PeekMed is a software system designed to help surgeons' specialists carry out the pre-operative planning in a prompt and efficient manner for several surgical procedures, based on their patients' imaging studies. The software imports diagnostics imaging studies such as x-rays, CT or magnetic resonance image (MRI). The import process can retrieve files from a CD ROM, a local folder or the PACS. In parallel, there is a database of digital representations related to prosthetic materials supplied by their producing companies. PeekMed allows health professional to digitally perform the surgical planning without adding any additional steps to that process. This software system requires no imaging study acquisition specification (no protocol). Experience in usage and a clinical assessment are necessary for a proper use of the software.
Device Story
PeekMed is a standalone 3D pre-operative planning software for surgeons; imports diagnostic imaging (X-ray, CT, MRI) from PACS, local folders, or CD-ROM; integrates a database of digital prosthetic templates. Surgeons use the software to perform 2D, 3D, or hybrid (2D/3D) surgical planning; allows interactive positioning, dimensioning, and rotation of prosthetic models; supports intersection of models. Output provides a visual overview of the surgical plan, including measurements and template placement. Used in clinical settings to assist surgical decision-making; benefits patient by facilitating efficient, accurate pre-operative planning. Operates on Windows or OS X workstations.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed product specifications and requirements were met. Validation included measuring images of three CE-marked implantable devices to confirm accuracy and repeatability of ruler and angle measurement functions. Usability was assessed via user surveys and satisfaction questionnaires.
Technological Characteristics
Standalone software application; runs on Windows or OS X; supports 2D, 3D, and hybrid (2D/3D) environments. Features include image import (DICOM/PACS), digital prosthetic template database, interactive measurement tools (ruler, angle), and model manipulation (rotation, positioning, intersection). Risk management performed per ISO 14971.
Indications for Use
Indicated for surgeons performing pre-operative planning for orthopedic surgical procedures (hip, knee, spine, upper limb, foot and ankle, trauma) in adult and pediatric patients using diagnostic imaging (X-ray, CT, MRI). Requires clinical assessment and user experience.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Food and Drug Administration (FDA) and the Department of Health & Human Services. The FDA logo features the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. To the left of the FDA logo is the symbol of the Department of Health & Human Services, which consists of a stylized human figure.
October 25, 2018
Peek Health, S.A. % Rachel Paul Senior Consultant, QA&RA Emergo Europe Consulting Prinsessegracht 20 THE HAGUE, 2514 AP NETHERLANDS
Re: K182464
Trade/Device Name: PeekMed Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: September 6, 2018 Received: September 10, 2018
Dear Rachel Paul:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the name Michael D. O'hara -S in a large font. Below the name is the text "Digitally signed by Michael D. O'hara -S". The image also contains the text "DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, 0.9.2342.19200300.100.1.1=1300 226759, cn=Michael D. O'hara -S Date: 2018.10.25 13:24:17-04'00'".
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182464
Device Name PeekMed
Indications for Use (Describe)
PeekMed is a software system designed to help surgeons' specialists carry out the pre-operative planning in a prompt and efficient manner for several surgical procedures, based on their patients' imaging studies.
The software imports diagnostics imaging studies such as x-rays, CT or magnetic resonance image (MRI). The import process can retrieve files from a CD ROM, a local folder or the PACS. In parallel, there is a database of digital representations related to prosthetic materials supplied by their producing companies.
PeekMed allows health professional to digitally perform the surgical planning without adding any additional steps to that process. This software system requires no imaging study acquisition specification (no protocol). Experience in usage and a clinical assessment are necessary for a proper use of the software.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# DEVICE NAME
# K_182464
### 1. Submission Sponsor
Peek Health, S.A.
Praça Conde de Agrolongo, nº123
4700-312
Braga
Portugal
Sara Silva
Chief Quality Officer (CQO)
# 2. Submission Correspondent
Emergo Europe Consulting
Prinsessegracht 20
The Hague
2514 AP
The Netherlands
Rachel Paul
Senior Consultant, QA & RA
Email: LST.AUS.Projectmanagement@ul.com
# 3. Date Prepared
September 6, 2018
#### 4. Device Identification
Trade/Proprietary Name: PeekMed
Common/Usual Name: Pre-operative planning software
{5}------------------------------------------------
| Classification Name: | Picture Archiving and Communications System (PACS) |
|-----------------------|----------------------------------------------------|
| Regulation Number: | 892.2050 |
| Product Code: | LLZ, System, Image Processing, Radiological |
| Device Class: | Class II |
| Classification Panel: | Radiology |
# 5. Legally Marketed Predicate Device(s)
| 510(K) No. | K073714 |
|----------------|-----------------------|
| Clearance date | 03/19/2008 |
| Device Name | TraumaCad version 2.0 |
| Manufacturer | Orthocrat, Ltd |
#### 6. Indications for Use Statement
PeekMed is a software system designed to help surgeons' specialists carry out the pre-operative planning in a prompt and efficient manner for several surgical procedures, based on their patients' imaging studies.
The software imports diagnostics imaging studies such as x-rays, CT or magnetic resonance image (MRI). The import process can retrieve files from a CD ROM, a local folder or the PACS. In parallel, there is a database of digital representations related to prosthetic materials supplied by their producing companies.
PeekMed allows health professional to digitally perform the surgical planning without adding any additional steps to that process. This software system requires no imaging study acquisition (no protocol). Experience in usage and a clinical assessment are necessary for a proper use of the software.
# 7. Device Description
PeekMed is a standalone software application that helps specialist doctors and surgeons to perform the pre-surgical planning for different procedures in a fast and effective way, based on the imaging studies of the patients.
PeekMed is a 3D pre-operative planning software for surgery. This software system allows surgeons to plan a surgery procedure simulating various environments, from hybrid (2D/3D) to 3D or 2D environments.
The software imports diagnostic imaging studies such as X-rays, Computed Tomography (CT) or Magnetic Resonance Image (MRI). The import process can retrieve files from a local folder or the Picture Archiving and Communications System (PACS) of the hospital/health center. In parallel, there is a database containing digital representations related to prosthetic materials supplied by their respective manufacturers. This offers the possibility of inserting templates of the materials to be used in the surgery in addition to the measurements, making a complete overview of the surgery.
{6}------------------------------------------------
In the case of a 3D or hybrid environment, the surgeon can resort to 3D model generated from a previous imaging study on the patient and 3D digital representations of the prosthetic material to be used during surgery, i.e. screws, fixation plates or full prosthesis, deriving from several producing companies.
# 8. Substantial Equivalence Discussion
The following table compares PeekMed to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance testing. The comparison of the devices provides more detailed information regarding the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
| Manufacturer | 510(K) Submitter | Predicate | Significant Differences |
|------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Peek Health, S.A. | Orthocrat, Ltd | |
| Device Trade Name | PeekMed | TraumaCad version 2.0 | |
| 510(k) Number | N/A | K140434 | N/A |
| Product Code | LLZ | LLZ | N/A |
| Regulation<br>Number | 21 CFR 892.2050 | 21 CFR 892.2050 | N/A |
| Regulation Name | System, Image<br>Processing, Radiological | System, Image Processing,<br>Radiological | N/A |
| Intended for use | PeekMed is a pre-<br>operative planning<br>software for surgery. | TraumaCad is a pre-<br>operative planning<br>software for surgery. | N/A |
| Indications for use | PeekMed is a software<br>system designed to help<br>surgeons' specialists<br>carry out the pre-<br>operative planning in a<br>prompt and efficient<br>manner for several<br>surgical procedures,<br>based on their patients'<br>imaging studies.<br>The software imports<br>diagnostics imaging<br>studies such as x-rays, CT | The TraumaCAD program<br>is indicated for assisting<br>healthcare professionals<br>in preoperative planning<br>of orthopedic surgery. The<br>device allows for<br>overlaying of prosthesis<br>templates on radiological<br>images and includes tools<br>for performing<br>measurements on the<br>image and for positioning<br>the template. Clinical | In more to 2D, PeekMed can offer a 3D pre-surgical planning or a hybrid 2D/3D environment in addition to isolated 2D and 3D. PeekMed has been tested and validated for 3D and hybrid planning. |
| Manufacturer | 510(K) Submitter | Predicate | Significant Differences |
| | Peek Health, S.A. | Orthocrat, Ltd | |
| Device Trade<br>Name | PeekMed | TraumaCad version 2.0 | |
| | or magnetic resonance<br>image (MRI). The import<br>process can retrieve files<br>from a CD ROM, a local<br>folder or the PACS. In<br>parallel, there is a<br>database of digital<br>representations related<br>to prosthetic materials<br>supplied by their<br>producing companies.<br>PeekMed allows health<br>professional to digitally<br>perform the surgical<br>planning without adding<br>any additional steps to<br>that process. This<br>software system requires<br>no imaging study<br>acquisition specification<br>(no protocol). Experience<br>in usage and a clinical<br>assessment are<br>necessary for a proper<br>use of the software. | judgments and experience<br>are required to properly<br>use the software. | |
| Subspecialties | PeekMed allows the<br>surgeon to perform the<br>pre-surgical planning<br>efficiently in the<br>following subspecialties:<br>- Hip<br>- Knee<br>- Spine<br>- Upper Limb<br>- Foot and Ankle | TraumaCad provides easy-<br>to-use solutions for<br>various orthopedic<br>subspecialties:<br>- Hip<br>- Knee<br>- Pediatric<br>- Trauma<br>- Deformity<br>- Analysis | Orthopedic subspecialties<br>supported by PeekMed<br>are also covered in<br>TraumaCad. The<br>majorities of the<br>procedures supported by<br>PeekMed are also<br>supported by TraumaCad.<br>The procedures that are<br>not common to both<br>devices have been tested |
| Manufacturer | 510(K) Submitter | Predicate | Significant Differences |
| | Peek Health, S.A. | Orthocrat, Ltd | |
| Device Trade<br>Name | PeekMed | TraumaCad version 2.0 | |
| | - Trauma | - Spine<br>- Foot and Ankle<br>- Upper Limb<br>The following<br>measurements can be<br>made using TraumaCad<br>measurements tools:<br>-Hip: Hip Deformity<br>Analysis, Polyethylene<br>Wear Wizard, Leg Length<br>Discrepancy Tool,<br>Acetabular Index, VCA<br>Angle of Lequesne, Cup<br>Version, Center of<br>Rotation, Stem Version | and validated through<br>PeekMed development. It<br>does not raise different<br>questions of safety or<br>effectiveness. |
| | For each subspecialty,<br>there are several<br>procedures:<br>-Hip: Hip Dysplasia<br>Correction, Limb Length<br>Discrepancy, Center of<br>Rotation – Ranawat<br>method, Acetabular<br>Angle, Total Hip<br>Arthroplasty<br><br>-Knee: Leg Deformity<br>Correction, AP Knee<br>Resection, AP Full Leg<br>resection, High Tibial<br>Osteotomy, ACL Tunnel<br>Reconstruction, Medial<br>Patellofemoral Ligament<br><br>-Spine: Single Cobb<br>Angle, Thoracic Kyphosis<br>Angle, Lumbar Lordosis<br>Angle, Sagittal Vertical<br>Axis, Pelvic Angles,<br>Coronal Balance, Sacro-<br>femoral Angle, Sagittal<br>Balance, Smith-Petersen<br>Osteotomy, Pedicle<br>Subtraction Osteotomy | -Knee: Limb Alignment<br>Analysis, Center Line<br>Finder, Simple Line, High<br>Tibial Osteotomy, Tibial<br>Cutting, Joint Line<br><br>-Spine: Vertebrae<br>Labeling, Cobb Angle,<br>Double Cobb Angle, Triple<br>Cobb Angle, Pelvic Radius<br>Angle, Sacral Obliquity, ,<br>Coronal Balance, Sagittal<br>Balance,<br>Spondylolysthesis,<br>Thoracic Kyphosis Angle,<br>Thoracic Trunk Shift, T1<br>Tilt Angle, Lumbar<br>Lordosis, Spine Slip Angle | |
| Manufacturer | 510(K) Submitter | Predicate | Significant Differences |
| | Peek Health, S.A. | Orthocrat, Ltd | |
| Device Trade | PeekMed | TraumaCad version 2.0 | |
| Name | | | |
| | -Upper Limb: Total<br>Shoulder Replacement, | Upper Limb: Center Line<br>Finder, Simple Line, Joint<br>Line | |
| | Clavicular Angle,<br>Shoulder Resurfacing | -Foot and ankle: Foot<br>Osteotomies Wizard, | |
| | -Foot and Ankle: Talar<br>Tilt, Hallux Valgus,<br>Moreau-Costa-Bertani<br>Internal Angle, Moreau-<br>Costa-Bertani External<br>Angle, Maestro Formula | Hallux Valgus Angle,<br>Hallux Valgus<br>Interphalangeus Angle,<br>Intermetatarsal Angle,<br>Distal Metatarsal Articular<br>Angle (DMAA), Proximal<br>Metatarsal Articular Angle<br>(PMAA), Talar Tilt | |
| | -Trauma: Diaphyseal<br>Shaft Fracture Angle,<br>Metaphyseal Shaft<br>Fracture Angle, Roof Arc | -Trauma: Limb Alignment<br>Analysis, Diaphyseal<br>Fracture Angulation,<br>Metaphyseal Fracture<br>Angulation, Center Line<br>Finder, Simple Line, Roof<br>Arc, Joint Line | |
| | | -Pediatric: Hip Deformity<br>Analysis, Limb Alignment<br>Analysis, Acetabular<br>Index, Reimer's Index, VCA<br>Angle of Lequesne | |
| Patient Population | Adults and pediatrics | Adults and pediatrics | N/A |
| End users | Surgeons | Surgeons | N/A |
| Computer | Personal Computer or<br>Workstation | Personal Computer or<br>Workstation | N/A |
| Manufacturer | 510(K) Submitter | Predicate | Significant Differences |
| | Peek Health, S.A. | Orthocrat, Ltd | |
| Device Trade<br>Name | PeekMed | TraumaCad version 2.0 | |
| Operating System | Windows or OS X | Windows | PeekMed can run through<br>Windows or OSX and has<br>been tested and validated<br>for the both systems. |
| Device availability | It can be set to start from<br>a workstation or<br>standalone for planning<br>procedures | It can be set to start from<br>a workstation or<br>standalone for planning<br>procedures | N/A |
| Images source | Receives medical images<br>from various sources<br>(including PACS) | Receives medical images<br>from various sources<br>(including PACS) | N/A |
| Data processing | The software processes<br>data in order to provide<br>an overlap and<br>dimensioning of digital<br>representations of the<br>prosthetic material | The software processes<br>data in order to provide<br>an overlap and<br>dimensioning of digital<br>representations of the<br>prosthetic material | N/A |
| Manufacturer | 510(K) Submitter | Predicate | Significant Differences |
| | Peek Health, S.A. | Orthocrat, Ltd | |
| Device Trade<br>Name | PeekMed | TraumaCad version 2.0 | |
| Digital overlap of<br>prosthetic material | Allows the overlap of<br>models and the<br>intersection of the<br>models | Allows the overlap of<br>models | The additional feature<br>from PeekMed of allowing<br>the intersection of the<br>models has been tested<br>and validated and does<br>not raise different<br>questions of safety or<br>effectiveness. |
| Interactive model<br>positioning | Yes | Yes | N/A |
| Interactive model<br>dimensioning | Yes | Yes | N/A |
| Model rotation | Yes | Yes | N/A |
| Support for digital<br>prosthetic<br>materials provided<br>by the<br>manufacturers | Yes | Yes | N/A |
| Automatic<br>calibration | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.