Quantib ND

K182564 · Quantib B.V. · LLZ · Dec 27, 2018 · Radiology

Device Facts

Record IDK182564
Device NameQuantib ND
ApplicantQuantib B.V.
Product CodeLLZ · Radiology
Decision DateDec 27, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Brain tissue segmentationDice index 0.96 ± 0.01; Absolute difference of relative volumes 1.7 ± 1.3 ppDataset A: 33 T1w MR images>1 (manual segmenters)
Cerebrospinal fluid segmentationDice index 0.78 ± 0.05; Absolute difference of relative volumes 1.8 ± 1.3 ppDataset A: 33 T1w MR images>1 (manual segmenters)
Intracranial volume segmentationDice index 0.98 ± 0.01Dataset A: 33 T1w MR images>1 (manual segmenters)
Hippocampus segmentationDice index 0.84 ± 0.03; Absolute difference of relative volumes 0.03 ± 0.02 ppDataset B: 89 T1w MR images>1 (manual segmenters)
Frontal lobe segmentationDice index 0.95 ± 0.01; Absolute difference of relative volumes 1.95 ± 0.90 ppDataset C: 13 T1w MR images>1 (manual segmenters)
Occipital lobe segmentationDice index 0.88 ± 0.03; Absolute difference of relative volumes 0.87 ± 0.75 ppDataset C: 13 T1w MR images>1 (manual segmenters)
Parietal lobe segmentationDice index 0.89 ± 0.03; Absolute difference of relative volumes 2.81 ± 1.13 ppDataset C: 13 T1w MR images>1 (manual segmenters)
Temporal lobe segmentationDice index 0.91 ± 0.01; Absolute difference of relative volumes 1.33 ± 0.76 ppDataset C: 13 T1w MR images>1 (manual segmenters)
Cerebellum segmentationDice index 0.98 ± 0.01; Absolute difference of relative volumes 0.47 ± 0.20 ppDataset C: 13 T1w MR images>1 (manual segmenters)
White matter hyperintensities segmentationDice index 0.61 ± 0.13; Absolute difference of relative volumes 0.2 ± 0.2 pp45 3D T1w images with corresponding T2w FLAIR images>1 (manual segmenters)

Indications for Use

Quantib™ ND is a non-invasive medical imaging processing application that is intended for automatic labeling, visualization, and volumetric quantification of segmentable brain structures from a set of magnetic resonance (MR) images. The Quantib™ ND output consists of segmentations, visualizations and volumetric measurements of brain structures and white matter hyperintensities. Volumetric measurements may be compared to reference centile data. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting. Quantib™ ND is a software application on top of Myrian®.

Device Story

Quantib™ ND is a post-processing software module for Myrian® platform; operates on standard Windows hardware. Inputs: 3D T1-weighted MR images; optional T2-weighted FLAIR MR images for white matter hyperintensity (WMH) segmentation. Operation: Uses atlas-based segmentation to label and quantify brain structures (hippocampus, lobes, cerebellum) and WMH. Baseline module analyzes single time-points; Follow-Up module performs longitudinal analysis of two time-points. Output: Segmentations, visualizations, and volumetric measurements; comparison of volumes to reference centile data. Used by radiologists and trained medical professionals in clinical settings to assist in quantitative reporting. Benefits: Provides objective, automated volumetric data to support clinical assessment of brain structures and longitudinal changes.

Clinical Evidence

Bench testing only. Validation compared automatic segmentations to manual ground truth across three datasets (A, B, C). Dataset A (33 scans): Brain tissue Dice 0.96, CSF Dice 0.78, ICV Dice 0.98. Dataset B (89 scans): Hippocampus Dice 0.84. Dataset C (13 scans): Lobe segmentations (Frontal, Occipital, Parietal, Temporal, Cerebellum) showed Dice indices ranging from 0.87 to 0.98. WMH analysis (45 scans) showed average Dice 0.61. No clinical prospective or retrospective studies were required.

Technological Characteristics

Software-based image processing application; runs on Windows OS as an add-on to Myrian®. Uses atlas-based segmentation algorithms. Complies with ISO 14971 (risk management), IEC 62304 (software lifecycle), and IEC 62366 (usability). Inputs: T1w and T2w FLAIR MR images. Outputs: 3D segmentations, volumetric measurements, and centile comparisons.

Indications for Use

Indicated for trained medical professionals to aid in the evaluation, assessment, and quantitative reporting of MR brain images. Used for automatic labeling, visualization, and volumetric quantification of brain structures and white matter hyperintensities in patients undergoing MR imaging.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in blue, followed by the words "U.S. Food & Drug Administration" in blue. December 27, 2018 Quantib BV Floor Van Leeuwen Ouality & Regulatory Manager Westblaak 106 ROTTERDAM, NL 3012 KM ZUID-HOLLAND Re: K182564 Trade/Device Name: Ouantib ND Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: November 12, 2018 Received: November 14, 2018 Dear Floor Van Leeuwen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael D. O'Hara Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K182564 Device Name Quantib™ ND 1.5 ## Indications for Use (Describe) Quantib™ ND is a non-invasive medical imaging processing application that is intended for automatic labeling, visualization, and volumetric quantification of segmentable brain structures from a set of magnetic resonance (MR) images. The Quantib™ ND output consists of segmentations, visualizations and volumetric measurements of brain structures and white matter hyperintensities. Volumetric measurements may be compared to reference centile data. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting. Quantib™ ND is a software application on top of Myrian®. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Quantib™ ND 1.5 510(k) Summary Image /page/3/Picture/1 description: The image shows the logo for Quantib. The logo consists of a blue atom-like structure on the left and the word "Quantib" in blue on the right. Below the word "Quantib" is the text "K182564". #### SUBMITTER 1 Quantib B.V. Westblaak 106 3012 KM Rotterdam Phone: (+31) 108 41 17 49 Contact Person: Floor van Leeuwen Date Prepared: September 17th, 2018 ## Device 2 Name of Device: Quantib™ ND 1.5 Common or Usual Name: Quantib™ ND Classification Name: System, image processing, radiology (892.2050) Regulatory Class: II Product Code: Picture archiving and communication system (LLZ) ### 3 PREDICATE DEVICE Device: Quantib™ Brain 1.3 Manufacturer: Quantib B.V. 510(k) Reg. No: K173939 This predicate has not been subject to a design-related recall. Classification Name: System, image processing, radiology (892.2050) Requlatory Class: II Product Code: Picture archiving and communication system (LLZ) #### Device description বা Quantib™ ND is a post-processing analysis module for Myrian®, which provides 3D image visualization tools that create and display user-defined views and streamlines interpretation and reporting. It is intended for automatic labeling, visualization, and volumetric quantification of identifiable brain structures from magnetic resonance images (a 3D T1-weighted MR image, with an additional T2-weighted FLAIR MR image for white matter hyperintensities (WMH) segmentation). The segmentation system relies on a number of atlases each consisting of a 3D T1-weighted MR image and a label map dividing the MR image into different tissue segments. {4}------------------------------------------------ Quantib™ ND provides quantitative information on both the absolute and relative volume of the seqmented regions. The automatic WMH segmentation is to be reviewed and if necessary. edited by the user before validation of the segmentation, after which volumetric information is accessible. Quantib ND consists of Quantib ND Baseline, which provides analysis of images of one time-point, and Quantib ND Follow-Up, which provides longitudinal analysis of images of two time-points. Quantib ND Follow-Up can only process images that have been processed by Quantib ND Baseline. Quantib ND is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the radiology specialist in quantitative reporting. ### 5 Indications for Use ## Indications for use Quantib™ ND 1.5 Quantib™ ND is a non-invasive medical imaging processing application that is intended for automatic labeling, visualization, and volumetric quantification of segmentable brain structures from a set of magnetic resonance (MR) images. The Quantib™ ND output consists of segmentations, visualizations and volumetric measurements of brain structures and white matter hyperintensities. Volumetric measurements may be compared to reference centile data. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting. Quantib™ ND is a software application on top of Myrian®. ## Indications for use predicate device (Quantib™ ND 1.3) Quantib™ Brain is a non-invasive medical imaging processing application that is intended for automatic labeling, visualization, and volumetric quantification of segmentable brain structures from a set of magnetic resonance (MR) images. The Quantib™ Brain output consists of segmentations, visualizations and volumetric measurements of grey matter (GM), white matter (WM), and cerebrospinal fluid (CSF). The output also visualizes and quantifies white matter hyperintensity (WMH) candidates. Users need to review and if necessary, edit WMH candidates using the provided tools, before validation of the WMHs. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting. Quantib™ Brain is a post-processing plugin for the GE Advantage Workstation (AW 4.7) or AW Server (AWS 3.2) platforms. ## Indications for use comparison The intended use of the device is partially equivalent to the intended use of the previously cleared predicate device [K173939] #### Comparison of technological characteristics 6 The following technological characteristics are the same for Quantib™ ND 1.5 and its predicate device Quantib™ Brain 1.3: - . Target users, anatomical site, and usage location - . Design - Standards met ● {5}------------------------------------------------ - . Required input The following technological characteristics are different: - . Indications for use and Intended use: The device is a software application on top of Myrian®; Seqmentation and quantification of separate lobes and hippocampus, comparison of whole brain, lobes, and hippocampus volumes to reference centile curves of a group representing the general population, and improved longitudinal volume change calculation using automated image registration are added. - . Human factors: Similar workflow, but implementation is slightly different based on existing workflow of underlying software packages - . Algorithm design For hippocampus segmentation, a slightly different refinement step is used than for other brain structures. Brain structures follow-up analysis is done using technological characteristics also applied elsewhere in Quantib ND and the predicate device. - . Performance: For measures reported by both devices, performance numbers show slight changes, attributable to the underlying software packages. Assessment of the performance of segmentation of separate lobes, hippocampus, and reference centile curves is added. - . Compatibility with the environment and other devices: Quantib™ ND is an add-on for Myrian, which can be installed on reqular hardware using Windows as operating system. - Reported measures: Whole brain measures (sum of Grey Matter [GM] and White Matter ● [WM]) are reported instead of GM and WM separately. To the reported measures the volumes of the following structures are added: Hippocampus, Frontal Lobe, Occipital Lobe, Parietal Lobe, Temporal Lobe, and Cerebellum. Reference centile curves are added, comparing these results to a group representing the general population. #### Performance data 7 #### 7.1 QUALITY AND SAFETY Quantib™ ND 1.5 was designed in compliance with the following process standards: - . ISO 14971 – Medical devices - Application of risk management to medical devices - IEC 62304 Medical device software Software life cycle processes . - IEC 62366 Medical devices Application of usability engineering to medical devices . The following quality assurance measures were applied to Quantib™ ND 1.5 development: - Risk and hazard analysis - Design reviews . - Unit level testing ● - Integration testing ● - System testing . - Performance testing . - . Usability engineering {6}------------------------------------------------ ## 7.2 ALGORITHM PERFORMANCE #### 7.2.1 Brain Structures To validate the quality of Quantib™ ND volume measurements and segmentations, these were compared to manual seqmentations of the same scan and their derived volumes. This analysis was performed for Brain Tissue, CSF, ICV, Hippocampus, Frontal Lobe, Occipital Lobe, Parietal Lobe, Temporal Lobe, and Cerebellum. For brain tissue, CSF, and ICV, the test set included 33 T1w MR images (Dataset A). The set was carefully selected to include data from multiple vendors and a series of representative scan settings. For each scan we selected six (6) slices for comparison. For the hippocampus the test set included 89 T1w images (Dataset B) and for the lobes the test set included 13 T1w MR images (Dataset C). For test sets B and C all slices were segmented manually for the comparison. The results are summarized below. | | Dataset | Dice index | Absolute difference of<br>the relative volumes [pp] | |-------------------|---------|-------------|-----------------------------------------------------| | Brain | A | 0.96 ± 0.01 | 1.7 ± 1.3 | | CSF | A | 0.78 ± 0.05 | 1.8 ± 1.3 | | ICV | A | 0.98 ± 0.01 | - | | Hippocampus total | B | 0.84 ± 0.03 | 0.03 ± 0.02 | | Hippocampus right | | 0.84 ± 0.03 | 0.01 ± 0.01 | | Hippocampus left | | 0.84 ± 0.03 | 0.01 ± 0.01 | Results of comparison between manual and automatic brain structure segmentation. Reported values are averages ± std. dev., computed over 6 segmented slices of 33 scans (Dataset B all slices were segmented. The Dice index provides a measure for overlap of manual and automatic segmentations (1 = perfect overlap). The absolute differences of the relative volumes are averages ± std. dev. in percentage points. | | Dataset | Dice index | Absolute difference of<br>the relative volumes [pp] | |----------------------|---------|-------------|-----------------------------------------------------| | Frontal lobe total | C | 0.95 ± 0.01 | 1.95 ± 0.90 | | Frontal lobe right | | 0.94 ± 0.02 | 1.02 ± 0.61 | | Frontal lobe left | | 0.94 ± 0.01 | 0.93 ± 0.50 | | Occipital lobe total | C | 0.88 ± 0.03 | 0.87 ± 0.75 | | Occipital lobe right | | 0.88 ± 0.03 | 0.43 ± 0.36 | | Occipital lobe left | | 0.87 ± 0.04 | 0.53 ± 0.53 | | Parietal lobe total | C | 0.89 ± 0.03 | 2.81 ± 1.13 | | Parietal lobe right | | 0.88 ± 0.04 | 1.45 ± 0.80 | | Parietal lobe left | | 0.88 ± 0.02 | 1.36 ± 0.56 | | Temporal lobe total | C | 0.91 ± 0.01 | 1.33 ± 0.76 | | Temporal lobe right | | 0.91 ± 0.02 | 0.72 ± 0.46 | | Temporal lobe left | | 0.91 ± 0.01 | 0.61 ± 0.39 | | Cerebellum total | C | 0.98 ± 0.01 | 0.47 ± 0.20 | {7}------------------------------------------------ | Cerebellum right | 0.97 ± 0.00 | 0.31 ± 0.13 | |------------------|-------------|-------------| | Cerebellum left | 0.97 ± 0.01 | 0.17 ± 0.11 | Results of comparison between manual and automatic brain structure segmentation of the lobes. Reported values are averages ± std. dev., computed over 13 scans of which all slices were segmented (Dataset C). The Dice index provides a measure for overlap of manual and automatic segmentations (1 = perfect overlap). The absolute differences of the relative volumes are averages ± std. dev. in percentage points. ## 7.2.2 White Matter Hyperintensities The test set for the White Matter Hyperintensities analysis included 45 3D T1w images, of which 7 contrast-enhanced, all with corresponding T2w FLAIR images. This set also represented various scan settings. WMHs were manually segmented on the T2w FLAIR images and compared to Quantib™ ND automatic segmentation output. The average Dice overlap between the manual segmentations and Quantib™ ND segmentations was 0.61 ± 0.13 (over all cases). The absolute difference of the relative volumes (for WMHs) was 0.2 ± 0.2 percentage points (over 38 cases without contrast-enhancement). #### CONCLUSIONS 8 By virtue of its intended use and physical and technological characteristics, Quantib™ ND 1.5 is substantially equivalent to a device that has been approved for marketing in the United States. The performance data shows that Quantib™ ND 1.5 is as safe and effective as the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...