Aplio a550 and a450, Software V2.8 Diagnostic Ultrasound System

K182427 · Canon Medical Systems Corporation · IYN · Nov 2, 2018 · Radiology

Device Facts

Record IDK182427
Device NameAplio a550 and a450, Software V2.8 Diagnostic Ultrasound System
ApplicantCanon Medical Systems Corporation
Product CodeIYN · Radiology
Decision DateNov 2, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric

Indications for Use

The Diagnostic Ultrasound Systems Aplio a550 Model CUS-AA550 and Aplio a450 Model CUS-AA450 are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs, trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial) and laparoscopic.

Device Story

Mobile diagnostic ultrasound systems; utilize wide array of probes (flat linear, convex linear, sector) with 2-20 MHz frequency range. Input: ultrasound signals from patient anatomy. Processing: system transforms signals into diagnostic images/flow analysis. Output: real-time ultrasound images/measurements displayed on integrated monitor. Used in clinical settings by trained healthcare professionals. Output assists in diagnosis of structures and dynamic processes. Benefits: non-invasive diagnostic imaging.

Clinical Evidence

No clinical data. Safety and efficacy demonstrated through bench testing, risk analysis, and software validation in accordance with FDA guidance and international standards (IEC 60601-1, IEC 60601-2-37, ISO 10993).

Technological Characteristics

Mobile diagnostic ultrasound system; electronic convex, sector, linear, and pencil scanning methods. 4-channel probe connectivity. 21.5-23 inch LED monitors. Frequency range 2-20 MHz. Complies with IEC 60601-1, IEC 60601-2-37, and ISO 10993. Software life cycle per IEC 62304.

Indications for Use

Indicated for visualization of structures and dynamic processes in fetal, abdominal, intra-operative (abdominal), pediatric, small organs, trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (adult/pediatric), peripheral vascular, transesophageal, musculo-skeletal (conventional/superficial), and laparoscopic applications.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Canon Medical Systems Corporation Orlando Tadeo, Jr. Sr. Manager, Regulatory Affairs 2441 Michelle Drive TUSTIN, CA 92780 November 2, 2018 Re: K182427 Trade/Device Name: Aplio a550 and a450, Software V2.8 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: September 4, 2018 Received: September 6, 2018 Dear Orlando Tadeo Jr.: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael D. O'Hara For Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K182427 Device Name Aplio a550 and a450, Software V2.8 Diagnostic Ultrasound System #### Indications for Use (Describe) The Diagnostic Ultrasound Systems Aplio a550 Model CUS-AA550 and Aplio a450 Model CUS-AA450 are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal), pediatric, small organs, trans-vaginal, trans-rectal, neonatal cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial) and laparoscopic. Type of Use (Select one or both, as applicable) | <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |-----------------------------------------------------------------------------------------------------------| | <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Transducer: Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application<br>Specific<br>(Tracks 3) | B | M | PWD | CWD | Color Doppler | Precision Imaging<br>Combined (Specify) | Apli Pure | Micro Pure | BEAM | Power | TDI | Elastography | SMI | Shear wave | 4D<br>(Volume color) | 3D Color<br>(Volume color) | STIC | STIC Color | Smart 3D | Fusion | Smart Navigation | 2D WMT | Other | [Note] | | |------------------------------------------------|---|---|-----|-----|---------------|-----------------------------------------|-----------|------------|------|-------|-----|--------------|-----|------------|----------------------|----------------------------|------|------------|----------|--------|------------------|--------|-------|--------|--| | Ophthalmic | | | | | | | | | | | | | | | | | | | | | | | | | | | Fetal | N | N | N | N | N | 2 | N | N | | N | | | N | | N | N | N | N | N | N | N | | 4,6 | | | | Abdominal | N | N | N | N | N | 2,3 | N | N | | N | | N | N | N | N | N | N | N | N | N | N | | 4,5,6 | | | | Intra-operative (Abdominal) | N | N | N | | N | 2 | N | N | | N | | | N | | | | | | N | N | | | 4 | | | | Intra-operative (Neuro) | | | | | | | | | | | | | | | | | | | | | | | | | | | Laparoscopic | N | N | N | | N | 2 | N | N | | N | | N | N | | | | | | | | | | | | | | Pediatric | N | N | N | N | N | 2,3 | N | N | | N | | | N | | N | N | N | N | N | N | N | | 4,6 | | | | Small Organ (Note 1) | N | N | N | | N | 2 | N | N | N | N | N | | N | N | N | | | | | N | N | N | | 4 | | | Neonatal Cephalic | N | N | N | N | N | 2,3 | N | N | | N | | | | | | | | | N | | | | | | | | Adult Cephalic | N | N | N | N | N | 3 | | | | N | | | | | | | | | | | | | | | | | Trans-rectal | N | N | N | | N | 2 | N | N | | N | | N | N | N | N | N | | | | N | N | | 4,6 | | | | Trans-vaginal | N | N | N | | N | 2 | N | N | | N | | N | N | N | N | N | | | | N | N | | 4,6 | | | | Trans-urethral | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal(Conventional) | N | N | N | | N | 2 | N | N | N | N | N | | N | N | | | | | N | N | N | | 4 | | | | Musculo-skeletal (Superficial) | N | N | N | | N | 2 | N | N | N | N | N | | N | N | | | | | N | N | N | | 4 | | | | Intravascular | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Adult | N | N | N | N | N | 3 | N | | | N | N | | | | | | | | | | | N | 5,6 | | | | Cardiac Pediatric | N | N | N | N | N | 3 | N | | | N | N | | | | | | | | | | | N | 5,6 | | | | Intravascular (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (Cardiac) | N | N | N | N | N | 3 | N | | | | N | | | | | | | | | | | N 6 | | | | | Intra-cardiac | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | Peripheral vessel | N | N | N | N | N | 2 | N | N | N | N | N | | N | N | | | | | N | N | N | | 4 | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix Prescription Use Only (Per 21 CFR 801.109) Note 1 Small organ includes thyroid, breast and testicle Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 Smart Sensor 3D Note 5 CHI (Per FDA approved contrast agent prescribing information) {4}------------------------------------------------ ### 510(k) Premarket Notification Aplio a550 and a450, Software V2.8 Diagnostic Ultrasound System #### Clinical Application Mode of Operation Specific CMD ama ടിമ്പിച്ചു. അമ്രാ Other [Note] IN B OLL LINM d (Tracks 3) IƏMO, АЕМ ЛЕӘЧ TIC Colo FAM រជាមួយនឹងស្រីននា mart 3D ğısədç) pəuiquq əhəli dəyə qalınmışdır. Bu mərkəzi və bir mənasının qalınmışdır. Bu mərkəzi və bir mənistəsini mərkəzi və bir mənistəsini mərkəzi və bir mənistəsinə mərkəzi və bir mənistəsi icro Pur ງ​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​ ശ്രാ വ uoisn DITERIARN TIEW suiseur I uoisipa. Ophthalmic Fetal P Abdominal P P P P 3 P P b Intra-operative (Abdominal) Intra-operative (Neuro) Laparoscopic Pediatric P P P P P 3 P P P Small Organ (Note 1) Neonatal Cephalic P P P P P 3 P P P P P P Adult Cephalic P 3 P Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Card.) Musculo-skeletal(Conventional) Musculo-skeletal (Superficial) Intravascular Other (Specify) Cardiac Adult P P P P P P P P 3 P Cardiac Pediatric P P P P P 3 P P P P Intravascular (Cardiac) Trans-esoph. (Cardiac) Intra-cardiac Other (Specify) Peripheral vessel Other (Specify) System: Aplio a550 and a450, Software V2.8 Transducer: PST-25BT Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K173090 Prescription Use Only (Per 21 CFR 801.109) Note 1 Small organ includes thyroid, breast and testicle Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 Smart Sensor 3D Note 5 CHI (Per FDA approved contrast agent prescribing information) {5}------------------------------------------------ System: Aplio a550, Aplio a450 system ver.1.3 Transducer:_ PST-30BT Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | Mode of Operation | | | | | | | | | | | | | | | | | | | | | | | | | | |---------------------------------|-------------------|---|-----|-----|-------------------------------------------|---------------|-----------|------------|------|-------|-----|--------------|-----|------------|----------------------|----------|------|------------|----------|--------|------------------|--------|-----------------|--|--|--| | Specific<br>(Tracks 3) | B | M | PWD | CWD | Precision Imaging<br>Combined (Specify) * | Color Doppler | Apli Pure | Micro Pure | BEAM | Power | TDI | Elastography | SMI | Shear wave | 4D<br>(Volume color) | 3D Color | STIC | STIC Color | Smart 3D | Fusion | Smart Navigation | 2D WMT | Other<br>[Note] | | | | | Ophthalmic | | | | | | | | | | | | | | | | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | | | | | | | | | | | | | | | | Abdominal | P | P | P | P | P | 3 | P | | | | P | | | P | | | | | | | | | 5 | | | | | Intra-operative (Abdominal) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | | | | | | | | | | | | | | | | Pediatric | P | P | P | P | P | 3 | P | | | | P | | | P | | | | | | | | | | | | | | Small Organ (Note 1) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Neonatal Cephalic | P | P | P | P | P | 3 | | | | | P | | | | | | | | | | | | | | | | | Adult Cephalic | P | P | P | P | P | 3 | | | | | P | | | | | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Conventional) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Adult | P | P | P | P | P | 3 | | | | | P | P | | | | | | | | | | P | 5 | | | | | Cardiac Pediatric | P | P | P | P | P | 3 | | | | | P | P | | | | | | | | | | P | 5 | | | | | Intravascular (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | | Peripheral vessel | | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K173090 Prescription Use Only (Per 21 CFR 801.109) Note 1 Small organ includes thyroid, breast and testicle Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 Smart Sensor 3D Note 5 CHI (Per FDA approved contrast agent prescribing information) {6}------------------------------------------------ Transducer:_ PST-50BT Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | Mode of Operation | | | | | | | | | | | | | | | | | | | | | | | | | |---------------------------------|-------------------|---|-----|-----|---------------|-------------------------------------------|------------|-----------|------|-------|-----|--------------|-----|------------|----------------------|----------------------------|------|------------|----------|--------|------------------|--------|-------|--------|--| | Specific<br>(Tracks 3) | B | M | PWD | CWD | Color Doppler | Precision Imaging<br>Combined (Specify) * | Micro Pure | Apli Pure | BEAM | Power | TDI | Elastography | SMI | Shear wave | 4D<br>(Volume color) | 3D Color<br>(Volume color) | STIC | STIC Color | Smart 3D | Fusion | Smart Navigation | 2D WMT | Other | [Note] | | | Ophthalmic | | | | | | | | | | | | | | | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | | | | | | | | | | | | | | | Abdominal | P | P | P | P | P | 3 | P | | | P | | | P | | | | | | | | | | | | | | Intra-operative (Abdominal) | | | | | | | | | | | | | | | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | | | | | | | | | | | | | | | Pediatric | P | P | P | P | P | 3 | P | | | P | | | P | | | | | | | | | | | | | | Small Organ (Note 1) | | | | | | | | | | | | | | | | | | | | | | | | | | | Neonatal Cephalic | P | P | P | P | P | 3 | | | | P | | | | | | | | | | | | | | | | | Adult Cephalic | P | P | P | P | P | 3 | | | | P | | | | | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Conventional) | | | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | | | | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Adult | P | P | P | P | P | 3 | | | | P | P | | | | | | | | | | P | | | | | | Cardiac Pediatric | P | P | P | P | P | 3 | | | | P | P | | | | | | | | | | P | | | | | | Intravascular (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | Peripheral vessel | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K173090 Prescription Use Only (Per 21 CFR 801.109) Note 1 Small organ includes thyroid, breast and testicle Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 Smart Sensor 3D Note 5 CHI (Per FDA approved contrast agent prescribing information) {7}------------------------------------------------ Transducer: PSI-70BT Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | Mode of Operation | | | | | | | | | | | | | | | | | | | | | | | | | | | |---------------------------------|-------------------|---|-----|-----|---------------|-------------------------------------------|-----------|------------|------|-------|-----|--------------|-----|------------|----------------------|----------|------|------------|----------|--------|------------------|--------|-------|--------|--|--|--| | Specific<br>(Tracks 3) | B | M | PWD | CWD | Color Doppler | Precision Imaging<br>Combined (Specify) * | Apli Pure | Micro Pure | BEAM | Power | TDI | Elastography | SMI | Shear wave | 4D<br>(Volume color) | 3D Color | STIC | STIC Color | Smart 3D | Fusion | Smart Navigation | 2D WMT | Other | [Note] | | | | | Ophthalmic | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Abdominal | P | P | P | P | P | 3 P | P | | | P | | | | P | | | | | | | | | | | | | | | Intra-operative (Abdominal) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Pediatric | P | P | P | P | P | 3 P | P | | | P | | | | P | | | | | | | | | | | | | | | Small Organ (Note 1) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Neonatal Cephalic | P | P | P | P | P | 3 P | P | | | P | | | | | | | | | | | | | | | | | | | Adult Cephalic | P | P | P | P | P | 3 | | | | P | | | | | | | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Conventional) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Adult | P | P | P | P | P | 3 P | P | | | P | P | | | | | | | | | | P | | 5 | | | | | | Cardiac Pediatric | P | P | P | P | P | 3 P | P | | | P | P | | | | | | | | | | P 5 | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Peripheral vessel | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K173090 Prescription Use Only (Per 21 CFR 801.109) Note 1 Small organ includes thyroid, breast and testicle Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 Smart Sensor 3D Note 5 CHI (Per FDA approved contrast agent prescribing information) {8}------------------------------------------------ #### System: Aplio a550 and a450, Software V2.8 Transducer:__ PVT-375BT Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application<br>Specific<br>(Tracks 3) | Mode of Operation | B | M | PWD | CWD | Precision Imaging<br>Combined (Specify) *<br>Color Doppler | Apli Pure | Micro Pure | BEAM | Power | TDI | Elastography | SMI | Shear wave | 4D<br>(Volume color)<br>3D Color | STIC | STIC Color | Smart 3D | Fusion | Smart Navigation | 2D WMT | Other<br>[Note] | |------------------------------------------------|-------------------|---|---|-----|-----|------------------------------------------------------------|-----------|------------|------|-------|-----|--------------|-----|------------|----------------------------------|------|------------|----------|--------|------------------|--------|-----------------| | Ophthalmic | | | | | | | | | | | | | | | | | | | | | | | | Fetal | | P | P | P | P | 2 | P | P | | P | | P | | | | | | P | | | | | | Abdominal | | P | P | P | P | 2 | P | P | | P | P | P | P | P | | | | P | P | P | | 4, 5 | | Intra-operative (Abdominal) | | | | | | | | | | | | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | | | | | | | | | | | | Pediatric | | P | P | P | P | 2 | P | P | | P | | | P | | | | | P | P | P | | 4 | | Small Organ (Note 1) | | | | | | | | | | | | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Conventional) | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Adult | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | | | | | | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | Peripheral vessel | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K173090 Prescription Use Only (Per 21 CFR 801.109) Note 1 Small organ includes thyroid, breast and testicle Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 Smart Sensor 3D Note 5 CHI (Per FDA approved contrast agent prescribing information) {9}------------------------------------------------ Transducer:_ PVT-375SC Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | Mode of Operation | | | | | | | | | | | | | | | | | | | | | | | |---------------------------------|-------------------|---|-----|-----|---------------|-------------------------------------------|-----------|------------|------|-------|-----|--------------|-----|------------|----------------------|----------------------------|------|------------|----------|--------|------------------|--------|--------------| | Specific<br>(Tracks 3) | B | M | PWD | CWD | Color Doppler | Precision Imaging<br>Combined (Specify) * | Apli Pure | Micro Pure | BEAM | Power | TDI | Elastography | SMI | Shear wave | 4D<br>(Volume color) | 3D Color<br>(Volume color) | STIC | STIC Color | Smart 3D | Fusion | Smart Navigation | 2D WMT | Other [Note] | | Ophthalmic | | | | | | | | | | | | | | | | | | | | | | | | | Fetal | P | P | P | | P | 2 | P | P | | P | | | P | | | | | P | | | | | 4 | | Abdominal | P | P | P | | P | 2 | P | P | | P | P | P | P | P | | | | P | P | P | P | | 4,5 | | Intra-operative (Abdominal) | | | | | | | | | | | | | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | | | | | | | | | | | | | Pediatric | P | P | P | | P | 2 | P | P | | P | | | P | | | | | P | P | P | P | | 4 | | Small Organ (Note 1) | | | | | | | | | | | | | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Conventional) | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Adult | | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | | | | | | | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | Peripheral vessel | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K173090 Prescription Use Only (Per 21 CFR 801.109) Note 1 Small organ includes thyroid, breast and testicle Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 Smart Sensor 3D Note 5 CHI (Per FDA approved contrast agent prescribing information) {10}------------------------------------------------ Transducer:_ PVT-382BT Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | Mode of Operation | | | | | | | | | | | | | | | | | | | [Note] | | | | | | |---------------------------------|-------------------|---|-----|-----|---------------|----------------------|-------------------|-----------|------------|------|-------|-----|--------------|-----|------------|----------------------|----------|------|------------|----------|--------|------------------|--------|-------|-----| | Specific<br>(Tracks 3) | B | M | PWD | CWD | Color Doppler | Combined (Specify) * | Precision Imaging | Apli Pure | Micro Pure | BEAM | Power | TDI | Elastography | SMI | Shear wave | 4D<br>(Volume color) | 3D Color | STIC | STIC Color | Smart 3D | Fusion | Smart Navigation | 2D WMT | Other | | | Ophthalmic | | | | | | | | | | | | | | | | | | | | | | | | | | | Fetal | P | P | P | P | P | 2 | P | P | P | | P | | | | P | | | P | P | P | | P | P | | 4 | | Abdominal | P | P | P | P | P | 2 | P | P | P | | P | | | | P | | | P | P | P | P | P | P | | 4,5 | | Intra-operative (Abdominal) | | | | | | | | | | | | | | | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | | | | | | | | | | | | | | | Pediatric | P | P | P | P | P | 2 | P | P | P | | P | | | | P | | | P | P | P | | P | P | | 4 | | Small Organ (Note 1) | | | | | | | | | | | | | | | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | | | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | | | | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Conventional) | | | | | | | | | | | | | | | | | | | | | | | | | | | Musculo-skeletal (Superficial) | | | | | | | | | | | | | | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Adult | | | | | | | | | | | | | | | | | | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | | | | |…
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...