K182653 · Applied Medical Resources Corp. · GEI · Nov 14, 2018 · General, Plastic Surgery
Device Facts
Record ID
K182653
Device Name
Voyant Maryland Fusion Device
Applicant
Applied Medical Resources Corp.
Product Code
GEI · General, Plastic Surgery
Decision Date
Nov 14, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Voyant Maryland Fusion device is a bipolar, electrosurgical device indicated for use with the Voyant Electrosurgical Generator in laparoscopic procedures where the ligation and division of vessels and tissue bundles is desired. The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device. The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.
Device Story
Bipolar electrosurgical handpiece; used with Voyant Electrosurgical Generator; performs vessel/tissue bundle sealing and division. Features pistol-grip handle, trigger for jaw closure, button for RF energy activation, and mechanical user-actuated blade for tissue division. Used in laparoscopic procedures by surgeons. Device captures tissue in jaws; delivers RF energy to seal; user manually actuates blade to divide. Benefits include effective hemostasis and tissue ligation in minimally invasive surgery.
Clinical Evidence
Chronic survival study performed to evaluate long-term seal quality, device performance, and potential adverse effects on adjacent structures. Vessels representative of indications were sealed and evaluated for hemostasis and hematoma. Results met predetermined acceptance criteria.
Technological Characteristics
Single-use bipolar electrosurgical handpiece; RF energy source; pistol-grip form factor with mechanical blade. Complies with IEC 60601 series for EMC and electrical safety. Software-controlled via device-specific script.
Indications for Use
Indicated for patients undergoing laparoscopic procedures requiring ligation and division of vessels up to 7mm and tissue bundles. Contraindicated for tubal coagulation or sterilization procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 14, 2018
Applied Medical Resources Corp. Mr. Andrew Nguyen Regulatory Affairs Specialist 22872 Avenida Empresa Rancho Santa Margarita, California 92688
Re: K182653
Trade/Device Name: Voyant Maryland Fusion Device (Model Numbers: EB015, EB017, EB215, EB216, EB217). Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: October 24, 2018 Received: October 25, 2018
Dear Mr. Nguyen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Long Long H. Chen Digitally sig Date: 2018.11.14 15:50:03 -5 -05'00' for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182653
#### Device Name
Voyant Maryland Fusion Device (Model Numbers: EB015, EB016, EB017, EB215, EB216, EB217)
#### Indications for Use (Describe)
The Voyant Maryland Fusion device is a bipolar, electrosurgical device indicated for use with the Voyant Electrosurgical Generator in laparoscopic procedures where the ligation of vessels and tissue bundles is desired.
The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device.
The device has not been shown to be effective for tubal coagulation for sterilization procedures, and should not be used for these procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| 510(k) Submitter: | Applied Medical Resources Corporation<br>22872 Avenida Empresa<br>Rancho Santa Margarita, CA 92688<br>(949) 713 – 8000 |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Andrew Nguyen<br>Regulatory Affairs Specialist I<br>Applied Medical Resources Corporation<br>Tel: (949) 713-5458<br>Fax: (949) 713-8205<br>Email: andrnguyen@appliedmedical.com |
| Preparation Date: | 24 September 2018 |
| Trade Name: | Voyant® Maryland Fusion Device<br>(Model Numbers: EB015, EB016, EB017, EB215, EB216, EB217) |
| Common Name: | Bipolar Electrosurgical Sealer-Divider |
| Classification: | General and Plastic Surgery Devices – Electrosurgical Cutting and Coagulation<br>Device and Accessories<br>Regulation: 21 CFR 878.4400<br>Device Class: Class II<br>Product Code: GEI |
| Legally Marketed<br>Device: | Voyant® 5mm Fusion Device<br>510(k)#: K172624<br>Produce Code: GEI |
| Device<br>Description: | The Voyant Maryland Fusion device (Model Numbers: EB015, EB016, EB017,<br>EB215, EB216, EB217) is an advanced bipolar instrument that uses RF energy,<br>provided by the Voyant Electrosurgical Generator (K141288, K182244), to seal<br>vessels up to and including 7mm in diameter. The device may also be used to<br>seal tissue bundles that can be captured in the device jaws. The device features<br>a mechanical, user-actuated blade for the division of sealed tissue. |
| Intended Use: | The Voyant Maryland Fusion device is a bipolar, electrosurgical device<br>indicated for use with the Voyant Electrosurgical Generator in laparoscopic<br>procedures where the ligation and division of vessels and tissue bundles is<br>desired.<br>The device can seal and divide vessels up to and including 7mm in diameter<br>and tissue bundles that can be captured in the jaws of the device. |
{4}------------------------------------------------
The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.
#### Summary of Technological Characteristics between Subject and Predicate Devices:
The subject and predicate devices are technologically similar in that both are single-use, electrosurgical hand piece devices designed to deliver RF energy to vessels and tissue captured between its jaws for tissue fusion. Both devices feature a pistol-grip style handle with a trigger for jaw closure and a button energy activation. The devices are equipped with a mechanical, user-actuated blade for the division of sealed tissue.
The subject device features an updated jaw design, software script, and an additional shaft length configuration.
### Discussion of Performance Testing:
The FDA guidance documents Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery (2016), Premarket Notification (510(k)) Submissions for Bipolar Electrosurgical Vessel Sealers for General Surgery (2016), and Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (2005) were considered in evaluating the subject device's electrical, software, and functional capabilities. The tests addressed below were utilized to demonstrate safety and efficacy of the subject device and substantial equivalence to the predicate device.
## EMC, Electrical Safety, and Mechanical Testing
The Voyant Maryland Fusion device (Model Numbers: EB015, EB017, EB215, EB216, EB217) was designed and evaluated in accordance with relevant standards of the IEC 60601 series for electromagnetic compatibility and electrical testing.
Simulated repeated-use and destructive jaw force testing were conducted to verify the mechanical integrity of the subject device's jaws.
#### System Testing
Burst pressure testing was conducted using the subject and predicate devices. Vessels representative of the devices' indications were sealed and the burst pressure for each vessel was recorded. The results of the study demonstrated that the subject has substantially equivalent performance to the predicate.
Thermal spread testing was performed to evaluate the thermal spread damage produced by the subject and predicate devices. Analysis of the measurements demonstrated that the subject has substantially equivalent performance to the predicate.
#### Clinical
A chronic survival study was performed to evaluate long-term seal quality, device performance, and the potential for an adverse effect on adjacent structures. Vessels representative of the devices' indications
{5}------------------------------------------------
were sealed and evaluated for hemostasis and signs of hematoma. The result of the study demonstrated that the subject device met the predetermined acceptance criteria. Software
Unit, integration, and system level software testing were conducted to evaluate the design, implementation, and performance of the device software script.
#### Conclusion:
The subject Voyant Maryland Fusion device (Model Numbers: EB015, EB016, EB017, EB215, EB216, EB217) is substantially equivalent in performance to the predicate Voyant 5mm Fusion device with respect to intended use (i.e. vessel sealing performance and local tissue effects) and does not raise any new issues of safety and efficacy.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.