KLS Martin Individual Patient Solutions (IPS) Planning System
Applicant
KLS-Martin L.P.
Product Code
PPT · Neurology
Decision Date
Jul 26, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4310
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The KLS Martin Individual Patient Solutions (IPS) Planning System is intended for use as a software system and image segmentation system for the transfer of imaging information from a computerized tomography (CT) medical scan. The input data file is processed by the IPS Planning System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, guides and case reports for use in the marking of cranial surgery. The IPS Planning System is also intended as a pre-operative software tool for simulating surgical treatment options.
Device Story
The KLS Martin IPS Planning System is a software and additive manufacturing platform for cranial reconstructive surgery. It takes patient CT scan data as input. Trained engineers use COTS software (Materialise Mimics, Geomagic Freeform Plus) to segment and manipulate images in collaboration with the physician to create virtual surgical plans. The system outputs digital models, case reports, and physical patient-specific anatomical models and marking guides. Physical outputs are produced via stereolithography (SLA) for resins/acrylics, selective laser melting for titanium, or selective laser sintering for polyamide. The device is used in a clinical setting to assist surgeons in pre-operative simulation and intra-operative bone marking. The system helps improve surgical precision and planning accuracy. All physical outputs are provided non-sterile and require user-performed steam sterilization.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including tensile and bending tests, biocompatibility testing per ISO 10993-1, steam sterilization validation per ISO 17665-1, and LAL endotoxin testing per AAMI ANSI ST72. Software verification and validation confirmed that all software requirements were implemented correctly.
Indicated for adult patients undergoing cranial surgery. Used as a software and image segmentation system to process CT scan data for the creation of patient-specific anatomical models, marking guides, and case reports to assist in surgical planning and bone marking.
Regulatory Classification
Identification
Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.
Predicate Devices
KLS Martin Individual Patient Solutions (IPS) Planning System (K181241)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
KLS-Martin L.P. Katie Rutland Senior Regulatory Affairs Specialist 11201 Saint Johns Industrial Parkway S Jacksonville, Florida 32246
Re: K182889
Trade/Device Name: KLS Martin Individual Patient Solutions (IPS) Planning System Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, And Their Accessories Regulatory Class: Class II Product Code: PPT Dated: July 17, 2019 Received: July 18, 2019
Dear Katie Rutland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Matthew Krueger Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182889
Device Name
KLS Martin Individual Patient Solutions (IPS) Planning System
#### Indications for Use (Describe)
The KLS Martin Individual Patient Solutions (IPS) Planning System is intended for use as a software system and image segmentation system for the transfer of imaging information from a computerized tomography (CT) medical scan. The input data file is processed by the IPS Planning System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, guides and case reports for use in the marking of cranial surgery. The IPS Planning System is also intended as a pre-operative software tool for simulating surgical treatment options.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, while the words "martin L.P." are in gray. The font is sans-serif and bolded.
11201 Saint Johns Industrial Pkwy S • Jacksonville, FL 32246 Phone: 904-641-7746 • 800-625-1557 • Fax: 904-641-7378 www.klsmartinnorthamerica.com
## Section 5 510(k) Summary 21 CFR 807.92
| Submitter: | KLS-Martin L.P.<br>11201 Saint Johns Industrial Pkwy S<br>Jacksonville, FL 32246 | |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Contact Person: | Katie Rutland<br>Senior Regulatory Affairs Specialist<br>Phone: 800-625-1557<br>Email: katie.rutland@klsmartin.com | |
| Date Prepared: | July 25, 2019 | |
| Trade Name: | KLS Martin Individual Patient Solutions (IPS) Planning<br>System | |
| Common Name: | System for the creation of patient specific anatomical<br>models, marking guides, and case reports | |
| Classification Name: | Powered Simple Cranial Drills, Burrs, Trephines, and Their<br>Accessories | |
| Regulatory Class: | II, 21 CFR 882.4310 | |
| Product Code: | PPT | |
| Predicate Devices: | KLS Martin Individual Patient Solutions (IPS) Planning<br>System (K181241) - Primary | |
| Reference Devices: | KLS Martin Individual Patient Solutions (K163579)<br>KLS Martin Individual Patient Solutions (K180962) | |
| Device Description: | The KLS Martin Individual Patient Solutions (IPS) Planning<br>System is a collection of software and associated additive<br>manufacturing (rapid prototyping) equipment intended to<br>provide a variety of outputs to support reconstructive cranial<br>surgeries. The system uses electronic medical images of the<br>patients' anatomy (CT data) with input from the physician,<br>to manipulate original patient images for planning and<br>executing surgery. The system processes the medical images<br>and produces a variety of patient specific physical and/or<br>digital output devices which include anatomical models,<br>guides, and case reports for use in the marking of cranial<br>bone in cranial surgery. | |
| Indications for Use: | The KLS Martin Individual Patient Solutions (IPS) Planning<br>System is intended for use as a software system and image<br>segmentation system for the transfer of imaging information<br>from a computerized tomography (CT) medical scan. The<br>input data file is processed by the IPS Planning System and<br>the result is an output data file that may then be provided as<br>digital models or used as input to a rapid prototyping portion<br>of the system that produces physical outputs including | |
anatomical models, guides, and case reports for use in the
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, while "martin" is in gray. The letters "L.P." are in black and are smaller than the other letters. The logo is simple and modern.
marking of cranial bone in cranial surgery. The IPS Planning System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options.
## Technological Characteristics/Substantial Equivalence Discussion:
The intended use of the subject device, KLS Martin Individual Patient Solutions (IPS) Planning System, is similar to the predicate device, KLS Martin Individual Patient Solutions (IPS) Planning System (K181241):
The subject and predicate device are intended for use as a software system and image segmentation system for the transfer of imaging from a computerized tomography (CT) medical scan. The input data file is processed through the virtual planning software systems and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs. These physical outputs can be anatomical models, guides, and case reports. All digital data and physical devices are used to aid the surgeon during surgery. They are both also intended as a pre-operative software tool for simulating surgical treatment options.
The indications for use statement for the subject device is nearly identical to the predicate device, differing only in anatomical region and output devices. The anatomical region for which the predicate device is intended is listed as maxillofacial, indicating that the device may be used only for maxillofacial applications. The subject device system is specific to only the cranial region and excludes maxillofacial applications. The change in anatomical region presented in the indications for use statement of the subject device system as compared to the predicate device does not change the therapeutic effects of the device. In addition, the subject device offers output devices that include anatomical models, marking guides, and case reports only. The predicate device offers output devices that include anatomical models, cutting guides, marking guides, splints, and case reports. Splints are not applicable to cranial reconstructive surgeries as they are only intended for orthognathic surgeries.
## Similarities to Predicate
Both the subject and primary predicate devices use image data obtained from a CT scan. They both use identical validated commercially off-the-shelf (COTS) software applications to transfer patient imaging from a DICOM format to a .STL format and manipulate the images to produce a final design file. In addition, both devices require trained employees/engineers who utilize the software applications to manipulate data and work with the physician to create the virtual planning session. The physician provides input for model manipulation and interactive feedback through viewing of digital models of system outputs that are modified by the trained employee/engineer during the planning session.
Design validation activities performed to verify the finished output device matches the initial input data (.STL) are identical to the predicate and reference devices, K181241 and K163579.
The subject and predicate device specifications are identical in device thickness, width, length, degree of curvature (in-plane and out-of-plane), screw hole spacing and number of screw holes. A comparison of the subject and predicate device specifications is provided in the table below. The marking guides are non-load bearing, non-implantable, and are only temporarily fixated for the duration of the surgery. A thickness range up to 5.0 mm for the
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for KLS Martin L.P. The text "KLS" is in red, while "martin" is in gray. The letters "L.P." are in black and are larger than the other letters in the logo.
cranial marking guides is to ensure the marking wall is thick enough and clearly visible for accurate marking of bone for resection.
Both systems use traditional (subtractive) and additive manufacturing methods to produce physical output devices that include patient specific anatomical models and guides. In addition, the systems produce digital models and case reports for the physician to use for planning the surgeries or to use during surgery.
Materials used in the manufacture of the output devices are identical in that both the subject and predicate devices use biocompatible polymers/acrylic resins, polyamide, and titanium (CP Titanium & Titanium Alloy). All output devices from both systems are provided non-sterile and must be sterilized by the end user prior to use. Validated sterilization studies performed to ensure a sterility assurance level (SAL) of 10- are identical for both the subject and predicate devices.
Both the subject and predicate devices are temporarily fixated with previously cleared screws of varying sizes and lengths to aid the surgeon in marking of bone more precisely and accurately.
## Differences to Predicate
The intended use of the subject device system is very similar to the predicate device, differing only in the anatomical region in which the device is intended for. The subject device is intended for cranial regions in cranial surgeries, while the predicate device is intended for maxillomandibular regions for maxillofacial surgeries.
The subject device utilizes the same commercially off-the-shelf (COTS) software applications for image segmentation and manipulation as the predicate device, except for the use of the software application IPS CaseDesigner and MathWorks® MATLAB, which are used to manipulate data for production of splints. The subject device system will not be outputting splints as these are intended for orthognathic surgeries.
The subject devices are temporarily fixated with previously cleared screws that range in diameter sizes of 1.0 mm - 2.7 mm with lengths ranging 2 mm - 22 mm. The predicate devices are temporarily fixated with the same previously cleared screws except for the 1.0 mm and 1.2 mm diameter screws.
The potential impact on substantial equivalence of each technological difference was addressed through risk analysis and verification and validation testing.
## Reference Devices
The KLS Martin Individual Patient Solutions (K163579 and K180962) have been included as reference devices to leverage performance testing regarding the material composition (material recycling process, degradation, tensile & bending), biocompatibility, manufacturing processes, as well as cleaning and sterilization for titanium output devices. The KLS Martin IPS Planning System titanium output devices are manufactured from identical materials, undergo the same manufacturing processes (traditional & additive), have the same biocompatibility, demonstrate similar performance characteristics, and are designed, verified, cleaned and sterilized using the same validated methods as the reference devices cleared in K163579 and K180962.
{7}------------------------------------------------
| Device Comparison Table | | | | |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | KLS Martin IPS Planning<br>System<br>(Subject Device) | KLS Martin IPS Planning<br>System<br>K181241<br>(Primary Predicate) | KLS Martin Individual Patient<br>Solutions<br>K163579<br>(Reference Device) | KLS Martin Individual Patient<br>Solutions<br>K180962<br>(Reference Device) |
| Indications for Use | The KLS Martin Individual<br>Patient Solutions (IPS) Planning<br>System is intended for use as a<br>software system and image<br>segmentation system for the<br>transfer of imaging information<br>from a computerized<br>tomography (CT) medical scan.<br>The input data file is processed<br>by the IPS Planning System and<br>the result is an output data file<br>that may then be provided as<br>digital models or used as input<br>to a rapid prototyping portion of<br>the system that produces<br>physical outputs including<br>anatomical models, guides, and<br>case reports for use in the<br>marking of cranial bone in<br>cranial surgery. The IPS<br>Planning System is also intended<br>as a pre-operative software tool<br>for simulating / evaluating<br>surgical treatment options. | The KLS Martin Individual<br>Patient Solutions (IPS) Planning<br>System is intended for use as a<br>software system and image<br>segmentation system for the<br>transfer of imaging information<br>from a medical scanner such as a<br>CT based system. The input data<br>file is processed by the IPS<br>Planning System and the result is<br>an output data file that may then<br>be provided as digital models or<br>used as input to a rapid<br>prototyping portion of the<br>system that produces physical<br>outputs including anatomical<br>models, guides, splints, and case<br>reports for use in maxillofacial<br>surgery. The IPS Planning<br>System is also intended as a<br>pre-operative software tool for<br>simulating/ evaluating surgical<br>treatment options. | KLS Martin Individual Patient<br>Solutions implant devices are<br>intended for use in the stabilization<br>and fixation of mandibular<br>fractures and mandibular<br>reconstruction. | KLS Martin Individual Patient<br>Solutions implant devices are<br>intended for use in the<br>stabilization and fixation of<br>mandibular fractures and<br>mandibular reconstruction. |
| Device Comparison Table | | | | |
| | KLS Martin IPS Planning System<br>(Subject Device) | KLS Martin IPS Planning System<br>K181241<br>(Primary Predicate) | KLS Martin Individual Patient Solutions<br>K163579<br>(Reference Device) | KLS Martin Individual Patient Solutions<br>K180962<br>(Reference Device) |
| Contraindications | 1. Obvious infections.<br>2. Hypersensitivity to foreign bodies.<br>3. Circulatory problems, systemic diseases, and metabolic disorders.<br>4. Insufficient or inadequate bone tissue.<br>5. Secondary diseases such as degenerative processes that may negatively influence the healing process.<br>6. Interventions carried out in a non-sterile environment (e.g. paranasal sinuses).<br>7. Regions exposed to inappropriate forces or excessive weight loads.<br>8. Patients unwilling or unable to follow instructions during the postoperative phase due to their mental, neurological, or physical condition.<br>9. Bone tumors located in the implant base region.<br>10. Obvious drug or alcohol abuse.<br>11. Significant changes to the patient's anatomy has occurred since the medical scan used for planning purposes was obtained. | 1. Obvious infections.<br>2. Hypersensitivity to foreign bodies.<br>3. Circulatory problems, systemic diseases, and metabolic disorders.<br>4. Insufficient or inadequate bone tissue.<br>5. Secondary diseases such as degenerative processes that may negatively influence the healing process.<br>6. Interventions carried out in a non-sterile environment (e.g. paranasal sinuses).<br>7. Regions exposed to inappropriate forces or excessive weight loads.<br>8. Patients unwilling or unable to follow instructions during the postoperative phase due to their mental, neurological, or physical condition.<br>9. Bone tumors located in the implant base region.<br>10. Obvious drug or alcohol abuse.<br>11. Significant changes to the patient's anatomy has occurred since the medical scan used for planning purposes was obtained. | 1. Obvious infections.<br>2. Hypersensitivity to foreign bodies.<br>3. Suspected sensitivity to the implant material.<br>4. Circulatory problems, systemic diseases and metabolic disorders.<br>5. Insufficient or inadequate bone tissue.<br>6. Secondary diseases such as degenerative processes that may negatively influence the healing process.<br>7. Interventions carried out in a non-sterile environment (e.g. paranasal sinuses).<br>8. Regions exposed to inappropriate forces or excessive weight loads.<br>9. Patients unwilling or unable to follow instructions during the postoperative phase due to their mental, neurological or physical condition.<br>10. Osteoporosis or osteomalacia or other structural bone damage preventing the stable fixation of implant components.<br>11. Bone tumors located in the implant base region.<br>12. Obvious drug or alcohol abuse. | 1. Obvious infections.<br>2. Hypersensitivity to foreign bodies.<br>3. Suspected sensitivity to the implant material.<br>4. Circulatory problems, systemic diseases and metabolic disorders.<br>5. Insufficient or inadequate bone tissue.<br>6. Secondary diseases such as degenerative processes that may negatively influence the healing process.<br>7. Interventions carried out in a non-sterile environment (e.g. paranasal sinuses).<br>8. Regions exposed to inappropriate forces or excessive weight loads.<br>9. Patients unwilling or unable to follow instructions during the postoperative phase due to their mental, neurological or physical condition.<br>10. Osteoporosis or osteomalacia or other structural bone damage preventing the stable fixation of implant components.<br>11. Bone tumors located in the implant base region.<br>12. Obvious drug or alcohol abuse. |
| Device Comparison Table | | | | |
| | KLS Martin IPS Planning<br>System<br>(Subject Device) | KLS Martin IPS Planning<br>System<br>K181241<br>(Primary Predicate) | KLS Martin Individual Patient<br>Solutions<br>K163579<br>(Reference Device) | KLS Martin Individual Patient<br>Solutions<br>K180962<br>(Reference Device) |
| Classification | 21 CFR 882.4310, Class II | 21 CFR 872.4120, Class II<br>21 CFR 892.2050, Class II | 21 CFR 872.4760, Class II | 21 CFR 872.4760, Class II |
| Product Code | PPT | DZJ, LLZ | JEY | JEY |
| Material | Anatomical Models:<br>Epoxy/Resin, Acrylic<br>Marking Guides: Polyamide,<br>Titanium Alloy (Ti-6Al-4V), CP<br>Titanium | Anatomical Models:<br>Epoxy/Resin, Acrylic<br>Cutting/Marking Guides:<br>Polyamide, Titanium Alloy (Ti-<br>6Al-4V), CP Titanium<br>Splints: methacrylate | Anatomical Models: Epoxy/Resin,<br>Acrylic<br>Implants: CP Titanium & Titanium<br>Alloy (Ti-6Al-4V) | Anatomical Models:<br>Epoxy/Resin, Acrylic<br>Cutting Guides: CP Titanium &<br>Titanium Alloy (Ti-6Al-4V)<br>Implants: CP Titanium &<br>Titanium Alloy (Ti-6Al-4V) |
| Manufacturing<br>Method | Epoxy/Resin, Acrylic:<br>Stereolithography (SLA)<br>CP Titanium: Traditional<br>(Subtractive)<br>Ti-6Al-4V: 3D (Additive:<br>Selective Laser Melting)<br>Polyamide: 3D (Additive;<br>Selective Laser Sintering) | Epoxy/Resin, Acrylic:<br>Stereolithography (SLA)<br>CP Titanium: Traditional<br>(Subtractive)<br>Ti-6Al-4V: 3D (Additive:<br>Selective Laser Melting)<br>Polyamide: 3D (Additive;<br>Selective Laser Sintering) | Epoxy/Resin, Acrylic:<br>Stereolithography (SLA)<br>CP Titanium: Traditional<br>(Subtractive)<br>Ti-6Al-4V: 3D (Additive:<br>Selective Laser Melting) | Epoxy/Resin, Acrylic:<br>Stereolithography (SLA)<br>CP Titanium: Traditional<br>(Subtractive)<br>Ti-6Al-4V: 3D (Additive:<br>Selective Laser Melting) |
| Software | Materialise Mimics (K073468)<br>Geomagic® Freeform Plus™ | Materialise Mimics (K073468)<br>Geomagic® Freeform Plus™<br>IPS CaseDesigner (K161634)<br>MathWorks® MATLAB | Materialise Mimics (K073468)<br>Geomagic® Freeform Plus™ | Materialise Mimics (K073468)<br>Geomagic® Freeform Plus™ |
| Sterilization | Non-sterile (Steam) | Non-sterile (Steam) | Non-sterile (Steam) | Non-sterile (Steam) |
| Device Comparison Table | | | | |
| | KLS Martin IPS Planning<br>System<br>(Subject Device)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.