K182846 · Pentax of America, Inc. · EOB · Jan 9, 2019 · Ear, Nose, Throat
Device Facts
Record ID
K182846
Device Name
PENTAX Medical EPK-3000 Video Imaging System
Applicant
Pentax of America, Inc.
Product Code
EOB · Ear, Nose, Throat
Decision Date
Jan 9, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.4760
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PENTAX Medical EPK-3000 Video Imaging System consists of PENTAX Medical VIDEO PROCESSOR EPK-3000, PENTAX Medical Video Naso-Pharyngo-Laryngoscope VNL-J10 Series, PENTAX Medical Digital Video Capture Modules, PENTAX Medical Laryngeal Strobe, and other ancillary equipment. 1) PENTAX Medical Video Processor EPK-3000 The PENTAX Medical EPK-3000 Video Processor is intended to be used with the PENTAX VNL8-J10, VNL11-J10, and VNL15-J10 endoscopes, PENTAX Medical Laryngeal Strobe, video monitors and other ancillary equipment for ENT endoscopic observation and nasopharyngoscopic (ENT) diagnosis, treatment and video observation with or without stroboscopy. The PENTAX Medical EPK-3000 Video Processor includes PENTAX i-Scan™, a digital, post-processing imaging enhancement technology. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. 2) PENTAX Medical Video Naso-Pharyngo-Laryngoscope VNL-J10 Series The PENTAX Medical Video Naso-Pharyngo-Laryngoscope VNL-J10 Series (VNL8-J10, VNL11-J10 and VNL15-J10) are intended to be used with a PENTAX EPK-3000 Video Processor, PENTAX Medical Laryngeal Strobe, documentation equipment, video monitor, endoscopic device and other ancillary equipment for optical visualization (via a video monitor) of, and, for VNL15-J10 only, therapeutic access to, nasal, pharyngeal, laryngeal and the upper airway anatomy.
Device Story
System comprises video processor, flexible naso-pharyngo-laryngoscopes, laryngeal strobe, and digital video capture modules. Inputs: optical signals from endoscope CCD sensor illuminated by 150W xenon lamp. Processor performs digital post-processing (i-Scan) to enhance mucosal texture/vessel visualization; outputs video to monitor. Strobe allows glottic action observation. Capture modules record/store video/audio data. Used in clinical ENT settings by physicians. System aids diagnosis and treatment; i-Scan serves as optional adjunct to white light endoscopy. Benefits include improved visualization of anatomy and documentation of procedures.
Clinical Evidence
Bench testing only. System interoperability verified through design verification and validation testing per FDA guidance. Electrical safety and EMC confirmed via IEC 60601-1-2 and ANSI/AAMI ES 60601-1. Photobiological safety of strobe lamp evaluated per IEC 62471 (Exempt risk group).
Technological Characteristics
System includes video processor, flexible endoscopes (VNL-J10 series), laryngeal strobe, and digital capture modules. Illumination: 150W xenon lamp. Imaging: CCD sensor. Connectivity: video output to monitor, digital capture interface. i-Scan: digital filter-based image enhancement (3 modes). Standards: IEC 60601-1-2 (EMC), ANSI/AAMI ES 60601-1 (Electrical Safety), IEC 62471 (Photobiological).
Indications for Use
Indicated for ENT endoscopic observation, diagnosis, and treatment of nasal, pharyngeal, laryngeal, and upper airway anatomy in patients requiring visualization or therapeutic access (VNL15-J10).
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Predicate Devices
PENTAX Medical EPK-3000 Video Imaging System (K172156)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 9, 2019
PENTAX of America, Inc. William Goeller Vice President, Quality and Regulatory Affairs 3 Paragon Drive Montvale, NJ 07645-1782
Re: K182846
Trade/Device Name: PENTAX Medical EPK-3000 Video Imaging System Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOB, PEA, EQL, OUG Dated: December 7, 2018 Received: December 10, 2018
Dear William Goeller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device. please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
# Srinivas Nandkumar -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182846
Device Name PENTAX Medical EPK-3000 Video Imaging System
Indications for Use (Describe)
PENTAX Medical EPK-3000 Video Imaging System consists of PENTAX Medical VIDEO PROCESSOR EPK-3000, PENTAX Medical Video Naso-Pharyngo-Laryngoscope VNL-J10 Series, PENTAX Medical Digital Video Capture Modules, PENTAX Medical Laryngeal Strobe, and other ancillary equipment.
1) PENTAX Medical Video Processor EPK-3000
The PENTAX Medical EPK-3000 Video Processor is intended to be used with the PENTAX VNL8-J10, VNL11-J10, and VNL15-J10 endoscopes, PENTAX Medical Laryngeal Strobe, video monitors and other ancillary equipment for ENT endoscopic observation and nasopharyngoscopic (ENT) diagnosis, treatment and video observation with or without stroboscopy.
The PENTAX Medical EPK-3000 Video Processor includes PENTAX i-Scan™, a digital, post-processing imaging enhancement technology. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling.
2) PENTAX Medical Video Naso-Pharyngo-Laryngoscope VNL-J10 Series
The PENTAX Medical Video Naso-Pharyngo-Laryngoscope VNL-J10 Series (VNL8-J10, VNL11-J10 and VNL15-J10) are intended to be used with a PENTAX EPK-3000 Video Processor, PENTAX Medical Laryngeal Strobe, documentation equipment, video monitor, endoscopic device and other ancillary equipment for optical visualization (via a video monitor) of, and, for VNL15-J10 only, therapeutic access to, nasal, pharyngeal, laryngeal and the upper airway anatomy.
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### PENTAX Medical EPK-3000 Video Imaging System Special 510(k)
Image /page/3/Picture/1 description: The image shows the Pentax Medical logo, which is red and grey. The word "PENTAX" is in red, and the word "MEDICAL" is in grey. The number "K182846" is in the upper right corner of the image. The logo is simple and clean.
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 221 CFR 807.92. All data included in this document is accurate and complete to the best of PENTAX Medical's knowledge.
#### Applicant: PENTAX Medical
HOYA Corporation PENTAX Division 3 Paragon Drive Montvale, New Jersey 07645-1782
Contact: William Goeller Vice President, Quality and Regulatory Affairs PENTAX Medical 3 Paragon Drive Montvale, New Jersey 07645-1782 Telephone: (201)571-2300 ext. 2318 FAX: (201)391-4189 Email: william.goeller@pentaxmedical.com
#### Date Prepared: October 5, 2018
Common Name: Endoscope, Accessories, Image Post-Processing For Color Enhancement/ Naso-Pharyngo-Laryngoscope / Laryngostroboscope / Video Capture Modules
### Name of the System: PENTAX Medical EPK-3000 Video Imaging System
Pentax Medical Video Processor EPK -3000 with PENTAX Medical Video Naso-Pharyngoscope VNL-J10 (K172156) with PENTAX Medical Laryngeal Strobe 9400 (listed class 1 device), with PENTAX Medical Digital Video Capture Modules 9310HD (listed class 1 devices)
#### Regulation Number:
21 CFR Part 876.1500 21 CFR Part 874.4760 21 CFR Part 874.4750 21 CFR Part 880.6310
#### Regulation Names:
Endoscope and accessories Nasopharyngoscope (flexible or rigid) and accessories Laryngostroboscope Medical device data system
Regulatory Class: Class II Product Code: PEA, EOB, EQL, OUG
#### Predicate Device: PENTAX Medical EPK-3000 Video Imaging System (K172156)
The proposed PENTAX modification is the combination of interoperable medical devices that will be used as one system. All the devices have been previously cleared and for the Class I devices, listed with FDA.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Pentax Medical logo in red and gray. The word "PENTAX" is in red, and the word "MEDICAL" is in gray. The number K182846 is in the upper right corner of the image. The logo is simple and modern.
The primary components of the new system include the following:
- PENTAX Medical Video Processor EPK-3000
- PENTAX Medical Video Naso-Pharyngo-Laryngoscope VNL-J10 Series (VNL8-J10, VNL11-J10, VNL15-J10)
- PENTAX Medical Laryngeal Strobe 9400
- PENTAX Medical Digital Video Capture Modules (two models are available: 9310HD/9372HD)
For the above listed system components the design was not altered following this combined end-to-end system configuration. The 510(k) is being submitted to account for the new system interoperability through the conduct of appropriate verification and risk management activities; and modifying labeling by specifying the relevant functional, performance, and interface characteristics.
The combined PENTAX Medical EPK-3000 Video Imaging System has the same fundamental technology and operating principles as the predicate device, as well as the same intended use. Both the PENTAX Medical EPK-3000 Video Imaging System and the predicate device are intended for illuminating and viewing the inside of the human body.
The indications for use were slightly modified by specifying "PENTAX Medical Laryngeal Strobe" instead of the "light sources (including strobe)" cited in the predicate's indications for use. "PENTAX Medical Digital Video Capture Modules" were also added to the Indications for Use, to specify "documentation equipment" mentioned in the predicate's indications for use.
#### The main differences between the subject devices and predicate devices are as follows:
- Adding a Laryngeal Stroboscope 9400 that allows observation of the glottic action of the vocal ● folds to the system configuration;
- . Adding a High-Definition Digital Video Capture Module (9310HD and 9372HD) to electronically record, display, transfer, and store digital video data of laryngeal structures, and audio data, for medical and pedagogical applications.
#### Device Description:
PENTAX Medical EPK-3000 Video Imaging System is used for ENT endoscopic observation and nasopharyngo-laryngoscopic (ENT) diagnosis, treatment, and video observation.
The System functions by receiving image signals from the image sensor in an endoscope, which are processed within a video processor and then output to a monitor. Brightness, color balance, and other properties of the displayed images can be adjusted using the system's control panel. The light from a xenon lamp at the distal end of the endoscope illuminates the body cavities of the patient through the endoscope connected to the video processor.
The primary components of the system include the following:
- PENTAX Medical Video Processor EPK-3000
- PENTAX Medical Video Naso-Pharyngo-Laryngoscope VNL-J10 Series (VNL-J10 Series)
- PENTAX Medical Laryngeal Strobe 9400
- PENTAX Medical Digital Video Capture Modules (two models are available: 9310HD/9372HD)
The PENTAX Medical EPK-3000 Video Imaging System is provided with the following accessories:
- Keyboard input device for the video processor
- Foot Switch used to remotely control processor functions
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Pentax Medical logo. The logo has the word "PENTAX" in red, and the word "MEDICAL" in gray. Above the logo is the number "K182846".
• White Balance Adjuster - used as the object of white balance feature
• Condenser Earth Cable - used to reduce high-frequency noise generated during high-frequency electro cautery device use with Pentax endoscopes
• Inlet Seal - prevents suctioned fluid from coming out of the instrument Channel Inlet during the use of suction function. During the reprocessing, it seals the instrument Channel Inlet in order to full the chemical solution inside the channel
• Bite Block - prevents patients from biting the endoscope insertion tube during an endoscopic examination
- Suction Control Valve intended to control suction
Additional accessories for reprocessing are provided with the device. These include Cleaning Adapter, Soaking Cap, Ventilation Cap, Cleaning Brush Kits, Endoscope Cleaning Brush Kits, and replacement O- Rings.
The PENTAX Medical Video Processor EPK-3000 is intended to be used with the PENTAX compatible endoscopes, light sources (including strobe), vide monitors and other ancillary equipment for ENT endoscopic observation and nasopharyngo-laryngoscopic (ENT) diagnosis, treatment, and video observation with or without stroboscopy.
PENTAX Medical Video Processor EPK-3000 functions with the PENTAX i-Scan technology, a digital filter-based image enhancement technique with three modes, i-Scan 1, 2, and 3. PENTAX i-Scan™ is intended to give the user an enhanced view of the texture of the mucosal surface and blood vessels.
White light is captured from a 150-Watt xenon lamp housed in the EPK-3000 Video Processor. All visualization is done with the white light mode first. White light illuminates the tissue and transfers the captured light through the video scope or a charged coupled device (CCD). Note that the white light visualization mode is always used first by the physician. The modification of red, green, blue (RGB) components for each pixel occurs when the i-Scan function is turned on in the EPK-3000 Video Processor. The resulting i- Scan image is then displayed on the observation monitor. The PENTAX Medical Video Processor EPK-3000 is compatible with PENTAX Medical Video Naso-Pharyngo-Laryngoscope VNL-J10 Series. The PENTAX Medical Video Naso-Pharyngoscope VNL-J10 Series are intended to be used with a PENTAX Video Processor (including Light source), documentation equipment, video monitor, and other ancillary equipment for optical visualization (via a video monitor) of, and/or therapeutic access to, nasal, pharyngeal, laryngeal and the upper airway anatomy. Three scopes are included in the VNL-J10 Series of scopes that is the subject of this submission: VNL8-J10, VNL11-J10, and VNL15-J10.
The VNL-J10 Series endoscopes are inserted transorally to visualize subjects under illumination transmitted from a video processor with a solid-state image sensor located at the distal end of the endoscope, and provide images of the target anatomy on the video monitor. The endoscopes are flexible which allows the insertion portion to shape according to the body cavity. They are also composed of an Insertion Portion, Control Body, PVE Connector, and Light Guide Plug. The VNL15-J10 can be used with endoscopic devices, each of which is introduced from the instrument channel inlet of the control body. Additionally, suctioning from the instrument channel at the distal end of the endoscope by pressing the suction control valve is available with this model. The VNL8-10 and the VNL11-J10 do not have an instrument channel.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for Pentax Medical. The word "PENTAX" is in red, and the word "MEDICAL" is in gray. The number "K182846" is in the upper right corner of the image.
#### Intended Use / Indications for use
PENTAX Medical EPK-3000 Video Imaging System consists of PENTAX Medical VIDEO PROCESSOR EPK-3000, PENTAX Medical Video Naso-Pharyngoscope VNL-J10 Series, PENTAX Medical Digital Video Capture Modules, PENTAX Medical Laryngeal Strobe, and other ancillary equipment.
#### 1) PENTAX Medical Video Processor EPK-3000
The PENTAX Medical EPK-3000 Video Processor is intended to be used with the PENTAX VNL8-J10, VNL11-J10, and VNL15-J10 endoscopes, PENTAX Medical Laryngeal Strobe, video monitors and other ancillary equipment for ENT endoscopic observation and nasopharyngo-laryngoscopic (ENT) diagnosis, treatment and video observation with or without stroboscopy.
The PENTAX Medical EPK-3000 Video Processor includes PENTAX i-Scan™, a digital, post-processing imaging enhancement technology. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling.
#### 2) PENTAX Medical Video Naso-Pharyngo-Laryngoscope VNL-J10 Series
The PENTAX Medical Video Naso-Pharyngo-Laryngoscope VNL-J10 Series (VNL8-J10, VNL11-J10 and VNL15-J10) are intended to be used with a PENTAX EPK-3000 Video Processor, PENTAX Medical Laryngeal Strobe, documentation equipment, video monitor, endoscopic device and other ancillary equipment for optical visualization (via a video monitor) of, and, for VNL15-J10 only, therapeutic access to, nasal, pharyngeal, laryngeal and the upper airway anatomy.
#### Summary of Technology Characteristics:
The PENTAX Medical EPK-3000 Video Imaging System is functionally equivalent to its predicate device, the PENTAX Medical EPK-3000 Video Imaging System cleared by FDA in 2018 (K172156). The only difference between the two devices is that two additional devices have been added to the subject system configuration, namely:
- . Laryngeal Stroboscope 9400 for observation of the glottic action of the vocal folds
- High-Definition Digital Video Capture Module (9310HD and 9372HD) for electronic recording, displaying, transferring, and storing of digital video and audio data of laryngeal or pharyngeal structures
The modified system configuration has been evaluated through performance testing and raise no issue of safety and effectiveness of the device as these differences have no effect on the performance, function or general intended use of the device.
#### Non-Clinical Performance Data
The PENTAX Medical EPK-3000 Video Imaging System has been successfully tested for its functions, performance and safety as per FDA recognized consensus standards. The following performance data are provided in support of the substantial equivalence determination.
#### System Interoperability Testing
Interoperability verification and validation testing of the end-to-end system has been informed by the FDA Guidance document "Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices" and has been verified and validated through Design Verification testing.
The acceptance criteria have been satisfied for all tests.
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the Pentax Medical logo in red and gray. The logo is stacked, with the word "PENTAX" in red on top and the word "MEDICAL" in gray on the bottom. The number K182846 is in the upper right corner of the image.
Cybersecurity risks have been assessed and mitigated according to the FDA Guidances for Industry and Staff "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" issued October 2, 2014, and "Postmarket Management of Cybersecurity in Medical Devices." issued December 28, 2016.
#### EMC and Electrical Safety
The acceptable level of electromagnetic compatibility (EMC) and electrical safety (ES) for the PENTAX Medical EPK-3000 Video Imaging System were confirmed by the following standards: IEC 60601-1-2:2007; ANSI/AAMI ES 60601-1:2005/(R)2012 and A1:2012, and A2:2010/(R)2012.
#### Photobiological Safety of Lamps
The photobiological safety of the Pentax Medical Laryngeal Strobe 9400 lamp systems was evaluated in accordance with IEC62471, Photobiological Safety of Lamps and Lamp Systems. It was determined that the lamp classification group was an Exempt risk group.
#### Substantial Equivalence Discussion:
The subject and predicate devices have the same intended use fundamental technology and there are no significant differences that raise different questions of safety and effectiveness. The differences in terms of design features, performance characteristics and constituent materials are not substantive.
#### Conclusion:
Accordingly, PENTAX Medical believes the PENTAX Medical EPK-3000 Video Imaging System is substantially equivalent to the identified predicate, the PENTAX Medical EPK-3000 Video Imaging System, cleared by FDA in 2018.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.