ICOONE h (also referred to as ICOONE LASER and ICOONE –h LASER)
K182453 · I-Tech Industries Srl · NUV · Jan 11, 2019 · General, Plastic Surgery
Device Facts
Record ID
K182453
Device Name
ICOONE h (also referred to as ICOONE LASER and ICOONE –h LASER)
Applicant
I-Tech Industries Srl
Product Code
NUV · General, Plastic Surgery
Decision Date
Jan 11, 2019
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ICOONE h is indicated for the relief of minor muscle aches and pain, relief of muscle spasms, temporary improvement of local blood circulation, and temporary reduction in the appearance of cellulite.
Device Story
Therapeutic massager; utilizes motorized rollers, suction, and light emission (LED 650nm, Laser 915nm) via Robosolo handpiece; operator-guided application to patient body; light sources activated via display (combined or independent); suction and light emission activate simultaneously upon handpiece activation; used in clinical settings by trained operators; output provides mechanical massage and light-induced heating; intended to relieve muscle aches/spasms, improve circulation, and reduce cellulite appearance.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (cytotoxicity, irritation, sensitization) and adherence to electrical safety and electromagnetic compatibility standards (IEC 60601-1, IEC 60601-1-2, IEC 60825-1).
Technological Characteristics
System includes motorized rollers, vacuum, and light sources (650nm LED, 915nm Laser). Power: 240/110 Vac. Safety standards: IEC 60825-1:2007, IEC 60601-1:2005, IEC 60601-2-22:2007, CEI EN 60601-1:2007/A11:2012, CEI EN 60601-1-2:2010. Biocompatibility: Cytotoxicity, irritation, sensitization. Light output: Continuous wave (CW), non-adjustable intensity.
Indications for Use
Indicated for relief of minor muscle aches/pain, muscle spasms, temporary improvement of local blood circulation, and temporary reduction in cellulite appearance in adult patients.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
I-tech Industries Srl % Jay Mansour Principal Mansour Consulting LLC 845 Aronson Lake Court Roswell. Georgia 30075
January 11, 2019
Re: K182453
Trade/Device Name: ICOONE h (also referred to as ICOONE LASER and ICOONE -h LASER) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: NUV Dated: December 7, 2018 Received: December 12, 2018
Dear Jay Mansour:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Neil R Ogden - Digitally signed by Neil R
S Ogden -S
Date: 2019.01.11 12:08:58
-05'00'
Rinite S Asher MD MBA FAC
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182453
Device Name
ICOONE h (also referred to as ICOONE LASER and ICOONE -h LASER)
### Indications for Use (Describe)
ICOONE h is indicated for the relief of minor muscle aches and pain, relief of muscle spasms, temporary improvement of local blood circulation, and temporary reduction in the appearance of cellulite.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510k Summary
# As required by 21 CFR 807.92 (c)
# 1- Date summary prepared: 1/11/2019
#### 2-Owner/Submitter/Sponsor/Applicant information:
I-Tech Industries S.r.l Via Cicogna 34/B 40068 San Lazzaro Di Savena Bologna Italy
#### 3-Device information:
Common/usual/classification name: Massager, Vacuum, Light Induced Heating
Device name: ICOONE h (also referred to as ICOONE LASER and ICOONE -h LASER)
| FDA 3 letter code | NUV |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| FDA regulation number: 21 CFR | 878.4810 |
| Regulation medical specialty | General & Plastic Surgery |
| Review panel | General & Plastic Surgery<br>Division of Surgical Devices (DSD)<br>General Surgery Devices Branch One - Light Based/Laser (GSDB1) |
| Class | II |
## 4- Substantial equivalency is claimed against the following predicate device(s):
| 510k<br>number | Trade or Proprietary or Model Name | Manufacturer | Primary predicate? |
|----------------|------------------------------------|--------------------|--------------------|
| K053611 | SmoothShapes | Biocellulase, Inc. | No |
| K061603 | SmoothShapes | Biocellulase, Inc. | Yes |
## 5- Description of the device:
The Icoone-h device is Therapeutic Massager machine attached to pivoting wheels, connected to a rolling stand, with a series of hand pieces equipped with motorized rollers, which are the core of the technology and that, opportunely guided by an operator, are applied to the patient's body.
lcoone-h Laser is also equipped with two light sources inside the Robosolo hand piece, each with the following wave lengths:
LED @ 650nm (50Mw) Laser @ 915nm (1W)
{5}------------------------------------------------
The light sources can be activated through the display, either combined or independently (only one of the two or both).
The sources are neither adjustable in intensity (always output at nominal value, as per specifications) nor in frequency (always continuous - CW).
Once selected, by turning on the Robosolo hand piece, both suctioning and light emission are activated at the same time.
The light is emitted via laser diodes or LEDs controlled by a dedicated power driver.
### 6- Intended use + indications for use
ICOONE h is indicated for the relief of minor muscle aches and pain, relief of muscle spasms, temporary improvement of local blood circulation, and temporary reduction in the appearance of cellulite.
### 7- Basis for a determination of substantial equivalency:
- (a) Indications for use:
The indication for use is identical to the predicates'.
- (b) Technological characteristics:
The technological characteristics are substantially equivalent to the predicate device, meeting the same technical standards.
| Device name | Icoone -h | Biocellulase<br>Smoothshapes | Biocellulase<br>Smoothshapes | Substantial<br>equivalency: same or<br>different |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| 510(k) number | TBD | K053611 | K061603 | |
| System<br>component | Laser, Massage, Suction, Rollers | Laser, Massage,<br>Suction, Rollers | Laser, Massage,<br>Suction, Rollers | Same |
| Mechanical<br>massage | yes | yes | yes | Same |
| Weight | 191.80 lb (87 kg) | 59.52 lb (27 kg) | 101.41 lb (46 kg) | Different (but similar) |
| Dimension | 37.40x80.71x19.68 inch (95x205x50 cm) | 31.47x19.68x13.78<br>inch (80x50x35 cm) | 21.26x23.23x51.97<br>inch (54x59x132 cm) | Different (but similar) |
| Light emitting<br>safety | IEC 60825-1: 2007 (in conjunction with IEC 60601-1: 2005)<br>IEC 60601-2-22: 2007 | 21 CFR 1040 | 21 CFR 1040 | Same (FDA accepts the<br>equivalency) |
| Electrical Safety | CEI EN 60601-1:2007 /A11: 2012 | IEC/EN 60601-1-1: 1995<br>IEC/EN 60601-1-1: 1995 | IEC/EN 60601-1-1: 1995<br>IEC/EN 60601-1-1: 1995 | Same<br>(our device is not a system<br>and IEC 60601-1 is not<br>applicable) |
| EMC | CEI EN 60601-1-2: 2010 | IEC 60601-1-2:1993,<br>2001 | IEC 60601-1-2:1993,<br>2001 | Same |
| Patient contact<br>material | Handpiece suction rollers | Handpiece suction<br>rollers | Handpiece suction<br>rollers | Same |
| Biocompatibility | Cytotoxicity, irritation, Sensitization | unknown | unknown | (As per requirements) |
| Indication for<br>use | Relieves of minor muscle aches and pains, Relieves muscle<br>spasms, Temporary improves local blood circulation,<br>Temporary reduces in appearance of cellulite | Relieves of minor muscle<br>aches and pains, Relieves<br>muscle spasms, Temporary<br>improves local blood<br>circulation, Temporary<br>reduces in appearance of<br>cellulite | Relieves of minor muscle<br>aches and pains, Relieves<br>muscle spasms, Temporary<br>improves local blood<br>circulation, Temporary<br>reduces in appearance of<br>cellulite | Same |
{6}------------------------------------------------
| Power Source | 240/110 Vac | 240/110 Vac | 240/110 Vac | Same |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|-------------------------------------|--------------------------------------------------------|
| IR power | max 1W | max 1W / 915nm<br>max 0.5 W / 650 nm | max 15W / 915 nm<br>max 1W / 650 nm | Different:<br>Similar to K053611<br>Safer than K061603 |
| Infrared<br>wavelengths | 650nm (LED)<br>915 nm (LASER) | 640 nm - 660 nm (LED)<br>880 nm - 940 nm (LASER) | 650nm (LED)<br>915 nm (LASER) | Same |
| Max. IR output<br>energy density | 26.9 W/m2 | unknown | unknown | (safe and effective) |
| Vacuum | Fractioned | Pulsed | Pulsed | Different (but similar) |
| Treated area | 3.15x2.36 inch (80x60mm - Robosolo head)<br>2.36x1.97 inch (60x50mm - Robotwin head) x2<br>1.97x1.38 inch (50x35mm - Robomini head)<br>0.031x0.031 inch (0.8x0.8 mm - Robomicro head with<br>applicator "D")<br>0.027x0.91 inch (0.7x23 mm - Robomicro head with<br>applicator "C")<br>0.59x1.14 inch (15x29 mm - Robomicro head with<br>applicator "B") | 1.57x1.57 inch<br>(40x40mm) | unknown | Different (but similar) |
- (c) Non-clinical tests- brief discussion:
The standards that we comply with (listed above) confirm the safety, fit and effectiveness.
- (d) Clinical tests- brief discussion:
Not applicable.
- (e) Non-clinical and clinical tests- conclusions drawn demonstrating that the device is as safe and as effective, and performs as well as or better than the predicate device(s):
ICOONE h performs as designed, in accordance with requirements. It is as safe and effective as the predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.