The Cardinal Health™ Poly-Coated Open-Back Protective Gown is intended to protect health care personnel and patients from the transfer of microorganisms, body fluids, and particulate material. Not intended for use in the operating room. The proposed Cardinal Health™ Poly-Coated Open-Back Protective Gown meets the barrier protection requirements of AAMI Level 4 per ANSI/AAMI PB70:2012. Liquid Barrier Performance and Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities, but has an open back which is non-protective. The Cardinal Health™ Poly-Coated Open-Back Protective Gowns is a single use, disposable medical device provided non-sterile.
Device Story
Protective gown; laminate construction (spunbond polypropylene coated with polyethylene); provides barrier protection for body and sleeves; features thumbhook or knit cuffs, neck removal feature, and integrated belt ties. Used in healthcare settings to protect personnel and patients from microorganisms, body fluids, and particulates; not for operating room use. Open-back design is non-protective. Single-use, disposable, non-sterile. Healthcare providers wear gown to reduce exposure risk; barrier performance meets AAMI Level 4 requirements for critical zones.
Clinical Evidence
Bench testing only. Performance evaluated per ANSI/AAMI PB70:2012, ASTM F1671 (viral barrier), ASTM D3776 (basis weight), ASTM D5034 (tensile strength), ASTM D5587 (tear strength), and CPSC 1610 (flammability). Biocompatibility testing performed per ISO 10993-1, -5, and -10. Results confirm non-cytotoxic, non-irritating, and non-sensitizing properties. Barrier performance met AAMI Level 4 requirements.
Technological Characteristics
Laminate material (spunbond polypropylene coated with polyethylene). Design includes thumbhook or knit cuffs, neck removal feature, and integrated belt ties. Barrier performance: AAMI Level 4 (ANSI/AAMI PB70:2012). Biocompatibility: ISO 10993 compliant. Non-sterile, single-use, disposable. Sizes: Universal and XX-Large.
Indications for Use
Indicated for health care personnel and patients to protect against transfer of microorganisms, body fluids, and particulate material. Not for use in the operating room. Single-use, disposable, non-sterile.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 8, 2021
Cardinal Health 200, LLC Caroline Miceli Manager, Regulatory Affairs 3651 Birchwood Drive Waukegan, Illinois 60085
Re: K182830
Trade/Device Name: Cardinal HealthTM Poly-Coated Open-Back Protective Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: QPC
Dear Caroline Miceli:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated April 26, 2019. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Dr. Ryan Ortega, Acting Assistant Director, Office of Surgical and Infection Control Devices, at 240-402-2303 or Ryan.ortega@fda.hhs.gov.
Sincerely,
## Bifeng Qian -S
For Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
April 26, 2019
Cardinal Health 200, LLC Caroline Miceli Manager, Regulatory Affairs 3651 Birchwood Drive Waukegan, Illinois 60085
Re: K182830
Trade/Device Name: Cardinal Health™ Poly-Coated Open-Back Protective Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYC Dated: March 29, 2019 Received: April 3, 2019
## Dear Caroline Miceli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{3}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Clarence W. Murray lii III -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
#### K182830
Device Name
Cardinal HealthTM Poly-Coated Open-Back Protective Gown
#### Indications for Use (Describe)
The Cardinal Health™ Poly-Coated Open-Back Protective Gown is intended to protect health care personnel and patients from the transfer of microorganisms, body fluids, and particulate material. Not intended for use in the operating room.
The proposed Cardinal Health™ Poly-Coated Open-Back Protective Gown meets the barrier protection requirements of AAMI Level 4 per ANSI/AAMI PB70:2012. Liquid Barrier Performance and Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities, but has an open back which is non-protective.
The Cardinal Health™ Poly-Coated Open-Back Protective Gowns is a single use, disposable medical device provided non-sterile.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### * DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
PSC Publishing Services (301) 443-6740 EF
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image features the logo for Cardinal Health. The logo consists of a stylized red bird-like design above the company name. The text "CardinalHealth" is written in a bold, sans-serif font, with the first letter capitalized and the rest in lowercase.
3651 Birchwood Drive Waukegan, IL 60085 www.cardinalhealth.com
## 510(k) SUMMARY Cardinal Health™ Poly-Coated Open-Back Protective Gown
| Manufacturer: | Cardinal Health 200, LLC<br>3651 Birchwood Drive<br>Waukegan, IL 60085 |
|-----------------------------|---------------------------------------------------------------------------------|
| Regulatory Affairs Contact: | Caroline Miceli<br>300 South Riverside Plaza<br>Suite 2010<br>Chicago, IL 60606 |
| Telephone Number: | (312) 270-2013 |
| Fax Number: | (312) 775-9324 |
| Date Summary Prepared: | April 25, 2019 |
| Trade Name: | Cardinal Health™ Poly-Coated Open-Back Protective Gown |
| Regulation Number: | 21 CFR §878.4040 |
| Device Class: | Class II |
| Regulation Name: | Surgical Apparel |
| Common/Classification Name: | Surgical Isolation Gown |
| Product Code: | FYC |
| Predicate Device: | Kool-Gard Backless Protective Gown (K952115) |
{6}------------------------------------------------
## Description
The Cardinal Health™ Poly-Coated Open-Back Protective Gown is a Class II medical device under the FDA product code of FYC, General & Plastic Surgery Panel, and Regulation 21 CFR 878.4040. The device description of the Cardinal Health™ Poly-Coated Open-Back Protective Gown is in accordance with the Guidance on Premarket Notification [510(k)] Submissions for Surgical Gowns and Surgical Drapes, issued on August 1, 1993 and Guidance for Industry and FDA Staff: Premarket Notification Requirements Concerning Gowns Intended for Use in Healthcare Settings, issued on December 9, 2015.
The Cardinal Health™ Poly-Coated Open-Back Protective Gown consists of a laminate (blue spunbond polypropylene coated with polyethylene) throughout the body of the gown (including sleeves). The sleeves have a thumbhook cuff formed into the sleeve or have a knit cuff sewn onto sleeve for keeping the sleeves in place on the wearer. The collar has a neck removal feature and the belt ties are integrated into the body. These gowns are non-sterile, single use, disposable gowns that are packaged in pastic bags and/or dispenser boxes. These gowns are offered in blue and come in both Universal and XX-Large sizes.
## Indications for Use
The Cardinal Health™ Poly-Coated Open-Back Protective Gown is intended to protect health care personnel and patients from the transfer of microorganisms, body fluids and particulate material. Not intended for use in the operating room. The proposed (subject device) Cardinal Health™ Poly-Coated Open-Back Protective Gown meets the barrier protection requirements of AAMI Level 4 per ANSI/AAMI PB70:2012, Liquid Barrier Performance and Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities, but has an open back which is non-protective. The Cardinal Health™ Poly-Coated Open-Back Protective Gown is a single use, disposable medical device provided non-sterile.
{7}------------------------------------------------
## Technological Characteristics
Shown below is the technological characteristics comparison of the subject device and the
predicate device.
| Element of<br>Comparison | Predicate Device:<br>K952115 Kool-Gard Backless Protective<br>Gown | Subject Device:<br>Cardinal Health™ Poly-Coated Open-<br>Back Protective Gown | Comparison |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications for Use | This device is intended for non-sterile use only, and is not intended for use in the Operating<br>Room. It is intended for use in areas where there<br>is the potential for medium to heavy fluid contact<br>in the front. | The proposed Cardinal Health™ Poly-Coated<br>Open-Back Protective Gown is intended to<br>protect health care personnel and patients from<br>the transfer of microorganisms, body fluids<br>and particulate material. Not intended for use<br>in the operating room<br>The proposed Cardinal Health™ Poly-Coated<br>Open-Back Protective Gown meets the barrier<br>protection requirements of AAMI Level 4 per<br>ANSI/AAMI PB70:2012, Liquid Barrier<br>Performance and Classification of Protective<br>Apparel and Drapes Intended for Use in<br>Health Care Facilities, but has an open back<br>which is non-protective.<br>The Cardinal Health™ Poly-Coated Open-<br>Back Protective Gown is a single use,<br>disposable medical device provided non-<br>sterile. | Similar |
| Barrier Protection<br>Level | ANSI/AAMI PB70:2012, Liquid Barrier<br>Performance and Classification of Protective<br>Apparel and Drapes Intended for Use in Health<br>Care Facilities was not in effect at this time. | AAMI Level 4 per ANSI/AAMI PB70:2012,<br>Liquid Barrier Performance and<br>Classification of Protective Apparel and<br>Drapes Intended for Use in Health Care<br>Facilities, but has an open back which is non-<br>protective. | Different |
| Directions for Use | None | None | Same |
| Material<br>Composition | Nonwoven polypropylene laminated with either<br>polypropylene or polyethylene film | Laminate (spunbond polypropylene coated<br>with polyethylene) | Similar |
| Design Features | Elastic at the cuff for keeping the sleeves in<br>place on the wearer<br>Belt Ties | Thumbhook cuff formed into sleeve or knit<br>cuff sewn onto sleeve for keeping the sleeves<br>in place on the wearer<br>Belt Ties integrated into body<br>Neck removal feature | Same |
## Technological Characteristics Comparison Table
{8}------------------------------------------------
| Sterility | Non-sterile | Non-sterile | Same |
|----------------|-----------------------------------|--------------------------------------|---------|
| Use | Single Use; Disposable | Single Use; Disposable | Same |
| Color and Size | Blue or Yellow<br>Size: Universal | Blue<br>Size: Universal and XX-Large | Similar |
## Non-Clinical Performance Testing
The proposed device was tested in accordance with Guidance on Premarket Notification [510(k)] Submission for Surgical Gowns and Surgical Drapes, August 1993, Guidance for Industry and FDA Staff - Premarket Notification Requirements Concerning Gowns Intended for Use in Healthcare Settings issued by the FDA on December 9, 2015 and ANSI/AAMI PB70:2003, Liquid Barrier Performance and Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities.
The table below contains a summary of the non-clinical performance testing of the device and it is compared to the predicate device. A complete listing of the name of each standard or test method used in this submission is also included after the table.
| Summary of Non-Clinical Performance Testing | | | |
|---------------------------------------------|--------------------------------------------------------------|-----------------------------------------------------------------------------|---------------------------------------------------------------------|
| Element of Comparison | Predicate Device:<br>K952115 | Subject Device: | |
| | Kool-Gard Backless<br>Protective Gown | Cardinal HealthTM Poly-Coated Open-Back Protective Gown | |
| | | Finished Good Test Results:<br>Mean (min/max)<br>Sleeve and Body material | Specification |
| Basis Weight | | | |
| IST 130.0-70 (R82) ** | 1.5 - 2.0 oz/yd2<br>(50.9 - 67.8 gsm) | N/A: ASTM D3776/D3776M-09<br>(2013) used in place of IST 130.0-<br>70 (R82) | N/A |
| ASTM D3776/D3776M-17 | N/A: Current method<br>not available at time of<br>predicate | 43.3 (39.9 / 46.1) gsm | 39.0 gsm min (average)<br>42.0 gsm target<br>45.0 gsm max (average) |
| Tensile Strength, CD * | | | |
| ASTM D1682-64 ** | > 5 lb/in | N/A: ASTM D5034-09 (2013) used<br>in place of ASTM D1682-64 | N/A |
{9}------------------------------------------------
| Summary of Non-Clinical Performance Testing | | | |
|---------------------------------------------------------|--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| Element of Comparison | Predicate Device:<br>K952115 | Subject Device: | Specification |
| | Kool-Gard Backless<br>Protective Gown | Cardinal Health™ Poly-Coated Open-Back Protective Gown<br><b>Finished Good Test Results:</b><br><b>Mean (min/max)</b><br>Sleeve and Body material | |
| ASTM D5034-17 | N/A: Current method<br>not available at time of<br>predicate | 14.2 (12.6 / 16.1) lb | 5.8 lb min (individual)<br>3.5 lb min (average) |
| Tear Strength, MD * | | | |
| ASTM D1117-77 ** | > 6 lb | N/A: ASTM D5587-15 used in<br>place of ASTM D1117-77 | N/A |
| ASTM D5587-15 | N/A: Current method<br>not available at time of<br>predicate | 6.0 (4.4 / 7.5) lb | 6.7 lb min (individual)<br>2.8 lb min (average) |
| Flammability | | | |
| CPSC, Part 1610 | Not reported in 510k<br>Summary | Class 1 | Class 1 |
| Hydrostatic head | | | |
| AATCC 127-1985 **<br>(w/synthetic blood) | > 45 inH20 material<br>8.0 - 10.11 inH20<br>(seam) | N/A: AATCC 127:2014 used in<br>place of AATCC 127:1985 | N/A |
| AATCC 127:2017 | N/A: Current method<br>not available at time of<br>predicate | >130 (130 / >130) cmH20<br>>51 (51 />51) inH20 | 140 cmH20 target<br>100 cmH20 min<br>(individual) |
| Viral barrier (resistance to<br>bacteriophage Phi-X174) | | | |
| ASTM F1671-13 | N/A: Current method<br>not available at time of<br>predicate | Pass | Pass |
Note - Individual Maximum (Ind. Max.): specification allows only one value at this level. Individual minimum (Ind. Min.): specification allows only one value at this level. The mean performance of the was compared to the mean specification values.
MD Grab tensile not specified, CD Grab tensile is limiting specification value; CD Trap Tear is not specified; MD Trap Tear is 米 the limiting specification value as it is the weaker direction.
** Predicate test methods for basis weight and grab tensile strength were accepted test methods at the time; subject device was tested on current standards.
{10}------------------------------------------------
| Summary of Non-Clinical Performance Testing | | |
|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Element of Comparison | Predicate Device:<br>Kool-Gard Backless Protective<br>Gown (K952115) | Subject Device:<br>Cardinal Health™ Poly-Coated Open-<br>Back Protective Gown |
| Liquid Barrier<br>Performance<br>Classification Properties | PB70 Performance Standard<br>was not available at the time of<br>the predicate submission.<br>Predicate device tested to<br>AATCC 127-1985 Hydrostatic<br>Pressure (with synthetic blood)<br>to establish protection level for<br>medium to heavy fluid contact<br>in the front. | All areas tested meet Level 4<br>performance requirements (ASTM<br>F1671), which is the highest<br>standardized level of barrier<br>performance<br>The Cardinal Health™ Poly-Coated<br>Open-Back Protective Gown meets the<br>barrier protection requirements of AAMI<br>Level 4 per ANSI/AAMI PB70:2012,<br>Liquid Barrier Performance and<br>Classification of Protective Apparel and<br>Drapes Intended for Use in Health Care<br>Facilities, but has an open back which is<br>non-protective. Testing was performed<br>in accordance with ASTM F1671 using<br>3 lots and 32 samples per lot in each<br>critical zone area. The critical zone areas<br>tested were the body or sleeve (same<br>material) and the heat sealed seam(s). |
| Biocompatibility | Certified, the materials used in<br>the device are the same as other<br>legally marketed devices for<br>same intended use. | Under the conditions of each study, the<br>Cardinal Health™ Poly-Coated Open-<br>Back Protective Gown is non-cytotoxic,<br>non-irritating, and is non-sensitizing and<br>has met the requirements per ISO-<br>10993-1. |
| Sterilization Modality | None, non-sterile | None, non-sterile |
{11}------------------------------------------------
## Test Standards
| Standard Number | Standards Organization | Standards Title | Version | Date |
|------------------------------|---------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|------------|
| AAMI PB70:2012 | Association for the<br>Advancement of<br>Medical<br>Instrumentation | Liquid Barrier Performance<br>Classification of Protective Apparel and<br>Drapes Intended for Use in Health Care<br>Facilities | 2012 | 8/1/2012 |
| ASTM-<br>F1671/F1671M-13 | ASTM<br>International | Standard Test Method for Resistance of<br>Materials Used in Protective Clothing to<br>Penetration by Blood-Borne Pathogens<br>Using Phi-X174 Bacteriophage<br>Penetration as a Test System | 2013 | 2013 |
| ASTM<br>D3776/D3776M-17 | ASTM<br>International | Test Methods for Mass Per Unit Area<br>(Weight) of Woven Fabric | 2017 | 2017 |
| ASTM D5034-17 | ASTM<br>International | Standard Test Method for Breaking<br>Strength and Elongation of Textile<br>Fabrics (Grab Test) | 2017 | 2017 |
| ASTM D5587-2015 | ASTM<br>International | Standard Test Method for Tearing<br>Strength of Fabrics by Trapezoid | 2015 | 2015 |
| AATCC 127-2017 | American<br>Association of<br>Textile Chemist<br>and Colorist | Water Resistance: Hydrostatic Pressure<br>Test | 2017 | 2017 |
| (CPSC), Part 1610 | Consumer<br>Product Safety<br>Commission | Standard for the Flammability of<br>Clothing Textiles | 2008 | 2008 |
| ISO 10993-<br>1:2009/(R)2013 | International<br>Organization of<br>Standardization | Biological Evaluation of Medical<br>Devices- Part 1: Evaluation and testing<br>within a risk management process | 2013 | 10/15/2009 |
| ISO 10993-5:2009 | International<br>Organization of<br>Standardization | Biological Evaluation of Medical<br>Devices- Part 5: Tests for In-Vitro<br>Cytotoxicity | 2009 | 6/1/2009 |
| ISO 10993-10:2010 | International<br>Organization of<br>Standardization | Biological Evaluation of Medical<br>Devices- Part 10: Tests for irritation and<br>delayed-type hypersensitivity | 2010 | 8/1/2010 |
## Conclusion
The Cardinal Health™ Poly-Coated Open-Back Protective Gown is as safe, as effective and performs as well as or better than the legally marketed predicate device, Kool-Gard Backless Protective Gown (K952115).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.