PYRENEES Cervical Plate System, BLUE RIDGE Cervical Plate System, OZARK Cervical Plate System, CAYMAN Thoracolumbar and Buttress Plate Systems
K182473 · K2m, Inc. · KWQ · Dec 12, 2018 · Orthopedic
Device Facts
Record ID
K182473
Device Name
PYRENEES Cervical Plate System, BLUE RIDGE Cervical Plate System, OZARK Cervical Plate System, CAYMAN Thoracolumbar and Buttress Plate Systems
Applicant
K2m, Inc.
Product Code
KWQ · Orthopedic
Decision Date
Dec 12, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PYRENEES and BLUE RIDGE Cervical Plate Systems: PYRENEES and BLUE RIDGE Cervical Plate Systems are indicated for use in anterior screw fixation to the cervical spine (C2-T1) for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (including fractures), spinal stenosis and tumors (primary and metastatic), failed previous fusions (pseudarthrosis) and deformity (defined as scoliosis, kyphosis or lordosis). OZARK Cervical Plate System: OZARK Cervical Plate System is indicated for use in anterior screw fixation to the cervical spine (C2-T1) for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (including fractures), spinal stenosis and tumors (primary and metastatic), failed previous fusions (pseudarthrosis) and deformity (defined as scoliosis, kyphosis or lordosis). CAYMAN Thoracolumbar and Buttress Plate Systems: The CAYMAN Buttress Plates are intended for use in spinal fusion procedures as a means to maintain the relative position of weak bony tissue such as allografts or autografts. The device is not intended for load bearing indications. The CAYMAN Thoracolumbar Plates are indicated for use via the lateral or anterolateral surgical approach in the treatment of thoracic and thoracolumbar (T1-L5) spine and for use as an anteriorly placed supplemental fixation device for the lumbosacral level below the bifurcation of the vascular structures (L5-S1). The Cayman Thoracolumbar Plate System is intended to provide temporary stabilization during fusion using autograph or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities and deformities: a) degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), b) pseudoarthrosis, c) spondylolysis, d) spondylolisthesis, e) fracture, f) neoplastic disease, g) unsuccessful previous fusion surgery, h) lordotic deformities of the spine, i) thoracolumbar or lumbar scoliosis, j) deformity (i.e., scoliosis, kyphosis, and/or lordosis) associated with deficient posterior elements such as that resulting from laminectomy.
Device Story
Spinal fixation systems consisting of plates and screws; provide temporary stabilization across spinal levels until fusion achieved. Used in surgical settings by orthopedic/neurosurgeons. Submission establishes MR Conditional labeling; adds optional sterile packaging for specific components. No changes to device design or materials. Benefits include stabilization of bony tissue during fusion procedures.
Clinical Evidence
Bench testing only. MR compatibility testing performed per ASTM F2503; results met acceptance criteria. Bacterial endotoxin testing (BET) performed per AAMI ST72:2011 for sterile-packaged devices; results met specified limits.
Indicated for skeletally mature patients requiring anterior screw fixation to the cervical spine (C2-T1) or lateral/anterolateral fixation to the thoracic/thoracolumbar spine (T1-L5) and lumbosacral level (L5-S1) for degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, tumors, pseudarthrosis, and spinal deformities (scoliosis, kyphosis, lordosis).
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
CAYMAN Thoracolumbar and Buttress Plate Systems (K131533)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 12, 2018
K2M Inc. % Karen E. Warden, Ph.D. President BackRoads Consulting, Inc. P.O. Box 566 Chesterland, Ohio 44026
Re: K182473
Trade/Device Name: PYRENEES Cervical Plate System, BLUE RIDGE Cervical Plate System, OZARK Cervical Plate System, CAYMAN Thoracolumbar and Buttress Plate Systems Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: November 23, 2018 Received: November 23, 2018
Dear Dr. Warden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ronald P. Jean -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182473
Device Name
PYRENEES Cervical Plate System, BLUE RIDGE Cervical Plate System
#### Indications for Use (Describe)
PYRENEES and BLUE RIDGE Cervical Plate Systems are indicated for use in anterior screw fixation to the cervical spine (C2-T1) for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (including fractures), spinal stenosis and tumors (primary and metastatic), failed previous fusions (pseudarthrosis) and deformity (defined as scoliosis, kyphosis or lordosis).
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182473
Device Name OZARK Cervical Plate System
#### Indications for Use (Describe)
OZARK Cervical Plate System is indicated for use in anterior screw fixation to the cervical spine (C2-T1) for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (including fractures), spinal stenosis and tumors (primary and metastatic), failed previous fusions (pseudarthrosis) and deformity (defined as scoliosis, kyphosis or lordosis).
Type of Use (Select one or both, as applicable)
| <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> |
|---------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182473
#### Device Name
CAYMAN Thoracolumbar and Buttress Plate Systems
#### Indications for Use (Describe)
The CAYMAN Buttress Plates are intended for use in spinal fusion procedures as a means to maintain the relative position of weak bony tissue such as allografts. The device is not intended for load bearing indications. The CAYMAN Thoracolumbar Plates are indicated for use via the lateral or anterolateral surgical approach in the treatment of thoracic and thoracolumbar (T1-L5) spine and for use as an anteriorly placed supplemental fixation device for the lumbosacral level below the bifurcation of the vascular structures (L5-S1). The Cayman Plate System is intended to provide temporary stabilization during fusion using autograph or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities and deformities: a) degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), b) pseudoarthrosis, c) spondylolisthesis, e) fracture, f) neoplastic disease, g) unsuccessful previous fusion surgery, h) lordotic deformities of the spine, i) thoracolumbar or lumbar scoliosis, j) deformity (i.e., scoliosis, kyphosis, and/or lordosis) associated with deficient posterior elements such as that resulting from laminectomy.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-family: Arial;"> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
| Date: | 6 September 2018 |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | K2M Inc.<br>600 Hope Pkwy. SE<br>Leesburg, Virginia 20175<br>Phone: 571.919.2000 |
| Sponsor Contact: | Nancy Giezen |
| 510(k) Contact: | Karen E. Warden, PhD<br>BackRoads Consulting<br>PO Box 566<br>Chesterland, OH 44026<br>Office: 440.729.8457 |
| Trade Names: | PYRENEES Cervical Plate System, BLUE RIDGE Cervical Plate<br>System, OZARK Cervical Plate System, CAYMAN Thoracolumbar and<br>Buttress Plate Systems |
| Common Name: | Spinal Fixation Systems |
| Regulatory Class: | Class II |
| Classification Name,<br>Regulation, Product<br>Code: | Appliance, fixation, spinal intervertebral body, 888.3060, KWQ |
| Device Description: | The previously cleared K2M PYRENEES, BLUE RIDGE, OZARK and<br>CAYMAN plate systems consist of a variety of plates and screws designed<br>to provide support across implanted levels of the cervical, thoracolumbar<br>and lumbosacral spine until fusion is achieved.<br>The primary purpose of this submission is to establish an MR Conditional<br>labeling claim for these implants. In addition, PYRENEES, BLUE RIDGE<br>and CAYMAN components that were previously provided non-sterile are<br>now being optionally offered as sterile packaged devices. |
| Intended Use: | PYRENEES and BLUE RIDGE Cervical Plate Systems:<br>PYRENEES and BLUE RIDGE Cervical Plate Systems are indicated for use<br>in anterior screw fixation to the cervical spine (C2-T1) for the following<br>indications: degenerative disc disease (DDD) (defined as neck pain of<br>discogenic origin with degeneration of the disc confirmed by history and<br>radiographic studies), spondylolisthesis, trauma (including fractures), spinal<br>stenosis and tumors (primary and metastatic), failed previous fusions<br>(pseudarthrosis) and deformity (defined as scoliosis, kyphosis or lordosis).<br>OZARK Cervical Plate System:<br>OZARK Cervical Plate System is indicated for use in anterior screw fixation<br>to the cervical spine (C2-T1) for the following indications: degenerative disc<br>disease (DDD) (defined as neck pain of discogenic origin with degeneration<br>of the disc confirmed by history and radiographic studies), spondylolisthesis,<br>trauma (including fractures), spinal stenosis and tumors (primary and<br>metastatic), failed previous fusions (pseudarthrosis) and deformity (defined<br>as scoliosis, kyphosis or lordosis).<br>CAYMAN Thoracolumbar and Buttress Plate Systems:<br>The CAYMAN Buttress Plates are intended for use in spinal fusion<br>procedures as a means to maintain the relative position of weak bony tissue<br>such as allografts or autografts. The device is not intended for load bearing<br>indications.<br>The CAYMAN Thoracolumbar Plates are indicated for use via the lateral or |
| | anterolateral surgical approach in the treatment of thoracic and<br>thoracolumbar (T1-L5) spine and for use as an anteriorly placed<br>supplemental fixation device for the lumbosacral level below the bifurcation<br>of the vascular structures (L5-S1). The Cayman Thoracolumbar Plate<br>System is intended to provide temporary stabilization during fusion using<br>autograph or allograft in skeletally mature patients in the treatment of the<br>following acute and chronic instabilities and deformities: a) degenerative<br>disc disease (defined as back pain of discogenic origin with degeneration of<br>the disc confirmed by patient history and radiographic studies), b)<br>pseudoarthrosis, c) spondylolysis, d) spondylolisthesis, e) fracture, f)<br>neoplastic disease, g) unsuccessful previous fusion surgery, h) lordotic<br>deformities of the spine, i) thoracolumbar or lumbar scoliosis, j) deformity<br>(i.e., scoliosis, kyphosis, and/or lordosis) associated with deficient posterior<br>elements such as that resulting from laminectomy. |
| Materials: | The PYRENEES, BLUE RIDGE and OZARK Cervical Plate Systems and<br>the CAYMAN Thoracolumbar and Buttress Plate Systems are manufactured<br>from titanium alloy (ASTM F136 and ASTM F1472), titanium (ASTM F67),<br>cobalt chrome (ASTM F1537) and nitinol (ASTM F2063). The materials<br>used in the manufacturing of the instruments are stainless steel with some<br>of the instruments incorporating aluminum, linen phenolic, Ultem, Radel,<br>polypropylene, acetal copolymer, or silicone in the handles or retractor<br>blades. |
| Primary Predicate: | OZARK Cervical Plate System (K2M Inc. – K172104) |
| Additional Predicates: | PYRENEES and BLUE RIDGE Cervical Plate Systems (K2M Inc. –<br>K162664 and K113329, respectively), CAYMAN Thoracolumbar and<br>Buttress Plate Systems (K2M Inc. – K131533) |
| Performance Data: | MR Compatibility testing per ASTM F2503 was performed. The test results<br>demonstrate that the PYRENEES, BLUE RIDGE and OZARK Cervical Plate<br>Systems and the CAYMAN Thoracolumbar and Buttress Plate Systems<br>performance met the prescribed acceptance criteria.<br>In addition, bacterial endotoxin testing (BET) was conducted in accordance<br>with AAMI ST72:2011. The test results demonstrate that the sterile devices<br>met the specified testing limit. |
| Technological<br>Characteristics: | The PYRENEES, BLUE RIDGE and OZARK Cervical Plate Systems and<br>the CAYMAN Thoracolumbar and Buttress Plate Systems possess the<br>same technological characteristics as their predicate devices; no changes<br>have been made to any of the devices. Therefore the fundamental scientific<br>technology of the PYRENEES, BLUE RIDGE and OZARK Cervical Plate<br>Systems and the CAYMAN Thoracolumbar and Buttress Plate Systems is<br>the same as previously cleared devices. |
| Conclusion: | The PYRENEES, BLUE RIDGE and OZARK Cervical Plate Systems and<br>the CAYMAN Thoracolumbar and Buttress Plate Systems possess the<br>same intended use and technological characteristics as the predicate<br>devices. Therefore the PYRENEES, BLUE RIDGE and OZARK Cervical<br>Plate Systems and the CAYMAN Thoracolumbar and Buttress Plate<br>Systems are substantially equivalent. |
# 510(k) Summary
{6}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.