The Codman Electrosurgical Irrigator is indicated to provide flow of irrigation fluid to the tips of bipolar irrigating forceps during electrosurgery.
Device Story
Reusable electrosurgical irrigator; delivers irrigation fluid to bipolar irrigating forceps tips during electrosurgery. System includes electrosurgical generator, interconnecting cable, and forceps with cord/tubing set. Operates via peristaltic pump; controlled by single irrigation knob; features LED display and pause button. Used in clinical settings by surgeons/nurses. Provides controlled fluid flow (slow drip to constant stream) to maintain surgical site visibility and tissue hydration; prevents tissue adhesion to forceps tips. Enhances surgical precision and safety during coagulation/cutting procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including design verification, flow accuracy testing, summative study, and nurse/surgeon design validation studies. Software verification and electrical safety/EMC testing (IEC 60601-1, IEC 60601-1-2) also performed.
Technological Characteristics
Reusable, non-sterile capital equipment. Peristaltic pump mechanism. Powered by 100-240 VAC, 50/60 Hz. Features LED display, single irrigation knob, and pause button. Complies with IEC 60601-1 and IEC 60601-1-2 standards. Software classified as Moderate Level of Concern.
Indications for Use
Indicated for patients undergoing electrosurgery requiring irrigation fluid delivery to the tips of bipolar irrigating forceps.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 27, 2018
Integra LifeSciences Corporation Ms. Jennifer Siu Sr. Regulatory Affairs Specialist 11 Cabot Boulevard Mansfield, Massachusetts 02048
Re: K182801
Trade/Device Name: Codman Electrosurgical Irrigator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: September 28, 2018 Received: October 2, 2018
Dear Ms. Siu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long H. Chen Chen -S Digitally signed by Long H. Date: 2018.11.27 14:59:06 -5 05'00'
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182801
Device Name Codman® Electrosurgical Irrigator
Indications for Use (Describe)
The Codman Electrosurgical Irrigator is indicated to provide flow of irrigation fluid to the tips of bipolar irrigating forceps during electrosurgery.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### Integra LifeSciences Corporation I. Submitter 11 Cabot Boulevard Mansfield, MA 02048
Contact Miss Jennifer Siu, Neuroscience, B.S. Sr. Regulatory Affairs Specialist jennifer.siu@integralife.com (781) 971-5692
Date of Submission: September 28, 2018
| II. Device | Device Proprietary Name | Codman® Electrosurgical Irrigator | |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | Common Name | Irrigator | |
| | Classification Name | Electrosurgical, Cutting & Coagulation & Accessories<br>(21 CFR 878.4400) | |
| | Regulatory Classification | II | |
| | Product Code | GEI | |
| | Review Panel | General & Plastic Surgery | |
| III. Predicate<br>Device | The predicate device for this submission is the Malis™ Irrigation<br>Module 1000 (K033499), which was cleared on November 28, 2003. | | |
| IV. Device<br>Description | The Codman Electrosurgical Irrigator is a reusable, non-sterile electrosurgical<br>device, for use in electrosurgery to deliver irrigation fluid to tissue. The<br>irrigator is part of an electrosurgical system consisting of an electrosurgical<br>generator connected via an interconnecting cable, and a pair of bipolar<br>irrigating forceps connected via a cord and tubing set. | | |
| V. Indications<br>for Use | The Codman Electrosurgical Irrigator is indicated to provide flow of irrigation<br>fluid to the tips of bipolar irrigating forceps during electrosurgery. | | |
| VI.<br>Comparison to<br>Predicate<br>Device | | The Codman Electrosurgical Irrigator is substantially equivalent to the<br>predicate device, the Malis Irrigation Module 1000. The subject device has the<br>same intended use and clinical utility, and similar design and operating<br>principles, accessories, and patient/user interface as the predicate device. The | |
{4}------------------------------------------------
table below provides a comparison between the subject device and the predicate device.
| Comparison of the Predicate and Subject Device | | |
|------------------------------------------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| | Predicate Device:<br>Malis Irrigation Module 1000<br>(K033499) | Subject Device:<br>Codman Electrosurgical<br>Irrigator<br>(This Submission) |
| FDA Product<br>Code | GEI | Same as predicate |
| Classification<br>Classification<br>Name | Class II - 21 CFR 878.4400 | Same as predicate |
| Classification<br>Name | Electrosurgical, Cutting &<br>Coagulation & Accessories | Same as predicate |
| Intended Use | To deliver irrigation fluid to the<br>tips of bipolar irrigating forceps | Same as predicate |
| Reusable | Yes | Same as predicate |
| Flow Settings | 0 - 20 | Same as predicate |
| Flow<br>Performance | Slow drip to constant stream | Same as predicate |
| Method of<br>Operation | Peristaltic pump | Same as predicate |
| Irrigator<br>Features | LED display<br>Single irrigation knob<br>Peristaltic pump<br>Cut/Coag irrigation toggle | LED display<br>Single irrigation knob<br>Peristaltic pump<br>Pause button |
| Material | Non-patient contacting | Same as predicate |
| Supply Voltage | 100-230 VAC | 100-240 VAC |
| Supply<br>Frequency | 50/60 Hz | Same as predicate |
| Supply Current | 0.60 A | 1.0 A |
| AC Powered | Yes | Same as predicate |
| Electrical Safety<br>Testing | IEC 60601-1<br>IEC 60601-1-2 | Same as predicate |
| Software Level of<br>Concern<br>Classification | Minor | Moderate |
| Non-Sterile | Yes | Same as predicate |
| Packaging | Polybags<br>Foam inserts<br>Mailer<br>Carton | Same as predicate |
| Accessories | Power cords<br>Interconnecting cable | Same as predicate |
#### The following performance data has been provided in support of the VII. Performance substantial equivalence determination. All testing was performed on final Data devices unless otherwise specified.
{5}------------------------------------------------
#### Bench Testing
Performance bench testing was conducted in alignment with FDA's guidance Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery issued August 15, 2016 and internal requirements.
| Performance Bench Test Results | |
|---------------------------------|------------|
| Test | Conclusion |
| Design Verification Testing | Pass |
| Flow Accuracy Testing | Pass |
| Summative Study | Pass |
| Nurse Design Validation Study | Pass |
| Surgeon Design Validation Study | Pass |
#### Software Testing
Software testing was conducted in accordance with FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices issued May 11, 2005.
| Software Verification Test Results | |
|------------------------------------------------------|------------|
| Test | Conclusion |
| Software Unit Verification Testing | Pass |
| Software Integration and System Verification Testing | Pass |
#### Electrical Safety/Electromagnetic Compatibility Testing
Electrical safety and EMC testing were conducted in accordance with IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance and IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests.
| Electrical Test Results | |
|-----------------------------------------------|------------|
| Test | Conclusion |
| IEC 60601-1:2005/A1:2012/C1:2012, Edition 3.1 | Pass |
| IEC 60601-1-2:2014, Edition 4.0 | Pass |
#### Sterilization
The Codman Electrosurgical Irrigator is a non-sterile capital equipment and is not intended to be sterilized. Therefore, sterilization is not applicable for this device. A validation was performed to support cleaning instructions in the device labeling (Instructions for Use) in accordance with FDA's Guidance Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling issued March 17, 2015.
#### Shelf-Life Testing
The Codman Electrosurgical Irrigator is a reusable, non-sterile device. Therefore, there is no expiry date and shelf-life is not applicable for this device.
{6}------------------------------------------------
### Biocompatibility Testing
The Codman Electrosurgical Irrigator is a non-patient contacting device. Therefore, biocompatibility is not applicable for this device.
#### Animal Studies
No animal studies were performed as appropriate verification and validation of the subject device was achieved based on the comparison to the predicate device and from the results of the bench, software, and electrical/safety testing.
#### Clinical Studies
No clinical studies were performed as appropriate verification and validation of the subject device was achieved based on the comparison to the predicate device and from the results of the bench, software, and electrical/safety testing.
#### VIII. Conclusion
Based upon the intended use, design, operating principles, patient/user interface, comparison to the predicate device, and testing conducted, it is concluded that the subject device, Codman Electrosurgical Irrigator, is substantially equivalent to the predicate device, Malis Irrigation Module 1000 (K033499), and therefore does not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.