K182743 · New Cos, Inc. · NWA · Oct 23, 2019 · General, Plastic Surgery
Device Facts
Record ID
K182743
Device Name
Patient-Specific Airway Stent
Applicant
New Cos, Inc.
Product Code
NWA · General, Plastic Surgery
Decision Date
Oct 23, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3720
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Patient-Specific Airway Stent is indicated for the treatment of adults ≥22 years of age with symptomatic stenosis of the airway. The silicone stent is intended for implantation into the airway by a physician using the recommended deployment system or an equivalent rigid bronchoscope and stent placement system that accepts the maximum stent diameter being placed. The stent is intended to be in the patient up to 12 months after initial placement.
Device Story
System comprises Web Software and patient-specific silicone Y-stent. Physician uploads patient CT scan to Web Software; software segments airway, calculates centerline, and allows physician to design stent dimensions (diameters 6-18mm, length up to 145mm). After physician approval, stent manufactured via silicone injection into 3D-printed mold; delivered nonsterile to medical center. Physician implants stent using rigid bronchoscope and common applicator system. Stent features anti-migration studs; branched Y-configuration. Device provides patient-specific anatomical fit compared to stock catalog stents. Benefits include improved airway patency for symptomatic stenosis. Software manages patient designs and CT scans; enables 3D prescription creation.
Clinical Evidence
Bench testing only. No clinical data. Testing included sterilization (SAL 10^-6), tear strength, radial compression, fatigue (1-year cycle), deployment testing, and extensive biocompatibility (ISO 10993-1: cytotoxicity, sensitization, irritation, toxicity, pyrogenicity, sub-chronic implantation, genotoxicity). Software verification/validation and human factors/usability testing performed. Migration testing conducted in procedural training model. Dimensional accuracy of airway segmentation and process validation confirmed.
Indicated for adults ≥22 years of age with symptomatic airway stenosis. Intended for implantation by a physician using a rigid bronchoscope/deployment system. Duration of use up to 12 months.
Regulatory Classification
Identification
The tracheal prosthesis is a rigid, flexible, or expandable tubular device made of a silicone, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the trachea or trachealbronchial tree. It may be unbranched or contain one or two branches. The metal tracheal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
Ultraflex Tracheobronchial Stent System – Sterile Uncovered (K121048)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 23, 2019
New COS Inc. Keith Grafmeyer Sr. Project Manager 7100 Euclid Ave, Ste 180 Cleveland, Ohio 44103
Re: K182743
Trade/Device Name: Patient-Specific Airway Stent Regulation Number: 21 CFR 878.3720 Regulation Name: Tracheal Prosthesis Regulatory Class: Class II Product Code: NWA Dated: October 17, 2019 Received: October 18, 2019
Dear Keith Grafmeyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James J. Lee, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182743
Device Name Patient-Specific Airway Stent
#### Indications for Use (Describe)
The Patient-Specific Airway Stent is indicated for the treatment of adults ≥22 years of age with symptomatic stenosis of the airway. The silicone stent is intended for implantation into the airway by a physician using the recommended deployment system or an equivalent rigid bronchoscope and stent placement system that accepts the maximum stent diameter being placed. The stent is intended to be in the patient up to 12 months after initial placement.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------|-----------------------------------------------|
|------------------------------------------------|-----------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
510(k) OWNER / MANUFACTURER: New COS, Inc. 7100 Euclid Ave Ste 180 Cleveland OH, 44103
| CONTACT PERSON: | Keith Grafmeyer<br>Sr. Product Manager<br>New COS, Inc.<br>Tel: (216) 800-5905 x 102<br>Email: kgrafmeyer@customorthopaedics.com |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------|
|-----------------|----------------------------------------------------------------------------------------------------------------------------------|
TRADE NAMES: Patient-Specific Airway Stent
October 23, 2019 PREPARATION DATE:
COMMON NAMES: Prosthesis, Tracheal, Preformed/Molded
| Product: Patient-Specific | Product<br>Code | Regulation and Classification<br>Name | Device<br>Class |
|--------------------------------------|-----------------|----------------------------------------|-----------------|
| Patient-Specific Airway Stent System | NWA | 21 CFR 878.3720<br>Tracheal Prosthesis | II |
#### PREDICATE DEVICE:
| Subject Device: Patient-Specific Airway<br>Stent | Corresponding Predicate Device(s) |
|--------------------------------------------------|-----------------------------------|
| Patient-Specific Airway Stent | K971509: ENDOXANE |
#### REFERENCE DEVICES:
| Device 510(k) Number | Device Name | Reason for Use |
|----------------------|----------------------------------------------------------------|----------------------------------------------------------|
| K073468 | Mimics | Performance<br>specification for airway<br>segmentation |
| K894380 | Dumon Tracheobronchial Stent | Performance<br>specification for<br>compression strength |
| K121048 | Ultraflex Tracheobronchial Stent<br>System – Sterile Uncovered | Performance<br>specification for<br>compression strength |
#### DEVICE DESCRIPTION:
The Patient-Specific Airway Stent is comprised of Web Software and the Patient-Specific Silicone Y-Stent. These two function together as a system to treat symptomatic stenosis of the airway. The Patient-Specific Silicone Stent is designed by a physician using a CT scan as a guide in the Web Software. The Web Software gives the user (physician) the ability to upload a scan, view the airway, and design a stent. The stent, after physician approval, is manufactured via silicone injection into a 3D-printed mold and delivered to the treating physician's medical center nonsterile.
{4}------------------------------------------------
The Patient-Specific Airway Stent includes the following general features:
- Deployed through a common applicator system in conjunction with an . appropriate rigid bronchoscope system
- Made of biocompatible, implant-grade silicone
- Steam sterilizable ●
- Anti-Migration studs on anterior surface ●
- Single-use only .
- Branched (Y-Stent) to prevent migration .
- Designed and approved in a proprietary software by a physician for an individual . patient
The Web Software has the following general features:
- Gives COS technicians the ability to segment the airway and automatically ● calculate a centerline
- Create, edit, and save patient-specific airway stents from segmented airway ●
- . Define exact stent diameters at specific locales
- Measure stent inner and outer diameters at any point along stent design .
- . CT scan as the imaging input modality
- Physician able to manage patient's stent designs and CT scans ●
- Physician able to approve stent design through software (create 3D prescription) .
### INTENDED USE AND INDICATIONS:
The Patient-Specific Airway Stent is indicated for the treatment of adults ≥22 years of age with symptomatic stenosis of the airway. The silicone stent is intended for implantation into the airway by a physician using the recommended deployment system or an equivalent rigid bronchoscope and stent placement system that accepts the maximum stent diameter being placed. The stent is intended to be in the patient up to 12 months after initial placement.
### BASIS OF SUBSTANTIAL EQUIVALENCE:
The Patient-Specific Airway Stent is substantially equivalent to the ENDOXANE stent (K971509) with regards to intended use and indications for use.
The materials used to design the Patient-Specific stent are substantially equivalent to the materials used to manufacture the Endoxane stents and the performance characteristics of the subject device have been tested to show substantial equivalency to the currently marketed predicate device. Biocompatibility testing has confirmed that the Patient-Specific Silicone Stent is acceptable for use as a medical device following the requirements stated in the FDA's Guidance on the Use of ISO 10993-1.
The primary technological difference between the subject and predicate device is the introduction of software to allow a physician to design a stent for a specific patient. The risks introduced by this difference have been mitigated in the software, rendering the subject device's risk profile substantially equivalent to the predicate. Software verification and validation was performed to confirm that the software fully satisfies all inputs.
{5}------------------------------------------------
# TECHNOLOGICAL CHARACTERISTICS SUMMARY:
The Patient-Specific Airway Stent is substantially equivalent to the predicate device in terms of technological characteristics including design, materials, and features.
## Identical Characteristics
- । Branched (Y) stent configurations (no more than 3 branches).
- Method of deployment through a common applicator system in coniunction with । an appropriate rigid bronchoscope system based on the maximum outer diameter being deployed.
- Uniform wall thickness option of 1.0mm.
- Visually smooth interior and exterior surfaces of device. |
- । Removable implant, if necessary (recommended to be replaced after 12 months of implantation).
- -Single-use implant only.
- Uniform material composed of biocompatible, implant-grade silicone proven | through performance and biocompatibility testing according to ISO 10993-1.
- Sterilizable silicone material via common steam cycle from AAMI ST79. -
- -Anti-migration studs on anterior surface of main branch of stent.
### Technological Differences
- (Subject) Stent is designed by a Physician using software for a specific patient vs. (Predicate) Stent is chosen from stock designs out of a catalog for a specific patient.
- (Subiect) Ability to vary the inner diameter of the stent at any location along the । stent from 6mm-18mm in 0.1mm increments (patient-specific design with software) vs. (Predicate) Discretely vary the inner diameter from 8-18mm in 1.0mm increments (stock stent design from catalog)
- (Subject) Ability to vary the length of the stent up to 145mm according to the । patient's anatomy vs. (Predicate) Ability to vary length up to 110mm based on available stent designs in a catalog
- (Subject) 3D-printed molds for silicone injection vs (Predicate) Machined molds । for silicone iniection
- (Subject) Secondary branch angle ≤ 100° vs (Predicate) Secondary branch | angle always 75°
- Subject device has lower flat-plate compression strength than the predicate | device
Any risks related to these technological differences have been mitigated to an acceptable level, rendering the subject device substantially equivalent to the predicate device. The technological characteristics, including all design mitigations, have been verified and validated.
### NON-CLINCICAL PERFORMANCE TESTING
Performance and functional testing were performed to ensure the Patient-Specific Airway Stent functions according to the inputs and demonstrates substantial equivalence to the predicate device.
- Sterilization Testing Confirms that the subject device can be steam sterilized to a SAL of 10-6, using a common cycle in medical centers.
- Tear Strength Testing - Supports the claim that the subject device's material is equivalent to the predicate's material in tear strength.
{6}------------------------------------------------
- Radial Compression Testing Compares the compression strength of the | subject device against the predicate device
- -Fatique Testing - Supports the claim that the subject device does not fatigue when cyclically compressed over the intended life of the implant (1 year)
- । Stent Deployment Testing – Supports the claim that the subject device is able to be deployed by a common applicator and rigid bronchoscope system
- । Biocompatibility Testing - Supports the claim that the subject device is acceptable to implant in terms of biocompatibility for up to one year; identical to the predicate device's permanent implantation duration.
- Cytotoxicity: MEM Elution Test с
- o Sensitization: Guinea Pig Maximization Test
- Irritation: Intracutaneous Reactivity O
- Toxicity: Acute Systemic Toxicity O
- Pyrogenicity: Material-Mediated Pyrogenicity O
- Subacute/Sub-Chronic Toxicity: Sub-Chronic Intramuscular Implantation o in Rabbits for 90days with Histopathology
- O Genotoxicity: Ames Bacterial Reverse Mutation Assay
- Genotoxicity: Mouse Lymphoma Assay o
- Genotoxicity: In Vivo Mouse Micronucleus Mutagenicity Assay O
- Chemical Characterization of Extractables and Leachables o
- Software Verification and Validation Testing Supports the claim that the । software functions as designed, including any design mitigations included to reduce risks associated with the addition of software and patient-specific stent creation into the subject device. Per FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, the software was classified as a Major level of concern.
- । Human Factors and Usability Testing for Web Software – Supports the claim that the software is as safe and as effective as the predicate device when used by its intended users in its intended use-environment.
- -Migration Testing - Bench testing in procedural training model to test migration forces required to dislodge an implanted patient-specific airway stent
- Accelerated Aging Testing Testing to evaluate the shelf-life of the patient- specific airway stents
- । Dimensional Testing of Airway Segmentation - Evaluated the accuracy of segmented airway rendering in proprietary segmentation software
- -Airway Segmentation Process Validation - Validated the process of segmenting an airway from a CT scan in proprietary segmentation software
### CONCLUSION
The Patient-Specific Airway Stent is substantially equivalent to the predicate device in intended use, indications for use, principle of operation, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.