K182874 · Spirosure, Inc. · MXA · Feb 13, 2019 · Clinical Toxicology
Device Facts
Record ID
K182874
Device Name
Fenom Pro Nitric Oxide Test
Applicant
Spirosure, Inc.
Product Code
MXA · Clinical Toxicology
Decision Date
Feb 13, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3080
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Fenom Pro™ Nitric Oxide Test is a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. FeNO is increased in some airway inflammatory processes, such as asthma, and often decreases in response to anti-inflammatory treatment. Measurement of FeNO by Fenom Pro™ is a method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy as an indication of therapeutic effect in patients with elevated FeNO levels. FeNO measurements are to be used as an adjunct to established clinical assessments. Fenom Pro™ is suitable for children, approximately 7-17 years, and adults 18 years and older. Testing using the Fenom Pro™ should only be done in a point-of-care healthcare setting under professional supervision. Fenom Pro™ should not be used in critical care, emergency care or in anesthesiology.
Device Story
Portable, non-invasive system for quantitative FeNO measurement; used in point-of-care settings by healthcare professionals. Input: patient exhaled breath via disposable mouthpiece. Process: breath sample passes through reactive filter (oxidizes interferents like CO, NH4, methanol) and humidity-removing conditioning cartridge; electrochemical sensor measures NO concentration via potential difference. Output: FeNO concentration displayed on GUI. Device provides real-time visual/audible feedback to guide patient breath flow rate per ATS/ERS guidelines. Benefits: monitors therapeutic effect of anti-inflammatory asthma treatment; adjunct to clinical assessment.
Clinical Evidence
Prospective, multi-center, open-label study (N=82) of uncontrolled asthma patients (7-79 years). Subjects measured FeNO at baseline and after 2 weeks of corticosteroid therapy. Results showed significant mean FeNO decrease (-51.4%, p<0.0001) correlating with symptom improvement (ACQ/pACQ -49.5%) and FEV1 increase (+7.7%). Clinical precision evaluated via intra-subject variability across multiple ranges.
Technological Characteristics
Solid-state potentiometric electrochemical sensor; reactive filter material for oxidation of CO, NH4, and methanol. Dimensions/form factor: main unit with touch screen, handpiece, and breath tube. Consumables: mouthpiece and breath conditioning cartridge. Factory calibrated. Power: 100-240V, 50-60Hz. Connectivity: standalone unit.
Indications for Use
Indicated for measuring fractional exhaled nitric oxide (FeNO) in human breath as an adjunct to clinical assessments for monitoring therapeutic effect of anti-inflammatory pharmacological therapy in asthma patients with elevated FeNO. Suitable for children (7-17 years) and adults (18+). Contraindicated for critical care, emergency care, and anesthesiology settings.
Regulatory Classification
Identification
A breath nitric oxide test system is a device intended to measure fractional nitric oxide in human breath. Measurement of changes in fractional nitric oxide concentration in expired breath aids in evaluating an asthma patient's response to anti-inflammatory therapy, as an adjunct to established clinical and laboratory assessments of asthma. A breath nitric oxide test system combines chemiluminescence detection of nitric oxide with a pneumotachograph, display, and dedicated software.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance entitled “Class II Special Controls Guidance Document: Breath Nitric Oxide Test System.” See § 862.1(d) for the availability of this guidance document.
Predicate Devices
NIOX MINO (k072816)
Submission Summary (Full Text)
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k182874
B. Purpose for Submission:
New Device
C. Measurand:
Breath Nitric Oxide
D. Type of Test:
Quantitative, electrochemical sensor
E. Applicant:
Spirosure, Inc.
F. Proprietary and Established Names:
Fenom Pro™ Nitric Oxide Test
G. Regulatory Information:
1. Regulation section:
21 CFR 862.3080, Breath nitric oxide test system
2. Classification:
Class II
3. Product code:
MXA
{1}
4. Panel:
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
Refer to indications for use below.
2. Indication(s) for use:
Fenom Pro™ Nitric Oxide Test is a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. FeNO is increased in some airway inflammatory processes, such as asthma, and often decreases in response to anti-inflammatory treatment. Measurement of FeNO by Fenom Pro™ is a method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy as an indication of therapeutic effect in patients with elevated FeNO levels. FeNO measurements are to be used as an adjunct to established clinical assessments. Fenom Pro™ is suitable for children, approximately 7-17 years, and adults 18 years and older.
Testing using the Fenom Pro™ should only be done in a point-of-care healthcare setting under professional supervision. Fenom Pro™ should not be used in critical care, emergency care or in anesthesiology.
3. Special conditions for use statement(s):
Fenom Pro™ may not be used by children under the age of approximately 7 years, including infants, as measurement requires patient cooperation. Fenom Pro™ may not be used by children under the age of 7 years, or by patients who are unable to understand and execute the instructions given by healthcare providers, as measurement requires patient cooperation.
Fenom Pro™ should not be used in critical care, emergency care, or in anesthesiology. All subjects should refrain from eating or drinking for at least 60 minutes before the FENOM test. Recent intake of nitrate rich food, such as lettuce, spinach, beets, walnuts, peanuts, and animal organs, can lead to increased FeNO levels.
Smoking reduces exhaled NO levels. Fenom Pro™ results obtained from subjects who smoke should only be considered after considering the subject's smoking history and the potential impact on NO levels.
For prescription use only.
{2}
4. Special instrument requirements:
None
I. Device Description:
The Fenom Pro™ Nitric Oxide Test (also referred to in short as Fenom Pro™) is a portable system for the non-invasive, quantitative measurement of the fraction of exhaled nitric oxide (NO) in expired human breath (FeNO).
The Fenom Pro™ Nitric Oxide Test is comprised of four major components (touch screen, handpiece, mouthpiece, breath conditioning cartridge). The main unit contains a touch screen interface for the use as well as houses the nitric oxide sensor and pneumatics needed to sample the patient’s breath. The patient interfaces with Fenom though the mouthpiece which is attached to the handpiece. The handpiece is connected to the main unit via a breath tube. The handpiece contains a breath conditioning cartridge which prepares the breath sample from the patient for proper analysis in the main unit. Both the mouthpiece and the breath conditioning cartridge are consumables.
The mouthpiece is a single patient use, disposable component that contains an anti-bacterial/anti-viral filter. The mouthpiece has an ergonomically designed oval interface to the patient to aid in creating a proper seal during the breath maneuver.
The breath conditioning cartridge is designed to remove humidity from breath samples. It is intended to be used 20 times before being disposed. It contains desiccant beads conditioned to a low dew point to ensure proper sample conditioning over the useful life of the cartridge. The device keeps track of the number of uses on the cartridge as well as independently measures the humidity of the breath sample after the cartridge to verify the gas sample humidity is within the device requirement for proper FeNO analysis.
J. Substantial Equivalence Information:
1. Predicate device name(s):
NIOX MINO
2. Predicate 510(k) number(s):
k072816
{3}
# 3. Comparison with predicate:
| Item | Candidate Device Fenom Pro™ Nitric Oxide Test k182874 | Predicate NIOX MINO k072816 |
| --- | --- | --- |
| Indications for Use | Fenom Pro™ Nitric Oxide Test is a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. FeNO is increased in some airway inflammatory processes, such as asthma, and often decreases in response to anti-inflammatory treatment. Measurement of FeNO by Fenom Pro™ is a method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy as an indication of therapeutic effect in patients with elevated FeNO levels. FeNO measurements are to be used as an adjunct to established clinical assessments. Fenom Pro™ is suitable for children, approximately 7-17 years, and adults 18 years and older. Testing using the Fenom Pro™ should only be done in a point-of-care healthcare setting under professional supervision. Fenom Pro™ should not be used in critical care, emergency care or in anesthesiology. | Same |
| Intended Use Setting | Point-of-care healthcare setting under professional supervision | Same |
| User age | Children 7-17 years, and adults | Same |
| Measurement principle | Electrochemical sensor technology | Same |
| Calibration | Factory Calibrated | Same |
| Measurement range | 10 – 200 ppb | 5 – 300 ppb |
{4}
| Item | Candidate Device
Fenom Pro™ Nitric Oxide Test
k182874 | Predicate
NIOX MINO
k072816 |
| --- | --- | --- |
| Detection Limit | 10 ppb | 5 ppb |
| Analysis Time | ~30 sec. | ~60 sec. |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition.
CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline.
CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures: Approved Guideline – Second Edition.
IEC 60601-1-2:2014 General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests and AIM 7351731 Medical Electrical Equipment and System Electromagnetic Immunity Test for RFID Readers.
L. Test Principle:
Fenom Pro™ Nitric Oxide Test uses solid-state electrochemical sensor technology sensitive to nitric oxide (NO) compounds. The solid state sensor is preceded by a reactive filter material that renders (oxidizes) potentially confounding species such as carbon monoxide (CO), ammonia (NH4), and methanol (CH4O) inactive, or inert, to the NO sensor. Fenom Pro™ provides visual and audible feedback during its use. A user performs a breath maneuver by exhaling into Fenom Pro™ Nitric Oxide Test, and the Fenom Pro™ Nitric Oxide Test graphical user interface (GUI) displays an indication of the user’s breath flow rate, such that the user can modulate their breath flow rate to be within the acceptable limits. An electrochemical potential difference between the electrodes of the sensor develops and is proportional to the amount of NO in the gas sample.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Nitric oxide was mixed in a balance gas to simulate breath samples. Samples NO concentrations were determined using a chemiluminescence device calibrated against a NIST traceable NO tank. Fenom Pro™ Nitric Oxide Test results were collected by multiple operators over 5 operating days, 2 sessions per day, 4 runs per session with two replicates for each concentration, across five different devices using the
{5}
concentrations 10, 25, 75 and 200 ppb (N =400 per concentration). Repeatability is an estimated of variation within one test run in one day. Within-device precision is an estimate of variation between test runs and days. The repeatability and within-device precision over the five days was determined for each concentration as shown in the tables below:
Repeatability
| NO concentration | 10 ppb | 25 ppb | 75 ppb | 200 ppb |
| --- | --- | --- | --- | --- |
| | SD (ppb) | SD (ppb) | CV (%) | CV (%) |
| Device 1 | 1.6 | 1.8 | 3.7 | 3.9 |
| Device 2 | 1.6 | 1.4 | 5.3 | 4.9 |
| Device 3 | 1.2 | 1.5 | 4.0 | 3.0 |
| Device 4 | 1.1 | 1.2 | 3.2 | 3.1 |
| Device 5 | 1.9 | 2.3 | 7.1 | 7.3 |
Within-device precision
| NO concentration | 10 ppb | 25 ppb | 75 ppb | 200 ppb |
| --- | --- | --- | --- | --- |
| | SD (ppb) | SD (ppb) | CV (%) | CV (%) |
| Device 1 | 1.5 | 2.3 | 5.9 | 6.8 |
| Device 2 | 1.5 | 1.7 | 6.2 | 6.2 |
| Device 3 | 1.2 | 1.6 | 4.3 | 3.4 |
| Device 4 | 1.2 | 1.2 | 4.0 | 4.5 |
| Device 5 | 2.3 | 3.3 | 7.5 | 8.5 |
## Clinical Precision
A multi-center, single visit, point-of care, inter-operator variability study was conducted to determine the repeatability of FeNO measured with the Fenom Pro™ Nitric Oxide Test. 3 different operators at each site instructed each subject to perform two valid Fenom Pro™ Nitric Oxide Test measurements. FeNO levels in this study were assessed for a total of 127 subjects at 4 different sites by a total of 12 operators.
The table below shows the intra-subject precision of Fenom Pro™ Nitric Oxide Test measurements within six different measurement ranges.
| Median Concentrations | N* | Within Subject Mean SD | 95% CI for SD | Within Subject Mean CV (%) | 95% CI for CV |
| --- | --- | --- | --- | --- | --- |
| 10 to <20 | 13 | 2.19 | 1.57, 3.61 | 14.23% | 10.21%, 23.49% |
| 20 to <30 | 30 | 2.44 | 1.94, 3.28 | 10.11% | 8.05%, 13.59% |
| 30 to <40 | 29 | 2.71 | 2.15, 3.66 | 8.11% | 6.44%, 10.97% |
| 40 to <50 | 11 | 4.74 | 3.31, 8.31 | 10.98% | 7.67%, 19.27% |
| >=50 | 44 | 5.57 | 4.60, 7.05 | 5.94% | 4.91%, 7.52% |
*Number of subjects, each providing two measurements.
{6}
An additional clinical precision evaluation, conducted with subjects under corticosteroid therapy, is provided in section 3.c.
# b. Linearity/assay reportable range:
Nitric oxide was mixed to create simulated breath gas to obtain 8 NO concentration levels ranging from 5-200 ppb (5, 10, 15, 30, 50, 100, 150, 200 ppb). Samples NO concentrations were determined using a chemiluminescence device calibrated against a NIST traceable NO tank. Five replicates were obtained with the Fenom Pro™ Nitric Oxide Test at each level, using two Fenom Pro™ Nitric Oxide Test devices.
Linearity
| Device | Slope | Intercept | R2 |
| --- | --- | --- | --- |
| Device 1 | 1.03 | 2.32 | 0.999 |
| Device 2 | 1.02 | 0.231 | 0.999 |
Effects of extreme temperature, relative humidity.
The effects of temperature, relative humidity (RH) were assessed using the following conditions: $15^{\circ}\mathrm{C} / 20\%$ RH, $15^{\circ}\mathrm{C} / 80\%$ RH, $30^{\circ}\mathrm{C} / 20\%$ RH, $30^{\circ}\mathrm{C} / 80\%$ RH, and ambient conditions at $22^{\circ}\mathrm{C} / 37\%$ RH. For each sample, nitric oxide was mixed in a balance gas of simulated breath and the concentration of NO was determined using a NO Gas Analyzer. Each concentration sample was measured using 2 Fenom Pro™ devices, with each concentration tested in five replicates per device per test condition. The results support the claimed operating conditions for the Fenom Pro™ Nitric Oxide Test: $15^{\circ}\mathrm{C} - 30^{\circ}\mathrm{C}$ , $20 - 80\%$ RH.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The Fenom Pro™ Nitric Oxide Test is manufacturer calibrated and does not require calibration by the user.
The device is pre-programmed to allow a defined number of tests for the nitric oxide sensor. Stability study results were provided to support the sensor life of 250 tests or 50 hours of use, whichever comes first. Replacement of the sensor is done by the manufacturer.
The breath conditioning cartridge is a replacement part of the device. Stability study results were provided to support the claimed cartridge shelf life of 3 months. Once installed, each cartridge is pre-programmed to allow a defined number of tests (12 hours of use or 20 uses, whichever comes first).
The shelf-life for the mouthpiece is 24 months. The mouthpiece is a single patient use, disposable component that contains an anti-bacterial/anti-viral filter.
{7}
# d. Detection limit:
The limit of detection for the Fenom Pro™ Nitric Oxide Test was determined based on CLSI EP17-A2. Two devices were tested at 0 ppb (60 replicates), 5 ppb (30 replicates) and 10 ppb (30 replicates) over three days. Nitric oxide samples were mixed in a balance gas of simulated breath. The limit of detection was calculated using the parametric option in CLSI EP17-A2, using the following formulas: $\mathrm{LoB} = \mu_{\mathrm{B}} + 1.645\sigma_{\mathrm{B}}$ and $\mathrm{LoD} = \mathrm{LoB} + 1.645\sigma_{\mathrm{S}}$ . The results of the limit of detection analysis support the claimed detection limit of 10 ppb.
# e. Analytical specificity:
# Endogenous substance interference
Interference testing was performed using simulated breath having 15, 30, and 50 ppb NO. Five sensors were used to evaluate endogenous interference. The substances and concentrations tested are summarized in the table below. A substance was defined as non-interfering if the response at the tested concentration was within $\pm 4$ ppb NO.
| Substance | Concentration tested | Concentration expected in exhaled breath | Sensor interference, equivalent to ppb NO |
| --- | --- | --- | --- |
| Acetaldehyde | 150 ppb | 100 ppb | 0.8 ppb |
| Acetone | 5,000 ppb | 10 ppb | 3.7 ppb |
| Acetonitrile | 150 ppb | 100 ppb | 120.8 ppb |
| Ammonia | 1,000 ppb | 0.5 ppb | 1.9 ppb |
| Carbon Dioxide | 8% vol | 8% vol | 2.6 ppb |
| Carbon Monoxide | 50 ppm | 50 ppm | 1.5 ppb |
| Ethanol | 165 ppm | 165 ppm | -0.2 ppb |
| Hydrogen | 50 ppm | 50 ppm | 0.5 ppb |
| Hydrogen Sulfide | 5 ppm | 1 ppm | -2.0 ppb |
| Isoprene | 1.5 ppm | 1 ppm | 2.3 ppb |
| Oxygen | 21% | 16% | -1.4 ppb |
| Hydrogen Peroxide | 500 ppm | 1ppb | 4.0 ppb |
| Nitrogen Dioxide | 13 ppb | 0ppb | 19.2 ppb |
# Exogenous substance interference
A clinical study was performed to investigate the influence of common orally consumed or used exogenous substances on FeNO results measured with Fenom Pro™ Nitric Oxide Test. Each subject performed one baseline measurement before the exogenous substance was introduced and one measurement 60 minutes post exposure or consumption. All subjects refrained from eating or drinking for 60 minutes before the FeNO testing. The results of the study are shown below and support that there is no significant effect of these exogenous substances on the
{8}
measurement of FeNO by Fenom Pro™ Nitric Oxide Test, if users allow one hour to pass before a FeNO measurement. The device labeling recommends that no food or beverage be consumed, and no smoking be done, for at least one hour before taking an FeNO measurement.
| Exogenous Substance | # of Subjects | Mean difference (ppb) | 95% Confidence Intervals (ppb) |
| --- | --- | --- | --- |
| Alcohol Free Mouthwash | 12 | -0.8 | [-3.5, 1.9] |
| Caffeinated Soda | 11 | -3.0 | [-6.2, 0.2] |
| Caffeine Free Soda | 10 | -1.7 | [-6.9, 3.5] |
| Menthol Lozenge | 15 | 0.9 | [-1.0, 2.8] |
| Mouthwash with Alcohol | 10 | -1.2 | [-7.4, 5.0] |
| Non-Menthol Lozenge | 10 | 0.8 | [-1.9, 3.5] |
| Toothpaste | 12 | -1.8 | [-4.0, 0.5] |
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable. A clinical study was conducted to validate the clinical performance of the Fenom Pro™ Nitric Oxide Test (see section 3.c.).
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
A multi-center, open label, non-randomized, prospective, single cohort study was
{9}
conducted to evaluate FeNO measured with the Fenom Pro™ Nitric Oxide Test, spirometry, and asthma symptoms in adult and pediatric subjects with uncontrolled asthma before and after corticosteroid treatment. A total of 82 males and females from 7 to 79 years of age with uncontrolled asthma (not taking asthma controller medications) and elevated FeNO were enrolled at 10 sites and completed the study according to the protocol. The demographic data for the subjects are presented below.
Subject demographics
| | | N (82 total) | % |
| --- | --- | --- | --- |
| Sex | Males | 46 | 56.1 |
| | Females | 36 | 43.9 |
| | | | |
| Age (years) | < 18 (Children) | 37 | 45.1 |
| | ≥ 18 (Adults) | 45 | 54.9 |
| | | | |
| Race | Caucasian | 60 | 73.2 |
| | African American | 13 | 15.9 |
| | Asian | 4 | 4.9 |
| | Two or more races | 1 | 1.2 |
| | Other | 4 | 4.9 |
| | | | |
| Ethnicity | Hispanic or Latino | 19 | 23.2 |
| | Not Hispanic or Latino | 62 | 75.6 |
| | Unknown/Not Provided | 1 | 1.2 |
Clinical Precision:
Each subject provided replicate FeNO measurements at Visit 1 (baseline) and Visit 2 (after approximately two weeks of corticosteroid therapy). The results are summarized in the below tables:
Visit 1
| Median NO concentration | N* | Within Subject Mean SD (ppb) | 95% CI for SD (ppb) | Within Subject Mean CV (%) | 95% CI for CV (%) |
| --- | --- | --- | --- | --- | --- |
| 10 to ≤ 20 | 6 | 1.28 | 0.8, 3.14 | 7.09 | 4.43, 17.40 |
| 20 to ≤ 30 | 5 | 4.97 | 2.98, 14.30 | 20.78 | 12.45, 59.75 |
| 30 to ≤ 40 | 5 | 2.04 | 1.22, 5.86 | 5.76 | 3.45, 16.55 |
| 40 to ≤ 50 | 13 | 3.03 | 2.18, 5.01 | 6.76 | 4.85, 11/16 |
| 50 to <75 | 23 | 3.69 | 2.86, 5.23 | 6.23 | 4.82, 8.82 |
| 75 to <100 | 12 | 6.39 | 4.53, 10.85 | 7.35 | 5.20, 12.47 |
| >=100 | 18 | 4.37 | 3.28, 6.54 | 3.09 | 2.32, 4.63 |
*Number of subjects
{10}
Visit 2
| Median concentration | N* | Within Subject Mean SD (ppb) | 95% CI for SD (ppb) | Within Subject Mean CV (%) | 95% CI for CV (%) |
| --- | --- | --- | --- | --- | --- |
| <10 | 2 | 0.29 | 0.13, 9.38 | 3.48 | 1.55, 111.03 |
| 10 to ≤ 20 | 12 | 1.18 | 0.83, 2.00 | 7.71 | 5.46, 13.08 |
| 20 to ≤ 30 | 20 | 1.88 | 1.43, 2.75 | 7.48 | 5.69, 10.93 |
| 30 to ≤ 40 | 18 | 1.81 | 1.36, 2.71 | 5.23 | 3.93, 7.85 |
| 40 to ≤ 50 | 11 | 2.23 | 1.56, 3.92 | 5.21 | 3.64, 9.15 |
| 50 to <75 | 15 | 2.93 | 2.15, 4.63 | 4.87 | 3.56, 7.68 |
| 75 to <100 | 3 | 5.22 | 2.72, 32.79 | 6.40 | 3.33, 40.25 |
| >=100 | 1 | 2.41 | - | 1.69 | - |
*Number of subjects
# Clinical Accuracy
Using the Fenom Pro™ Nitric Oxide Test, FeNO was measured in all subjects at baseline and after 2 weeks of corticosteroid treatment (follow-up). Testing was performed by 42 trained technicians. At study entry, the adult subjects ( $\geq$ 18 years of age) had FeNO levels $>30$ ppb and children subjects ( $<$ 18 years of age) had FeNO levels $>25$ ppb. The changes in FeNO (Mean $\pm$ SD) in these patients are summarized in the below table.
| | Baseline FeNO (ppb) | Follow-up FeNO (ppb) | Change (ppb) | p-value* |
| --- | --- | --- | --- | --- |
| All subjects (N = 82) | 81.8 ± 49.2 | 39.5 ± 24.5 | -42.3 ± 45.5 | < 0.0001 |
$^{*}\mathrm{p}$ -value for statistical significance of change vs baseline.
FeNO and various asthma outcome measures [Asthma Control Questionnaire (ACQ/pACQ (pediatric ACQ)), and Forced Expiratory Volume at 1 second (FEV1)] are presented in Tables below. The majority of subjects showed improvement in FEV1 $(66.7\%)$ and ACQ/pACQ $(78.6\%)$ and a decrease in FeNO levels at visit 2, as compared to their baseline measurements taken at visit 1.
FeNO / Symptoms / FEV1 Comparison Table (all subjects)
| Measurement Variables | % Change Baseline to Follow-Up | p-value* |
| --- | --- | --- |
| FeNO | -51.4% | < 0.0001 |
| Symptoms (ACQ/pACQ scores) | -49.5% | < 0.0001 |
| Spirometry (FEV1) | +7.7% | < 0.0001 |
FeNO / Symptoms / FEV1 Comparison Table (Adults $> = 18$
| Measurement Variables | % Change Baseline to Follow-Up | p-value* |
| --- | --- | --- |
| FeNO | -50.7% | < 0.0001 |
{11}
FeNO / Symptoms / FEV1 Comparison Table (Children <18)
| Measurement Variables | % Change Baseline to Follow-Up | p-value* |
| --- | --- | --- |
| FeNO | -52.3% | < 0.0001 |
| Symptoms (ACQ/pACQ scores) | -50.3% | < 0.0001 |
| Spirometry (FEV1) | 5.2% | < 0.0001 |
*p-value for statistical significance of change vs baseline.
The relationship between the percent change in FeNO and the percent change in pre- and post-bronchodilator FEV1 from visit 1 to visit 2 was investigated. The magnitude of the FeNO change and degree of improvements in ACQ and FEV1 are different because the scale and precision of these metrics varies. The data presented in the table below indicate that FEV1 and symptom score (ACQ) are different metrics and are not directly correlated with FeNO as determined by Fenom Pro™ Nitric Oxide Test.
| Correlation between change in FENO and change in FEV1 and ACQ in the ITT population (adults and children combined) | | |
| --- | --- | --- |
| | | FeNO |
| Change in FEV_{1} | R-square correlation | -0.14 |
| | P-value | 0.2161 |
| Change in asthma symptom score (ACQ) | R-square correlation | 0.21 |
| | P-value | 0.0637 |
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
The expected values are provided from the literature. In the labeling the sponsor states, "Given that physiological and environmental factors can affect FeNO levels, in clinical practice, 'healthy' FeNO levels need to be established on an individual basis. However, most healthy individuals will have NO levels in the range 5-35 ppb (children slightly lower 5-25 ppb) when measured at 50 mL/s. (ATS/ERS Recommendations for Standardized Procedures for the Online and Offline Measurement of Exhaled Lower Respiratory Nitric Oxide and Nasal Nitric Oxide, 2005. Am J Respir Crit Care Med. 2005;171:912-930)."
{12}
N. Instrument Name:
Fenom Pro™ Nitric Oxide Test
O. System Descriptions:
1. Modes of Operation:
A user will perform a breath maneuver by exhaling into Fenom Pro™ Nitric Oxide Test, and the Fenom Pro™ Nitric Oxide Test graphical user interface (GUI) will display an incentive or indication of the user’s breath flow rate, such that the user can modulate their breath flow rate to be within the acceptable limits recommended by the ATS and ERS guidelines.
The Fenom Pro™ Nitric Oxide Test device evaluates the breath maneuver and determines if the breath maneuver was performed according to the ATS and ERS guidelines with regards to breath flow rate and duration. The breath sample is presented through the reactive filter that transforms interferent species into inert species for the sensor and then fluidly presents the gas sample to the sensor. An electrochemical potential difference between the electrodes of the sensor develops and is proportional to the amount of NO in the gas sample.
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes _______ or No __x__
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes _______ or No __x__
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes __x__ or No _______
3. Specimen Identification:
The patient performs the test in real-time and is identified by patient name.
4. Specimen Sampling and Handling:
The user obtains a breath sample by having the subject exhale into the device.
{13}
5. Calibration:
The manufacturer performs calibration for each Fenom Pro™ Nitric Oxide Test device. No calibration is required by the operator.
6. Quality Control:
Fenom Pro™ Nitric Oxide Test provides internal controls as well as an External Quality Control program for the user to verify the reliability of measurements.
The External Quality Control (QC) program consists of two parts: One positive control from a qualified staff member with a stable FeNO value providing a normal biological FeNO sample and a negative control consisting of a NO free gas sample which is generated by the device from ambient air.
The labeling states that QC Users must be over 18 years of age, non-smoker, no known airway diseases or chronic cold, preferably no allergies or asthma, and stable FeNO values between 10-40 ppb. The device has built in QC rules and the user will see the Passed or Failed instructions after the daily QC is performed.
The process consists of performing four QC measurements, one per day, within 7 days. The negative control result must be <10ppb. The positive control result must be with process control limits for control charts with subgroup size one as defined in "Understanding Statistical Process Control" by Donald L. Wheeler and David S. Chambers. The control limits are established as follows: the most recent result (V0) is compared to a rolling average (AVG) and the prior result (V1). The control limit is a multiple of the rolling absolute range (RAR), which is the average difference of successive results. The control limits are:
|V₀ – AVG| ≤ RAR * 2.66, or 5ppb, whichever is greater
|V₀ – V₁| ≤ RAR * 3.268, or 5ppb, whichever is greater
When the QC has passed both criteria, the device is ready for clinical use. A QC test must be performed each day of use. If the QC tester does not perform a QC test in 30 days, the qualification is suspended and the QC tester needs to re-qualify.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Analytical Accuracy
An accuracy study was performed to compare performance of the Fenom Pro™ Nitric Oxide Test device to a chemiluminescence device calibrated against a NIST traceable NO tank. Four Fenom Pro™ Nitric Oxide Test devices were tested over six concentrations (10 ppb, 25 ppb, 75 ppb, 100 ppb, 200 ppb, and 300 ppb) with ten replicates each for a total of 240 tests. For each concentration, nitric oxide was mixed in a balance gas of simulated breath.
{14}
The results of the study are shown below:
| Mean Bias (%) | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | 10 ppb | 25 ppb | 75 ppb | 100 ppb | 200 ppb | 300 ppb |
| Device 1 | -1.1 ppb
(-11.0 %) | -2.6 ppb
(-10.4%) | -0.5 | 1.1 | -6.4 | -2.5 |
| Device 2 | -0.7 ppb
(-7.0%) | -1.6 ppb
(-6.4%) | -5.4 | -2.7 | -7.7 | -5.0 |
| Device 3 | 3.2 ppb
(32.0%) | 3.0 ppb
(12.0%) | -0.1 | -1.5 | -4.6 | -5.2 |
| Device 4 | -0.9 ppb
(-9.0%) | -1.2 ppb
(-4.8%) | -7.0 | -4.6 | -7.3 | -3.2 |
## Electrical Safety testing
The sponsor provided documentation describing conformity assessment to the relevant electrical safety standard (IEC 60601-1).
## Electromagnetic Compatibility testing
The sponsor provided documentation describing conformity assessment to the relevant EMC standard IEC 60601-1-2:2014
## Biocompatibility testing
Fenom PRO Nitric Oxide Test device will be contacting patient’s gas pathway via air. In addition, the mouthpiece component will have direct surface contact. The contact duration is limited (<24 hours). Based on the intended patient contact types and duration, the sponsor provided demonstrated acceptable cytotoxicity, intracutaneous reactivity, sensitization, and acute systemic toxicity for the device.
## Effects of extreme temperature, relative humidity.
The operating conditions for the device is 15°C-30°C, 20-80% RH. The effects of temperature, relative humidity (RH) were assessed using the following conditions: 15°C/20% RH, 15°C/80% RH, 30°C/20% RH, 30°C/80% RH, and ambient conditions at 22°C/37% RH. For each sample, nitric oxide was mixed in a balance gas of simulated breath and the concentration of NO was determined using a NO Gas Analyzer. Each concentration sample was measured using 2 Fenom Pro™ devices, with each concentration tested in five replicates per device per test condition. The readings from the candidate device meet the acceptance criteria of within +/-5 ppb at 15 ppb, and +/-10% at 75 and 200ppb. The results support the recommended operating conditions of 5°C-30°C (59°F to 86 °F), 20-80% RH.
## Effects of altitude
The sponsor also tested device performance at the following atmospheric pressure range of 106 to 92 kPa, equivalent to -1500 feet to +2500 feet above sea level, and show that the device performance are not affected. The labeling states the operation condition of the device is from Sea level to 6500 ft.
{15}
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.