K182545 · International Medical Industries, Inc. · FPA · Jan 11, 2019 · General Hospital
Device Facts
Record ID
K182545
Device Name
Tamper Evident Cap with Male Luer Lock
Applicant
International Medical Industries, Inc.
Product Code
FPA · General Hospital
Decision Date
Jan 11, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The Tamper Evident Cap with Male Luer Lock is indicated for use as a cap for female Luer ports on medical devices such as manifolds, stopcocks or set.
Device Story
The Tamper Evident Cap with Male Luer Lock (MTEC) is a sterile, single-use device designed to cover female Luer ports on medical devices; it provides a physical barrier and visual evidence of tampering if removed after installation. The device is manufactured from polypropylene and polystyrene resins. It is intended for use in clinical settings where intravascular administration sets are utilized. The device is operated by healthcare providers to secure open Luer ports. By providing evidence of access, the device helps maintain the sterility of the fluid path and alerts clinicians to potential unauthorized access or contamination, thereby supporting patient safety during intravenous therapy.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via ISO 80369-7:2016 (separation force, unscrewing torque, ease of assembly, stress cracking, liquid leakage, resistance to overriding), shelf-life evaluation, and biocompatibility testing per ISO 10993-1.
Technological Characteristics
Injection-molded polypropylene and polystyrene components. Male Luer lock interface. Sterile (Ethylene Oxide). Compliant with ISO 80369-7:2016 for small-bore connectors. Biocompatibility per ISO 10993-1 (External Communicating Device, Blood Path, Indirect Contact, Limited Duration).
Indications for Use
Indicated for use as a cap for female Luer ports on medical devices (manifolds, stopcocks, or sets) to provide evidence of access.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 11, 2019
International Medical Industries, Inc. Kathleen Lavender Director, Quality and Compliance 2981 Gateway Drive Pompano Beach, Florida 33069
Re: K182545
Trade/Device Name: Tamper Evident Cap with Male Luer Lock Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: December 10, 2018 Received: December 12, 2018
Dear Kathleen Lavender:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182545
#### Device Name
Tamper Evident Cap with Male Luer Lock
#### Indications for Use (Describe)
The Tamper Evident Cap with Male Luer Lock is indicated for use as a cap for female Luer ports on medical devices such as manifolds, stopcocks or set.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary: K182545
| Manufacturer: | International Medical Industries, Inc. |
|-----------------------------|---------------------------------------------------------|
| Address: | 2981 Gateway Drive<br>Pompano Beach, Florida 33069 |
| Corresponding official: | Kathleen Lavender<br>Director of Quality and Compliance |
| Telephone Number: | 954.917.9570 x 283 |
| Email: | klavender@imiweb.com |
| Summary Date: | January 10, 2019 |
| Trade Name: | Tamper Evident Cap with Male Luer Lock |
| Common or Usual Name: | I.V. Administration Set |
| Regulation Number: | 21 CFR 880.5440 |
| Regulation Name: | Intravascular Administration Set |
| Product Code: | FPA |
| Class: | II |
| Panel: | General Hospital |
| Primary Predicate Device: | Dual Luer Lock Cap (K101385) |
| Secondary Predicate Device: | Tamper Evident Cap with Luer Lock (K861276) |
#### Device Description:
The Tamper Evident Cap with Male Luer Lock (MTEC) is a sterile, single use device which will be used to cover female Luer ports on medical devices and provide evidence of access. MTEC is provided in two packaging configurations; a 10-up tray and single unit blisters. Both configurations are offered in multiple colors and all the components used in the manufacture of MTEC are comprised of polystyrene and polypropylene resin.
{4}------------------------------------------------
| Feature of the device: | Device:<br>K182545<br>Tamper Evident Cap<br>with male Luer Lock | Primary Predicate:<br>K101385<br>Baxter Dual Luer<br>Lock Cap | Secondary Predicate:<br>K861276<br>Tamper Evident Cap<br>with Luer Lock | Discussion/Comment |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | FPA | FPA | FPA | No Change |
| Classification | Class II | Class II | Class II | No Change |
| Number of uses | Single use Rx only | Single use Rx only | Single use Rx only | No Change |
| Materials | Device is an injection<br>molded polypropylene<br>component. | Device is an injection<br>molded polypropylene<br>component. | Device is an injection<br>molded polypropylene<br>component. | Same: No Change |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Same: No Change |
| Biocompatibility | Biocompatibility<br>assessment of the Dual<br>Luer Lock Cap has<br>been conducted based<br>on ISO 10993-1 -<br>Biological Evaluation<br>of Medical Devices. | Biocompatibility<br>assessment of the Dual<br>Luer Lock Cap has<br>been conducted based<br>on ISO 10993-1 -<br>Biological Evaluation<br>of Medical Devices. | Biocompatibility<br>assessment of the Dual<br>Luer Lock Cap has been<br>conducted based on ISO<br>10993-1 - Biological<br>Evaluation of Medical<br>Devices. | Same: No Change |
| Indication for Use: | The Tamper Evident<br>Cap with Male Luer<br>Lock is indicated for<br>use as a cap for female<br>Luer ports on medical<br>devices such as<br>manifolds, stopcocks<br>or sets. | The Dual Luer Lock<br>Cap is indicated for<br>use as a cap for male<br>or female Luer ports<br>on medical devices<br>such as manifolds,<br>stopcocks or sets. | The Tamper Evident<br>Cap with Female Luer<br>Lock is indicated for use<br>as a cap for male Luer<br>ports on medical devices | Device is pending<br>premarket notification.<br>Predicate is<br>commercially available<br>as Baxter Dual Luer<br>Lock Cap<br>Not subject to recall<br><br>The primary difference<br>is that the primary<br>predicate device has<br>both male and female<br>Luer connection points<br>whereas the proposed<br>device only has a male<br>Luer lock connection<br>point.<br>The primary difference<br>between the secondary<br>predicate device and the<br>proposed device is that<br>the secondary predicate<br>only has a female Luer<br>lock connection point. |
| Design Feature:<br>Tamper<br>Evidence | Device is designed to<br>cover an open Luer<br>port after the Luer port<br>has been accessed and<br>will provide evidence<br>of tampering if<br>removed after | Device is designed to<br>cover an open Luer<br>port after the Luer port<br>has been accessed. | Device is designed to<br>cover an open Luer port<br>after the Luer port has<br>been accessed and will<br>provide evidence of<br>tampering if removed<br>after installation. | The proposed product is<br>designed with a primary<br>function of a standard<br>male Luer lock cap with<br>a secondary function as<br>a tamper evident<br>(evidence of access) |
{5}------------------------------------------------
# Technological Characteristics:
The MTEC device is an assembly of injection molded components with a polypropylene Luer lock tip cap. The intended use and function of the proposed device is identical to the primary predicate device with respect to functionality. Both the proposed device (tip cap) and the primary predicate device are comprised of polypropylene resin. There are two design differences between the proposed device and the primary predicate device. The primary difference is that the primary predicate device has both male and female Luer connection points whereas the proposed device only has a male Luer lock connection point. The second difference is that the proposed device has a tamper evident design feature. This design feature is identical to the secondary predicate device.
# Performance Data:
Bench testing was performed to demonstrate that the proposed MTEC device is substantially equivalent to both the primary predicate and secondary predicate devices.
| Test Name | Test Description |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 80369-7:2016 | Design verification to ISO 80369-7:2016 Small-bore connectors for<br>liquids and gases in healthcare applications -- Part 7: Connectors for<br>intravascular or hypodermic applications |
| Shelf Life Evaluation | Design verification of the shelf life. |
| Sterilization Product<br>Adoption | AAMI TIR 28:2009/(R)2013 Product adoption and process<br>equivalence for ethylene oxide sterilization |
| Packaging | ISO 11607-1:2006 Packaging for terminally sterilized medical<br>devices -- Part 1: Requirements for materials, sterile barrier systems<br>and packaging systems |
| Sterility | ISO 11135:2014 Sterilization of health-care products -- Ethylene<br>oxide -- Requirements for the development, validation and routine<br>control of a sterilization process for medical devices |
| Residuals | ISO 10993-7:2008 Biological evaluation of medical devices -- Part 7:<br>Ethylene oxide sterilization residuals |
| Biocompatibility | ISO 10993-1, “Biological evaluation of medical devices – Part 1:<br>Evaluation and testing within a risk management process”; Guidance<br>for Industry and Food and Drug Administration Staff, Jun 16, 2016<br>and "Guidance for Industry and FDA Staff – Intravascular<br>Administration Sets Premarket Notification Submissions [510(k)],”<br>July 11, 2008, as recognized by the FDA. |
{6}------------------------------------------------
### Discussion of Non-Clinical Testing:
International Medical Industries conducts risk analyses and design verification tests that are based on the result of these analyses. All test results meet the acceptance criteria and support that the device is appropriately designed for the intended use.
The MTEC device meets the bench testing requirements of ISO 80369-7:2016 Small-bore connectors for liquids and gases in healthcare applications -- Part 7: Connectors for intravascular or hypodermic applications. The primary predicate device was tested to ISO 594-2:1998 which has since been replaced by ISO 80369-7:2016.
Compliance with ISO 80369-7:2016 demonstrates that The MTEC device meets the Bench performance testing requirements as follows:
- * Separation Force
- * Unscrewing torque
- * Ease of assembly * Stress cracking
- * Liquid leakage test
- * Resistance to overriding
### Proposed device vs. primary predicate device
The MTEC device meets the bench testing requirements of ISO 80369-7:2016 Small-bore connectors for liguids and gases in healthcare applications -- Part 7: Connectors for intravascular or hypodermic applications. The primary predicate device was tested to ISO 594-2:1998 which has since been replaced by ISO 80369-7:2016. Compliance with ISO 80369-7:2016 demonstrates that the proposed device is substantially equivalent to the primary predicate device K101385.
### Proposed device vs. secondary predicate device
The MTEC device's tamper evident design feature was tested and successfully passed all internal specifications. It is substantially equivalent to the secondary predicate device K861276.
{7}------------------------------------------------
### Biocompatibility Testing
Biocompatibility testing of the Tamper Evident Cap with Male Luer Lock was conducted per Guidance Document "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"; Guidance for Industry and Food and Drug Administration Staff, Jun 16, 2016 and "Guidance for Industry and FDA Staff - Intravascular Administration Sets Premarket Notification Submissions [510(k)]," July 11, 2008, as recognized by the FDA. Biocompatibility testing was conducted in accordance with the cited guidance and standards as required for an External Communicating Device, Blood Path, Indirect Contact, Limited Duration.
# Conclusion:
Review of the in-vitro performance test data as well as comparison of the device classification, indications for use, operating principle, technological characteristics, sterility, and biocompatibility demonstrate that the subject device, Tamper Evident Cap with Male Luer Lock (MTEC) is substantially equivalent to the primary predicate device, the Baxter Dual Luer Lock Cap (K101385). Any technological differences between the subject and the predicate devices do not raise any issues of safety and effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.