K182664 · Agency For Medical Innovations GmbH · KOA · Dec 7, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K182664
Device Name
FiXcision
Applicant
Agency For Medical Innovations GmbH
Product Code
KOA · Gastroenterology, Urology
Decision Date
Dec 7, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4730
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The A.M.I. FiXcision device is intended to be used by qualified physicians to provide access of internal structures and for manipulating soft tissue (cutting) for treatment of simple anal fistulae.
Device Story
FiXcision is a single-use, manual surgical instrument for coloproctology. It consists of a stainless steel probe, guide, base plate, and a circular cutter (stainless steel/PVC-U) with a silicone protective cap. Used by physicians in a clinical setting to treat simple anal fistulae. The probe passes through the fistula tract; the guide compresses tissue; the base plate acts as a stop for the circular cutter. The tubular blade enables circumferential tissue removal. The device provides a mechanical means for tissue excision, aiding the physician in fistula treatment. Benefits include precise tissue removal during surgical intervention.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including design validation (wear of cutting edge, tensile force, probe bending, guide strength), biocompatibility (ISO 10993-5, 10993-10), sterilization, shelf-life validation, and usability testing per IEC 62366-1.
Technological Characteristics
Manual surgical instrument. Materials: stainless steel, PVC-U, silicone. Components: probe, guide, base plate, circular cutter. EO-sterilized. Single-use. Mechanical cutting principle via tubular blade. No energy source. No software.
Indications for Use
Indicated for qualified physicians to treat simple anal fistulae by providing access to internal structures and manipulating (cutting) soft tissue.
Regulatory Classification
Identification
A manual gastroenterology-urology surgical instrument and accessories is a device designed to be used for gastroenterological and urological surgical procedures. The device may be nonpowered, hand-held, or hand-manipulated. Manual gastroenterology-urology surgical instruments include the biopsy forceps cover, biopsy tray without biopsy instruments, line clamp, nonpowered rectal probe, nonelectrical clamp, colostomy spur-crushers, locking device for intestinal clamp, needle holder, gastro-urology hook, gastro-urology probe and director, nonself-retaining retractor, laparotomy rings, nonelectrical snare, rectal specula, bladder neck spreader, self-retaining retractor, and scoop. A manual surgical instrument that is intended specifically for use as an aid in the insertion, placement, fixation, or anchoring of surgical mesh during urogynecologic procedures are classified under § 884.4910 of this chapter.
Predicate Devices
Flexible Cutting Scissors / V. Mueller O.V. (K760715)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 7, 2018
Agency for Medical Innovations GmbH % Allison C. Komiyama, Ph.D., R.A.C. Principal Consultant AcKnowledge Regulatory Strategies, LLC 2251 San Diego Ave., Suite B-257 San Diego, CA 92110
Re: K182664
Trade/Device Name: FiXcision Regulation Number: 21 CFR§ 876.4730 Regulation Name: Manual Gastroenterology-Urology Surgical Instrument and Accessories Regulatory Class: I Product Code: KOA Dated: September 21, 2018 Received: September 25, 2018
Dear Allison C. Komiyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jeffrey W. Cooper -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182664
Device Name FiXcision
Indications for Use (Describe)
The A.M.I. FiXcision device is intended to be used by qualified physicians to provide access of internal structures and for manipulating soft tissue (cutting) for treatment of simple anal fistulae.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
Image /page/3/Picture/1 description: The image shows the logo for A.M.I. The logo is in blue and has three gray dots underneath the letters. The letters are large and bold. There is a registered trademark symbol next to the last I in the logo.
K182664 Page 1 of 3
# 510(k) Summary K182664
#### DATE PREPARED
December 5, 2018
## MANUFACTURER AND 510(k) OWNER
Agency for Medical Innovations GmbH Im Letten 1, 6800 Feldkirch, Vorarlberg, Austria Telephone: +43 5522 90505-0 Fax: +43 5522 90505-4006 Official Contact: Martin Hohlrieder, Ph.D., CTO
#### REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., R.A.C. Michelle Rubin-Onur, Ph.D. AcKnowledge Regulatory Strategies, LLC Telephone: +1 (619) 458-9547 Email: akomiyama@acknowledge-rs.com
### PROPRIETARY NAME OF SUBJECT DEVICE
FiXcision
#### COMMON NAME
Surgical Instruments, Gastroenterology-Urology, Manual (And Accessories)
#### DEVICE CLASSIFICATION
Manual gastroenterology-urology surgical instrument and accessories (21 CFR 876.4730, Product Code KOA, Class I)
#### PREMARKET REVIEW
ODE/DRGUD/GEDB Gastroenterology Devices Branch
#### INDICATIONS FOR USE
The A.M.I. FiXcision device is intended to be used by qualified physicians to provide access of internal structures and for manipulating soft tissue (cutting) for treatment of simple anal fistulae.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for A.M.I. The letters are in blue and are large and bold. There are three gray circles below the letters, one below each letter. There is a registered trademark symbol to the right of the last letter.
DEVICE DESCRIPTION
FiXcision is an EO-sterilized, single use, multi-part, manual cutting instrument for use during anal fistula surgery in coloproctology. FiXcision enables circumferential tissue removal for the treatment of simple anal fistulae. The four components include the following:
- Probe: Component made of stainless steel that passes through the fistula tract.
- Guide: Component made of stainless steel that compresses tissue within the fistula tract.
- Base Plate: Component made of stainless steel that acts as a stop for the Circular ● Cutter.
- Circular Cutter: Component made of stainless steel and polyvinyl chloride (PVC-U). The Circular Cutter cuts tissue from fistula tract. The distal end of the Circular Cutter has a protective cap made of silicone that is removed prior to use.
#### PREDICATE DEVICE IDENTIFICATION
The FiXcision device is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|------------------------------------------------------------------------|-------------------|
| K760715 | Flexible Cutting Scissors / V. Mueller O.V.<br>Baxter Healthcare Corp. | ✓ |
#### SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for FiXcision. The following nonclinical testing has been performed in order to demonstrate equivalence to the predicate device:
- Sterilization and shelf life validation
- . Biocompatibility testing:
- Cytotoxicity per ISO 10993-5 .
- . Sensitization per 10993-10
- Irritation per 10993-10
- . Performance testing (bench):
- . Design validation
- Wear of the cutting edge of the Circular Cutter .
- . Tensile force of the Circular Cutter
- . Bending test for the Probe
- 트 Strength of screwed end of the Guide
- Usability per IEC 62366-1
#### EQUIVALENCE TO PREDICATE DEVICE
The subject device has equivalent intended use and similar technological characteristics as the device cleared in K760715. Unlike the predicate device that had a general intended use for cutting, FiXcision has specific indications to only be used on patients who have simple anal fistulae. Unlike the straight blade(s) of the predicate, FiXcision
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for A.M.I. The letters are in blue and are large and bold. There are three gray circles below the letters, and a registered trademark symbol is next to the last letter.
cuts tissue using a tubular blade. This technological characteristic has undergone testing to ensure the device is substantially equivalent to the predicate.
#### CONCLUSION
Based on the testing performed, including biocompatibility, sterilization and shelf life validation, and non-clinical performance testing (bench), it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The similar intended use, technological characteristics, and performance characteristics for the proposed FiXcision device are assessed to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.