
The TRAUS Dental Handpieces, TRAUS CRB46LN and TRAUS CRB46NN are intended for a wide range of dental procedures including:
A. Implant placement , including
- 1. Preparation of the osteotomy site
- 2. Bone contouring , osteoplasty
- B. Periodontal surgeries
- 1. Bone contouring & alveoplasty around living teeth
- 2. Removal of exostosis
- C. Bone grafting
- 1. Preparation of the donor site (for e.g. symphysis and ascending rames etc.)
- 2. Harvesting autogen living bone
- 3. Sinus elevation & grafting of alveolar sockets
D. Removal and sectioning of teeth and teeth bone for e.g. impacted third molars and complicated extractions
Type of Use (Select one or both, as applicable)
| <span> <b> <span style="font-size: 16px;">☑</span> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|----------------------------------------------------------------------------------------------------------------------------|
| <span> <b> <span style="font-size: 16px;">☐</span> </b> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for SAESHIN. The logo features a blue geometric shape above the company name, which is also in blue. Below the logo is the company's address, which includes "echeon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu".
# 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-Tel: +82 53 587 2345 / Fax: +82 53 580 0939
#### 510(k) Summary
[As required by 21 CFR 807.92]
#### 1. 510(k) Number
K182892
# 2. Date Prepared [21 CFR 807.92(a)(1)]
November 20, 2019
## 3. Submitter's Information [21 CFR807.92(a)(1)]
- . Name of Applicant: Saeshin Precision Co., Ltd.
- # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Address: -Daegu, 42921, Republic of Korea
- Contact Name: Jong Woo, Choi (Mr.) / Quality Manager .
- Telephone No. : +82 53 587 2345 -
- -Fax No. : +82 53 580 0939
- Email Address : ksqc@saeshin.com -
- . Registration Number: 3007958831
## 4. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| • | Trade Name: | TRAUS Dental Handpieces |
|---|----------------------------|----------------------------------------------------------|
| • | Classification Name: | Dental Handpieces and Accessories |
| • | Common Name: | Handpiece, Contra- And Right-Angle Attachment,<br>Dental |
| • | Classification Panel: | Dental |
| • | Classification Regulation: | 21 CFR 872.4200 |
| • | Product Code: | EGS |
| • | Device Class: | I |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo and contact information for a company called SAESHIN. The logo is a blue geometric shape. The address is # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 42921, Korea. The telephone number is +82 53 587 2345 and the fax number is +82 53 580 0939.
# 5. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicate devices within this submission are shown as follows:
- 510(k) Number: K171436 ●
- Applicant: Saeshin Precision Co., Ltd. .
- Common Name: . Dental Handpieces and Accessories
- Device Name: STRONG Dental Handpieces
There are no significant differences between the Model TRAUS Dental Handpieces and the predicate device. It is substantially equivalent to these devices in device design, composition of materials and technical specifications.
## 6. Description of the Device [21 CFR 807.92(a)(4)]
The TRAUS Dental Handpieces; TRAUS CRB46LN and TRAUS CRB46NN are gear driven hand-held dental handpieces with Gear Ratio of 20:1. They can be driven by torque adjustable electrical motors for surgery treatment. They are attached to drive via ISO 3964 coupling. The head clamp accepts instrument complying with ISO 1797-1. They have contra-angle attachment intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
# 7. Indications for Use [21 CFR 807.92(a)(5)]
The TRAUS Dental Handpieces, TRAUS CRB46LN and TRAUS CRB46NN are intended for a wide range of dental procedures including:
- A. Implant placement, including
- 1. Preparation of the osteotomy site
- 2. Bone contouring, osteoplasty
- B. Periodontal surgeries
- 1. Bone contouring & alveoplasty around living teeth
- 2. Removal of exostosis
- C. Bone grafting
- 1. Preparation of the donor site (for e.g. symphysis and ascending rames etc.)
- 2. Harvesting autogen living bone
- 3. Sinus elevation & grafting of alveolar sockets
- D. Removal and sectioning of teeth and teeth bone for e.g. impacted third molars and complicated extractions
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for SAESHIN. The logo consists of a blue geometric shape above the word "SAESHIN" in a stylized blue font. The geometric shape is made up of several smaller shapes, creating a three-dimensional effect. The overall design is modern and corporate.
# 8. Technological Characteristics [21 CFR 807.92(a)(6)]
| | | TRAUS CRB46LN | TRAUS CRB46NN |
|-------------------|-----------------------------------|-----------------|-----------------|
| Revolu-<br>tion | Max Speed<br>in Rpm | 40,000 rpm | 40,000 rpm |
| | After the gear<br>speed reduction | 2,000 rpm | 2,000 rpm |
| Gear Ratio | | 20:1 | 20:1 |
| Weight | | approx.87g | approx.88g |
| Size | | Ø20 x 94.8mm | Ø20 x 94.8mm |
| Articles | | Low speed angle | Low speed angle |
| Standard Coupling | | ISO 3964 | ISO 3964 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo and contact information for a company called SAESHIN. The logo is a blue geometric shape. The address is # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 42921, Korea. The telephone number is +82 53 587 2345 and the fax number is +82 53 580 0939.
#### Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92] 9.
When compared to the predicate device (K171436), the STRONG Dental Handpieces presented in this submission has the same:
- Indications for Use .
- Device Design .
- Composition of materials .
- · Technical Specifications
| Parameter | Proposed<br>TRAUS Dental Handpieces (Model Nos.:<br>TRAUS CRB46LN, TRAUS CRB46NN) | Predicate<br>STRONG Dental Handpieces (Model Nos.:<br>ACL(B)-611, ACL(B)-621, ACL(B)-631,<br>ACL(B)-651, ACL(B)-04C) |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Unknown | K171436 |
| Manufacturer | Saeshin Precision Co., Ltd | Saeshin Precision Co., Ltd. |
| Indications for Use | The TRAUS Dental Handpieces are indicated for wide range of dental procedures.<br>A. Implant placement, including<br>1. Preparation of the osteotomy site<br>2. Bone contouring, osteoplasty<br>B. Periodontal surgeries<br>1. Bone contouring & alveoplasty around living teeth<br>2. Removal of exostosis<br>C. Bone grafting<br>1. Preparation of the donor site (for e.g. symphysis and ascending rames etc.)<br>2. Harvesting autogen living bone<br>3. Sinus elevation & grafting of alveolar sockets | The Strong Dental Handpieces are indicated for wide range of dental procedures.<br>A. Implant placement, including<br>1. Preparation of the osteotomy site<br>2. Bone contouring, osteoplasty<br>B. Periodontal surgeries<br>1. Bone contouring & alveoplasty around living teeth<br>2. Removal of exostosis<br>C. Bone grafting<br>1. Preparation of the donor site (for e.g. symphysis and ascending rames etc.)<br>2. Harvesting autogen living bone<br>3. Sinus elevation & grafting of alveolar sockets<br>D. Removal and sectioning of teeth and teeth |
| Parameter | Proposed<br>TRAUS Dental Handpieces (Model Nos.:<br>TRAUS CRB46LN, TRAUS CRB46NN) | Predicate<br>STRONG Dental Handpieces (Model Nos.:<br>ACL(B)-61I, ACL(B)-62I, ACL(B)-631,<br>ACL(B)-651, ACL(B)-04C) |
| | D. Removal and sectioning of teeth and teeth<br>bone for e.g. impacted third molars and<br>complicated extractions | bone for e.g. impacted third molars and<br>complicated extractions |
| Device Design | | |
| Operational Mode | Gear | Gear |
| Gear Ratio | 20:1 | 20:1, 16:1, 64:1, 32:1, 1:1 |
| Length | 94.8 mm | 96.9, 96.9, 99.5, 96.9, 83.3 mm |
| Diameter | Ø 20 | Ø 19.6 |
| Dia. of tool | 2.35 mm | 2.35 mm |
| Max. Overall Length of<br>Rotary Instrument | 30 mm | 30 mm |
| Head Height | 14 mm | 14.45,14.2 mm |
| Head Diameter | Ø 9.6 mm | Ø 8, Ø 7.7 mm |
| Shank | By ISO 1797-1 | By ISO 1797-1 |
| Length of Shank | 9 – 12 mm | 9 – 12 mm |
| Type of Chuck | Push-Button Locking | Push-Button Locking |
| Coupling Dimension | By ISO 3964 | By ISO 3964 |
| Type of Connector | Not Applied | Not Applied |
| Accessories | Spray Nozzle | Spanner |
| Composition of Materials<br>/Surface treatment | | |
| Gear | SUS420F/Vacuum Heat Treatment | SUS420F/Vacuum Heat Treatment |
| Shank | N/A | SUS304 |
| Head and Nozzle | SUS303F/Hard Chrome coated | SUS303F/Hard Chrome coated |
| Chuck | SUS420F/Vacuum Heat Treatment | SUS420F/Vacuum Heat Treatment |
| Handle | SUS303F/Hard Chrome coated | AL6061/Anodizing |
| Pipe | SUS304/Hard Chrome coated | SUS303F/Hard Chrome coated |
| | | N/A (ACL(B)-04C) |
| Patient-Contacting | Chuck, Head | Chuck, Head |
| Operator-Contacting | Handle, Head | Handle, Head |
| Technical Specification | | |
| Chuck Design | Type1 Push-Button Locking by ISO 1797-1 | Type1 Push-Button Locking by ISO 1797-1 |
| Parameter | Proposed<br>TRAUS Dental Handpieces (Model Nos.:<br>TRAUS CRB46LN, TRAUS CRB46NN) | Predicate<br>STRONG Dental Handpieces (Model Nos.:<br>ACL(B)-61I, ACL(B)-62I, ACL(B)-631,<br>ACL(B)-65I, ACL(B)-04C) |
| Bur Extraction Force(N)<br>Max. Torques<br>Max. Water Pressure<br>Max. Speed in rpm<br>Shank Conformance<br>Coupling Dimension<br>Lubricant<br>Chemical Composition<br>Biocompatible<br>Delivery system | 45 N<br>80 Ncm<br>N/A<br>40,000 rpm<br>By ISO 1797-1<br>By ISO 3964<br>N/A<br>510k# | 45 N<br>50 Ncm<br>N/A<br>50,000 rpm, 35,000 rpm<br>By ISO 1797-1<br>By ISO 3964<br>N/A |
| Sterilization | Steam Heat 132 °C / 4minutes | Steam Heat 132 °C / 4minutes |
| Operating principle | The TRAUS Dental Handpieces are gear driven<br>hand-held dental handpieces with gear ratio of<br>20:1. It can be driven by torque adjustable<br>electrical motors for surgery treatment. It is<br>attached to drive via ISO 3964 coupling. The<br>head clamp accepts instrument complying with<br>ISO 1797-1. They have contra angle attachment<br>for difficult to reach areas intended to prepare<br>dental cavities for restorations, such as fillings,<br>and for cleaning teeth. | The STRONG Dental Handpieces are gear<br>driven hand-held dental handpieces with gear<br>ratio of 20:1, 16:1, 64:1, 32:1 and 1:1. It can be<br>driven by torque adjustable electrical motors<br>for surgery treatment. It is attached to drive via<br>ISO 3964 coupling. The head clamp accepts<br>instrument complying with ISO 1797-1. They<br>have contra angle attachment for difficult to<br>reach areas intended to prepare dental cavities<br>for restorations, such as fillings, and for<br>cleaning teeth. |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for SAESHIN. The logo consists of a geometric shape made up of blue triangles above the company name, which is also in blue. Below the company name is the text "Dasa-eup, Dalseong-gun, Daegu".
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for SAESHIN, which is written in a stylized blue font. Above the company name is a geometric blue logo. Below the logo is the address: # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu.
23921, Korea
Tel: +82 53 587 2345 / Fax: +82 53 580 0939
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the contact information for a company called SAESHIN. The address is # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 42921, Korea. The telephone number is +82 53 587 2345, and the fax number is +82 53 580 0939.
# 10. Summary of Non-Clinical Performance Data
#### Biocompatibility
Patient contact materials of the TRAUS Dental Handpieces are as follows:
#### 1) TRAUS CRB46LN, TRAUS CRB46NN
| Part | Material | Surface<br>treatment | Patient<br>contact | Duration of Contact<br>by ISO 10993-1 | Bio-<br>compatibility |
|--------------------|----------|--------------------------|--------------------|---------------------------------------|-----------------------|
| Head and<br>nozzle | SUS303F | Hard Chrome<br>Coated | Yes | Limited (<24 hours) | Yes |
| Chuck | SUS420F | Vacuum heat<br>treatment | Yes | Limited (<24 hours) | Yes |
- Cytotoxicity per ISO 10993-5
- . Sensitization per ISO 10993-10
- . Irritation per ISO 10993-10
#### Bench Testing
The bench tests were performed in order to ensure the performance of the TRAUS Dental Handbieces: TRAUS CRB46LN and TRAUS CRB46NN verify conformity to ISO 14457 and demonstrate substantial equivalence to the predicates.
TRAUS CRB46LN and TRAUS CRB46NN samples were compliant with ISO 14457: 2017 Dentistry - Handpieces and Motors and demonstrated substantial equivalence to the predicates.
#### I Sterilization
The TRAUS Dental handpieces, TRAUS CRB46LN and TRAUS CRB46NN are provided nonsterile and labeled for sterilization by the end user, the Instructions for use include the following parameters:
| Sterilization Type | Temperature | Time | Load<br>Characteristics | Dry Time |
|------------------------------------------|-------------|--------|-------------------------|----------|
| Steam Sterilization<br>(Pre-vacuum Type) | 132 °C | 4 min. | Sterilization Bag | 30 min. |
The sterilization method was validated per :
- . Reprocessing validation per the FDA Guidance Document entitled, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" dated March 17, 2015
- Cleaning Process Validation per ASTM F2314 and ISO 15883-5
- Sterilization Validation per ISO 17665-1 and ISO 17665-2
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo and contact information for a company called SAESHIN. The logo is a blue geometric shape. The contact information includes the address: #52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 42921, Korea, as well as the telephone number: +82 53 587 2345 and fax number: +82 53 580 0939.
# 11.Conclusion [21 CFR 807.92(b)(3)]
In accordance with the Federal Food, Drug and Cosmetic, 21 CFR Part 807, and based on the information provided in this premarket notification Saeshin Precision Co., Ltd. Concludes that the TRAUS Dental Handpieces are substantially equivalent to predicate devices as described herein.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.