← Product Code [OLO](/productcode/OLO) · K182731

# Corelink Navigation Instruments (K182731)

_Corelink, LLC · OLO · Dec 13, 2019 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K182731

## Device Facts

- **Applicant:** Corelink, LLC
- **Product Code:** [OLO](/productcode/OLO.md)
- **Decision Date:** Dec 13, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

CoreLink Navigation Instruments are intended to be used in the preparation and placement of CoreLink screws (Tiger and Entasis) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

## Device Story

CoreLink Navigation Instruments are non-sterile, reusable manual surgical instruments; fabricated from surgical stainless steel. Designed to interface with Medtronic StealthStation® System; instruments enable surgeon to track position of surgical tools relative to patient anatomy during spinal procedures. Input consists of physical interaction with patient anatomy; system uses CT/MR models, fluoroscopy, or digitized landmarks to provide real-time navigation. Used in OR by surgeons; output is visual guidance on navigation system display. Assists in precise screw placement; potentially reduces surgical error and improves patient outcomes in spinal surgery.

## Clinical Evidence

Bench testing only. Testing performed to ASTM F2554-10 standards to verify functionality, compatibility with the Medtronic StealthStation® System, and substantial equivalence to predicate instruments.

## Technological Characteristics

Manual surgical instruments; material: surgical stainless steel; compatibility: Medtronic StealthStation® System; navigation interface: standard optical/electromagnetic tracking markers; sterilization: reusable (requires reprocessing); standards: ASTM F2554-10.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Medtronic Navigated Reusable Instruments ([K161210](/device/K161210.md))
- Globus Navigation Instruments ([K153203](/device/K153203.md))
- Alphatec Navigation Instruments ([K153603](/device/K153603.md))
- SI-BONE iFuse Navigation ([K172268](/device/K172268.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

December 13, 2019

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. The words "U.S. FOOD & DRUG" are in a larger font than the word "ADMINISTRATION".

Corelink LLC Meredith May Vice President 4628 Northpark Dr. Colorado Springs, Colorado 80918

Re: K182731

Trade/Device Name: Corelink Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: November 12, 2019 Received: November 13, 2019

Dear Meredith May:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K182731

Device Name CoreLink Navigation Instruments

#### Indications for Use (Describe)

CoreLink Navigation Instruments are intended to be used in the preparation and placement of CoreLink screws (Tiger and Entasis) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

Type of Use (Select one or both, as applicable)

| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

CoreLink Navigation Instruments 510K Traditional 510(k) Premarket Notification

## 510(k) Summary: CoreLink's Navigation Instruments

In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular, 21 CFR Part 807.92, the following summary of information is provided:

| Submitter:                        | CoreLink LLC<br>2072 Fenton Logistics Park Blvd.<br>St. Louis, MO 63026                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person                    | Steven D. Mounts<br>(888) 349-7808 x2205                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Date Prepared                     | December 13, 2019                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Trade Name                        | CoreLink Navigation Instruments                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Class                             | Class II                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Classification Name<br>and Number | Orthopedic Stereotaxic Instrument, 21 CFR 882.4560                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Common Name                       | Navigation Instruments                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Product Code / Panel              | OLO, Orthopedic                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Predicate Device                  | Medtronic Navigated Reusable Instruments, K161210 (Primary)<br>Globus Navigation Instruments, K153203<br>Alphatec Navigation Instruments, K153603<br>SI-BONE iFuse Navigation, K172268                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Special Controls                  | N/A                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Description                | CoreLink Navigation Instruments are non-sterile, reusable manual surgical instruments made of surgical stainless steel. These instruments are designed to interface with the already-cleared Medtronic StealthStation® Navigation System to assist surgeons in precisely locating anatomical structures.                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Indications for Use               | CoreLink Navigation Instruments are intended to be used in the preparation and placement of CoreLink screws (Tiger and Entasis) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System, which is indicated for any medical condition in which the use of |                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                   |                                                                                                                                                                                                                                                                                                                                                                                                                             | stereotactic surgery may be appropriate, and where<br>reference to a rigid anatomical structure, such as<br>vertebra, can be identified relative to a CT or MR based<br>model, fluoroscopy images, or digitized landmarks of the<br>anatomy.                                                                                                                                                                                                                     |
| Summary of the                    | Technological<br>Characteristics                                                                                                                                                                                                                                                                                                                                                                                            | The CoreLink Navigation Instruments are designed to<br>interface with the already-cleared Medtronic<br>StealthStation® Navigation System. CoreLink Navigation<br>Instruments have similar designs as the predicate<br>instruments and incorporate all necessary design<br>features to enable navigation and use with the<br>StealthStation®. Like the predicate devices, the subject<br>CoreLink Navigation Instruments are made of surgical<br>stainless steel. |
| Performance Data                  |                                                                                                                                                                                                                                                                                                                                                                                                                             | Testing to ASTM F2554-10 was performed as well as<br>testing to ensure CoreLink's Navigation Instruments are<br>acceptable for their intended use, functionality and<br>compatibility with the Medtronic StealthStation® System<br>and to ensure substantial equivalence to the predicate<br>instruments.                                                                                                                                                        |
| Conclusion                        |                                                                                                                                                                                                                                                                                                                                                                                                                             | Based on the indications for use, technological<br>characteristics, required performance testing and<br>comparison to the predicate device, CoreLink's<br>Navigation Instruments have been shown to be<br>substantially equivalent to legally marketed predicate<br>device and is safe and effective for its intended use                                                                                                                                        |

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/device/K182731](https://fda.innolitics.com/device/K182731)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
