← Product Code [PNR](/productcode/PNR) · K182649

# Enteral Feeding Syringes with Enfit Connectors (K182649)

_Jiangsu Kangbao Medical Equipment Co., Ltd. · PNR · Nov 9, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K182649

## Device Facts

- **Applicant:** Jiangsu Kangbao Medical Equipment Co., Ltd.
- **Product Code:** [PNR](/productcode/PNR.md)
- **Decision Date:** Nov 9, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

Enteral Feeding Syringe with ENFit Connector is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians in all age groups.

## Device Story

Enteral feeding syringes with ENFit connectors; manual fluid dispenser/measuring device. Components: barrel, plunger, piston. Used in clinical or home settings by clinicians or laypersons to deliver fluids into gastrointestinal system of patients unable to swallow. Sterile, single-use; ethylene oxide sterilized. ENFit connectors ensure compatibility and prevent misconnection. Healthcare providers use device to administer nutrition or medication; output is controlled fluid delivery. Benefits include safe, accurate enteral administration.

## Clinical Evidence

No clinical data. Bench testing only. Compliance verified via ISO 7886-1 (manual syringes), ISO 80369-3 (enteral connectors), ISO 80369-20 (test methods), and ISO 10993-5/10 (biocompatibility). Performance testing included dimension analysis, dosing accuracy, and connector incompatibility.

## Technological Characteristics

Materials: Polypropylene (barrel/plunger), Polyisoprene (piston). Manual operation. Sizes: 1ml to 100ml. ENFit small-bore connectors per ISO 80369-3. Sterilization: Ethylene Oxide (SAL 10^-6). Single-use. No software/electronics.

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- NeoConnector Oral/Enternal Feeding Syringes with ENFit Connector ([K161039](/device/K161039.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 9, 2018

Jiangsu Kangbao Medical Equipment Co., Ltd % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 China

Re: K182649

> Trade/Device Name: Enteral Feeding Syringes with Enfit Connectors Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: PNR Dated: September 19, 2018 Received: September 24, 2018

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Daniel G. Walter Jr -S 2018.11.09 16:55:01 -05'00'

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

K182649

Device Name

Enteral Feeding Syringes with Enfit Connectors

Indications for Use (Describe)

Enteral Feeding Syringe with ENFit Connector is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians in all age groups.

Type of Use (Select one or both, as applicable)

| <span style="font-family: sans-serif;"> <span style="font-size: 12px;"> <span style="font-weight: bold;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)       </span> </span> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <span style="font-size: 12px;"> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> </span>                                   |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: K182649

- 1. Date of Preparation: 11/06/2018
- 2. Sponsor Identification

Jiangsu Kangbao Medical Equipment Co., Ltd 78#, North Suzhong Road, Baoying, Yangzhou, 225800, China.

Establishment Registration Number: 3009742443

Contact Person: Qian Li Position: Marketing Manager Tel: +86-571-87987813 Email: ruffyhz(@aliyun.com

- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Mr. Chengyu Wang (Alternative Contact Person)

Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 001-3609253199 Email: info@mid-link.net

{4}------------------------------------------------

K182649 Page 2 of 4

K182649

#### 4. Identification of Proposed Device

Trade Name: Enteral Feeding Syringes with Enfit Connectors Model: 1ml, 2.5ml, 5ml, 10ml, 20ml, 60ml and 100ml Classification Name: Gastrointestinal tube and accessories Classification: II Product Code: PNR Regulation Number: 21 CFR 876.5980 Review Panel: Gastroenterology/Urology

Indications for Use:

Enteral Feeding Syringe with ENFit Connector is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians in all age groups.

Device Description

The proposed device, Enteral Feeding Syringes with ENfit Connectors, is provided sterile, single use. It has 3 components: (1) barrel, (2) plunger, and (3) piston.

The Enteral Feeding Syringes with ENfit Connectors is available in various size, including 1ml, 2.5ml, 5ml, 10ml, 20ml, 60ml and 100ml. In addition, it contains low dose enteral specific connectors.

The Enteral Feeding Syringes with ENfit Connectors is sterilized by Ethylene Oxide Gas to achieve a SAL of 106 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of 5 years.

- 5. Identification of Predicate Device
Predicate Device

510(k) Number: K161039

Manufacturer: NeoMed, Inc.

Product Name: NeoConnector Oral/Enternal Feeding Syringes with ENFit Connector (12ml to 100ml) NeoConnector Low Dose Tip Oral/Enternal Feeding Syringes with ENFit Connector (0.5ml to 6ml)

- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed 

{5}------------------------------------------------

device complies with the following standards:

- > ISO 7886-1:2017 Sterile hypodermic syringes for single use - Part 1: Syringes for manual use;
- > ISO 80369-3:2016 Small-bore connectors for liquids and gases in healthcare applications — Part 3: Connectors for enteral applications
- > ISO 80369-20:2015 Small-bore connectors for liquids and gases in healthcare applications — Part 20: Common test methods
- > ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
- > ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

In addition, the dimension analysis, dosing accuracy test and connector incompatibility test have been conducted to demonstrate the proposed device complies with these requirement.

- 7. Clinical Test Conclusion
No clinical study is included in this submission.

- 8. Substantially Equivalent (SE) Comparison

| ITEM                | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Predicate Device                                                                                                                                                                                                                                                                                                                                          |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code        | PNR                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | K161039<br>PNR                                                                                                                                                                                                                                                                                                                                            |
| Regulation Number   | 21 CFR 876.5980                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 21 CFR 876.5980                                                                                                                                                                                                                                                                                                                                           |
| Class               | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | II                                                                                                                                                                                                                                                                                                                                                        |
| Indications for Use | Enteral Feeding Syringe with ENFit<br>Connector is indicated for use as a<br>dispenser, a measuring device, and a<br>fluid transfer device. It is used to deliver<br>fluids into the gastrointestinal system of<br>a patient who is physically unable to eat<br>and swallow. The enteral syringes are<br>intended to be used in clinical or home<br>care settings by users ranging from<br>laypersons (under the supervision of a<br>clinician) to clinicians in all age groups. | The device is indicated for use as a<br>dispenser, a measuring device, and a fluid<br>transfer device. It is used to deliver fluids<br>into the body orally or enterally. It is<br>intended to be used in clinical or home care<br>setting by users ranging from clinicians to<br>laypersons (under the supervision of a<br>clinician) in all age groups. |
| Configuration       | Piston;<br>Plunger;<br>Barrel with ENFit connector                                                                                                                                                                                                                                                                                                                                                                                                                               | Piston;<br>Plunger;<br>Barrel with ENFit connector                                                                                                                                                                                                                                                                                                        |

#### Table 2 Substantially Equivalent Comparison

{6}------------------------------------------------

| Single Use                             | Yes                                            | Yes                                      |
|----------------------------------------|------------------------------------------------|------------------------------------------|
| Operation Mode                         | Manual Use                                     | Manual Use                               |
| Label/Labeling                         | Complied with 21 CFR part 801                  | Complied with 21 CFR part 801            |
| Size                                   | 1ml, 2.5ml, 5ml, 10ml, 20ml, 60ml and<br>100ml | 0.5~100ml                                |
| Performance                            | Complies with:                                 | Complies with:                           |
|                                        | ISO 7886-1                                     | ISO 7886-1                               |
|                                        | ISO 80369-3                                    | ISO 80369-3                              |
| Patient-contact component and material |                                                |                                          |
| Barrel                                 | Polypropylene (PP)                             | Unknown                                  |
| Plunger                                | Polypropylene (PP)                             | Unknown                                  |
| Piston                                 | Polyisoprene                                   | Unknown                                  |
| Color additive                         | Pigment                                        | Unknown                                  |
| Biocompatibility                       |                                                |                                          |
| Cytotoxicity                           | No Cytotoxicity                                | No Cytotoxicity                          |
| Skin Irritation                        | No Irritation to Skin                          | No Irritation to Skin                    |
| Sensitization                          | No significant evidence of sensitization       | No significant evidence of sensitization |
| Sterilization                          |                                                |                                          |
| Method                                 | EO Sterilized                                  | EO Sterilized                            |
| SAL                                    | 10⁻⁶                                           | 10⁻⁶                                     |
| Endotoxin Limit                        | 20 EU per device                               | 20 EU per device                         |

The specifications of proposed device and predicate device are different, however, both of them comply with the same recognized standards. Therefore the differences on syringe size will not raise new problems on safety and effectiveness of the proposed device. In addition, the patient-contact components of the proposed device are same as those of the predicate device, although the patient-contacting materials of the predicated device are unknown, the proposed device and the predicated devices are biocompatible and conforms to ISO 10993 series standards. Therefore, the proposed device is claimed to be substantially equivalent to the predicate devices.

- Substantially Equivalent (SE) Conclusion 9.
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K182649](https://fda.innolitics.com/device/K182649)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
