Miha Bodytec II

K182519 · Miha Bodytec GmbH · NGX · Jun 7, 2019 · Physical Medicine

Device Facts

Record IDK182519
Device NameMiha Bodytec II
ApplicantMiha Bodytec GmbH
Product CodeNGX · Physical Medicine
Decision DateJun 7, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5850
Device ClassClass 2
AttributesTherapeutic

Intended Use

miha bodytec II is a machine with electronic muscle stimulation based on EMS technology. Regarding its use, the device is specifically designed as an addition to other sports and for training muscles. It must be used for healthy muscles and clients, not be used for rehabilitation purposes. miha bodytec II is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. miha bodytec II is not intended to be used in conjunction with therapy or treatment of medical conditions of any kind. None of the miha bodytec II training programs are designed for injured or ailing muscles and its use on such muscles is contraindicated.

Device Story

Transcutaneous electrical muscle stimulation (EMS) device; stimulates motor nerves via electrical impulses transmitted through electrodes. System includes control unit, i-body electrode vest, straps, and belt. Used in professional sports settings; operated by manufacturer-trained professional trainers. Trainer selects programs (impulse familiarization, invigoration, endurance, relax) and adjusts intensity per channel via control unit UI. Athlete wears biocompatible undergarments under electrode accessories; electrodes moistened with tap water. Device generates symmetric biphasic pulses; triggers action potentials on motoneurons to generate mechanical muscle fiber responses. Trainer instructs athlete on exercises during stimulation. RFID transponder cards store training results and individual settings. Benefits include improved muscle performance through targeted muscle work. Safety features include short-circuit/no-load monitoring, emergency stop, and firmware-based error detection.

Clinical Evidence

Bench testing only. Includes electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), biocompatibility (ISO 10993-5/10), and software validation (IEC 62304). Performance testing included oscilloscope waveform analysis under various loads (500Ω, 2kΩ, 10kΩ), thermographic inspections, and shock/vibration testing. Electrode dispersion and shelf-life testing (ASTM F 1980-16) confirmed performance and durability.

Technological Characteristics

Powered muscle stimulator; 10-channel symmetric biphasic waveform. Materials: Aluminum control unit; stainless steel multifilament yarn electrodes. Connectivity: Wired (D-Sub 25-pin/proprietary magnetic connectors). Power: 100-240V AC. Software: Firmware-driven, IEC 62304 compliant. Standards: IEC 60601-2-10, IEC 60601-1-11, ISO 10993, ISO 14971.

Indications for Use

Indicated for healthy individuals to stimulate healthy muscles to improve or facilitate muscle performance. Contraindicated for use on injured or ailing muscles or in conjunction with therapy/treatment of medical conditions.

Regulatory Classification

Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 7, 2019 miha bodytec GmbH % Rhonda Alexander Sr. Consultant, Leadership & Regulatory Development IUVO Consulting, LLC P.O. Box 56483 Virginia Beach, Virginia 23456 Re: K182519 Trade/Device Name: miha bodytec II Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: May 3, 2019 Received: May 6, 2019 Dear Rhonda Alexander: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Vivek Pinto Assistant Director, Acute Injury Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K182519 Device Name miha bodytec II ### Indications for Use (Describe) miha bodytec II is a machine with electronic muscle stimulation based on EMS technology. Regarding its use, the device is specifically designed as an addition to other sports and for training muscles. It must be used for healthy muscles and clients, not be used for rehabilitation purposes. miha bodytec II is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. miha bodytec II is not intended to be used in conjunction with therapy or treatment of medical conditions of any kind. None of the miha bodytec II training programs are designed for injured or ailing muscles and its use on such muscles is contraindicated. The miha bodytec II electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles. The various types of muscle work that the miha bodytec II can impose on the stimulated muscles are able to improve or facilitate muscle performance. The miha bodytec II may therefore a technique of muscle training. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary ## for # miha bodytec II This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92 #### 1 Sponsor | Sponsor: | miha bodytec GmbH<br>Siemensstr. 1<br>86368 Gersthofen<br>Germany | |-----------------|-------------------------------------------------------------------------------------------------| | Contact Person: | Dr. Rhonda Alexander<br>IUVO Consulting, LLC<br>(757) 582-4337<br>ralexander@iuvoconsulting.com | Date Prepared: May 2, 2019 ### 2 Device Name and Classification Proprietary Name: miha bodytec II Common/Usual Name: Powered muscle stimulator Classification Name: Stimulator, Muscle, Powered, For Muscle Conditioning (21 CFR 890.5850, Product Code NGX) ### Predicate Device 3 E-Fit EF-1280, K133225 Predicate Device: #### Intended Use 4 miha bodytec II is a machine with electronic muscle stimulation based on EMS technology. Regarding its use, the device is specifically designed as an addition to other sports and for training muscles. It must be used for healthy muscles and clients, not be used for rehabilitation purposes. {4}------------------------------------------------ miha bodytec II is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. miha bodytec II is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the miha bodytec II training programs are designed for injured or ailing muscles and its use on such muscles is contraindicated. The miha bodytec II electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles. The various types of muscle work that the miha bodytec II can impose on the stimulated muscles are able to improve or facilitate muscle performance. The miha bodytec II may therefore be considered a technique of muscle training. ### 5 Device Description and Function miha bodytec II is a transcutaneous electrical muscle stimulation (EMS) device which stimulates motor nerves by means of electrical impulses transmitted by electrodes. These excitations of motor nerves are transmitted to the muscle fibers where they stimulate a muscular response. Depending on the parameters of the electrical impulses (pulse frequency, pulse intensity, pulse duration, pulse width, pulse rise, pause time, total session duration), different types of muscle work can be imposed on the stimulated muscles. miha Bodytec II consists of a control unit mounted on a stand for the selection of programs, setting the parameters and starting/stopping the device, the i-body® electrode vest for applying electrodes to the upper body, i-body® straps for applying electrodes to the arm and legs and the i-body® belt for applying electrodes to the buttocks. The control unit and the i-body® accessories are connected via cables and must be worn on top of optionally available undergarment. miha bodytec II must be used in a professional sport setting. The device must be operated by a trainer who has received a full training from the manufacturer. Before the trainer selects the accessories incl. electrodes in the correct size, applies the athlete i.e. by wearing the electrode vest and connects the straps und belt via cable to the vest and to the control unit. The trainer can choose between several training programs on the control unit for impulse familiarization, invigoration basic/advanced, muscular endurance and body relax. The intensity can be adjusted by the trainer at the UI of the control unit separately for each channel. Complete body training which addresses all muscle groups is possible with up to 10 pairs of electrodes. Each athlete receives an RFID transponder card for storing training results and individually adjusted programs. Once the training is started, the control unit generates and transmits the electrical signals to the electrodes via cable. miha bodytec II uses bipolar pulses and supplies all channels equally during all programs. {5}------------------------------------------------ During pulse application, the trainer instructs the athlete on specific exercises to perform. The training can be stopped anytime by pressing the multi-function / stop button. ### Predicate Device Comparison б # 6.1 General | Characteristic | New Device | Predicate Device | Similar / Different | |-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number | K182519 | K133225 | - | | Device Name,<br>Model | miha bodytec II | E-Fit EF-1280 | - | | Manufacturer | miha bodytec GmbH | Fit-Pro KFT Ltd. | - | | Indications for<br>Use | miha bodytec II is a machine<br>with electronic muscle<br>stimulation based on EMS<br>technology. Regarding its<br>use, the device is specifically<br>designed as an addition to<br>other sports and for training<br>muscles. It must be used for<br>healthy muscles and clients,<br>not be used for<br>rehabilitation purposes. | E-fit EF-1280 is a<br>machine with<br>electronic muscle<br>stimulation based on<br>EMS technology.<br>Regarding its use, the<br>device is specifically<br>designed as an<br>addition to other<br>sports and for training<br>muscles. It must be<br>used for only healthy<br>muscles and clients,<br>not for rehabilitation<br>purposes. | There are no differences<br>between the subject<br>device and the<br>predicate(s) with respect<br>to indications and<br>intended use. | | | miha bodytec II is intended<br>to stimulate healthy muscles<br>in order to improve or<br>facilitate muscle<br>performance. miha bodytec<br>II is not intended to be used<br>in conjunction with therapy<br>or treatment of medical<br>diseases or medical<br>conditions of any kind. None<br>of the miha bodytec II<br>training programs are<br>designed for injured or ailing<br>muscles and its use on such<br>muscles is contraindicated. | The E-Fit EF-1280<br>intended to stimulate<br>healthy muscles in<br>order to improve or<br>facilitate muscle<br>performance. The E-<br>Fit EF-1280 is not<br>intended to be used<br>in conjunction with<br>therapy or treatment<br>of medical diseases or<br>medical conditions of<br>any kind. None of the<br>E-Fit EF-1280 training<br>programs is designed<br>for injured or ailing<br>muscles and its use | - | | The miha bodytec II | | on such muscles is | | | electrical impulses allow the | | contraindicated. | | | triggering of action | | | | | potentials on motoneurons | | | | | of motor nerves | The E-Fit EF-1280 | | | | (excitations). These | electrical impulses | | | | excitations of motoneurons | allow the triggering of | | | | are transmitted to the | action potentials on | | | | muscle fibers via the motor | motoneurones of | | | | endplate where they | motor nerves | | | | generate mechanical muscle | (excitations). These | | | | fiber responses that | excitations of | | | | correspond to muscle work. | motoneurones are | | | | Depending on the | transmitted to the | | | | parameters of the electrical | muscle fibers via the | | | | impulses (pulse frequency, | motor endplate | | | | duration of contraction, | where they generate | | | | duration of rest, total | mechanical muscle | | | | session duration), different | fiber responses that | | | | types of muscle work can be | correspond to muscle | | | | imposed on the stimulated | work. Depending on | | | | muscles. | the parameters of the | | | | | electrical impulses | | | | | (pulse frequency, | | | | | duration of | | | | | contraction, duration | | | | | of rest, total session | | | | The various types of muscle | duration), different | | | | work that the miha bodytec | types of muscle work | | | | II can impose on the | can be imposed on | | | | stimulated muscles are able | the stimulated | | | | to improve or facilitate | muscles. | | | | muscle performance. The | | | | | miha bodytec II may | | | | | therefore be considered a | The various types of | | | | technique of muscle | muscle work that the | | | | training. | E-Fit EF-1280 can | | | | | impose on the | | | | | stimulated muscles | | | | | are able to improve or | | | | | facilitate muscle | | | | | performance. The E- |…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...