Pigtail Ureteral Catheter Set, Sof-Flex AQ Coated Pigtail Ureteral Catheter
Applicant
Cook Incorporated
Product Code
EYB · Gastroenterology, Urology
Decision Date
Jun 17, 2019
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Indicated for drainage or irrigation in the urinary tract as well as access, advancement or exchange of wire guides.
Device Story
Pigtail Ureteral Catheter Sets and Sof-Flex® AQ® catheters facilitate urinary tract drainage, irrigation, and wire guide manipulation. Sets include polyurethane catheter (5-7 Fr) and PTFE-coated stainless steel wire guide; Sof-Flex® AQ® is a hydrophilic-coated polyurethane catheter (8.2-10 Fr). Distal ends feature pigtail loops with multiple sideports for fluid flow; proximal ends have polyamide female Luer lock adapters. Used by clinicians in clinical settings for urological procedures. Catheters provide access and drainage; wire guides assist in catheter placement or exchange. Radiopaque markers aid visualization. Devices are sterile, single-use, and intended to perform similarly to existing urological catheters, providing safe urinary tract access and fluid management.
Polyurethane catheter tubing; polyamide female Luer lock adapter; stainless steel wire guides with PTFE coating. Distal pigtail loop with sideports. Radiopaque. Sterilized via EtO (SAL 10^-6). Sizes: 5, 6, 7, 8.2, 10 Fr. Lengths: 56 cm or 70 cm.
Indications for Use
Indicated for drainage or irrigation in the urinary tract as well as access, advancement, or exchange of wire guides in patients requiring urological catheterization.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 17, 2019
Cook Incorporated Chelsea Woods Regulatory Affairs Specialist 750 Daniels Way Bloomington, IN 47404
Re: K182695
Trade/Device Name: Pigtail Ureteral Catheter Set Pigtail Ureteral Catheter Sof-Flex® AO® Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: EYB Dated: May 9, 2019 Received: May 10, 2019
#### Dear Chelsea Woods:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit/tray have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Glenn B. Bell, Ph.D. Assistant Division Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182695
Device Name Pigtail Ureteral Catheter Set Pigtail Ureteral Catheter Sof-Flex® AQ®
Indications for Use (Describe)
Indicated for drainage or irrigation in the urinary tract as well as access, advancement, or exchange of wire guides.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K182695 Page 1 of 4 H-002-1
Image /page/3/Picture/1 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in white, bold, sans-serif font. Below "COOK" is the word "MEDICAL" in white, sans-serif font, set against a dark red background.
OOK INCORPORATED 750 DANIELS WAY OOMINGTON, IN 47404 USA 2235 TOLL FREE: 800.457,4500 WW.COOKMEDICAL.COM
### 2.0 510(k) Summary
### Pigtail Ureteral Catheter Set and Pigtail Ureteral Catheter Sof-Flex® AQ® 21 CFR §807.92 Date Prepared: May 8, 2019
#### Submitted By:
Submission: Applicant: Primary Contact: Secondary Contact: Applicant Address:
Primary Contact Phone: Secondary Contact Phone: Contact Fax:
#### Device Information:
Trade Name:
Device Common Name: Regulation Name: Regulation Number: Product Code: Device Class Review Panel:
Traditional 510(k) Premarket Notification Cook Incorporated Chelsea Woods Karthik Pillai Cook Incorporated 750 Daniels Way Bloomington, IN 47404 (812) 335-3575 x104007 (812) 335-3575 x104929 (812) 332-0281
Pigtail Ureteral Catheter Set Pigtail Ureteral Catheter Sof-Flex® AQ® Catheter, Ureteral, Gastro-Urology Urological Catheter and accessories 21 CFR §876.5130 EYB Class II Gastroenterology/Urology
#### Predicate Device:
The predicate device is the AQ® Hydrophilic Urological Catheter manufactured by Cook Urological, Inc, which was cleared under K962004.
#### Predicate Device:
- AQ® Hydrophilic Urological Catheter (K962004) ■
#### Reference Device:
- I Open-End Ureteral Catheter Sof-Flex® (K171662)
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Cook Medical. The logo is a red rectangle with the word "COOK" in white, sans-serif font on the top half. Below the word "COOK" is a smaller red rectangle with the word "MEDICAL" in white, sans-serif font. The logo is simple and clean, and the red and white colors are eye-catching.
COOK INCORPORATED
750 DANIELS WAY
BLOOMINGTON, IN 47404 USA
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
## Device Description:
The Pigtail Ureteral Catheter Sets consist of a catheter and wire guide. The catheters are constructed from polyurethane tubing and have a proximal female Luer lock adapter constructed from polyamide. The catheters are available in 5, 6, and 7 French (Fr) and a length of 70 centimeters (cm). The distal end of the catheter has a single pigtail loop with six sideports. The catheter has ink marks denoting depth and body orientation. The stainless steel wire guides are coated with polytetrafluoroethylene (PTFE) and available in diameters of 0.038 and 0.045 inches (in) with a length of 145 cm.
The Pigtail Ureteral Catheter Sof-Flex® AO® is a catheter with a hydrophilic coating. The catheters are constructed from polyurethane tubing and have a female Luer lock adapter constructed from polyamide. The catheters are available in 8.2 and 10 Fr and have a length of 56 cm. The distal end of the catheter has a single pigtail loop with 5 evenly spaced sideports around the pigtail loop. There are also 6 sideports spaced at 1 cm intervals just proximal to the pigtail loop on the same side of the catheter. For the next 10 cm after the sideports, on the straight portion of the catheter, there are no sideports. After which, there are 9 sideports placed at 1 cm intervals spiraled around the catheter across 8 cm.
The subject devices are provided sterile and are intended for one time use.
## Indications for Use:
Indicated for drainage or irrigation in the urinary tract as well as access, advancement or exchange of wire guides.
## Comparison to Predicate Device:
The subject devices have similar indications for use, methods of operation, and fundamental technological characteristics as the predicate device. Differences between the subject devices and the predicate device include a difference in indications for use, length, and coating. Characteristics of the subject devices that differ from the predicate
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Cook Medical. The word "COOK" is in white, bold, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font on a red background that is shaped like a banner.
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA PHONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
device are supported by testing and analysis. The substantial equivalence comparison of the subject devices to the predicate device is provided in the table below.
| | Predicate Device | Subject Devices | |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|
| | AQ Hydrophilic Urological<br>Catheter | Pigtail Ureteral Catheter<br>Set | Pigtail Ureteral Catheter<br>Sof-Flex® AQ® |
| 510(k)<br>Number | K962004 | Subject of this Submission | Subject of this Submission |
| Manufacturer | Cook Urological, Inc (Merged<br>with Cook Incorporated) | Cook Incorporated | |
| Regulation<br>Number | 876.5130 - Urological Catheter<br>and accessories | Identical to Predicate | |
| Product Code | EYB - Catheter, Ureteral, Gastro-<br>Urology | Identical to Predicate | |
| Device Class | II | Identical to Predicate | |
| Indications<br>for Use | The AQ Hydrophilic Urological<br>Catheters are intended for<br>drainage, irrigation and/or<br>retrograde pyelogram. The<br>hydrophilic coating will allow the<br>catheters to become lubricious<br>which will reduce friction. | Indicated for drainage or irrigation in the urinary tract as<br>well as access, advancement or exchange of wire<br>guides. | |
| Catheter<br>Diameter (Fr) | 3 - 10 | 5, 6, 7 | 8.2, 10 |
| Catheter<br>Length (cm) | 70 | 70 | 56 |
| Catheter<br>Material | Polyurethane, polyethylene, and<br>vinyl | Polyurethane | |
| Distal End | Pigtail Coil, Straight, Flexi-Tip,<br>Whistle Tip, Round Tip,<br>Echotip®, Angled Tip, Spiral Tip,<br>Cone Tip | Pigtail Coil | |
| Sideports | Sideports in coil and shaft | Identical to Predicate | |
| Ink Marks | Orientation Marker | Orientation Marker,<br>Depth Marker | None |
| Coating | Hydrophilic | None | Hydrophilic |
| Radiopaque<br>Catheter | Yes | Identical to Predicate | |
| Proximal<br>Fitting | Luer Lock Connector | Identical to Predicate | |
| Wire Guide<br>Compatibility<br>(in) | 0.025, 0.038, 0.045 | 0.038 and 0.045 | 0.038 |
| Set<br>Components | Catheter and Wire Guide,<br>Catheter Only | Catheter and Wire Guide | Catheter Only |
| Packaging | Tyvek Pouch | Identical to Predicate | |
| Sterilization | EtO | Identical to Predicate | |
| SAL | 10-6 | Identical to Predicate | |
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, uppercase letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, uppercase letters on a red background. The logo is simple and clean, and the red and white colors are eye-catching.
COOK INCORPORATED
750 DANIELS WAY
BLOOMINGTON, IN 47404 USA
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
## Performance Data:
The following testing was performed in order to demonstrate that the subject devices, the Pigtail Ureteral Catheter Sets and Pigtail Ureteral Catheter Sof-Flex® AQ®, met applicable design and performance requirements.
- I Biocompatibility
- Sterilization I
- Radiopacity ■
- l Visual Inspection
- I Dimensional Evaluation
- Wire Guide Compatibility I
- Leakage ■
- Flow Rate I
- Kink Radius l
- l Curl Restoration
- Curl Retention Strength ■
- I Tensile Strength
- Dynamic Friction l
## Conclusion:
The results of these tests support a conclusion that Pigtail Ureteral Catheter Sets and the Pigtail Ureteral Catheter Sof-Flex® AQ® will perform as intended. The subject devices do not raise new questions of safety or effectiveness as compared to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.